Skip to main content

Split-VMAT technique to control the deep inspiration breath hold time for breast cancer radiotherapy

Abstract

Background

To improve split-VMAT technique by optimizing treatment delivery time for deep-inspiration breath hold (DIBH) radiotherapy in left-sided breast cancer patients, when automatic beam-interruption devices are not available.

Methods

Ten consecutive patients were treated with an eight partial arcs (8paVMAT) plan, standard of care in our center. A four partial arcs (4paVMAT) plan was also created and actual LINAC outputs were measured, to evaluate whether there was a dosimetric difference between both techniques and potential impact on the delivered dose. Subsequently, ten other patients were consecutively treated with a 4paVMAT plan to compare the actual treatment delivery time between both techniques. The prescribed dose was 40.05 Gy/15 fractions on the PTV breast (breast or thoracic wall), lymph nodes (LN) and intramammary lymph node chain (IMN). Treatment delivery time, PTVs coverage, conformity index (CI), organs at risk (OAR) dose, monitor units (MU), and gamma index were compared.

Results

Both split-VMAT techniques resulted in similar dose coverage for the PTV Breast and LN, and similar CI. For PTV IMN we observed a 5% increased coverage for the volume receiving ≥ 36 Gy with 4paVMAT, with an identical volume receiving ≥ 32 Gy. There was no difference for the OAR sparing, with the exception of the contralateral organs: there was a 0.6 Gy decrease for contralateral breast mean (p ≤ 0.01) and 1% decrease for the volume of right lung receiving ≥ 5 Gy (p = 0.024). Overall, these results indicate a modest clinical benefit of using 4paVMAT in comparison to 8paVMAT. An increase in the number of MU per arc was observed for the 4paVMAT technique, as expected, while the total number of MU remained comparable for both techniques. All the plans were measured with the Delta4 phantom and passed the gamma index criteria with no significant differences. Finally, the main difference was seen for the treatment delivery time: there was a significant decrease from 8.9 to 5.4 min for the 4paVMAT plans (p < .05).

Conclusions

This study is mainly of interest for centers who are implementing the DIBH technique without automatic beam-holding devices and who therefore may require to manually switch the beam on and off during breast DIBH treatment. Split-VMAT technique with 4 partial arcs significantly reduces the treatment delivery time compared to 8 partial arcs, without compromising the target coverage and the OAR sparing. The technique decreases the number of breath holds per fraction, resulting in a shorter treatment session.

Introduction

In the last 15 years, considerable efforts have been made to minimize cardiac and lung toxicity of postoperative radiotherapy for left-sided breast cancer. The implementation of techniques such as intensity modulated radiotherapy and deep inspiration breath hold (DIBH) allowed for a better sparing of these organs at risk (OARs) [1,2,3,4,5,6]. Literature suggests that the combination of volumetric modulated arc therapy (VMAT) and DIBH can even further decrease the mean heart dose and the ipsilateral lung dose for left-sided breast cancer radiotherapy including regional lymph nodes and IMN [7,8,9].

Different treatment planning solutions to combine DIBH and VMAT are described in the literature, including the use of multiple small partial arcs (split-VMAT) that mimic tangential fields; a full 360° arc; a hybrid plan combining tangential fields and partial arcs; and a single or double partial arc totaling between 190° and 250° [6,7,8,9,10,11,12,13,14]. In the last few years, multiple dosimetric studies have been published comparing these different VMAT treatment designs for left-sided DIBH, showing better results for split-VMAT techniques, regardless whether nodal areas had to be treated as well [13, 15,16,17,18].

