Influence of target dose heterogeneity on dose sparing of normal tissue in peripheral lung tumor stereotactic body radiation therapy

Abstract

Objective

To evaluate the influence of target dose heterogeneity on normal tissue dose sparing for peripheral lung tumor stereotactic body radiation therapy (SBRT).

Methods

Based on the volumetric-modulated arc therapy (VMAT) technique, three SBRT plans with homogeneous, moderate heterogeneous, and heterogeneous (HO, MHE, and HE) target doses were compared in 30 peripheral lung tumor patients. The prescription dose was 48 Gy in 4 fractions. Ten rings outside the PTV were created to limit normal tissue dosage and evaluate dose falloff.

Results

When MHE and HE plans were compared to HO plans, the conformity index of the PTV was increased by approximately 0.08. The median mean lung dose (MLD), V5, V10, V20 of whole lung, D2%, D1cc, D2cc of the rib, V30 of the rib, D2% and the maximum dose (Dmax) of the skin, and D2% and Dmax of most mediastinal organs at risk (OARs) and spinal cord were reduced by up to 4.51 Gy or 2.8%. Analogously, the median Dmax, D2% and mean dose of rings were reduced by 0.71 to 8.46 Gy; and the median R50% and D2cm were reduced by 2.1 to 2.3 and 7.4% to 8.0%, respectively. Between MHE and HE plans there was little to no difference in OARs dose and dose falloff beyond the target. Furthermore, the dose sparing of rib V30 and the mean dose of rings were negatively correlated with the rib and rings distance from tumor, respectively.

Conclusions

For peripheral lung tumor SBRT, target conformity, normal tissue dose, and dose falloff around the target could be improved by loosening or abandoning homogeneity. While there was negligible further dose benefit for the maximum target dose above 125% of the prescription, dose sparing of normal tissue derived from a heterogeneous target decreased as the distance from the tumor increased.

Background

Worldwide, lung cancer remains the leading cause of cancer incidence and mortality [1]. Also, the lung is a common site of metastases for various other cancers. Radiotherapy is an alternative treatment method for inoperable patients. Yet, local control of conventional fraction radiotherapy has historically been poor due to insufficient total radiation doses [2]. Stereotactic body radiation therapy (SBRT) can very precisely deliver a high dose of radiation to an extracranial target, using either a single dose or a small number of fractions [3]. The clinical benefits of SBRT have been demonstrated by several prospective studies for both early-stage, non-small cell lung cancer (NSCLC) and pulmonary oligo-metastases [4,5,6].

Meanwhile, a high dose per fractional treatment is more likely to cause late effects, which can be very destructive and result in significant dysfunction to the treated tissues. It can be devastating to quality of life and even deadly [7]. Some lung SBRT studies indicate that toxicities in different organs at risk (OARs), including lung [8], chest wall [9,10,11], skin [12], esophagus [13], great vessels [14], trachea, and bronchus [15], are significantly correlated with the radiation dose of corresponding OARs [8, 16]. Therefore, it is of great importance to reduce radiation dose to normal tissues in SBRT. Specialized treatment planning is required to deliver the dose with as high as possible a target conformity and steep dose gradients beyond the target [17].

Because heterogeneous radiotherapy plans allow for an equal or lower dose of OARs compared to homogeneous plans [18, 19], clinicians prefer to loosen or abandon homogeneity of the target dose with SBRT. Hot spots within target volumes are generally viewed to be clinically desirable, as long as there is no spillage into normal tissues [20]. However, there is still no consensus on the dose heterogeneity level inside the target. RTOG recommends a selection of the prescription isodose surface set to a value between 60 and 90% of the dose at the center of the planning target volume (PTV) or the maximum dose (Dmax) [21]. Therefore, there is a lot of variation in target heterogeneity for SBRT plans, and which level of dose heterogeneity is the most optimal for SBRT in lung tumor remains unclear. Furthermore, the quality of a radiotherapy plan is typically dependent on the experience and preference of its planners [22,23,24,25]. Consequently, previous dosimetry studies of SBRT plans are almost all based on manual planning [23,24,25,26], which makes it difficult to guarantee unbiased comparison. In the present study, we developed an automatic planning program for SBRT to reduce reliance on the experience and preference of planners and to evaluate the relationship between dose sparing of normal tissue with different levels of dose heterogeneity.

