Calibration and evaluation of the relative biological effectiveness for carbon-ion radiotherapy in a new relative to a clinically applied treatment planning system

Abstract

Background

The study objective was to validate the relative biological effectiveness (RBE) in RayStation for carbon-ion radiotherapy (CIRT) using the Syngo treatment planning system as reference.

Methods

Local effect model I was established in RayStation (Ray-LEM) with the same parameters as in LEM I in Syngo (Syngo-LEM). Three cube plans covering most of the tumors treated at our center were generated with Syngo-LEM. Ray-LEM re-calculated the Syngo plans and compared the RBEs to the Syngo counterparts. The results showed that RayStation RBE was smaller than Syngo RBE. To ensure that Ray-LEM reproduced Syngo RBE, the observed deviations were used to scale the maximum RBE (RBEmax) in Ray-LEM. After this calibration, we further compared the RayStation RBE to Syngo RBE using additional plans in both homogeneous phantoms and patients, to ensure that the calibrated Ray-LEM reproduced Syngo RBE even with more complex planning features.

Results

The calibration increased the RBEmax by 2.3% to raise the Ray-LEM RBE. The target mean RBE deviations in the phantom evaluation plans were median: 0.0 (minimum: − 1.1 to maximum: 0.7) %, and the target mean RBE deviations of the clinical target volumes of 16 patient cases were − 0.4 (− 1.5 to 0.2) %.

Conclusions

The residual RBE difference between RayStation and Syngo was found to be ≤ 1.0%. Thus, we can propose to use RayStation for clinical CIRT treatment planning. However, the potential differences due to the absorbed beam model warrants further exploration.

Backgroud

Clinical carbon-ion radiotherapy (CIRT) originated from Japan in 1994 and has been shown to have physical and biological advantages over proton or photon radiotherapy [1]. Following the pioneers’ steps, our center established CIRT in 2015 with the Syngo (V13C, Siemens, Germany) treatment planning system (TPS). Unlike the constant relative biological effectiveness (RBE) of 1.1 used for proton radiotherapy [2], Syngo integrates the local effect model I (LEM I) [3] and a pencil beam algorithm (PBA) to provide RBE-weighted doses for clinical CIRT treatment planning. LEM I was designed to calculate a variable RBE value of carbon-ion in combination with the active scanning technique according to the local particle spectra [4]. Although several versions of LEMs have been developed [5,6,7], LEM I is the only LEM applied in a clinical TPS (LEM is hereafter referred to as LEM I). So far, our center has treated more than 5000 patients using LEM implemented in Syngo (Syngo-LEM) [8]. Preliminary results supported the effectiveness of CIRT with LEM [9,10,11,12,13,14,15]. However, the performance of Syngo has gradually declined with increasing clinical demands for CIRT. In 2019, our center deployed RayStation (V10B, RaySearch, Sweden) in the CIRT system as an alternative to Syngo TPS. Although RayStation is suggested to provide much faster and more robust treatment planning using the same LEM (Ray-LEM) and PBA for CIRT for clinical implementation, the RBE-weighted dose calculation in RayStation warrants comprehensive evaluation.

For CIRT planning, RBE-weighted and corresponding absorbed dose distributions given by RayStation should be consistent with Syngo. However, our preliminary evaluation showed that RayStation gave slightly smaller RBEs than Syngo. Therefore, we needed to examine Ray-LEM and the corresponding fragment spectra referring to Syngo. Inaniwa et al. [16] performed a similar study using a modified microdosimetric kinetic model (MKM). They matched the dose distributions from two TPSs based on a cube plan. Inspired by their methodology, we first compared the RBE distributions between RayStation and Syngo. Then, based on the difference, we determined a calibration factor and applied it to Ray-LEM. As a result, RayStation could reproduce the RBE of the Syngo TPS.