The increasing use of surface guided radiation therapy (SGRT) systems has allowed for a reduction localization uncertainty during treatment delivery. When using DIBH techniques, SGRT enables monitoring of the patient movements and respiration, optimizing position reproducibility and minimizing internal target motion, hence increasing the accuracy of the treatment of specific anatomic sites [19,20,21]. This system can be linked to the treatment machine to trigger automatic beam delivery or beam hold when respiration is within predefined respiratory phases [20]. When SGRT or any other monitoring device is not present or linked to the beam hold, RTTs have to manually interrupt the beam by observing the patient’s respiration. This can be uncomfortable for the RTTs and might create insecurity with the technique, mainly during the implementation of the technique. Hence, we decided to apply split-VMAT arcs with planned stops to systematize the treatment for both patients and RTTs, allowing them to minimize unplanned beam interruptions during treatment.

We designed a technique using 8 partial arcs (8paVMAT) with a 20 s maximum treatment time per arc. This solution led to a long treatment time per fraction given 8 separate DIBH were necessary to be able to deliver all the arcs. Based on feedback from patients and RTTs revealing that a majority of patients were able to comfortably hold their breath more than 20 s, we decided to reduce the number of arcs to four arcs with an average delivery time of 30 s per arc (4paVMAT). In case the breath hold had to be interrupted during the treatment, the RTTs could stop the beam manually and start the treatment again where it was interrupted.

In this work, we compare the 4paVMAT technique with the 8paVMAT and its benefits in terms of treatment time.

Methods and materials

Treatment planning techniques

Two different planning techniques are compared in this study: a 8 partial arcs VMAT (8paVMAT) and a 4 partial arcs VMAT (4paVMAT).In an 8paVMAT plan, 30° overlapping arcs mimic tangential fields, where the start/stop angle was between 300°/20° for the medial arcs and between 80°/180° for the lateral ones. In 4paVMAT, the angle of the arcs was increased to 50° keeping the start/stop angle between 300°/20° for the medial arcs and between 90°/180° for the lateral arcs, allowing the arcs to overlap in both techniques (see Fig. 1).

Fig. 1
figure 1

Axial caption of the treatment arcs set up for one of the selected patients. On the left the 8paVMAT is displayed and on the right the 4paVMAT

All plans were generated with Monaco v5.11 treatment planning system (Elekta AB, Stockholm, Sweden), with 6MV photons. The number of control points (CP) per arc was also increased in the 4paVMAT setting. The gantry angle increment was decreased, to increase the modulation. This value correlates with the number of sectors during the first optimization step in which the multileaf collimator moves continuously from one side to the other and then changes its direction.

The collimator angles were 15° and 345° for the different arcs, in order to be parallel to the chest and to the heart. A virtual bolus of 5 mm was used for the planning and removed at the end of the optimization to take into account the breast movement and possible swelling between fractions [22,23,24]. For all patients, 40.05 Gy in 15 fractions was prescribed to the breast or thoracic wall, LN and IMN. All plans were normalized at 95% of the breast PTV to receive 95% of the prescribed dose (38.05 Gy).

Table 1 shows the different parameters used for both templates during the optimization.

Table 1 Parameters defined for 8paVMAT and 4paVMAT plans

Dosimetric and QA comparisons

  1. (a)

    Patient selection

Left sided breast cancer patients necessitating intramammary lymph node (IMN) irradiation or patients with a challenging anatomy for which VMAT was deemed necessary were included. All patients were treated using DIBH with real-time monitoring using SGRT.

Ten breast cancer patients, consecutively treated in our center with VMAT and DIBH using an 8paVMAT plan, were selected for the present study. For every patient, an additional 4paVMAT plan was subsequently created in order to investigate target coverage and dose to the OAR.

  1. (b)

    Plan comparison

The twenty plans were evaluated and compared. PTV’s coverage, OARs dose and number of monitor units (MU) were compared between plans. The D90% coverage for all PTVs and the conformity index (CI) were compared. The CI is defined as CI = TVRI/TV * TVRI/VRI, where TVRI is the target volume covered by the reference isodose (95% of the prescribed dose), TV is the target volume and VRI is the volume encompassed by the reference isodose [25]. For OARs, mean heart doses, V17Gy and V5Gy to the heart, mean contralateral breast doses, V17Gy for ipsilateral lung, mean lungs doses, V5Gy to the contralateral lung and body were compared. The choice of these constraints was based on an internal protocol which regroups constraints found in the literature [26,27,28,29,30]. The total MU, and the minimum and maximum MU per arc were also compared. “DVH Analytics” was used for plan comparison [31].

