Skip to main content

Dosimetric comparison between RapidArc and HyperArc techniques in salvage stereotactic body radiation therapy for recurrent nasopharyngeal carcinoma

Abstract

Background

To evaluate dosimetric differences of salvage irradiations using two commercially available volumetric modulated arc therapy (VMAT) stereotactic body radiation therapy (SBRT) techniques: RapidArc (RA) and HyperArc (HA), for recurrent nasopharyngeal carcinoma (NPC) after initial radiation therapy.

Methods

Ten patients with recurrent NPC status previously treated with radiation therapy were considered suitable candidates for salvage SBRT using VMAT approach. Two separate treatment plans were created with HA and RA techniques for each case, with dosimetric outcomes compared with respect to tumor target coverage and organs-at-risk (OARs) sparing. Furthermore, the cumulative radiobiological effects to the relevant OARs from the original radiotherapy to the respective salvage SBRT plans were analyzed in terms of biologically effective dose (BED).

Results

Treatment with HA exhibited similar target dose coverage as with RA, while delivering a higher mean dose to the targets. Using RA technique, the mean maximal doses to optic apparatus and the mean brain dose were reduced by 1 to 1.5 Gy, comparing to HA technique. The conformity index, gradient radius, and intermediate dose spillage in HA plans were significantly better than those in RA. With HA technique, the volume of brain receiving 12 Gy or more was reduced by 44%, comparing to RA technique. The cumulative BEDs to spinal cord and optic apparatus with RA technique were 1 to 2 Gy3 less than those with HA. HA technique significantly reduced the volume within body that received more than 100 Gy.

Conclusions

With better dose distribution than RA while maintaining sufficient target dose coverage, HA represents an attractive salvage SBRT technique for recurrent NPC.

Background

Radiation therapy with or without concurrent chemotherapy represents the standard treatment for nasopharyngeal cancer (NPC) and leads to a 5-year local control rate of greater than 85% [1,2,3,4,5,6,7,8,9]. However, local recurrence still represents a major source of morbidity and mortality in patients with advanced stage NPC [10, 11]. Because of the invasive nature of the malignancy which may involve many critical tissues at skull base, radiation therapy often remains to be the main salvage treatment modality once the disease recurs locally.

The main feature of stereotactic body radiation therapy (SBRT) is the delivery of relatively few fractions of ultra-high dose coverage tightly conformed to the intended target volumes, while minimizing dosages to adjacent critical organs at risk (OARs). Many studies have documented that SBRT may improve tumor control, reduce toxicity, and improve quality of life in patients with recurrent head and neck cancer, including NPC [12,13,14,15,16].

RapidArc (RA, Varian Medical System, Palo Alto, CA, USA) is an isocentric-coplanar volumetric-modulated arc radiotherapy (VMAT) technique that can deliver highly conformal, intensity-modulated radiation therapy (IMRT) doses via a single arc or multiple rotations of the gantry of a linear accelerator [17]. RA enables treatment plans with an improved dosimetric outcome as compared to multifield IMRT while reducing the treatment time per fraction in the SBRT setting [18].

HyperArc (HA, Varian Medical System) is a relatively new isocentric VMAT technique developed specifically for non-coplanar, multileaf collimator (MLC)-based stereotactic radiotherapy with automated treatment optimization and dose delivery [19, 20]. HA has been demonstrated as a novel stereotactic radiosurgery (SRS) technique for single or multiple brain metastasis [21,22,23]. To our best knowledge, however, few studies have explored the ability of HA to generate high-quality treatment plans for extracranial lesions – especially as pertains to SBRT which is gaining popularity as a salvage measure for recurrent head & neck malignancies. In such case, since SBRT for re-irradiation often utilizes few fractions of relatively high dosage while the initial radiotherapy mainly follows conventional fractionation scheme, the combined biologic or clinical effect at any anatomic site of interest (i.e. critical OARs) cannot be inferred from the simple summation of physical dosages received in the sequentially separate treatment courses. Rather, corrections using the concept of biologically effective dose (BED) or equivalent dose to 2-Gy per fraction (EQD2) may be used in order to analyze the ultimate dosimetric consequences. This would be of crucial importance when offering curative SBRT to re-irradiate recurrent NPC where the anatomic region is filled with critical normal structures. Upon inverse planning for SBRT, the “dose” constraints in the conventional dose-volume histogram (DVH) should likewise be converted to BED in order to facilitate meaningful comparisons for dosimetric outcome evaluations.

The aim of the current study was to evaluate the dosimetric differences between the RA and HA SBRT techniques for the salvage treatment of recurrent NPC after initial primary radiation therapy, factoring radiobiologically-corrected cumulative doses to critical OARs as part of DVH objective parameters upon IMRT inverse planning.

