Cochlear sparing in LINAC-based radiosurgery for vestibular schwannoma: a dosimetric comparison of dynamic conformal arc, IMRT and VMAT treatment plans

Abstract

Background

Stereotactic radiosurgery (SRS) is the preferred treatment for vestibular schwannoma (VS) in patients with preserved hearing and tumour diameter < 3 cm. Emerging evidence suggests restricting cochlear dose could preserve hearing. This retrospective replanning study aims to compare dynamic conformal arc therapy (DCAT), intensity modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT) plans for superiority of cochlear dose sparing without compromising tumour coverage.

Methods

Eligibility criteria included sporadic VS, serviceable hearing and availability of CT and MRI for planning. The original gross tumour volume and brainstem OAR volume were retained; the cochlea was newly contoured on the planning CT scan (bone window). Each case was replanned using the three above techniques, prescribing 12 Gy to the 80% isodose line. No dose constraint was applied to the cochlea.

Results

Eighteen patients were replanned. Mean tumour volume was 2.25 cc. Tumour coverage and tumour mean dose (DCAT: 14.2, IMRT: 14.6, VMAT: 14.5 Gy) were comparable. Paddick and RTOG conformity indices were better for DCAT (0.66 and 1.6) and VMAT (0.69 and 1.5) compared to IMRT (0.56 and 1.9). DCAT had superior gradient index (3.0) compared to VMAT (3.4) and IMRT (3.4). VMAT delivered the lowest mean brainstem maximum dose (8.3 Gy) and decreased the mean cochlear dose (3.4 Gy) by 2.3 and 2.1 Gy, and the mean cochlear maximum dose (3.6 Gy) by 2.4 and 2.5 Gy relative to DCAT and IMRT, respectively.

Conclusion

LINAC-based SRS treatment using VMAT can achieve better cochlear dose sparing than DCAT or IMRT while maintaining tumour coverage.

Background

Vestibular schwannoma (VS) is a benign tumour arising from the Schwann cells of the vestibulocochlear nerve. It makes up 10% of all intracranial neoplasms and 80% of tumours at the cerebellopontine angle [1]. The early symptoms are hearing loss and tinnitus in the affected ear, but larger tumours may also be associated with gait imbalance. With improvements in screening for hearing loss and increased access to MRI, the incidence of VS is increasing [2] and these tumours are often detected while the patient still has useful hearing. Management options include observation, microsurgery or radiosurgery.

Stereotactic radiosurgery (SRS) has become the preferred treatment for VS in patients with preserved hearing and tumour diameter < 3 cm [3, 4]. Originally, SRS was performed using Gamma Knife (GK), but SRS can now be delivered using conventional LINAC or CyberKnife with parity in treatment outcomes [5, 6]. Moreover, LINAC-based SRS has grown technically more advanced with its ability to use intensity modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT) techniques, with the possibility of better sparing of organs-at-risk (OAR), especially the cochlea. As there is accumulating evidence supporting dose sparing of the cochlea to preserve hearing [7,8,9,10,11], choosing a LINAC-based treatment technique that is superior in reducing the cochlear dose is prudent. Moreover, dose sparing must be achieved without compromising tumour coverage, since limiting the dose to the cochlea at the expense of tumour coverage is currently not recommended [12, 13].

The purpose of this retrospective dosimetric study is to compare plans generated using dynamic conformal arc therapy (DCAT), IMRT and VMAT techniques for LINAC-based SRS of VS. The primary aim is to compare cochlear dose sparing achieved using these techniques without compromising tumour coverage. The secondary aim is to assess tumour-related factors that affect dose sparing of the cochlea.

Methods

Patient selection

Patients included in this treatment planning study were previously treated at the Royal Adelaide Hospital between 2000 and 2015. Selection criteria were unilateral VS, pre-treatment pure-tone average (PTA) less than 50 decibels (dB) and availability of CT and MRI for planning. Patients were excluded if they had bilateral VS (neurofibromatosis type 2), previous treatment with radiotherapy or surgery or received SRS for recurrent VS. Hospital Research Ethics Committee approval was obtained for this study.