  1. (iii)

    Plan measurements

Actual LINAC output was measured to evaluate whether there was a dosimetric difference between both techniques and if it would have an impact on the delivered dose.

For Quality Assurance (QA), the gamma index was used, which provides a numerical quality value that serves as a measure of disagreement in the regions that fail the acceptance criteria [32]. All the plans were measured on an Elekta InfinityTM equipped with an AgilityTM head with the Delta4 + phantom using global gamma evaluation with 3 %/3 mm criteria above a 20% of maximum dose threshold for 95% of measured points.

In cases where patients cannot hold their breath long enough, the RTTs interrupt the beam manually and restart the treatment where it was stopped. To ensure correct treatment delivery in case of beam interruption, two QA measurements were acquired for 4paVMAT: one without interruptions and another interrupting every arc.

Treatment delivery time comparison

The ten patients of the dosimetric study were treated with the 8paVMAT plan and treatment times were recorded. After validation of the 4paVMAT technique by comparing DVH parameters and QA in these patients, ten new consecutive patients were included, planned and treated with the 4paVMAT technique. Treatment times were compared with the 8paVMAT treatment times.

Contouring and treatment details

The clinical target volumes (CTVs) were delineated according to ESTRO guidelines and a margin of 5mm was added around the CTV to obtain the planning target volume (PTV) for breast or thoracic wall, lymph nodes (LNs) and IMN. The PTVs were cropped at 3mm from the surface of the skin. SGRT with the IDENTIFY system from Humediq (Varian Medical Systems, Palo Alto, CA, USA) was used to optimize set up and monitor DIBH.

Statistical analysis

The statistical analysis was performed using a Mann-Whitney U test for treatment delivery time, a Wilcoxon signed-rank test for all the other parameters, and Friedmann test for the gamma index criteria evaluation, at a significance level under 0.05.

Results

Patient characteristics are shown in Table 2. Patients underwent breast conserving surgery or mastectomy and the mean age for 8paVMAT and 4paVMAT treated groups is 45.7 and 60.6 years old, respectively.

Table 2 Treated volumes for the patients selected

Table 3 shows the dosimetric results from the two different techniques.

Table 3 Dosimetric parameters results obtained for 8paVMAT and 4paVMAT (data are shown as mean values with one standard deviation, and range between brackets)

Both techniques met the mandatory dose constraints for OAR and target coverage. We can observe similar results for the PTV breast and PTV LN coverage, and conformity index. A significant difference is observed for the PTV IMN 36 Gy coverage: there is a 5.4% coverage increase with 4paVMAT plans. For the OARs, both techniques showed similar results for the main constraints: there was no difference in mean heart dose, V5Gy and V17Gy to the heart, mean lung doses, V17Gy to the ipsilateral lung, V30Gy to the humeral head and V5Gy to the body. Regarding the doses to the contralateral organs, there is an average decrease of 0.6 Gy (p < 0.01) for the mean contralateral breast dose and 1.2% (p = 0.024) for the volume of contralateral lung receiving > 5 Gy, with the 4paVMAT plans. Figure 2 shows the mean dose-volume histograms (DVH) for contralateral lung and contralateral breast. There is no difference in the total number of MU between both treatments, but we see a significant, yet logical increase of the minimum and maximum MU per arc with 4paVMAT—see Fig. 3 (p < 0.05).