Methods

Study groups

Ten patients with recurrent NPC who had been treated with initial primary radiation therapy and subsequent salvage SBRT were enrolled. Their clinical and dosimetric characteristics are shown in Table 1. The RA and HA treatment plans were created retrospectively for each patient to meet a previously set and radiobiologically sound salvage SBRT planning criteria (Table 2).

Table 1 Characteristics of the recurrent nasopharyngeal cancer patients
Table 2 Planning objectives for the target and organs at risk

Initial primary radiation therapy and the salvage SBRT treatment plan criteria

The initial primary radiation therapy plan was generated as previously described [11]. In short, the planning target volume (PTV) was extended 0 to 3 mm from the clinical target volume (CTV). The dose prescription was 70 Gy in 35 fractions for the gross tumor and enlarged lymph nodes, 63 Gy in 35 fractions for the bilateral upper neck, and 56 Gy in 35 fractions for the low-risk region. Simultaneously integrated boost (SIB) IMRT was given to all NPC patients. The treatment plans were generated using the Eclipse treatment planning system (ver. 8.60, Varian Medical Systems, Palo Alto, CA, USA).

For salvage SBRT, the CTV was defined as the locally recurrent nasopharyngeal tumor. The planning target volumes (PTV) were extended 0–2 mm from the corresponding CTVs. The prescription dose and constraints for OARs were based on the initial radiation treatment, radiobiologically adjusted tolerance dose and physician’s ultimate decision. The median prescription dose of the 10 enrolled patients was 36.75 Gy in 5 fractions. A minimum of 95% of the prescription dose was assumed to cover 95% of the PTV. The priority of the treatment planning was sparing of OAR following by target coverage. The details of the planning objectives for the target and OARs are listed in Table 2.

RapidArc and HyperArc treatment plans

Computed tomography data sets and target volume/normal organ contours from the 10 enrolled patients were transferred to the Eclipse treatment planning system (ver. 15.5, Varian Medical Systems). The virtual Encompass (QFix, Avondale, PA, USA) mask was added only for the HA plans. The corresponding HA and RA plans were then generated according to patient-specific target dose prescription and OAR constraints. 6 MV flattening filter-free photon beams were used, with 1400 MU/min dose rate from a Varian TrueBeam (Varian Medical Systems) linear accelerator equipped with 120-leaf high-definition MLC (with a dynamic beam aperture and a spatial resolution of 2.5 mm leaf width × 32 pairs at the center, 5 mm width × 28 pairs in the peripheral leaves, and maximum static field size 40 cm × 22 cm). In the HA plan, the isocenter was positioned automatically at the center of the selected target structures. The collimator angle and field size were designed to minimize OAR dosages. In addition, the arc fields were also automatically arranged: one full or half coplanar arc with a couch rotation of 0° and up to three partial noncoplanar arcs with couch rotations of 315°, 45°, and 90° (or 270°) [21, 22], respectively (Fig. 1).

Fig. 1
figure 1

Beam arrangements and arc trajectories of HyperArc (a) and RapidArc (b). a One partial coplanar arc, 0 degrees; three noncoplanar arcs, 45, 90, and 315 degrees; b two coplanar arcs

In the RA plan, the same isocenter (at the center of the selected target) was set as the HA plan. These two-arc technique (counterclockwise rotation from 179° to 181° and clockwise rotation backwards) was applied for all RA treatment plans (Fig. 1). For all HA and RA plans, the optimization and dose calculation were done per the Photon Optimizer and Anisotropic Analytical Algorithm (ver. 15.5.1, Varian Medical Systems). Jaw tracking optimization was applied to the HA and RA plans. The normal tissue objective optimizers were SRS NTO (ver. 15.5.1, Varian Medical Systems) for HA and automatic NTO (ver. 15.5.1, Varian Medical Systems) for RA.

Cumulative dose to organs at risk

The dosimetric data of the two treatment courses, including the initial radiation therapy and the salvage SBRT (RA and HA treatment plans, including CT images, structure sets, and radiation doses), were exported from Eclipse to Velocity (ver. 3.2.1, Varian Medical Systems) for each enrolled patient, except one patient whose initial treatment plan had been generated by the Pinnacle treatment planning system (Philips Radiation Oncology Systems, Fitchburg, WI, USA). The cumulative doses to the OARs from the two treatment courses were calculated by the deformable multipass registration method using Velocity [24]. The BED of the OARs were also calculated by the following formula [25, 26]: BED = n × d (1 + d/α/β), where n stands for number of treatment fractions, d is dose per fraction in Gy, and α/β ratio is assumed to be 3 Gy for all OARs. Thus, the relation between BED and EQD2 for the OARs in this study is given as: BED = EQD2 (1 + 2/3) = 1.66 × EQD2, with unit of Gy3.

