Optimization of dose distributions of target volumes and organs at risk during stereotactic body radiation therapy for pancreatic cancer with dose-limiting auto-shells

Abstract

Background

To identify optimization of dose distributions of target volumes and decrease of radiation doses to normal tissues during stereotactic body radiation therapy (SBRT) for pancreatic cancer with dose-limiting auto-shells.

Methods

With the same prescription dose, dose constraints of normal organs and calculation algorithm, treatment plans of each eligible patient were re-generated with 3 shells, 5 shells and 7 shells, respectively. The prescription isodose line and beam number of each patient in 3-shell, 5-shell and 7-shell plan remained the same. Hence, a triplet data set of dosimetric parameters was generated and analyzed.

Results

As the increase of shell number, the conformal index, volumes encompassed by 100% prescription isodose line and 30% prescription isodose line significantly decreased. The new conformal index was higher in 3-shell group than that in 5-shell and 7-shell group. A sharper dose fall-off was found in 5-shell and 7-shell group compared to 3-shell group. And the tumor coverage in 7-shell was better than that of 3-shell and 5-shell. Lower D5cc of the intestine, D10cc of the stomach, Dmax of the spinal cord and smaller V10 of the spleen was confirmed in 7-shell group compared to 3-shell group.

Conclusions

More conformal dose distributions of target volumes and lower radiation doses to normal organs could be performed with the increase of dose-limiting auto-shells, which may be more beneficial to potential critical organs without established dose constraints.

Background

High doses may be required to achieve local control attributable to inherent radio-resistance of pancreatic cancer. However, the abutting organs at risk (OAR), including the stomach, duodenum and bowel, limit the prescription dose because of radio-sensitivity. Due to the advances in the radiotherapy, the nonisocentric and noncoplanar technique of stereotactic body radiation therapy (SBRT) could deliver precise radiation with high local dose conformation, precise target localization [1] with motion compensation strategies [2, 3] and facilitate quick dose fall-off outside the tumor volume [4, 5].

Nevertheless, treatment planning is strongly user-dependent since optimization parameters were preselected [6, 7]. Hence, a well-balanced plan should be based on the ALARA (as-low-as-reasonably-achievable) principle, and is generally preferable with extreme quality in only a few categories as long as any strict critical structure limits are not violated. Owing to the normal tissues in close proximity to the pancreas, a plan with trade-offs, if clinical justified, among all planning objectives is required.

Traditionally, collimators are nontrivial for controlling dose conformity and dose fall-off. While multiple fixed collimators may allow better conformity and fewer hot spots around the tumor boundary than a single one, but it is time-consuming and a large number of collimators could even impair plan quality. The Iris collimator can reduce treatment time [8], but may not necessarily lead to better dosimetric plan quality [9]. Also, a larger tumor volume indicated a larger aperture collimator, contributing to a larger penumbra and slow dose fall-off, detrimental to normal organs.

Apart from the collimator, shell structures are also used for optimizing dose conformity and reducing skin entry doses as well as hot spots outside the target area. Furthermore, in addition to constraints of OARs, auto-shell is required to control dose gradients around the target volume, but without too much compromise of target volume coverage. Therefore, the selection of the suited number of shell structures and distances between each shell is pivotal.

As a result, the aim of our study was to implement dose-limiting auto-shells in the SBRT for pancreatic cancer to minimize doses of OAR and non-critical organs and further optimize dose distributions of target volumes with Sequential Multi-Objective Optimization.

Methods

Eligibility

Medical records of patients with pancreatic cancer receiving SBRT in our center from October 2016 to April 2017 were reviewed. All patients received clinical examinations and abdominal CT or MRI scanning. The inclusion criteria were as follows: diagnose confirmed by pathological examinations; locally advanced pancreatic cancer; patients with resectable or borderline resectable cancer intolerant of surgical resections; age ranging from 18 to 75 years old; tumor located in the pancreatic head with maximum diameter < 5 cm; ECOG ≤2; all patients receiving the same prescription dose. The following exclusion criteria were used: patients with a history of radiotherapy prior to the SBRT; metastatic pancreatic cancer.