Methods

Local effect model (LEM)

LEM assumes that the biological effect of ionizing radiation on a cellular scale only depends on the mean number of killing events per cell, which is given by the total local energy deposition, no matter whether the energy has been deposited by photons or ions [17]. The local dose distribution can be simulated by the track structure model, while the averaged biological effect can be simulated using the Monte-Carlo method to account for its stochastic nature [4]. Syngo-LEM follows the same concepts while simplifying the processes for better efficiency. In CIRT planning, the initial slopes of the cell-survival curves of chordoma, given by the parameter \(\alpha_{T, E}\) averaged over all ion types T and energies E, were pre-calculated and tabulated as RBEmax, i.e., the ratios of \(\alpha_{T, E}\) values to the \(\alpha_{X}\) value of photon radiation. As LEM was first applied in the treatment of skull-base tumors, \(\left( {{\upalpha }/{\upbeta }} \right)_{X}\) was set at 2 Gy to account for late reactions of the central nervous system [18]. Although some centers set a different value for the certain tumors [19], our center used the same \(\left( {{\upalpha }/{\upbeta }} \right)_{X}\) of 2 Gy for both tumors and normal tissues. In mixed carbon-ion irradiations, Syngo-LEM calculates the dose-weighted \(\alpha_{T, E}\) using the table based on corresponding fragment spectra [20] and determines the final RBE through a Monte-Carlo simulation.

Configuration of Ray-LEM

All model parameters in Ray-LEM were set to the same values as in Syngo-LEM. The \(\alpha_{T, E}\) table in Syngo-LEM was converted to RBEmax-values and imported into Ray-LEM [21]. The fragment spectra in Syngo were processed by RaySearch to derive a file compatible with RayStation as consistent as possible. However, the remaining differences were that the clinically used number of Monte-Carlo samples was 500 in Syngo but 1000 in RayStation. In addition, the step size in-depth in the spectra was changed from a dose gradient-dependent step size in Syngo to a fixed value of 0.1 cm in RayStation, and the energy resolution was also different for different ion types. Because Ray-LEM was presumed to be consistent with Syngo-LEM, the aforementioned parameters in Syngo-LEM were used in Ray-LEM, to minimize the potential difference from Syngo-LEM and to clarify the origin.

Calibration of Ray-LEM using Syngo-LEM as reference

The flowchart of the calibration shows in Fig. 1. In step 1, we defined the cube targets for calibration. Following the procedure in Inaniwa et al. [16], we generated three 6.0 × 6.0 × 6.0 cm3 cube targets in a WP. The depths of the center of the target volumes were 4.0, 10.0, and 16.0 cm, respectively. Depths and sizes of the target volumes aimed to cover most of the tumors treated at our center.

Fig. 1
figure 1

The flowchart of the calibration of Ray-LEM

In step 2, based on the cube targets, we generated three single beam treatment plans in Syngo. The RBE-weighted prescription was 4.0 Gy (RBE) with LEM. The dose grid size was 0.3 cm.

In step 3, Syngo-optimized plans were imported into RayStation, where dose re-calculations were performed based on Ray-LEM. By recalculation of the treatment plan with RayStation based on the fluence distribution, we obtained a new RBE-weighted and new absorbed dose distribution. As only the ratio of both reflects the RBE distribution, the RBE-weighted dose distribution of RayStation was normalized to that of Syngo. This implies that the absorbed dose distribution of RayStation was scaled as well. In this way, the RBE distribution is conserved and the deviations in RBE could now be read directly from the deviations in absorbed doses. Step 4 compared RayStation RBE to that of Syngo RBE. If the absorbed doses of RayStation were not the same as those of Syngo, we went to step 5. In this step, the average deviation of the absorbed doses was applied as a correction to the RBE distribution by applying it to the RBEmax-values of the input database for all depths and ion types. Steps 3–5 were iteratively performed until the averaged RayStation RBE was the same as in Syngo RBE. In the following, this procedure is termed as calibration of Ray-LEM.

This calibration procedure was chosen since we cannot access the RBE (and its deviation) in the TPS directly and to obtain the RBE deviation from the absorbed dose deviations, identical RBE-weighted doses are required.

Validation of Ray-LEM after calibration

The above calibration was derived using three simple geometric fields. However, real treatment plans may differ in field number, target shapes, prescriptions, and depths. Therefore, a variety of additional treatment plans differing in these features were re-calculated with RayStation after being generated with the Syngo. Similarly, the RBE-weighted dose of RayStation was normalized to that of Syngo, and the absorbed dose was scaled accordingly (to access RBE deviations). Residual deviations in absorbed dose were then used to characterize residual deviations in RBE and thus the accuracy of the calibration for the general cases. A similar approach was followed by Inaniwa et al. [16] and Fossati et al. [22]. Table 1 lists the details of the plan parameters.