Fig. 2
figure 2

The mean DVH for right lung (left figure) and contralateral breast (right figure) using 8paVMAT and 4paVMAT. The colored shadows show the interquartile ranges (IQRs) from the mean values

Fig. 3
figure 3

Boxplots of MU for 8paVMAT and 4paVMAT (total MU of all arcs, maximum MU per arc, minimum MU per arc)

The main difference was seen in treatment delivery time with 8.9 min and 5.4 min for 8paVMAT and 4paVMAT, respectively (p < 0.01)—Fig. 4. Regarding the QA measurements, all the plans passed the gamma index criteria, whether the arcs were interrupted or not. The results can be seen in Fig. 5.

Fig. 4
figure 4

Mean treatment delivery time in minutes for 8paVMAT and 4paVMAT

Fig. 5
figure 5

Measurement results with the gamma index for 8paVMAT and 4paVMAT with and without treatment interruption

Discussion

Previous dosimetric studies have shown exciting achievements regarding the constraint goals for OAR with split-VMAT techniques [13, 15,16,17, 33]. However, very few studies focused on the split-VMAT technique itself [13].

In our institute, the combination of VMAT and DIBH for left-sided breast treatment was introduced using split-VMAT with 8 partial arcs. This design was chosen to systematize the patient’s treatment and to decrease the number of unplanned beam interruptions. We aimed to decrease the treatment time per fraction and decided to compare DIBH 8paVMAT with 4paVMAT.

The combination of the optimization parameters of the 4paVMAT resulted in an adequate plan with a decreased treatment delivery time per fraction. Our analysis reveals that 8paVMAT and 4paVMAT demonstrate equivalent coverage of the PTV Breast and LN, and the CI. Although there is no difference for the PTV IMN regarding the volume receiving 32 Gy, there is a significant 5% increase of the volume receiving 36 Gy, when using 4paVMAT.

Our results are consistent with other dosimetric studies regarding target coverage and OARs sparing [7, 13, 34, 35]. Table 4 compares published data with our results.

Table 4 Comparison of average dose parameters for VMAT with DIBH between different dosimetric studies

The main differences may be due to differences in prescription, structure margins or OAR priority.

For the OARs, there was no significant difference between 4 and 8paVMAT with respect to the mean heart doses, heart V5Gy, heart V17Gy, ipsilateral lung V17Gy, mean lung doses, V30Gy of the humeral head PRV and V5Gy to the body.

In our study, there was an improvement of 0.6 Gy in the mean contralateral breast doses (p < 0.01) for the 4paVMAT plan. There was also a significant 1% decrease for the volume of contralateral lung receiving 5 Gy. One possible explanation for this is the increased scattered dose with 8paVMAT because of the number of arcs [36]. Yet, further studies should be performed to confirm these results. These are the small dosimetric benefits of the 4paVMAT template.

Many concerns have been raised regarding the low dose bath to peripheral organs and the increased risk for radiation induced malignancy or the still unknown effects of such doses. Our results show the volume of contralateral lung receiving 5 Gy of 3.6% and 2.4% for 8paVMAT and 4paVMAT, respectively. These results show that the low dose volumes can be comparable between VMAT and 3D CRT techniques, the latter being well known for a very low to almost no dose to the contralateral lung.

The observed increase in treatment time per arc resulted, as expected, from an increase of the minimum and maximum number of MU per arc with 4paVMAT, however no difference in the total number of MUs between both techniques was found.

During our QA checks, beam interruption was simulated for 4paVMAT technique for all arcs to ensure that the treatment maintained a good deliverability (4paVMAT int). Figure 5 shows that all the checks, even those with a beam interruption, passed the global gamma index evaluation, using 3%/3 mm criteria above a 20% of maximum dose threshold for more than 95% of the measured points. This proves the linear accelerator to reliably deliver the dose, even in case of beam interruptions.

The main goal of this study was to decrease the treatment delivery time per fraction. To the best of our knowledge, no study has been published to compare the treatment delivery time for different VMAT techniques combined with DIBH for left-sided breast radiotherapy.