Plan evaluation statistics

Plan evaluation criteria

The parameters used to evaluate the quality of the planned dose distributions for both the HA and RA plans were target coverage, sparing of OARs, and dosimetric parameters mainly recommended by the report from the AAPM Task Group 101 [27].

Dosimetric parameters and treatment efficiency

The treatment plans were evaluated by comparing the dosimetric parameters derived from the DVHs for target coverage and sparing of OARs. D2 stands for the dose to 2% of the CTV or PTV, and D98 stands for the dose to 98% of the CTV or PTV, each describing the maximum and minimum dose for the target volumes, respectively. The conformity index (CI), as previously described [28, 29], was defined as follows:

(prescription isodose volume × target volume)/ (volume of the target covered by the prescription isodose volume) 2.

The homogeneity index (HI) was determined as the ratio of the highest dose received by 5% of the PTV to the lowest dose received by 95% of the PTV [27].

The intermediate dose spillage was determined as the ratio of the volume of 50% of the prescription isodose curve to the PTV. High dose spillage was calculated as the ratio of the volume outside the PTV that received > 105% of the prescription dose to the PTV volume (V [V105% - PTV] / [PTV]).

Additionally, we calculated the gradient radius as the difference between the equivalent sphere radii of the volume of 50% of the prescription isodose curve and the prescription isodose volume [30]. Monitor units (MUs) and the delivery time were used to assess treatment efficiency.

Statistical analysis

The dosimetric endpoints of the target volumes, OARs, CIs, HI, intermediate and high dose spillage, the gradient radius, MUs, and the physical dose and BED to OARs were analyzed using the Wilcoxon signed rank test (SPSS, ver. 19, IBM, NY, USA). All tests were 2-tailed, and a p-value < 0 .05 was considered statistically significant.

Results

Target volume coverage

A detailed comparison of the dosimetric parameters of HA and RA plans for the recurrent NPC patients treated with salvage SBRT is shown in Table 3. HA plans had higher CTV and PTV coverages than RA plans, but both differences were not significant (CTV: 97.71% vs. 95.72%, p = 0.214; PTV: 94.12% vs. 92.05%, p = 0.333). HA plans did exhibit higher mean dose to targets than RA plans that achieved statistical significance (CTV: 38.09 Gy vs. 37.53 Gy, p = 0.007; PTV: 37.75 Gy vs. 37.02 Gy, p = 0.005). The isodose curves in the respective HA and RA plans for a patient case are presented in Fig. 2. As one can see, the 50% isodose line conformed better around the target region in the HA plan compared to that in the RA plan, while the 20% isodose line in the RA plan was mostly spread out in the lateral and anterior directions.

Table 3 Comparison of dosimetric parameters between salvage HyperArc and RapidArc for recurrent nasopharyngeal cancer patients
Fig. 2
figure 2

Isodose curves for the applied HyperArc and RapidArc plans. Clinical target volume: orange; planning target volume: pink. a-c The HyperArc plan. d-f The RapidArc plan. Color wash dose level, from 20% of the prescription dose to 100% of the prescription dose

Sparing of organs at risk

Table 3 also includes dose-volume parameters for sparing of OARs in salvage HA vs. RA treatments. All requirements for critical organ-dose constraints for re-irradiation alone were satisfied. The maximum and mean doses to brainstem and spinal cord as generated in HA plans were not significantly different from those generated in RA plans. Significant differences were observed between HA and RA plans in the maximal doses to the optic apparatus (specifically optic nerve_left, p = 0.005; optic chiasm, p = 0.013) and in the mean brain dose (p = 0.009). Using RA technique, the mean maximal doses to the optic nerves and optic chiasm and mean brain dose were reduced by 1 to 1.5 Gy. With HA technique, the volume of brain receiving 12 Gy and more (V12) was decreased by 44% (HA vs. RA: 3.79 c.c. vs. 8.54 c.c., p = 0.022), while the mean maximal dose to the brain was significantly less than the RA technique (HA vs. RA: 24.40 Gy vs. 26.46 Gy, p = 0.017). Figure 3 shows the difference in mean dose-volume histograms between HA and RA plans for OAR sparing.