Table 1 Parameters of the evaluation plans in WPs

In addition, we randomly selected 16 patients with typical target volumes and dose prescriptions used at our center. These included eight head and neck cases (two eye cancers, two parotid cancers, two nasopharynx carcinomas (NPCs), and two lacrimal cancers), two non-small-cell-lung-cancer cases, four abdominal cases (two liver cancers and two pancreatic cancers), and two prostate cases. Their CIRT treatment plans from Syngo were also re-calculated based on the same CT image sets and the corresponding stopping power conversion curve. After calibration, the mean RBEs of the clinical target volumes (CTVs) in RayStation were compared with those in Syngo in the same way as for the cube plans.

Results

Calibration

Figure 2 shows the absorbed and RBE-weighted depth dose distributions (DDDs) of both TPS after normalization of the RBE-weighted doses as well as the corresponding local differences on the central axis of three calibration plans. The figures in the left panel are the Ray-LEM results before calibration while the figures in the right panel are the Ray-LEM results after calibration. Table 2 shows the corresponding target mean doses. The correction by the RBEmax calibration factor was + 2.3%. The dose distributions after calibrations are in good agreement with those from Syngo plans. Although there are some fluctuations, the local difference of both RBE-weighted and absorbed doses agree well, which indicates the remaining RBE difference is small. The largest difference was found for the plan centered at 4.0 cm (Fig. 2a) from the surface of WP, where the target mean absorbed doses after calibration is still 0.01 Gy lower than that from Syngo.

Fig. 2
figure 2

RBE-weighted (DRBE-Ray) and absorbed (Dabs-Ray) DDDs at the central axis of the calibration plans after calibrations and the corresponding RBE-weighed (DRBE-Syngo) and absorbed (Dabs-Syngo) DDDs from Syngo. Local differences are also displayed in the figures. Absorbed DDDs before calibration (Dabs-not Cal) in the left panel are shown for comparisons. The centers of targets of (ac) are 4.0, 10.0, and 15.0 cm from the surface of WPs, respectively

Table 2 Target mean RBE-weighted and absorbed doses of the three calibration plans

Phantom evaluation

After RBE calibration was applied and the RBE-weighted dose normalization was performed, the RBE difference of the target mean RBE between RayStation and Syngo was in median: 0.0 (minimum: − 1.1 to maximum: 0.7) %. Detailed results are summarized in Additional file 1: Table S1. Figures 3 and 4 display the results of the plans whose targets were centered at 11.0 cm depth. Furthermore, Fig. 3 shows the comparisons of different target geometries, while Fig. 4 evaluates different prescriptions and beam arrangements.

Fig. 3
figure 3

Comparison of RBE-weighted (DRBE-Ray) and absorbed (Dabs-Ray) doses of the recalculated plans from Raystation and the corresponding RBE-weighted (DRBE-Syngo) and absorbed (Dabs-Syngo) doses from Syngo, together with the local differences. The prescription doses of (ad) are 3.0, 3.5, 4.0, and 4.5 Gy (RBE) respectively. The beam arrangements of (c, e, and f) are a single beam, two orthogonal beams, and two opposed beams

Fig. 4
figure 4

Comparison of RBE-weighted (DRBE-Ray) and absorbed (Dabs-Ray) doses of the re-calculated plans from RayStation and the corresponding RBE-weighted (DRBE-Syngo) and absorbed (Dabs-Syngo) doses from Syngo. Local differences are also displayed in the figures. Among them, the centers of targets of (ad) are 3.0, 7.0, 11.0, and 15.0 cm from the surface of WPs, respectively. Target sizes of (e, c, fh) are 4.0, 6.0, 8.0, 10.0, and 12.0 cm, respectively. The target shapes of (c and i) refer to cubic and spherical target volumes, respectively

Figure 3a–d show the absorbed and RBE-weighted DDDs as well as the corresponding local differences on the central axis of four cube plans. Their targets centered at the depths of 3.0 cm (a), 7.0 cm (b), 11.0 cm (c), and 15.0 cm (d), respectively. In Fig. 3a, the proximal of DDDs have a spike, which may be considered clinically by using a range shifter. At 11 cm depth, the cube sizes of Fig. 3e, c, f–h vary from 4.0 cm (e), 6.0 cm (c), 8.0 cm (f), and 10.0 cm (g) to 12.0 cm (h). In addition, Fig. 3i shows the result for a spherical target volume.