Robotic radiosurgery system and treatment planning system

SBRT was delivered via CyberKnife, an image-guided frameless stereotactic robotic radiosurgery system (Accuray Corporation, Sunnyvale CA). The treatment planning process was carried out with a dedicated treatment planning system, Multiplan version 4.0.2 (Accuray Inc.).

The sequential method was applied for all plans in the investigation. The inverse treatment-planning algorithm was performed to maximize the minimum dose to target volume or the mean dose, known as “optimize coverage (OCO)” in the system. The upper bounding constraints of the OAR were restricted to the following doses during the optimization (volume of interest limits, VOI limits): spinal cord: 3Gy; stomach: 15Gy; intestine: 14Gy; duodenum: 14Gy. In order to minimize the maximum doses to the critical structures, the above limits were stricter than those reported in the American Association of Physicists in Medicine guidelines in TG-101 [10]. The optimization of monitor unit (MU) was performed as follows: total MU: 90,000; max MU per beam: 500; max MU per node: 1500. This could reduce isodose lines showing up as streaks in the direction of beam entry points, and hot spots in the vicinity of the beam entry points just below the skin surface [11].

Target delineations

The procedure of CyberKnife was similar to our previous study [12, 13]. Gross tumor volume (GTV) was delineated as a radiographically evident gross disease by contrast CT acquired from the portal-venous phase. At the discretion of the physician, clinical target volume (CTV) encompassing areas of the potential subclinical disease spread was also designated. In most cases, the CTV equaled GTV. A 2–5 mm expansion margin was included to determine the planning target volume (PTV). When the tumor was adjacent to critical organs, the expansion of PTV outside of CTV in this direction should be avoided. Therefore, the margin expansion was allowed to be nonuniform. References of normal tissue constraints were according to the American Association of Physicists in Medicine guidelines in TG-101 [10].

Auto-shells planning

Each patient treatment plan was re-generated with 3 shells, 5 shells and 7 shells, respectively. All the patients received the same prescription dose (37.5Gy/5f). At least 90% of PTV should be covered by the prescription dose. The prescription isodose line was limited to 70–80%, which would restrict the tumor Dmax. The prescription isodose line of each patient in 3-shell, 5-shell and 7-shell plan was the same. Shells were determined based on the expansions of PTV margins (Details shown in Table 1). The dose limitations between every two shells were assigned (Table 2) in order to obtain desired dose fall-off without compromising coverage significantly. The distance from the PTV margin to each shell was 5 mm, 15 mm and 30 mm in 3-shell group, respectively; 2 mm, 5 mm, 15 mm, 30 mm and 60 mm in 5-shell group, respectively; 2 mm, 5 mm, 10 mm, 15 mm, 30 mm, 60 mm and 100 mm in 7-shell group, respectively. Beam reduction in the system was utilized to restrict beam numbers, remove low-weighted beams and reoptimize with the remaining beams to keep the plan quality, which resulted in ensuring the same number of beams in the three plans. The plans were calculated by Monte Carlo algorithm with high resolution and the uncertainty was 1%.

Table 1 Generation of shells based on the expansions of PTV margins
Table 2 Dose limitations between every two shells