Fig. 3
figure 3

Comparison of the mean dose-volume histograms for the HyperArc (HA) and RapidArc (RA) techniques. a-d Organs at risk (OAR); X axis, relative dose, the percentage of the prescription dose; Y axis, ratio of OAR volumes, the volume percentage of the OAR

Comparison of dosimetric parameters

Conformity and homogeneity indices

Figure 4 shows the box plot of the dosimetric parameters with regard to CI, gradient radius, high dose spillage, and intermediate dose spillage for both salvage treatment techniques. The mean CI and HI are shown in Table 3. The HA plans showed a higher degree of conformity (CI: HA, 1.22 vs. RA, 1.42, p = 0.007) while achieving slightly more homogenous dose distribution than RA plans without statistical significance (HI: HA, 1.11 vs. RA, 1.17, p = 0.241).

Fig. 4
figure 4

Boxplots of the dosimetric parameters for the HyperArc and RapidArc plans. a conformity index; b gradient radius; c high dose spillage; d intermediate dose spillage. Boxes, median value and upper and lower quartiles; Whiskers, maximum and minimum values within 1.5 interquartile range; Dots, outliers

Dose gradient

Intermediate dose spillage was determined by the ratio of the volume of 50% of the prescription isodose curve to the PTV. For such measure of dose gradient, HA showed significantly faster dose fall-offs than RA technique did (HA vs. RA: 3.79 vs. 6.05, p = 0.005). Similar high dose spillages were observed in HA than in the RA plans (HA vs. RA: 0.03 vs. 0.08, p = 0.445). The mean gradient radius in HA plans (0.83 cm) was significantly shorter than that in RA plans (1.21 cm).

Cumulative biological doses to organs at risk

For biologically meaningful comparisons, Table 4 provides the individual BED values to the OARs for the initial primary treatment and either salvage SBRT, as well as the cumulative BED for the entire sequential treatment course with either salvage technique (P + HA vs. P + RA). For salvage re-irradiation alone, there was significantly less BED to the optic apparatus in RA plans (BED to optic nerve_left: HA > RA, p = 0.005; BED to chiasm: HA > RA, p = 0.013), but a significant reduction of the brain BED in HA plans (BED to brain, HA < RA, p = 0.028). For the combined treatment, the cumulative BEDs to the optic apparatus (optic chiasm, optic nerves, and left eye) in the P + RA plans were indeed significantly less than those in the P + HA plans, while the cumulative BED to the brain in the P + HA plans was significantly less than that in the P + RA plans (p = 0.038).

Table 4 Biologically effective dose to the OARs from initial primary treatment, and salvage HyperArc or RapidArc SBRT for recurrent nasopharyngeal cancer patients

Treatment efficiency

Across all fractions of the salvage SBRT treatment, HA plans generated an average of 21,390 MUs, as compared to 13,467 MUs generated by RA plans (p = 0.0037). The median estimated delivery time from dry runs for RA plans was 3:26 min per fraction (range: 2:31 min. to 5:52 min.) and that for HA plans was 5:17 min per fraction (range: 4:59 min. to 6:40 min.). The median difference in the time of delivery was less than 2 min (111 s), without considering patient alignment and imaging acquisition.

Discussion

There are only a few reports that have mentioned the use of modern VMAT technique for recurrent nasopharyngeal cancer in the salvage SBRT setting [31]. Here, we report the dosimetric results of 10 recurrent NPC patients by creating salvage SBRT treatment plans and demonstrate the feasibility of such salvage re-irradiation approaches by comparing the noncoplanar VMAT (HA) with the coplanar VMAT (RA) technique. Our study was prompted by the lack of data concerning the treatment of recurrent NPC using SBRT with the goal of minimizing the cumulative radiation dose to pertinent OARs.

In general, RA is able to produce a high-quality treatment plan and achieve fast delivery of SRS or SBRT [18, 32]. The RA plans in this study were generated with Eclipse treatment plan system (Varian Medical System, Palo Alto, CA) for a commercial LINAC system (TrueBeam, Varian Medical System, Palo Alto, CA) equipped with high-definition MLCs to deliver radiotherapy beams in a stereotactic, isocentric coplanar fashion. Such VMAT technique has been reported to result in a tightly conformed homogeneous dose distribution over intended target volumes while reducing low-dose areas in the periphery [32, 33]. Here, we demonstrated that RA plans consistently reduced the radiation dose to multiple OARs simultaneously with good coverage of the targets.

There are also VMAT techniques other than the RA approach. The HA technique allows for a partial or full coplanar arc and up to 3 noncoplanar arcs in an isocentric fashion. In the current study, we demonstrated that the most distinguishing advantage of HA over RA is being able to consistently achieve better CI, gradient radius and intermediate dose spillage. As a result of the improved CI and dose fall-off with the HA technique in comparison with RA, a higher mean target dose was consistently achieved in HA plans without compromising the CTV or PTV coverage. Therefore, HA can produce a very high-quality treatment plan with excellent dosimetric parameters.