Planning methodology

The original treatment plans of the selected patients were retrieved from the archive. The gross tumour volume (GTV) and brainstem OAR volumes were retained. As it was not a routine practice to contour cochlea in our institution, the cochlea was newly contoured using the bone window (3000–4500/600–800 HU) of the planning CT scan by two investigators independently [14] (Fig. 1). Any contour variation was assessed and resolved. No GTV to planning target volume (PTV) margin expansion was used. For each patient, the SRS was replanned using three techniques: DCAT (Brainlab iPlan RT 4.5), IMRT (Brainlab iPlan RT 4.5) and VMAT (Brainlab Elements RT Cranial SRS 1.5) (Fig. 2), prescribing 12 Gy to the 80% isodose line. The plans were accepted only if tumour coverage was > 99.9%. The IMRT and DCAT planning was performed by a senior radiation therapist and the VMAT by a senior radiation physicist who had no knowledge of the dosimetric outcomes of the IMRT and DCAT plans.

Fig. 1
figure 1

Example of contouring cochlea (light blue), GTV (orange), and brainstem (green)

Fig. 2
figure 2

Isodose distribution for DCAT (top), IMRT (middle), VMAT (bottom) for a representative case. Yellow line = GTV; dark blue line = cochlea; green line = brainstem; PTV expansion = 0 mm; Colourwash threshold is set between 2 and 16 Gy

All treatment plans were generated using a Varian TrueBeam STx 6FFF beam model in the respective planning system. Both DCAT and VMAT plans were derived from standard non-coplanar 3 arc templates, with arc entry biased according to the laterality of the target. The Brainlab Elements VMAT planning system additionally provided an automatic arc angle optimisation driven by the locations of the PTV and OARs. The IMRT plans were generated using 9 beams, arranged according to target laterality. All plans were optimised for target coverage with sufficiently low priority on OARs to avoid compromising target coverage. Doses were computed using a 1 mm dose grid.

Plan comparison

The plan quality was compared using the following dosimetric parameters: target coverage (D98% and D2%), conformity index (CI) (RTOG, where lower is better, and Paddick, where higher is better), gradient index (GI, lower is better) and homogeneity index (HI) [15, 16]. OAR dose sparing achieved by each technique was compared using Dmax (D0.03cc) and Dmean for the cochlea, and Dmax (D0.03cc) and D1cc for the brainstem. Tumour characteristics—volume, dimensions, distance from the fundus of the internal auditory canal and Ohata stage [17]—were also recorded to assess their association with cochlear dose.

Statistical analysis

A linear mixed-effects model was employed to assess significant difference between the techniques for various dosimetric outcome parameters. A random effect was used to adjust for clustering on patient, as each patient underwent multiple techniques. Linear regression models were used to assess the association between various cochlear dosimetric parameters and tumour characteristics. The statistical software used was SAS 9.4 (SAS Institute Inc., Cary, NC, USA).

Results

Eighteen eligible patients were identified. The mean tumour volume was 2.25 cc (SD 2.3), median 1.38 cc, range 0.34–7.7 cc. Other tumour characteristics are provided in Table 1, and descriptive statistics (plan quality parameters and OAR dose) in Table 2.

Table 1 Tumour characteristics
Table 2 Descriptive statistics of dosimetric parameters

From a plan quality perspective, all plans had adequate target coverage (Table 2). The mean dose to the tumour volume, D2% and D98% were very similar across all groups. The mean RTOG CI for DCAT, IMRT and VMAT was 1.6 (SD 0.2), 1.9 (0.4) and 1.5 (0.2), respectively. The mean Paddick CI for DCAT, IMRT and VMAT was 0.66 (0.089), 0.56 (0.079) and 0.69 (0.079), respectively. The difference between the mean Paddick and RTOG CIs of the DCAT and VMAT plans was not statistically significant but were superior to IMRT (Tables 2 and 3). The DCAT plans achieved better GI, with a mean of 3.0 (0.5) compared with 3.4 (0.5) for IMRT and 3.4 (0.4) for VMAT, and this difference was statistically significant (P < 0.0001).

Table 3 Mean difference in plan quality and OAR dosimetric parameters between treatment techniques

VMAT was superior in sparing the OARs (brainstem and cochlea). For VMAT, the mean brainstem Dmax (D0.03cc), cochlear Dmean and cochlear Dmax (D0.03cc) were 8.3 Gy (4.2), 3.4 Gy (1.7) and 3.6 Gy (1.8), respectively. The difference between all cochlear dosimetric parameters was statistically superior in favour of VMAT. The mean cochlear dose was 2.3 and 2.1 Gy lower for VMAT than DCAT and IMRT, respectively. The mean cochlear Dmax (D0.03cc) was 2.4 and 2.5 Gy lower for VMAT than DCAT and IMRT, respectively. Additionally, the mean brainstem D1cc was 1.1 and 0.6 Gy lower for VMAT than DCAT and IMRT, respectively (Tables 2 and 3).

