Skip to main content

Table 5 Subgroup analysis by treatment technique: mean difference in cochlear D0.03cc and cochlear Dmean by distance from fundus

From: Cochlear sparing in LINAC-based radiosurgery for vestibular schwannoma: a dosimetric comparison of dynamic conformal arc, IMRT and VMAT treatment plans

Technique

Cochlear dosimetric parameter

Mean difference per distance (Gy/mm) (95% CI)

Global P value

DCAT

Cochlea D0.03cc

− 0.59 (− 0.95, − 0.22)

0.0035

IMRT

Cochlea D0.03cc

− 1.0 (− 1.44, − 0.57)

0.0002

VMAT

Cochlea D0.03cc

− 0.73 (− 1.04, − 0.43)

0.0001

DCAT

Cochlea Dmean

− 0.54 (− 0.89, − 0.20)

0.0038

IMRT

Cochlea Dmean

− 0.92 (− 1.30, − 0.54)

0.0001

VMAT

Cochlea Dmean

− 0.67 (− 0.95, − 0.38)

0.0001

  1. D0.03cc = maximum dose at 0.03 cc; Dmean = mean dose