Statistical analysis

Parameters selected for evaluation of doses of target volumes included conformity index (CI), new conformity index (nCI), heterogeneity index (HI), gradient index (GI), coverage, volumes encompassed by 100% isodose line (100% PD-V), volumes encompassed by 50% isodose line (50% PD-V) and volumes encompassed by 30% isodose line (30% PD-V). Crucially, the formula of calculation of CI was: \( \mathrm{CI}=\frac{prescription\ isodose\ volume\ \left( PIV,{cm}^3\right)}{tumor\ volume\ encompassed\ prescription\ isodose\ line\ \left( TIV,{cm}^3\right)} \). The nCI was calculated as follows: \( \mathrm{nCI}=\frac{CI}{coverage} \), in which coverage was defined as the ratio of target volumes covered with prescription dose to the target volume. The HI was determined by the following formula: \( \mathrm{HI}=\frac{D_{max}}{R_{xDose\kern0.5em }} \), in which RxDose was the prescription dose. The definition of GI was the ratio of the volume of half the prescription isodose to the volume of the prescription isodose [14]. Evaluations of doses to OAR included the maximum dose (Dmax) and volume doses. Dmax was defined as the dose of a 0.035-cc or less. The dose volume of the intestine, stomach and spinal cord was the dose of a 5-cc volume (D5cc), 10-cc volume (D10cc) and 0.35-cc volume (D0.35cc). Doses of a 5-cc and a 10-cc volume were investigated in the case of the duodenum. In addition to these OAR, doses to the spleen were also studied, including the mean dose (Dmean), the dose of a half spleen volume (D50%) and the volume receiving 5Gy (V5) and 10Gy (V10). However, 2 patients had splenectomy. Hence, dose evaluations were only performed in 18 patients.

Comparisons between parameters in different shell groups were based on analysis of variance (ANOVA) for a randomized block design. Fisher’s least significance difference test (LSD) was performed when significant differences were found in the three groups after analysis of variance. Results were deemed significant at P <  0.05. Statistical analyses were performed using IBM SPSS version 22.0 (SPSS Inc., Armonk, NY).

Results

Patient characteristics

A total of 20 patients were enrolled in the investigation. The median age was 64.5 years old. The prescription dose was 37.5Gy/5f (BED10 = 65.625Gy, biological effective dose, α/β = 10). The median GTV and PTV was 26.95cm3 and 36.33cm3, respectively. The beam number ranged from 150 to 210 and the median was 186.

Evaluations of dose distributions of PTV

With the increase of shell number, the CI, 100% PD-V and 30% PD-V significantly decreased (CI: 3-shell vs 5-shell: 1.12 ± 0.05 vs 1.10 ± 0.04, P <  0.001; 5-shell vs 7-shell: 1.10 ± 0.04 vs 1.09 ± 0.03, P = 0.024; 3-shell vs 7-shell: 1.12 ± 0.05 vs 1.09 ± 0.03, P <  0.001) (100% PD-V: 3-shell vs 5-shell: 38.92 ± 12.31 vs 38.45 ± 12.85, P = 0.036; 5-shell vs 7-shell: 38.45 ± 12.85 vs 37.73 ± 12.62, P = 0.002; 3-shell vs 7-shell: 38.92 ± 12.31 vs 37.73 ± 12.62, P <  0.001) (30% PD-V: 3-shell vs 5-shell: 344.53 ± 119.93 vs 312.90 ± 110.18, P <  0.001; 5-shell vs 7-shell: 312.90 ± 110.18 vs 299.15 ± 106.41, P = 0.006; 3-shell vs 7-shell: 344.53 ± 119.93 vs 299.15 ± 106.41, P <  0.001). Additionally, there was significant difference between 3-shell and 5-shell, 3-shell and 7-shell with respect to nCI and GI. Furthermore, the 50% PD-V was smaller in 7-shell compared with that of 3-shell and 5-shell. And the tumor coverage in 7-shell was better than that of 3-shell and 5-shell (Table 3).

Table 3 Evaluations of dose distributions of PTV

Evaluations of doses to OAR

The doses to normal tissue were shown in Table 4. After restriction of doses to OAR with different shells, D5cc of the intestine, D10cc of the stomach and Dmax of the spinal cord was lower in 7-shell than those in 3-shell (D5cc of the intestine: 3-shell vs 7-shell: 15.50 ± 2.10 vs 14.95 ± 2.51, P = 0.003; D10cc of the stomach: 3-shell vs 7-shell: 14.32 ± 1.74 vs 13.97 ± 2.23, P = 0.020; Dmax of the spinal cord: 3-shell vs 7-shell: 7.12 ± 1.20 vs 6.63 ± 0.89, P = 0.046), though only a trend of decrease of D10cc of the stomach and Dmax of the spinal cord was found when the number of shell increased.