The linear mixed-effects model did not show any significant association between cochlear dosimetric parameters and tumour volume or dimensions but did show significant association with the fundus distance (Table 4). Hence, linear regression modelling was used to test the association between cochlear dose parameters and the fundus distance for each treatment technique (Table 5). The results showed statistically significant associations regardless of the treatment technique. For DCAT, IMRT and VMAT, for every 1 mm increase in fundus distance, the cochlear Dmax and Dmean decreased by 0.59 and 0.54 Gy, 1.0 and 0.92 Gy, and 0.73 and 0.67 Gy, respectively (Table 5).

Table 4 Mean difference in cochlear D0.03cc and cochlear Dmean by tumour characteristics
Table 5 Subgroup analysis by treatment technique: mean difference in cochlear D0.03cc and cochlear Dmean by distance from fundus

Discussion

Dose to the cochlea is increasingly recognised as likely a critical factor in preventing hearing loss [7,8,9,10,11]. However, cochlear dose sparing can be challenging since vestibular schwannomas are very often anatomically closely related to the cochlea. Indeed, of the various tumour characteristics in our study, the distance from the fundus was the only significant predictor of the cochlear dose regardless of the treatment technique, which is not surprising as the fundus distance is a surrogate for the tumour's proximity to the cochlea and will be inversely related to the cochlear dose. For SRS, although not supported by strong evidence, the current recommended cochlear dose limit is 4 Gy [18, 19]. Even for the currently recommended marginal dose of 11–14 Gy [3, 20, 21], it is still a significant challenge to restrict the dose to cochlea < 4 Gy. Moreover, for better preservation of hearing, restricting cochlear dose as low as < 3 Gy has been proposed [22]. Restricting the dose to the cochlea to such a low dose can be achieved only by means of a very steep dose gradient.

On the other hand, minimizing cochlear dose may not, by itself, be sufficient to preserve hearing as there are other factors that play a role in hearing outcome after SRS [23]. Dose to the cochlear nerve, transient tumour expansion and vascular injury from radiotherapy have all been speculated to contribute to the hearing loss. Hence, cochlear dose sparing at the expense of tumour coverage should be generally avoided. A practical approach will be to maintain tumour coverage and apply the ALARA principle to restrict dose to cochlea. Herein lies the importance of choosing the best treatment technique that facilitates maximum dose reduction to the cochlea and optimum tumour coverage.

The published dosimetric studies have mostly compared LINAC-based SRS with Gamma Knife SRS [24, 25]. We are not aware of any other study that compared all three DCAT, IMRT and VMAT techniques for LINAC-based SRS for VS, particularly from the standpoint of cochlear dose sparing. This comparison is very relevant as many centres have access to only LINAC-based SRS, with increasing availability of IMRT and VMAT techniques to deliver the treatment.

We were able to show that, although each technique generated high-quality plans that met the planning objectives, VMAT plans were superior to DCAT or IMRT. VMAT produced a more conformal plan than IMRT and was on par with DCAT. However, the superiority of VMAT was unequivocal in sparing OAR as it delivered the lowest brainstem and cochlear doses. Compared to IMRT and DCAT, the mean difference between cochlear dosimetric parameters was at least 2 Gy less for VMAT. We believe such a decrease may be of substantial advantage when the aim is to achieve maximum dose sparing to the cochlea. DCAT was able to achieve better GI compared to VMAT, but this by itself is of minimal significance since the higher GI for VMAT is likely a trade-off for achieving better OAR dose sparing.