Methods

Patient eligibility

A total of 30 patients, who were treated with SBRT at West China Hospital between April 2011 and March 2017 for peripheral lung cancer or pulmonary oligo-metastases, were enrolled. The characteristics of the patients are shown in Table 1.

Table 1 The characteristics of patients

Treatment planning

Each patient was immobilized in a stereotactic body frame (SBF, Elekta Oncology System, Sweden) in the supine position with arms raised above the head. The intravenous contrast-enhanced 4D-CT (SOMATOM Definition AS+, SIEMENS, 120 kVp, 90 mAs) covering the total lung volume was obtained. All the CTs were reconstructed in 3 mm slice thickness and transferred to the RayStation treatment planning system (TPS) (RaySearch Laboratories, v4.7). The details of localization, simulation, immobilization, delineation of target and OARs volumes, and prescription dose constraints were discussed in RTOG 0915 [21].

An automatic SBRT planning software for automatically creating the planning auxiliary structures, adding beams and initial objectives and constraints, adjusting parameters and optimization was developed based on the RayStation TPS. The program simulated the process of manual plan design, and included adding plan objectives and constraints, adjusting the parameters in the process of optimization. Objectives and constraints were set according to the prescription. Parameter adjustment was based on each optimized objective value, to ensure the objective value was in the range of 10 to 30 times of tolerance (tolerance = 0.0001). The total number of iterations per patient was arbitrarily set to 10. The minimum precision of automatic adjustment was 2 cGy, which is far more precise than that of manual adjustment (more than 50 cGy in most cases). Using the same initial planning conditions for one randomly selected patient, 10 automated and 10 manual SBRT plans were generated to assess the repeatability. Variation coefficient was calculated as follows: standard deviation/mean value × 100%. Greater than 2% was considered to be poor repeatability.

The prescription dose to PTV was 48 Gy in 4 fractions. Based on the average CT of each patient, three SBRT plans with homogeneous, moderate heterogeneous, and heterogeneous target doses (the HO, MHE, and HE plans) were generated and customized to the accelerator (Elekta Versa HD, Elekta Oncology, UK) with a 6-MV photon beam, in which the PTV dose was controlled in 90–110%, 90–125% and 90–∞% of the prescription dose, respectively. VMAT technique was used for all plans, while the collapsed cone algorithm was performed to compute the final dose. The Dmax in the HO and MHE plans were based on the selection of the prescription isodose surface set to 90% and 80% of Dmax, respectively [21]. Given the uncertainty of the minimum dose (Dmin), at least D99% (DV is the absorbed dose that covers a specified volume V) must be ≥ 90% of the prescription dose. Two coplanar full arcs, the same initial objectives and constraints, the collapsed cone algorithm, and a 2 mm grid were used. Ten rings (5 mm width for each ring) outside the PTV were created to limit the dose of normal tissue and evaluate dose falloff.

Plan analysis

All automatic plans were independently assessed by three radiation oncologists using a previously described standard [27], including that target doses meet prescription goals and doses of OARs reach or are below the dose-volume limits. Using a 3-point scale, the best target coverage or OAR sparing would score 3 points and the worst would score 1 point. The overall score for each plan was the weighted summation of all the scores (33% for the weights of target coverage and 67% for the weights of OAR sparing). The mean value of three overall scores by three oncologists for each plan was regarded as each plan’s final score.