Figures 4a–d show the absorbed and RBE-weighted DDDs as well as the corresponding local differences at the central axis of the cube plans. Their prescriptions were 3.0 (a), 3.5 (b), 4.0 (c), and 4.5 Gy (RBE) (d), where Fig. 4c is the same as Fig. 3c while other cube plans use new prescriptions. Figure 4c, e, f show the DDDs together with the corresponding local differences at the center axis of the cube plans with a single beam (c), two orthogonal beams (e), and two opposed beams (f).

Evaluation of patients’ plans

Similarly, the comparison of patient cases was also conducted based on the calibrated Ray-LEM and the RBE-weighted doses were again scaled to match those of Syngo. The averaged/median clinical RBE-weighted prescriptions per fraction of 16 patients were 3.5 (2.75–9.0) Gy(RBE). The mean RBE deviations of CTVs between RayStation and that of Syngo were − 0.4 (− 1.5 to 0.2) %. Table 3 gives detailed information.

Table 3 Comparison of Ray-LEM and Syngo-LEM in 16 patient cases

Discussion

In this study, Ray-LEM was established with the same parameters and the same fragment and RBEmax input data were used for Ray-LEM and Syngo-LEM. Three cube plans were generated to quantify the potential deviations and to derive a scaling factor to correct the RBE of Ray-LEM to that of Syngo. After that, we applied the treatment plans within the WPs of different conditions and patient cases to evaluate the calibrated Ray-LEM. The results showed that the remaining difference was minimal.

Syngo was utilized to treat over 5000 patients in our center and proved to be effective in various diseases [9,10,11,12,13,14,15]. To make the former experience also valid for the future CIRT practice, we choose to calibrate the RBE of RayStation to that of Syngo, instead of scaling the current dose prescriptions. In this way, we can achieve the delivery of the same RBEs in both TPS. Furthermore, we also performed analysis on patients with different tumors and sites. Based on the comprehensive evaluation, we could adopt RayStation for the clinical CIRT planning and the same clinical results are expected as for Syngo TPS.

Without calibration, the local difference curves of three calibration cubes implied that RayStation underestimates the RBE. As the underestimation in the beam entrances and target regions was uniform, we implemented a single calibration factor. Furthermore, our ongoing analysis showed that the linear energy (LET) of the fragment spectra in RayStation was overestimated. While carbon-ions are the main contributor to the dose averaged LET (LETd) and thus to the RBE, the ion types from Z = 1–5 also contribute to the LETd in the distal part of the SOBP. Our approach could therefore not mitigate the RBE deviation in this region.

The largest difference after calibration was found in the tail behind the SOBP. Three factors likely played a role there. The main reason was the absorbed dose difference in this region. Our unpublished data showed that absorbed doses from RayStation were closer to the measurements than for Syngo. Second, the rebinned spectra by RayStation were not exactly the same as those in Syngo, which caused additional RBE differences. Third, the number of random sampling in Syngo-LEM was 500, which was a compromise between statistical accuracy and optimization speed. Therefore, the calculated RBE-weighted doses would have over 3% uncertainty [4]. Because the calculation speed of RayStation improved, the number of simulated events was set to 1000, which especially improved the accuracy in the tail behind the SOBP.

The present study generated calibration and evaluation plans in Syngo. RayStation only re-calculated the dose distribution instead of re-optimizing the dose distribution to avoid the absorbed dose differences due to different optimizers. After calibration, the evaluation plans whose targets centered at 11 cm from the surface of WPs in RayStation were re-optimized based on the same RBE-weighted doses. The target mean absorbed dose deviation to the Syngo-optimized plans was − 0.7% (− 2.8 to 7.1%). The detailed results are provided Additional file 1: Table S1. These results indicate that optimization may lead to different absorbed dose distributions based on the same RBE-weighted doses.