Table 4 Evaluations of dose to OAR

Crucially, in addition to the conventional OARs, the dose-volume parameters of the spleen were further investigated. There was no significant difference between the three groups regarding Dmean, V5 and D50%. Nevertheless, V10 was much smaller in 7-shell group than that in 3-shell group.

Given that one patient with a large tumor abutting to the spleen, invading the splenic hilar, the radiation dose was much higher than that for other patients. Therefore, the further analysis of V10 precluded the patient was performed. It was also shown that V10 was smaller in 7-shell group compared with 3-shell group (3-shell: 0 ± 1.31; 5-shell: 0 ± 1.30; 7-shell: 0 ± 0.82; 3-shell vs 5-shell, P = 0.367; 5-shell vs 7-shell: P = 0.156; 3-shell vs 7-shell: P = 0.024). An exemplary image showing the comparison of dose contributions between the three shells, five shells and seven shells plan was illustrated (Fig. 1).

Fig. 1
figure 1

a three-shell plan and d the dose distribution, b five-shell plan and e the dose distribution, c seven-shell plan and f the dose distribution

Discussion

Complex tumor shapes or lesions adjacent to critical organs were challenging to treatment plans, which however did not indicate multiple calibration of structures or complicated optimization steps [15]. Therefore, even a high-quality treatment plan, short and simple optimization scripts may be appropriate [14]. Besides, manual tuning parameters combined with iterative optimization with different scripts were based on trade-offs between planning objectives, which were determined by clinical preferences and should be prioritized. Hence, a well-balanced plan required emphasis on several planning constraints and aims without any compromise of other vital goals. In our study, all of the treatment plans were generated with pre-selection of constraints, some of which were even stricter than those recommended, and optimization of PTV coverage (OCO), in order to meanwhile comply with ALARA. All other constraints remained unchangeable to explore the influence of shell structures on dose distributions.

The results of our study indicate that with the increase of the shells, CI, volumes encompassed by different isodose lines, doses to OAR with a certain volume and V10 of the spleen decreased while dose conformity was better. In the Multiplan® System, auto-shells are generated at the expansion of different diameters based on the PTV. Typically, the generation of auto-shells could be performed with a unique dilation value per direction, or with one value to all directions for symmetric dilations. Auto-shells are used to control dose distribution both close to and far away from a target. An auto-shell close to a target can modify conformity, while a shell distant to a target can be used to control low dose distributions or remote hot spots. Additionally, for those normal tissues not contoured as OAR, which were deemed possibly invulnerable to the radiation, hot spots could be avoided by auto-shells.

In previous studies, it was suggested that a shell structure of 1-3 mm was used to limit the high dose conformity, while two auto-shells with a distance between each other of 5-10 mm for the optimization of the prescription dose conformity and 15-20 mm for the optimization of proximate dose fall-off. Additionally, a shell structure of 30-50 mm was employed to limit the skin doses and hot spots outside the proximate target region [9]. Therefore, the number and distance of shell structures should be adjusted according to the target volume, clinical justifications and planning goals. Furthermore, Schlaefer et al. [15] concluded that an additional shell structure with a distance of 5 mm to the PTV should be used to control dose gradient around the PTV. In this study, the distances between each shell were similar to those reported.

