Skip to main content

Table 3 Comparison of Ray-LEM and Syngo-LEM in 16 patient cases

From: Calibration and evaluation of the relative biological effectiveness for carbon-ion radiotherapy in a new relative to a clinically applied treatment planning system

List

Site

Fraction

DRBE/fx [ Gy(RBE)]

RBE Deviation (%)

P1

Eye

5

9.0

 − 1.2

P2

Eye

5

9.0

 − 1.5

P3

Parotid

20

3.5

0.2

P4

Parotid

20

3.5

0.2

P5

NPC

5

3.5

0.2

P6

NPC

5

3.5

0.0

P7

Lacrimal

18

3.5

0.0

P8

Lacrimal

18

3.5

 − 0.2

P9

Lung

10

6.5

 − 0.3

P10

Lung

8

8.0

 − 0.5

P11

Liver

10

2.75

0.0

P12

Liver

10

6.0

 − 0.5

P13

Pancreas

18

3.0

 − 0.4

P14

Pancreas

18

3.0

 − 0.4

P15

Prostate

12

3.8

 − 1.0

P16

Prostate

12

3.8

 − 1.20