Our study only involved the fragment spectra of our facility. It is noteworthy that the beam quality of beamlines in other facilities may yield different fragment spectra. This may cause some RBE differences even when based on the same LEM. However, we investigated various conditions corresponding to very different fragment spectra, so it seems likely that the results would be similar. Only a few patients were evolved for evaluation in the present study. Hence, further evaluations using more patients with different tumors are warranted.

Conclusions

In this study, we first established Ray-LEM using the same parameters and the same fragment and RBEmax input data as the clinical applied Syngo-LEM. The comparison to Syngo RBE showed that Ray-LEM underestimated RBE. This difference was corrected by increasing the RBEmax in Ray-LEM by 2.3% to that of Syngo. After calibration, we comprehensively evaluated Ray-LEM referring to Syngo-LEM by using both phantom plans and patient cases. The results showed that the residual RBE difference between the two TPSs was ≤ 1.0%. Based on this result, it is justified to adopt the Ray-LEM for future clinical CIRT and similar clinical results can be expected as for the Syngo TPS.

Availability of data and materials

The datasets used and analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

CIRT:

Carbon-ion radiotherapy

Ray-LEM:

Local effect model I in RayStation

Syngo-LEM:

LEM I in Syngo

WP:

Water phantom

RBEmax:

The maximum value of relative biological effect

TPS:

Treatment planning system

LEM I:

Local effect model I

PBA:

Pencil beam algorithm

MKM:

Modified microdosimetric kinetic model

CTV:

Clinical target volume

DDD:

Depth dose distribution

DRBE-Ray:

RBE-weighted doses in RayStation

Dabs-Ray:

Absorbed doses in RayStation

DRBE-Syngo:

RBE-weighted doses in Syngo

Dabs-Syngo:

Absorbed doses in Syngo

LET:

Linear energy transfer

LETd:

Dose averaged LET

References

  1. Tsujii H, Kamada T, Shirai T, Noda K, Tsuji H, Karasawa K. Carbon-Ion radiotherapy. 2013.

  2. Paganetti H, Niemierko A, Ancukiewicz M, Gerweck LE, Goitein M, Loeffler JS, et al. Relative biological effectiveness (RBE) values for proton beam therapy. Int J Radiat Oncol. 2002;53(2):407–21.

    Article  Google Scholar 

  3. Scholz M, Kellerer AM, Kraft-Weyrather W, Kraft G. Computation of cell survival in heavy ion beams for therapy. Radiat Environ Biophys. 1997;36(1):59–66.

    Article  CAS  Google Scholar 

  4. Kramer M, Scholz M. Treatment planning for heavy-ion radiotherapy: calculation and optimization of biologically effective dose. Phys Med Biol. 2000;45(11):3319–30.

    Article  CAS  Google Scholar 

  5. Thilo ES, Michael S. Cluster effects within the local effect model. Radiat Res. 2007;167(3):319–29.

    Article  Google Scholar 

  6. Elsasser T, Weyrather WK, Friedrich T, Durante M, Iancu G, Kramer M, et al. Quantification of the relative biological effectiveness for ion beam radiotherapy: direct experimental comparison of proton and carbon ion beams and a novel approach for treatment planning. Int J Radiat Oncol Biol Phys. 2010;78(4):1177–83.

    Article  Google Scholar 

  7. Friedrich T, Scholz U, Elsasser T, Durante M, Scholz M. Calculation of the biological effects of ion beams based on the microscopic spatial damage distribution pattern. Int J Radiat Biol. 2012;88(1–2):103–7.

    Article  CAS  Google Scholar 

  8. https://www.sphic.org.cn/

  9. Gao J, Hu J, Guan X, Yang J, Hu W, Kong L, et al. Salvage carbon-ion radiation therapy for locoregionally recurrent head and neck malignancies. Sci Rep. 2019;9(1):4259.

    Article  Google Scholar 

  10. Cai X, Du Y, Wang Z, Li P, Yu Z, Zhang Q, et al. The role of carbon ion radiotherapy for unresectable locally recurrent rectal cancer: a single institutional experience. Radiat Oncol. 2020;15(1):209.

    Article  CAS  Google Scholar 

  11. Hu J, Huang Q, Gao J, Guan X, Hu W, Yang J, et al. Clinical outcomes of carbon-ion radiotherapy for patients with locoregionally recurrent nasopharyngeal carcinoma. Cancer-Am Cancer Soc. 2020;126(23):5173–83.