Balik et al. [25] compared GK and VMAT plans for eight VS patients, prescribing 12–13 Gy, and showed no significant difference in plan quality parameters or OAR doses. In this study, the cochlear Dmean was 7.23 (3.13) Gy for VMAT plans, which is higher than the 3.4 (1.7) Gy in our study, probably due to the planning dose constraint they applied to cochlea (Dmax of 12 Gy). Abacioglu et al. [26] compared GK and RapidArc (TrueBeam Linac) SRS plans for 6 VS cases and prescribed 12.5 Gy. Of the two plan optimisation strategies they used, the second strategy is of interest since this aimed for maximum OAR sparing. The cochlear Dmean was 3.3 (0.8) and 4.1 (0.9) for GK and RapidArc, respectively. Our VMAT plans achieved a slightly lower cochlear Dmean, but it is comparable given our lower marginal dose of 12 Gy. A larger study by Kim et al. [24], comparing GK and VMAT plans of 19 patients, also did not show any significant difference in the cochlear doses. They showed that VMAT using 5 coplanar arcs produced superior Paddick CI for targets larger than 0.5 cc compared to GK, but the mean cochlear dose (6.0 [1.4] Gy) was higher. Their prescription dose was 12 Gy and they also used a maximum dose of 12 Gy as the cochlear dose constraint. Dutta et al. [27] compared DCAT (Brainlab) with CyberKnife and prescribed 13–15 Gy. They found that DCAT delivered a significantly higher cochlear Dmean compared to CyberKnife (6.9 [0.7] vs 5.4 [0.6]). Our study also demonstrated that VMAT reduced the mean cochlear dose by at least 2 Gy compared to DCAT and delivered lower cochlear Dmean than this study. This supports the use of VMAT in institutions that do not have access to CyberKnife, although further studies comparing these two techniques may quantify this.

Sharma et al. [28] created 5 different sets of static conformal field, DCAT and IMRT SRS plans for their cohort of 8 patients and, of these, the two DCAT and IMRT plans that prescribed 12 Gy to the 80% isodose line are of interest here. Unlike our study, IMRT was able to significantly lower the mean cochlear dose (5.0 [0.98] Gy), but not the maximum cochlear dose (9.8 [1.8] Gy), compared with DCAT. While their mean dose was slightly lower than that achieved in our IMRT plans (5.5 [2.3] Gy), we could not compare the maximum doses as they did not mention how the maximum dose was defined.

In contrast to our findings, Lagerwaald et al. [29] failed to show any difference in reducing the maximum dose to critical organs, including the cochlea and brainstem, when they compared VMAT with single arc and 5 arc DCAT for three cases. A plausible explanation for this failure of VMAT to achieve superior sparing of critical organs could be due to tumour location in the small sample. Our study showed that distance from fundus was a significant predictor of the ability to spare the cochlea regardless of the treatment technique. The 18 cases we selected may have increased the variability of tumour size and location, and thereby allowed the benefits of VMAT to be more evident.

Our study has some limitations. First, the VMAT planning was performed using newer planning software than that used to produce DCAT and IMRT plans. Second, we did not set a cochlear dose constraint, which could have achieved better results, not only for IMRT and DCAT but also for VMAT; we chose not to do this in order to maintain uniformity between treatment planning techniques and to allow maximum dose sparing with least risk of compromising tumour coverage. Lastly, although we demonstrated the dosimetric superiority of VMAT, we did not compare other factors like planning time and treatment delivery time.

Conclusion

Our study supports the use of VMAT as the preferred technique for LINAC-based SRS for vestibular schwannoma where hearing preservation is a goal. LINAC-based SRS treatment planning using the VMAT technique (Brainlab) can achieve better cochlear dose sparing than DCAT or IMRT while maintaining tumour coverage.

Data availability

Research data are stored in an institutional repository and will be shared upon reasonable request to the corresponding author.

Abbreviations

SRS:

Stereotactic radiosurgery

VS:

Vestibular schwannoma

DCAT:

Dynamic conformal arc therapy

IMRT:

Intensity modulated radiation therapy

VMAT:

Volumetric modulated arc therapy

OAR:

Organs-at-risk

PTA:

Pure-tone average

GTV:

Gross tumour volume

PTV:

Planning target volume

CI:

Conformity index

GI:

Gradient index

References

  1. Babu R, Sharma R, Bagley JH, Hatef J, Friedman AH, Adamson C. Vestibular schwannomas in the modern era: epidemiology, treatment trends, and disparities in management. J Neurosurg. 2013;119(1):121–30.

    Article  Google Scholar 

  2. Evans DG, Moran A, King A, Saeed S, Gurusinghe N, Ramsden R. Incidence of vestibular schwannoma and neurofibromatosis 2 in the North West of England over a 10-year period: higher incidence than previously thought. Otol Neurotol. 2005;26(1):93–7.

    Article  Google Scholar 

  3. Tsao MN, Sahgal A, Xu W, De Salles A, Hayashi M, Levivier M, et al. Stereotactic radiosurgery for vestibular schwannoma: International Stereotactic Radiosurgery Society (ISRS) Practice Guideline. J Radiosurg SBRT. 2017;5(1):5–24.