Yoon et al. [16] has reported clinical practice of different shells for brain metastases. In their study, the original CyberKnife plans were produced using one to three dose-limiting auto-shells at prescription dose level and low dose levels of 10 to 30% of PD. The modified CyberKnife plans were generated by five dose-limiting shells at prescription dose level, intermediate dose level of 50% of prescription dose, and low dose levels with an optimal shell-dilation size based on their experience. It was shown that there was no difference in CI and tumor coverage between two plans, while the normal tissue volume receiving 50% of prescription dose was significantly decreased in the modified plans compared with the original plans by 1.28-fold, indicating that a steeper dose fall-off could be achieved with the increase of auto-shells.

The previous results were similar to ours. The GI was higher in 5-shell and 7-shell group compared with 3-shell group. The underlying reason may be that the allowed distance for dose fall-off (37.5Gy to 26Gy) from the PTV margin to the first shell was 5 mm in 3-shell group. While in 5-shell and 7-shell group, the dose fall-off (37.5Gy to 26Gy) was divided into two processes: the dose decreased from 37.5Gy to 32Gy within a short distance of 2 mm (from the PTV margin to the first shell), and from 32Gy to 26Gy within a distance of 3 mm (from the first shell to the second shell).

Additionally, it was elucidated that V10 of the spleen was significantly lower in 7-shell group than that in 3-shell group, which may suggest that the radiation dose to the spleen was lower in 7-shell group.

Conclusions

Better conformity, lower radiation doses and smaller low-dose areas of normal organs could be achieved with the increase of dose-limiting auto-shells, though maybe at the cost of tumor coverage. Crucially, for organs without established dose constraints, milder radiation-induced toxicities may be attributable to auto-shells. Typically, protection of the spleen from excessive unintentional irradiation with auto-shells may reduce the incidence of lymphocytopenia, which may be beneficial to prognosis. Therefore, auto-shells could be an option to customize individual plans to ensure adequate normal organs sparing when clinically indicated.

Abbreviations

100% PD-V:

Volumes encompassed by 100% isodose line

30% PD-V:

Volumes encompassed by 30% isodose line

50% PD-V:

Volumes encompassed by 50% isodose line

ALARA:

As-low-as-reasonably-achievable

ANOVA:

Analysis of variance

CI:

Conformity index

CTV:

Clinical target volume

GI:

Gradient index

GTV:

Gross tumor volume

HI:

Heterogeneity index

LSD:

Fisher’s least significance difference test

MU:

Monitor unit

nCI:

new conformity index

OAR:

Organs at risk

OCO:

Optimize coverage

PTV:

Planning target volume

SBRT:

Stereotactic body radiation therapy

VOI:

Volumes of interest

References

  1. Leksell L. The stereotaxic method and radiosurgery of the brain. Acta Chir Scand. 1951;102:316–9.

    CAS  PubMed  Google Scholar 

  2. Colombo F, Benedetti A, Pozza F, Zanardo A, Avanzo RC, Chierego G, et al. Stereotactic radiosurgery utilizing a linear accelerator. Appl Neurophysiol. 1985;48:133–45.

    CAS  PubMed  Google Scholar 

  3. Hartmann GH, Schlegel W, Sturm V, Kober B, Pastyr O, Lorenz WJ. Cerebral radiation surgery using moving field irradiation at a linear accelerator facility. Int J Radiat Oncol Biol Phys. 1985;11:1185–92.

    Article  CAS  Google Scholar 

  4. Dieterich S, Cleary K, D’Souza W, Murphy M, Wong KH, Keall P. Locating and targeting moving tumors with radiation beams. Med Phys. 2008;35:5684–94.

    Article  Google Scholar 

  5. Chung HT, Kim DG. Modern radiosurgery equipment for treating brain metastases. Prog Neurol Surg. 2012;25:236–47.

    Article  Google Scholar 

  6. Hoegele W, Loeschel R, Merkle N, Zygmanski P. An efficient inverse radiotherapy planning method for VMAT using quadratic programming optimization. Med Phys. 2012;39:444–54.