One of the concerns for salvage re-irradiation approach is the potential toxicity to critical normal structures from the accumulative radiation dose in time. Few studies have performed detailed dosimetric analyses to analyze the added toxicities due to overlapped dose coverage from the initial primary radiotherapy and salvage SBRT. Again, at any anatomic site of interest the combined biologic effects of sequentially separated radiation treatment courses of distinct fractionation schemes cannot be inferred from the simple summation of total radiation doses deposited. We have thus employed the concept of BED (or EQD2) to guide our salvage SBRT planning (Table 2) and provided the dosimetric comparisons between RA & HA techniques separately as well as in combination with the initial primary treatment (Table 4).

The cumulative dosage, either in terms of physical or biologically effective dose, to OARs only showed minor differences between HA and RA plans. In certain OARs, such as the optic apparatus and temporal lobe, the cumulative dose from the primary radiotherapy and salvage SBRT plans were significantly different between the RA and HA plans, but the difference was of a small magnitude (1–4 Gy, or 0.03–2.15Gy3). The difference in OAR doses between HA and RA was seen to be a result of the different arc trajectories. Enabling the spread-out of radiation dose is a characteristic of noncoplanar beam delivery. At the very least, more entrance and exit doses from the noncoplanar arcs would lead to wider low-dose regions. The radiobiological significance of OARs embedded within such a low-dose bath remains unknown. The current algorithm of HA technique is very effective in controlling high and intermediate dose spreading and fall-off while maintaining conformity. For example, the brain V12, an indicator of brain necrosis, was significantly smaller with HA than with RA technique (HA vs. RA: 3.79 c.c. vs. 8.54 c.c., p = 0.022). The maximal dose of the brain in HA was also significantly less than in RA plans (HA vs. RA: 24.40Gy vs. 26.46Gy, p = 0.017).

Neurologic organs are relatively easy to identify and their treatment induced side effects are easier to evaluate. However, toxicities to soft tissues such as trismus or muscle fibrosis, along with the dose-toxicity relationships, are more difficult to evaluate but could nonetheless truly impact patient’s quality of life. It has been reported that patients who were retreated with a cumulative external beam dose of > 100 Gy had a high incidence of severe complications [34]. In our study, HA significantly decreased the body volume that received > 100 Gy as compared to RA technique.

There are still some limitations to this current study. First, the 10 enrolled patients had diverse treatment plan criteria for the initial primary radiation treatments before the subsequent salvage SBRT treatments. This limitation reflected the reality of heterogeneous clinical presentation of recurrent NPC cases. While we were unable to optimize the initial treatment plans that had been delivered, we could maximize the quality of the salvage treatment plans. Both HA and RA techniques showed excellent target dose coverages while sparing OARs. Second, in the current study, we only summed the BEDs of the initial primary radiotherapy and salvage SBRT treatment, without considering the factor of possible normal tissue repairs during the time in-between which might also influence the degree of ultimate toxicity expression. For neurological tissues, for example, the tolerance to cumulative dose may be enhanced with the increase in the time interval between the initial treatment and re-irradiation [35, 36]. It may be challenging to analyze these kinetic parameters since the efficiency of such latent repair often remains imprecisely known for various OARs of interest. In theory, the very fact that BED is “additive” which we try to illustrate in the current study also reflects the feasibility of incorporating such normal tissue protective effect – but only if future research could provide more robust quantitative data. It would be analogous to what has been done by past investigators in accounting for the effect of “accelerated repopulation” for acute responding tissues [37]. Regardless, by ignoring the plausible “protective” effect for normal tissue repair during a prolonged time interval, the dose constraints as listed in Table 2 would represent even more conservative estimates independent from the chosen reirradiation technique of either RA or HA approach. Third, there have not been clinically significant outcomes observed with either HA or RA technique so far. It remains unknown what degree of late toxicity might be caused by small increments of total radiation doses or BEDs for certain OARs. With increasing understanding and skills about how to prevent severe treatment-related toxicities [38, 39], the high-quality VMAT treatment planning might continue to help improve the therapeutic ratio of salvage SBRT treatment for recurrent NPC cases.

Conclusions

The novel dosimetric distributions in conformality, homogeneity and low dose spillage make the HA technique an attractive SBRT option for the salvage treatment of recurrent NPC. It should be noted that certain OARs under the arc trajectory of HA (e.g., optic apparatus) that had received substantial amount of dose during the initial course of primary radiation treatment might accumulate more radiation dose than the same OARs would with RA technique. Further clinical studies using HA for recurrent NPC would be necessary to confirm the therapeutic benefits and the toxicity profiles.