The mean treatment delivery time was significantly decreased by around 40% with 4paVMAT, enabling us to spare on average 3.5 min beam-on time per fraction (Fig. 4). This also reduced the number of unplanned beam interruptions for both RTTs and patients, and proved to be in total a faster and more convenient delivery solution. In a time slot of 20 min for DIBH and 15 min for normal treatments, this means a gain of almost 20% which makes it possible to treat up to 5 more patients on an 8 h treatment day.

A limitation of the present study is the fact that the delivery time was measured in different patients for 4paVMAT and 8paVMAT plans. However, since the differences in treated volumes were small (Table 2), we expect that this would have little impact on the average treatment time.

Finally, this new delivery technique was successfully adopted in the department. The 4paVMAT did not lead to any issue regarding the breath holds by any of the patients. Whenever the patients were not able to hold their breath long enough, the RTTs interrupted the beam manually and started the treatment again where it had been stopped.

Conclusion

This study is mainly of interest for centers who are implementing the DIBH technique without automatic beam-interruption devices, and who therefore require to manually switch the beam on and off during breast DIBH treatment. We provide a solution using a pre-planned number of beam interruptions during the treatment.

The 4paVMAT technique provides a faster radiotherapy delivery option than 8paVMAT for DIBH treatment of breast cancer including regional LN and IMN, without clinically important dosimetric differences for targets coverage and OAR sparing. With potentially half the number of breath holds per fraction, this technique enabled us to decrease the treatment time by about 3.5 minutes per fraction. QA measurements showed 4paVMAT to be correctly delivered, even in case of beam interruption; hence, it was adopted in our department as the new standard for VMAT treatment of left-sided breast cancer with DIBH.

Availability of data and materials

Research data are stored in an institutional repository and could be shared upon request to the corresponding author.

Abbreviations

8paVMAT:

A Eight partial arcs

4paVMAT:

A four partial arcs

LN:

Lymph nodes

IMN:

Intramammary lymph node chain

CI:

Conformity index

MU:

Monitor units

OAR:

Organs at risk

DIBH:

Deep inspiration breath hold

VMAT:

Volumetric modulated arc therapy

CP:

Number of control points

SGRT:

Surface guided radiation therapy

QA:

Quality Assurance

CTVs:

Clinical target volumes

PTV:

Planning target volume

TVRI :

Target volume covered by the reference isodose (95% of the prescribed dose)

TV:

Target volume

VRI :

Volume encompassed by the reference isodose

DVH:

Dose-volume histograms

IQRs:

Interquartile ranges

References

  1. Veldeman L, Madani I, Hulstaert F, De Meerleer G, Mareel M, De Neve W. Evidence behind use of intensity-modulated radiotherapy: a systematic review of comparative clinical studies. Lancet Oncol. 2008;9(4):367–75.

    Article  Google Scholar 

  2. De Neve W, De Gersem W, Madani I. Rational Use of Intensity- Modulated Radiation Therapy: The Importance of Clinical Outcome. Seminars in Radia- tion Oncology. 2012;22(1):40–9.

    Article  Google Scholar 

  3. Hayden AJ, Rains M, Tiver K. Deep inspiration breath hold technique reduces heart dose from radiotherapy for left-sided breast cancer. J Med Imaging Radiat Oncol. 2012;56(4):464–72.

    Article  Google Scholar 

  4. Smyth LM, Knight KA, Aarons YK, Wasiak J. The cardiac dose-sparing benefits of deep inspiration breath-hold in left breast irradiation: a systematic review. J Med Radiat Sci. 2015;62(1):66–73.

    Article  Google Scholar 

  5. Simonetto C, Eidemüller M, Gaasch A, Pazos M, Schönecker S, Reitz D, Kääb S, Braun M, Harbeck N, Niyazi M, Belka C, Corradini S. Does deep inspiration breath-hold prolong life? Individual risk estimates of ischaemic heart disease after breast cancer radiotherapy. Radiother Oncol. 2019;131:202–7.

