From: Calibration and evaluation of the relative biological effectiveness for carbon-ion radiotherapy in a new relative to a clinically applied treatment planning system

Fig. 4

Comparison of RBE-weighted (DRBE-Ray) and absorbed (Dabs-Ray) doses of the re-calculated plans from RayStation and the corresponding RBE-weighted (DRBE-Syngo) and absorbed (Dabs-Syngo) doses from Syngo. Local differences are also displayed in the figures. Among them, the centers of targets of (ad) are 3.0, 7.0, 11.0, and 15.0 cm from the surface of WPs, respectively. Target sizes of (e, c, fh) are 4.0, 6.0, 8.0, 10.0, and 12.0 cm, respectively. The target shapes of (c and i) refer to cubic and spherical target volumes, respectively