Availability of data and materials

Not applicable.

Abbreviations

SBRT:

Stereotactic body radiation therapy

RA:

RapidArc

HA:

HyperArc

OARs:

Organs at risk

NPC:

Nasopharyngeal cancer

VMAT:

Volumetric modulated arc radiotherapy

MLC:

Multileaf collimator

PTV:

Planning target volume

CTV:

Clinical target volume

EQ D2:

Equivalent dose to 2 Gy per fraction

BED:

Biologically effective dose

CI:

Conformity index

HI:

Homogeneity index

MUs:

Monitor units

SRS:

Stereotactic radiosurgery

References

  1. Lee AW, Poon YF, Foo W, Law SC, Cheung FK, Chan DK, et al. Retrospective analysis of 5037 patients with nasopharyngeal carcinoma treated during 1976-1985: overall survival and patterns of failure. Int J Radiat Oncol Biol Phys. 1992;23:261–70.

    Article  CAS  Google Scholar 

  2. Teo P, Yu P, Lee WY, Leung SF, Kwan WH, Yu KH, et al. Significant prognosticators after primary radiotherapy in 903 nondisseminated nasopharyngeal carcinoma evaluated by computer tomography. Int J Radiat Oncol Biol Phys. 1996;36:291–304.

    Article  CAS  Google Scholar 

  3. Lee N, Xia P, Quivey JM, Sultanem K, Poon I, Akazawa C, et al. Intensity-modulated radiotherapy in the treatment of nasopharyngeal carcinoma: an update of the UCSF experience. Int J Radiat Oncol Biol Phys. 2002;53:12–22.

    Article  Google Scholar 

  4. Wolden SL, Chen WC, Pfister DG, Kraus DH, Berry SL, Zelefsky MJ. Intensity-modulated radiation therapy (IMRT) for nasopharynx cancer: update of the memorial Sloan-Kettering experience. Int J Radiat Oncol Biol Phys. 2006;64:57–62.

    Article  Google Scholar 

  5. Ng WT, Lee MC, Hung WM, Choi CW, Lee KC, Chan OS, et al. Clinical outcomes and patterns of failure after intensity-modulated radiotherapy for nasopharyngeal carcinoma. Int J Radiat Oncol Biol Phys. 2011;79:420–8.

    Article  Google Scholar 

  6. Lee N, Harris J, Garden AS, Straube W, Glisson B, Xia P, et al. Intensity-modulated radiation therapy with or without chemotherapy for nasopharyngeal carcinoma: radiation therapy oncology group phase II trial 0225. J Clin Oncol. 2009;27:3684–90.

    Article  Google Scholar 

  7. Lee NY, Zhang Q, Pfister DG, Kim J, Garden AS, Mechalakos J, et al. Addition of bevacizumab to standard chemoradiation for locoregionally advanced nasopharyngeal carcinoma (RTOG 0615): a phase 2 multi-institutional trial. Lancet Oncol. 2012;13:172–80.

    Article  CAS  Google Scholar 

  8. Tham IW, Hee SW, Yeo RM, Salleh PB, Lee J, Tan TW, et al. Treatment of nasopharyngeal carcinoma using intensity-modulated radiotherapy-the national cancer Centre Singapore experience. Int J Radiat Oncol Biol Phys. 2009;75:1481–6.

    Article  Google Scholar 

  9. Peng G, Wang T, Yang KY, Zhang S, Zhang T, Li Q, et al. A prospective, randomized study comparing outcomes and toxicities of intensity-modulated radiotherapy vs. conventional two-dimensional radiotherapy for the treatment of nasopharyngeal carcinoma. Radiother Oncol. 2012;104:286–93.

    Article  Google Scholar 

  10. Lee AW, Sze WM, Au JS, Leung SF, Leung TW, Chua DT, et al. Treatment results for nasopharyngeal carcinoma in the modern era: the Hong Kong experience. Int J Radiat Oncol Biol Phys. 2005;61:1107–16.

    Article  Google Scholar 

  11. Lin YW, Chen CC, Lin LC, Lee SP. The impact of reduced-volume, intensity-modulated radiation therapy on disease control in nasopharyngeal carcinoma. PLoS One. 2015;10:e0125283.

    Article  Google Scholar 

  12. Wu SX, Chua DT, Deng ML, Zhao C, Li FY, Sham JS, et al. Outcome of fractionated stereotactic radiotherapy for 90 patients with locally persistent and recurrent nasopharyngeal carcinoma. Int J Radiat Oncol Biol Phys. 2007;69:761–9.