    Google Scholar 

  12. Chen J, Mao J, Ma N, Wu KL, Lu J, Jiang GL. Definitive carbon ion radiotherapy for tracheobronchial adenoid cystic carcinoma: a preliminary report. BMC Cancer. 2021;21(1):734.

    Article  CAS  Google Scholar 

  13. Chen J, Lu JJ, Ma N, Zhao J, Chen C, Fan M, et al. Early stage non-small cell lung cancer treated with pencil beam scanning particle therapy: retrospective analysis of early results on safety and efficacy. Radiat Oncol. 2019;14(1):16.

    Article  Google Scholar 

  14. Yang J, Gao J, Qiu X, Hu J, Hu W, Wu X, et al. Intensity-modulated proton and carbon-ion radiation therapy in the management of head and neck sarcomas. Cancer Med. 2019;8(10):4574–86.

    Article  CAS  Google Scholar 

  15. Guan X, Gao J, Hu J, Hu W, Yang J, Qiu X, et al. The preliminary results of proton and carbon ion therapy for chordoma and chondrosarcoma of the skull base and cervical spine. Radiat Oncol. 2019;14(1):206.

    Article  CAS  Google Scholar 

  16. Inaniwa T, Kanematsu N, Matsufuji N, Kanai T, Shirai T, Noda K, et al. Reformulation of a clinical-dose system for carbon-ion radiotherapy treatment planning at the National Institute of Radiological Sciences, Japan. Phys Med Biol. 2015;60(8):3271–86.

    Article  Google Scholar 

  17. Scholz M, Kraft G. Track structure and the calculation of biological effects of heavy charged particles. Adv Space Res. 1996;18(1–2):5–14.

    Article  CAS  Google Scholar 

  18. Schulz-Ertner D, Karger CP, Feuerhake A, Nikoghosyan A, Combs SE, Jäkel O, et al. Effectiveness of carbon ion radiotherapy in the treatment of skull-base chordomas. Int J Radiat Oncol Biol Phys. 2007;68(2):449–57.

    Article  Google Scholar 

  19. Liermann J, Naumann P, Weykamp F, Hoegen P, Debus J, Herfarth K. Effectiveness of carbon ion radiation in locally advanced pancreatic cancer. Front Oncol. 2021;11:708884.

    Article  Google Scholar 

  20. Parodi K, Mairani A, Brons S, Hasch BG, Sommerer F, Naumann J, et al. Monte Carlo simulations to support start-up and treatment planning of scanned proton and carbon ion therapy at a synchrotron-based facility. Phys Med Biol. 2012;57(12):3759–84.

    Article  CAS  Google Scholar 

  21. Karger CP, Peschke P. RBE and related modeling in carbon-ion therapy. Phys Med Biol. 2017;63(1):01TR02.

    Article  Google Scholar 

  22. Fossati P, Molinelli S, Matsufuji N, Ciocca M, Mirandola A, Mairani A, et al. Dose prescription in carbon ion radiotherapy: a planning study to compare NIRS and LEM approaches with a clinically-oriented strategy. Phys Med Biol. 2012;57(22):7543.

    Article  Google Scholar 

Download references

Acknowledgements

We appreciate the advice and help from our dear colleagues and also appreciate the generosity of Raysearch Company, who provided the Raystation for evaluation.

Funding

Not applicable.

Author information

Authors and Affiliations

Authors

Contributions

JF and H designed the study. WW and ZJ performed the study. W and SF helped analyzed the results. WW and JF wrote the manuscript. All authors reviewed and approved the manuscript.

Corresponding authors

Correspondence to Jingfang Zhao or Hao Shen.

Ethics declarations

Ethical approval and consent to participate

The Ethical Review Board of Shanghai Proton and Heavy Ion Center approved this study, and the written informed consent could be obtained from all patients prior to inclusion into the study.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1

. Table S1: All the treatment plans used for the evaluation of the Ray-LEM.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Wang, W., Huang, Z., Sun, W. et al. Calibration and evaluation of the relative biological effectiveness for carbon-ion radiotherapy in a new relative to a clinically applied treatment planning system. Radiat Oncol 17, 219 (2022). https://doi.org/10.1186/s13014-022-02181-5

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13014-022-02181-5

Keywords