    Article  CAS  Google Scholar 

  7. Ziegenhein P, Kamerling CP, Bangert M, Kunkel J, Oelfke U. Performance-optimized clinical IMRT planning on modern CPUs. Phys Med Biol. 2013;58:3705–15.

    Article  Google Scholar 

  8. Echner GG, Kilby W, Lee M, Earnst E, Sayeh S, Schlaefer A, et al. The design, physical properties and clinical utility of an iris collimator for robotic radiosurgery. Phys Med Biol. 2009;54:5359–80.

    Article  CAS  Google Scholar 

  9. Blanck O, Wang L, Baus W, Grimm J, Lacornerie T, Nilsson J, et al. Inverse treatment planning for spinal robotic radiosurgery: an international multi-institutional benchmark trial. J Appl Clin Med Phys. 2016;17:313–30.

    Article  Google Scholar 

  10. Benedict SH, Yenice KM, Followill D, Galvin JM, Hinson W, Kavanagh B, et al. Stereotactic body radiation therapy: the report of AAPM task group 101. Med Phys. 2010;37:4078–101.

    Article  Google Scholar 

  11. Li T, Ozhasoglu C, Burton S, Flickinger J, Heron DE, Huq MS. A method to improve dose gradient for robotic radiosurgery. J Appl Clin Med Phys. 2015;16:333–9.

    Article  Google Scholar 

  12. Zhu X, Ju X, Cao F, Fang F, Qing S, Shen Y, et al. Safety and efficacy of stereotactic body radiation therapy combined with S-1 simultaneously followed by sequential S-1 as an initial treatment for locally advanced pancreatic cancer (SILAPANC) trial: study design and rationale of a phase II clinical trial. BMJ Open. 2016;6:e013220.

    Article  Google Scholar 

  13. Zhu X, Li F, Ju X, Cao F, Cao Y, Fang F, et al. Prognostic role of stereotactic body radiation therapy for elderly patients with advanced and medically inoperable pancreatic cancer. Cancer Med. 2017; https://www.doi.org/10.1002/cam4.1164.

  14. Treuer H, Hoevels M, Luyken K, Visser-Vandewalle V, Wirths J, Kocher M, et al. Intracranial stereotactic radiosurgery with an adapted linear accelerator vs. robotic radiosurgery: comparison of dosimetric treatment plan quality. Strahlenther Onkol. 2015;191:470–6.

    Article  Google Scholar 

  15. Schlaefer A, Schweikard A. Stepwise multi-criteria optimization for robotic radiosurgery. Med Phys. 2008;35:2094–103.

    Article  CAS  Google Scholar 

  16. Yoon K, Cho B, Kwak J, Lee D, Kwon D, Ahn S, et al. Optimization of Cyberknife treatment planning using a dose-limiting auto-shells method for brain metastases. Med Phys. 2016;43:3371.

    Article  Google Scholar 

Download references

Acknowledgements

We appreciate Dr. Jiuhong Chen for her precise revisions and constructive comments.

Funding

This study was sponsored by China Health Promotion Foundation (THC2015001) and Youth Fund of Changhai Hospital (CH201709).

Availability of data and materials

Not applicable.

Author information

Authors and Affiliations

Authors

Contributions

CYS re-generated the plans. ZXF drafted the manuscript. JXP and ZHJ revised the manuscript. LYM, YCS and SYJ checked the plans. ZXF, DZT and GXL performed statistical analyses. ZHJ designed the study. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Huojun Zhang.

Ethics declarations

Ethics approval and consent to participate

The study was approved by the institutional review board of our hospital.

Consent for publication

The consents for publication of data have been obtained from patients.

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Cao, Y., Zhu, X., Ju, X. et al. Optimization of dose distributions of target volumes and organs at risk during stereotactic body radiation therapy for pancreatic cancer with dose-limiting auto-shells. Radiat Oncol 13, 11 (2018). https://doi.org/10.1186/s13014-018-0956-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13014-018-0956-7

Keywords