    Article  Google Scholar 

  13. Vargo JA, Moiseenko V, Grimm J, Caudell J, Clump DA, Yorke E, et al. Head and neck tumor control probability: radiation dose-volume effects in stereotactic body radiation therapy for locally recurrent previously-irradiated head and neck cancer: report of the AAPM working group. Int J Radiat Oncol Biol Phys. 2018;S0360-3016(18):30107.

    Google Scholar 

  14. Vargo JA, Ward MC, Caudell JJ, Riaz N, Dunlap NE, Isrow D, et al. A multi-institutional comparison of SBRT and IMRT for definitive Reirradiation of recurrent or second primary head and neck cancer. Int J Radiat Oncol Biol Phys. 2018;100:595–605.

    Article  Google Scholar 

  15. Wild E, Bangert M, Nill S, Oelfke U. Noncoplanar VMAT for nasopharyngeal tumors: plan quality versus treatment time. Med Phys. 2015;42:2157–68.

    Article  Google Scholar 

  16. Dizman A, Coskun-Breuneval M, Altinisik-Inan G, Olcay GK, Cetindag MF, Guney Y. Reirradiation with robotic stereotactic body radiotherapy for recurrent nasopharyngeal carcinoma. Asian Pac J Cancer Prev. 2014;15:3561–6.

    Article  Google Scholar 

  17. Otto K. Volumetric modulated arc therapy: IMRT in a single gantry arc. Med Phys. 2008;35:310–7.

    Article  Google Scholar 

  18. Roa DE, Schiffner DC, Zhang J, Dietrich SN, Kuo JV, Wong J, et al. The use of RapidArc volumetric-modulated arc therapy to deliver stereotactic radiosurgery and stereotactic body radiotherapy to intracranial and extracranial targets. Med Dosim. 2012;37:257–64.

    Article  Google Scholar 

  19. Clark GM, Popple RA, Prendergast BM, Spencer SA, Thomas EM, Stewart JG, et al. Plan quality and treatment planning technique for single isocenter cranial radiosurgery with volumetric modulated arc therapy. Pract Radiat Oncol. 2012;2:306–13.

    Article  Google Scholar 

  20. Clark GM, Popple RA, Young PE, Fiveash JB. Feasibility of single-isocenter volumetric modulated arc radiosurgery for treatment of multiple brain metastases. Int J Radiat Oncol Biol Phys. 2010;76:296–302.

    Article  Google Scholar 

  21. Ruggieri R, Naccarato S, Mazzola R, Ricchetti F, Corradini S, Fiorentino A, et al. Linac-based VMAT radiosurgery for multiple brain lesions: comparison between a conventional multi-isocenter approach and a new dedicated mono-isocenter technique. Radiat Oncol. 2018;13:38.

    Article  Google Scholar 

  22. Ohira S, Ueda Y, Akino Y, Hashimoto M, Masaoka A, Hirata T, et al. HyperArc VMAT planning for single and multiple brain metastases stereotactic radiosurgery: a new treatment planning approach. Radiat Oncol. 2018;13:13.

    Article  Google Scholar 

  23. Ueda Y, Ohira S, Yamazaki H, Mabuchi N, Higashinaka N, Miyazaki M, et al. Dosimetric performance of two linear accelerator-based radiosurgery systems to treat single and multiplebrain metastases. Br J Radiol. 2019;92:20190004.

    Article  Google Scholar 

  24. Yuan J, Lee R, Dusenbery KE, Lee CK, Mathew DC, Sperduto PW, et al. Cumulative doses to brain and other critical structures after multisession gamma knife stereotactic radiosurgery for treatment of multiple metastatic tumors. Front Oncol. 2018;8:65.

    Article  Google Scholar 

  25. Barendsen GW. Dose fractionation, dose rate and iso-effect relationships for normal tissue responses. Int J Radiat Oncol Biol Phys. 1982;8:1981–97.

    Article  CAS  Google Scholar 

  26. Fowler JF. The linear-quadratic formula and progress in fractionated radiotherapy. Br J Radiol. 1989;62:679–94.

    Article  CAS  Google Scholar 

  27. Benedict SH, Yenice KM, Followill D, Galvin JM, Hinson W, Kavanagh B, et al. Stereotactic body radiation therapy: the report of AAPM task group 101. Med Phys. 2010;37:4078–101.

    Article  Google Scholar 

  28. Nakamura JL, Verhey LJ, Smith V, Petti PL, Lamborn KR, Larson DA, et al. Dose conformity of gamma knife radiosurgery and risk factors for complications. Int J Radiat Oncol Biol Phys. 2001;51:1313–9.