    Google Scholar 

  4. Wolbers JG, Dallenga AH, Mendez Romero A, van Linge A. What intervention is best practice for vestibular schwannomas? A systematic review of controlled studies. BMJ Open. 2013;3(2): e001345.

    Article  Google Scholar 

  5. Benghiat H, Heyes G, Nightingale P, Hartley A, Tiffany M, Spooner D, et al. Linear accelerator stereotactic radiosurgery for vestibular schwannomas: a UK series. Clin Oncol (R Coll Radiol). 2014;26(6):309–15.

    Article  CAS  Google Scholar 

  6. Mahboubi H, Sahyouni R, Moshtaghi O, Tadokoro K, Ghavami Y, Ziai K, et al. CyberKnife for treatment of vestibular schwannoma: a meta-analysis. Otolaryngol Head Neck Surg. 2017;157(1):7–15.

    Article  Google Scholar 

  7. Régis J, Tamura M, Delsanti C, et al. Hearing preservation in patients with unilateral vestibular schwannoma after gamma knife surgery. Prog Neurol Surg. 2008;21:142–51.

    Article  Google Scholar 

  8. Patel KS, Ng E, Kaur T, et al. Increased cochlear radiation dose predicts delayed hearing loss following both stereotactic radiosurgery and fractionated stereotactic radiotherapy for vestibular schwannoma. J Neurooncol. 2019;145(2):329–37.

    Article  Google Scholar 

  9. Lin RH, Wang TC, Lin CD, et al. Predictors of hearing outcomes following low-dose stereotactic radiosurgery in patients with vestibular schwannomas: a retrospective cohort review. Clin Neurol Neurosurg. 2017;162:16–21.

    Article  Google Scholar 

  10. Tamura M, Carron R, Yomo S, et al. Hearing preservation after gamma knife radiosurgery for vestibular schwannomas presenting with high-level hearing. Neurosurgery. 2009;64(2):289–96.

    Article  Google Scholar 

  11. Schumacher AJ, Lall RR, Lall RR, et al. Low-dose gamma knife radiosurgery for vestibular schwannomas: tumor control and cranial nerve function preservation after 11 Gy. J Neurol Surg Part B Skull Base. 2017;78(1):2–10.

    Google Scholar 

  12. Frischer JM, Gruber E, Schöffmann V, Ertl A, Höftberger R, Mallouhi A, et al. Long-term outcome after Gamma Knife radiosurgery for acoustic neuroma of all Koos grades: a single-center study. J Neurosurg. 2019;130(2):388–97.

    Article  Google Scholar 

  13. Jacob JT, Carlson ML, Schiefer TK, Pollock BE, Driscoll CL, Link MJ. Significance of cochlear dose in the radiosurgical treatment of vestibular schwannoma: controversies and unanswered questions. Neurosurgery. 2014;74(5):466–74.

    Article  Google Scholar 

  14. Pacholke HD, Amdur RJ, Schmalfuss IM, Louis D, Mendenhall WM. Contouring the middle and inner ear on radiotherapy planning scans. Am J Clin Oncol. 2005;28(2):143–7.

    Article  Google Scholar 

  15. Torrens M, Chung C, Chung HT, Hanssens P, Jaffray D, Kemeny A, et al. Standardization of terminology in stereotactic radiosurgery: Report from the Standardization Committee of the International Leksell Gamma Knife Society: special topic. J Neurosurg. 2014;121(Suppl):2–15.

    Article  Google Scholar 

  16. Paddick I. A simple scoring ratio to index the conformity of radiosurgical treatment plans. Technical note J Neurosurg. 2000;93(Suppl 3):219–22.

    Article  Google Scholar 

  17. Ohata K, Tsuyuguchi N, Morino M, Takami T, Goto T, Hakuba A, et al. A hypothesis of epiarachnoidal growth of vestibular schwannoma at the cerebello-pontine angle: surgical importance. J Postgrad Med. 2002;48(4):253–8.

    CAS  Google Scholar 

  18. Carlson ML, Vivas EX, McCracken DJ, Sweeney AD, Neff BA, Shepard NT, et al. Congress of neurological surgeons systematic review and evidence-based guidelines on hearing preservation outcomes in patients with sporadic vestibular schwannomas. Neurosurgery. 2018;82(2):E35–9.