The Heterogeneity Index (HI, HI = (D2%-D98%)/D50%) was calculated according to ICRU report 83 [28], and the Conformity Index (CI, CI = (TVPIV × TVPIV)/(TV × PIV) [N.B. TVPIV = PTV volume within the prescription isodose volume, TV = PTV volume, and PIV = prescription isodose volume]) was calculated according to the Paddick Index equation [29]. The tumor to OAR distance (defined as the minimum distance from the tumor edge to OAR edge) was measured. Differences were analyzed by the Friedman test (among the three plans) or Wilcoxon signed ranks test (between two plans), and P < 0.05 (2-tailed, Friedman) and P < 0.017 (α/3, 2-tailed, Wilcoxon) were considered to be statistically significant. The correlation between two variables was analyzed by Spearman's rank correlation, and P < 0.05 (2-tailed) was considered to be statistically significant.

Results

All of the 10 automatic plans generated for repeatability evaluation had the same dose distribution, including the dose of PTV and OARs and dose falloff (all variation coefficients < 2%). On the contrary, repeatability of the manual plans varied greatly. Variation coefficients for many parameters were greater than 5%, and those of mediastinal OARs were even more than 10%.

According to the evaluation of three radiation oncologists, MHE plans achieved the highest final scores regarding target coverage and OAR sparing (median, 15.83; range, 13.89–16.78), which was superior than HO plans (median, 14.56; range, 12.67–15.67; P < 0.001) and HE plans (median, 15.05; range, 13.56–16.11; P = 0.001) (Additional file 1: Figure 1).

Figures 1, 2 and 3 present the dose differences of PTV, OARs, and the rings for the three kinds of plans. The statistical dose comparisons are shown in the Additional file 2: Tables 1–3. The number of patients exceeding the dose threshold are shown in Table 2. All dosimetric parameters met the prescription dose constraints except R50% and R100% (ratio of 50% and 100% prescription isodose volume to the PTV volume). The monitor units (MUs) of the three plan types were as follows (by order of magnitude): HO plans < MHE plans < HE plans. The maximum difference of MUs was found between HO and HE plans, where the median increase was 400 (range, 170–895) MUs.

Fig. 1
figure 1

Pairwise comparisons of PTV dose. Each data point was derived from subtracting the value achieved with one plan from the one achieved with another plan. Horizontal bars indicate median values. PTV = planning target volume; Dmax = maximum dose; Dmin = minimum dose; DV = absorbed dose that covers a specified fractional volume V; HO, MHE, and HE plans = homogeneous, moderate heterogeneous, and heterogeneous plans

Fig. 2
figure 2

Pairwise comparisons of OARs doses. Each data point was derived from subtracting the value achieved with one plan from the one achieved with another plan. Horizontal bars indicate median values. OARs = organs at risk; TPBT = trachea and proximal bronchial tree; MLD = mean lung dose; ESO = esophagus; GV = great vessels; SC = spinal cord; Dmax = maximum dose; D1cc = minimum absorbed dose that covers 1 cc of the volume; D2cc = minimum absorbed dose that covers 2 cc of the volume; VD = volume that receives at least the absorbed dose D Gy; HO, MHE, and HE plans = homogeneous, moderate heterogeneous, and heterogeneous plans. *percentage

Fig. 3
figure 3

Pairwise comparisons of Rings dose. Each data point was derived from subtracting the value achieved with one plan from the one achieved with another plan. Horizontal bars indicate median values. Dmean = mean dose; HO, MHE, and HE plans = homogeneous, moderate heterogeneous, and heterogeneous plans

Table 2 Number of patients exceeding the dose threshold

PTV dose

D50% and HIs of three kinds of plans, by order of magnitude, were as follows: HO plans < MHE plans < HE plans. For CIs, the values calculated from both MHE and HE plans were greater than those calculated from HO plans, where the median differences were approximately 0.080. There was no statistical difference in CIs between MHE and HE plans (P = 0.165).