    Article  Google Scholar 

  6. Russo S, Esposito M, Hernandez V, Saez J, Rossi F, Paoletti L, Pini S, Bastiani P, Reggiori G, Nicolini G, Vanetti E, Tomatis S, Scorsetti M, Mancosu P. Does deep inspiration breath hold reduce plan complexity? Multicentric experience of left breast cancer radiotherapy with volumetric modulated arc therapy. Phys Med. 2019;59:79–85.

    Article  Google Scholar 

  7. Osman SOS, Hol S, Poortmans PM, Essers M. Volumetric modulated arc therapy and breath-hold in image-guided locoregional left-sided breast irradiation. Radiother Oncol. 2014;112(1):17–22.

    Article  Google Scholar 

  8. Swamy ST, Radha CA, Kathirvel M, Arun G, Subramanian S. Feasibility study of deep inspiration breath-hold based volumetric modulated arc therapy for locally advanced left sided breast cancer patients. Asian Pac J Cancer Prev. 2014;15(20):9033–8.

    Article  Google Scholar 

  9. Jensen CA, Roa AMA, Johansen M, Lund JÅ, Frengen J. RobustnessofVMATand3DCRTplanstowardsetuperrorsinradiationtherapyoflocallyadvancedleft-sidedbreastcancerwithDIBH. Phys Med. 2018;45:198–204.

    Article  Google Scholar 

  10. Pham TT, Ward R, Latty D, Owen C, Gebski V, Chojnowski J, Kelly C, Ahern V, Tiver K, Stuart K, Wang W. Left-sided breast cancer loco-regional radiotherapy with deep inspiration breath-hold: does volumetric-modulated arc radiotherapy reduce heart dose further compared with tangential intensity-modulated radiotherapy? J Med Imaging Radiat Oncol. 2016;60(4):545–53.

    Article  Google Scholar 

  11. Sakka M, Kunzelmann L, Metzger M, Grabenbauer GG. Cardiac dose-sparing effects of deep-inspiration breath-hold in left breast irradiation: is IMRT more beneficial than VMAT? Strahlenther Onkol. 2017;193(10):800–11.

    Article  Google Scholar 

  12. Dubouloz A, Nouet P, Koutsouvelis N, Dipasquale G, Jaccard M, Fargier-Bochaton O, Rouzaud M. 20 treating breast cancer with VMAT in deep inspiration breath hold: the Geneva experience. Phys Med. 2018;56:12–3.

    Article  Google Scholar 

  13. Boman E, Rossi M, Haltamo M, Skyttä T, Kapanen M. Anewsplit arcVMATtechniqueforlymphnodepositivebreastcancer. Phys Med. 2016;32(11):1428–36.

    Article  Google Scholar 

  14. Akira SAKUMI, Kenshiro SHIRAISHI, Tsuyoshi ONOE, Kentaro YAMAMOTO, Akihiro HAGA, Kiyoshi YODA, Kuni OHTOMO, and Keiichi NAKAGAWA. Single-Arc Volumet- ric Modulated Arc Therapy Planning for Left Breast Cancer and Regional Nodes. Journal of Radiation Research, 53(1):151–153, 2012.

  15. Jeulink M, Dahele M, Meijnen P, Slotman BJ, Verbakel WFAR. Is there a preferred IMRT technique for left-breast irradiation? J Appl Clin Med Phys. 2015;16(3):197–205.

    Article  Google Scholar 

  16. Virén T, Heikkilä J, Myllyoja K, Koskela K, Lahtinen T, Seppälä J. Tangentialvolumetricmodulatedarctherapytechniqueforleft-sidedbreast cancer radiotherapy. Radiat Oncol. 2015;10(1):1–8.

    Article  Google Scholar 

  17. Pasler M, Lutterbach J, Björnsgard M, Reichmann U, Bartelt S, Georg D. VMAT techniques for lymph node-positive left sided breast cancer. Z Med Phys. 2015;25(2):104–11.