    Article  Google Scholar 

  19. Hanna GG, Murray L, Patel R, Jain S, Aitken KL, Franks KN, et al. UK consensus on normal tissue dose constraints for stereotactic radiotherapy. Clin Oncol (R Coll Radiol). 2018;30(1):5–14.

    Article  CAS  Google Scholar 

  20. Bhandare N, Jackson A, Eisbruch A, Pan CC, Flickinger JC, Antonelli P, et al. Radiation therapy and hearing loss. Int J Radiat Oncol Biol Phys. 2010;76(3 Suppl):S50–7.

    Article  Google Scholar 

  21. Goldbrunner R, Weller M, Regis J, Lund-Johansen M, Stavrinou P, Reuss D, et al. EANO guideline on the diagnosis and treatment of vestibular schwannoma. Neuro Oncol. 2020;22(1):31–45.

    Article  Google Scholar 

  22. Baschnagel AM, Chen PY, Bojrab D, Pieper D, Kartush J, Didyuk O, et al. Hearing preservation in patients with vestibular schwannoma treated with Gamma Knife surgery. J Neurosurg. 2013;118(3):571–8.

    Article  Google Scholar 

  23. Tolisano AM, Hunter JB. Hearing preservation in stereotactic radiosurgery for vestibular schwannoma. J Neurol Surg B Skull Base. 2019;80(2):156–64.

    Article  Google Scholar 

  24. Kim H, Potrebko P, Rivera A, Liu H, Eldredge-Hindy HB, Gunn V, et al. Tumor volume threshold for achieving improved conformity in VMAT and Gamma Knife stereotactic radiosurgery for vestibular schwannoma. Radiother Oncol. 2015;115(2):229–34.

    Article  Google Scholar 

  25. Balik S, Chao S, Neyman G. Gamma Knife and volumetric modulated arc therapy stereotactic radiosurgery plan quality and OAR sparing comparison for pituitary adenomas and vestibular schwannomas. J Radiosurg SBRT. 2018;5(3):237–47.

    Google Scholar 

  26. Abacioglu U, Ozen Z, Yilmaz M, Arifoglu A, Gunhan B, Kayalilar N, et al. Critical appraisal of RapidArc radiosurgery with flattening filter free photon beams for benign brain lesions in comparison to GammaKnife: a treatment planning study. Radiat Oncol. 2014;9:119.

    Article  Google Scholar 

  27. Dutta D, Balaji Subramanian S, Murli V, Sudahar H, Gopalakrishna Kurup PG, Potharaju M. Dosimetric comparison of Linac-based (BrainLAB®) and robotic radiosurgery (CyberKnife ®) stereotactic system plans for acoustic schwannoma. J Neurooncol. 2012;106(3):637–42.

    Article  Google Scholar 

  28. Sharma SD, Chadha P, Talapatra K, Mahtre V, Kumar AP, Balasubramaniam A. Critical neurological structure sparing radiosurgery of vestibular schwannoma: dosimetric comparison of different techniques and dose prescription methods. J Cancer Res Ther. 2014;10(1):29–37.

    Article  Google Scholar 

  29. Lagerwaard FJ, Meijer OW, van der Hoorn EA, Verbakel WF, Slotman BJ, Senan S. Volumetric modulated arc radiotherapy for vestibular schwannomas. Int J Radiat Oncol Biol Phys. 2009;74(2):610–5.

    Article  Google Scholar 

Download references

Acknowledgements

The authors would like to acknowledge Brainlab, for loaning Elements RT Cranial SRS 1.5 to the Royal Adelaide Hospital. Brainlab had no other input to the conduct or reporting of this study.

Funding

The authors declare that no funds, grants, or other support were received during the preparation of this manuscript.

Author information

Authors and Affiliations

Authors

Contributions

All authors contributed to the study conception and design and data analysis. All authors were involved in writing the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Jeremy Khong.

Ethics declarations

Ethics approval and consent to participate

This is a planning comparison study. The Hospital Research Ethics Committee has confirmed that no ethical approval is required. Consent for research participation and publication was obtained at the time of consent for the original course of treatment.

Competing interests

The authors have no relevant financial or non-financial interests to disclose.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Khong, J., Govindaraj, R., Ramm, D. et al. Cochlear sparing in LINAC-based radiosurgery for vestibular schwannoma: a dosimetric comparison of dynamic conformal arc, IMRT and VMAT treatment plans. Radiat Oncol 18, 2 (2023). https://doi.org/10.1186/s13014-022-02188-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13014-022-02188-y

Keywords