    Article  CAS  Google Scholar 

  29. Feuvret L, Noel G, Mazeron JJ, Bey P. Conformity index: a review. Int J Radiat Oncol Biol Phys. 2006;64:333–42.

    Article  Google Scholar 

  30. Hong LX, Garg M, Lasala P, Kim M, Mah D, Chen CC, et al. Experience of micromultileaf collimator linear accelerator based single fraction stereotactic radiosurgery: tumor dose inhomogeneity, conformity, and dose fall off. Med Phys. 2011;38:1239–47.

    Article  Google Scholar 

  31. Alongi F, Clerici E, Pentimalli S, Mancosu P, Scorsetti M. Initial experience of hypofractionated radiation retreatment with true beam and flattening filter free beam in selected case reports of recurrent nasopharyngeal carcinoma. Rep Pract Oncol Radiother. 2012;17:262–8.

    Article  Google Scholar 

  32. Mayo CS, Ding L, Addesa A, Kadish S, Fitzgerald TJ, Moser R. Initial experience with volumetric IMRT (RapidArc) for intracranial stereotactic radiosurgery. Int J Radiat Oncol Biol Phys. 2010;78:1457–66.

    Article  Google Scholar 

  33. Zhuang M, Zhang T, Chen Z, Lin Z, Li D, Peng X, et al. Advanced nasopharyngeal carcinoma radiotherapy with volumetric modulated arcs and the potential role of flattening filter-free beams. Radiat Oncol. 2013;8:120.

    Article  Google Scholar 

  34. Pryzant RM, Wendt CD, Delclos L, Peters LJ. Re-treatment of nasopharyngeal carcinoma in 53 patients. Int J Radiat Oncol Biol Phys. 1992;22:941–7.

    Article  CAS  Google Scholar 

  35. Kirkpatrick JP, van der Kogel AJ, Schultheiss TE. Radiation dose-volume effects in the spinal cord. Int J Radiat Oncol Biol Phys. 2010;76:S42–9.

    Article  Google Scholar 

  36. Das S, Patro K, Mukherji A. Recovery and tolerance of the organs at risk during re-irradiation. Journal of Current Oncology. 2018;1:23–8.

    Article  Google Scholar 

  37. Withers HR, Maciejewski B, Taylor JM, Hliniak A. Accelerated repopulation in head and neck cancer. Front Radiat Ther Oncol. 1988;22:105–10.

    Article  CAS  Google Scholar 

  38. Liu S, Lu T, Zhao C, Shen J, Tian Y, Guan Y, et al. Temporal lobe injury after re-irradiation of locally recurrent nasopharyngeal carcinoma using intensity modulated radiotherapy: clinical characteristics and prognostic factors. J Neuro-Oncol. 2014;119:421–8.

    Article  CAS  Google Scholar 

  39. Gebhardt BJ, Vargo JA, Ling D, Jones B, Mohney M, Clump DA, et al. Carotid dosimetry and the risk of carotid blowout syndrome after Reirradiation with head and neck stereotactic body radiation therapy. Int J Radiat Oncol Biol Phys. 2018;101:195–200.

    Article  Google Scholar 

Download references

Acknowledgements

None.

Funding

No funding was received.

Author information

Authors and Affiliations

Authors

Contributions

Study concepts: HW Ho, CC Yang, SP Lee, YW Lin. Study design: HW Ho, SP Lee, YW Lin. Data acquisition: HW Ho, HM Lin, YW Lin. Quality control of data and algorithms: HM Lin, CC Yang, SC Wang. Data analysis and interpretation: HW Ho, HM Lin, SP Lee, YW Lin. Statistical analysis: HM Lin, SC Wang. Manuscript preparation: HM Lin, SC Wang, SP Lee. Manuscript editing: HW Ho, CC Yang, YW Lin, SP Lee. Manuscript review: HW Ho, YW Lin, SP Lee. The authors read and approved the final manuscript.

Corresponding author

Correspondence to Yu-Wei Lin.

Ethics declarations

Ethics approval and consent to participate

This study was approved by the Institutional Review Board at Chi Mei Medical Center. The institutional review board waived the need for written informed consent from the participants because this was a retrospective electric treatment plan-review study.

Consent for publication

Not applicable.

Competing interests

None of the authors of this manuscript have conflicts of interest, except the corresponding author (YW Lin) has received Speaker honorarium from Varian Medical Systems.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Ho, HW., Lee, S.P., Lin, HM. et al. Dosimetric comparison between RapidArc and HyperArc techniques in salvage stereotactic body radiation therapy for recurrent nasopharyngeal carcinoma. Radiat Oncol 15, 164 (2020). https://doi.org/10.1186/s13014-020-01602-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13014-020-01602-7

Keywords