    Article  Google Scholar 

  18. Fogliata A, Seppälä J, Reggiori G, Lobefalo F, Palumbo V, De Rose F, Franceschini D, Scorsetti M, Cozzi L. Dosimetric trade-offs in breast treatment with VMAT technique. Br J Radiol. 2017;90(1070):20160701.

    Article  Google Scholar 

  19. Worm E, Offersen B, Poulsen PR, Askholm E, Yates ES, Hansen R. [OA013] Treatment accuracy in surface guided deep-inspiration breath-hold radiotherapy for left-sided breast cancer. Phys Med. 2018;52(2018):5–6.

    Article  Google Scholar 

  20. Hoisak JDP, Pawlicki T. The role of optical surface imaging systems in radiation therapy. Semin Radiat Oncol. 2018;28(3):185–93.

    Article  Google Scholar 

  21. Freislederer P, Kugele M, Ollers M, Swinnen A, Sauer TO, Bert C, Giantsoudi D, Corradini S, Batista V. Recent advanced in surface guided radiation therapy. Radiat Oncol. 2020;15(1):1–11.

    Article  Google Scholar 

  22. Tyran M, Tallet A, Resbeut M, Ferre M, Favrel V, Fau P, Moureau-Zabotto L, Darreon J, Gonzague L, Benkemouche A, Varela-Cagetti L, Salem N, Farnault B, Acquaviva MA, Mailleux H. Safetyandbenefitofusingavirtualbolusduringtreatmentplanning for breast cancer treated with arc therapy. J Appl Clin Med Phys. 2018;19(5):463–72.

    Article  Google Scholar 

  23. Rossi M, Boman E, Kapanen M. Optimal selection of optimization bolus thickness in planning of vmat breast radiotherapy treatments. Med Dosim. 2019;44(3):266–73.

    Article  Google Scholar 

  24. Lizondo M, Latorre-Musoll A, Ribas M, Carrasco P, Espinosa N, Coral A, Jornet N. Pseudo skin flash on VMAT in breast radiotherapy: optimization of virtual bolus thickness and HU values. Phys Med. 2019;63:56–62.

    Article  Google Scholar 

  25. Feuvret L, Noël G, Mazeron JJ, Bey P. Conformityindex: a review. Int J Radiat Oncol Biol Phys. 2006;64(2):333–42.

    Article  Google Scholar 

  26. Patel AK, Ling DC, Richman AH, Champ CE, Huq MS, Heron DE, Beriwal S. Hypofractionated whole-breast irradiation in large-breasted women—is there a dosimetric predictor for acute skin toxicities? Int J Radiat Oncol Biol Phys. 2019;103(1):71–7.

    Article  Google Scholar 

  27. Chen G-P, Liu F, White J, Vicini FA, Freedman GM, Arthur DW, Li XA. A planning comparison of 7 irradiation options allowed in rtog 1005 for early-stage breast cancer. Med Dosim. 2015;40(1):21–5.

    Article  CAS  Google Scholar 

  28. Offersen BV, Boersma LJ, Kirkove C, Hol S, Aznar MC, Sola AB, Kirova YM, Pignol JP, Remouchamps V, Verhoeven K, Weltens C, Arenas M, Gabrys D, Kopek N, Krause M, Lundstedt D, Marinko T, Montero A, Yarnold J, Poortmans P. ESTROconsensusguidelineontargetvolumedelineationforelective radiation therapy of early stage breast cancer. Radiother Oncol. 2015;114(1):3–10.

    Article  Google Scholar 

  29. Offersen BV, Boersma LJ, Kirkove C, Hol S, Aznar MC, Sola AB, Kirova YM, Pignol JP, Remouchamps V, Verhoeven K, Weltens C, Arenas M, Gabrys D, Kopek N, Krause M, Lundstedt D, Marinko T, Montero A, Yarnold J, Poortmans P. ESTRO consensus guideline on target volume delineation for elective radiation therapy of early stage breast cancer, version 1.1. Radiother Oncol. 2016;118(1):205–8.

    Article  Google Scholar 

  30. Taylor C, Duane FK, Dodwell D, Gray R, Wang Z, Wang Y, Peto R, McGale P, Correa C, Aznar MC, Ewertz M, Anderson SJ, Bergh J, Jagsi R, Pierce L, Pritchard KI, Whelan T, Swain S. Estimating the risks of breast cancer radiotherapy: evidence from modern radiation doses to the lungs and heart and from previous randomized trials. J Clin Oncol. 2017;35(15):1641–9.

    Article  Google Scholar 

  31. Cutright D, Gopalakrishnan M, Roy A, Panchal A, Mittal BB. DVH analytics: a DVH database for clinicians and researchers. J Appl Clin Med Phys. 2018;19(5):413–27.

    Article  Google Scholar 

  32. Depuydt T, Van Esch A, Huyskens DP. A quantitative evaluation of IMRT dose distributions: refinement and clinical assessment of the gamma evaluation. Radiother Oncol. 2002;62(3):309–19.

    Article  Google Scholar 

  33. Tsai PF, Lin SM, Lee SH, Yeh CY, Huang YT, Lee CC, Hong JH. Thefeasibilitystudyofusingmultiplepartialvolumetric-modulated arcs therapy in early stage left-sided breast cancer patients. J Appl Clin Med Phys. 2012;13(5):62–73.

    Article  Google Scholar 

  34. Ranger A, Dunlop A, Hutchinson K, Convery H, Maclennan MK, Chantler H, Twyman N, Rose C, McQuaid D, Amos RA, Griffin C, DeSouza NM, Donovan E, Harris E, Coles CE, Kirby A. A dosimetric comparison of breast radiotherapy techniques to treat locoregional lymph nodes including the internal mammary chain. Clin Oncol. 2018;30(6):346–53.

    Article  CAS  Google Scholar 

  35. Rossi M, Boman E, Kapanen M. Contralateral tissue sparing in lymph node-positive breast cancer radiotherapy with VMAT technique. Med Dosim. 2019;44(2):117–21.

    Article  Google Scholar 

  36. Poeta S, Stock M, Jourani Y, De Brouwer TB, Vandekerkhove C, Simon S. EP-1377: IMRT and VMAT peripheral dose measurements. Radiother Oncol. 2015;115:S743.

    Article  Google Scholar 

Download references

Acknowledgements

The authors thank the following persons for their input: Dr. Florian Charlier from the Bordet Radiation Oncology department, and the physicists and dosimetrists from the Medical Physics department of the Institut Jules Bordet.

Funding

None.

Author information

Authors and Affiliations

Authors

Contributions

SP: corresponding author, designed the study, performed analysis of the data and wrote the manuscript. YJ, ADC, RVDB, assisted in designing the study and in analyzing the data and provided intensive feedback on the manuscript, for multiple versions. DVG and NR provided intensive feedback on the manuscript, for multiple versions. All authors have read and approved the final manuscript.

Corresponding author

Correspondence to Sara Poeta.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

Dr. Poeta has nothing to disclose. Dr. Jourani has nothing to disclose. Dr. De Caluwé reports grants from AstraZeneca, outside the submitted work. Dr. Van den Begin has nothing to disclose. Dr. Van Gestel reports being a member of the Advisory board or receiving honoraria from Sanofi, Accuray, Takeda, Merck-Pfizer and Novartis. Dr. Reynaert has nothing to disclose.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Poeta, S., Jourani, Y., De Caluwé, A. et al. Split-VMAT technique to control the deep inspiration breath hold time for breast cancer radiotherapy. Radiat Oncol 16, 77 (2021). https://doi.org/10.1186/s13014-021-01800-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13014-021-01800-x

Keywords