OARs dose

For most OARs, the doses in MHE and HE plans were lower than those in HO plans, where the median doses were reduced by up to 4.51 Gy or 2.8%. On the contrary, for dose parameters of OARs between MHE and HE plans, the differences existed only in V10, V20, MLD of the whole lung, D2%, D2cc and V30 of the rib, D2% of the skin, and D2% of the esophagus, where the differences were less than 0.26 Gy or 0.3%.

Lung

The MLD and V20 of the whole lung in MHE and HE plans were lower than those in HO plans, where the median dose reduction was approximately 0.50 Gy (range, 0.16–2.03 Gy) and 1% (0.3–4.2%), respectively. Also, the numbers of patients whose MLD exceeded 6 and 4 Gy and V20 exceeded 12%, 10% and 4% in MHE and HE plans were less than those in HO plans. Moreover, the V5 and V10 of the whole lung in MHE and HE plans were lower than those in HO plans, where the median dose reduction ranged from 1.6 to 2.8%.

OARs of chest wall

For the rib, the median D2%, D2cc and D1cc were reduced by 3.85, 3.50 and 3.99 Gy (MHE plans vs. HO plans), and 4.51, 4.34 and 4.29 Gy (HE plans vs. HO plans). The median reductions of the rib V30 in MHE and HE plans were both 2.2% compared to HO plans. Furthermore, the dose sparing of V30 was negatively correlated with the tumor to rib distance (ρ = − 0.685, P < 0.001, MHE plans and HO plans; ρ = − 0.680, P < 0.001, HE plans and HO plans). For skin, the median Dmax in MHE and HE plans were reduced by 1.94 and 2.28 Gy, respectively, compared to HO plans.

Mediastinal OARs and spinal cord

For most mediastinal OARs including esophagus, heart, great vessels, trachea, and proximal bronchial tree (TPBT), the doses in MHE and HE plans were lower than those in HO plans, where the median dose improvements were 0.87–1.95 Gy. Similarly, the median Dmax of the spinal cord in MHE and HE plans was reduced by approximately 1.50 Gy compared to those in HO plans.

Dose falloff

The median R50% reductions were 2.1 (range, 0.6–10.2, P < 0.001, MHE plans vs. HO plans) and 2.3 (0.4–13.1, P < 0.001, HE plans vs. HO plans). For HE plans, only 6 patients with a PTV volume of less than 7.5 cc did not meet the R50% requirements of RTOG 0915, while for MHE and HO plans these patient numbers were 11 and 22, respectively. The Dmax (% of dose prescribed) 2 cm from the PTV in any direction (D2cm) for MHE and HE plans was lower than those for HO plans, where the median D2cm reductions were 7.4% for MHE plans vs. HO plans (0.3% to 21.9%) and 8.0% for HE plans vs. HO plans (− 0.4% to 22.6%). However, the median differences of R50% and D2cm between MHE and HE plans were lower than 0.3 and 0.9%, respectively.

For Dmax, D2% and the mean dose (Dmean) of all rings (excluding Dmax and D2% of Ring1), the median dose reductions ranged from 0.80 to 8.46 Gy (HE plans vs. HO plans) and 0.71 to 6.60 Gy (MHE plans vs. HO plans), while the value was only 0.17 to 1.20 Gy (HE plans vs. MHE plans). The Dmax of Ring1 in MHE and HE plans were higher than those in HO plans, where the median differences were 2.66 and 4.30 Gy. No matter Dmax, D2% or Dmean, the maximum differences were both in Ring2 or Ring3, then the differences got smaller from Ring2 or Ring3 to Ring10. Furthermore, the median Dmean and the median Dmean reductions in both MHE and HE plans vs. HO plans were negatively correlated with the distance from tumor to ring (ρ ranged from − 0.964 to − 1.000, P < 0.001).

Discussion

The goal of this study is to characterize the relationship between the OARs sparing, dose falloff, and dose heterogeneity level of the target. To our knowledge, this is the first time that, based on an automatic planning program, three heterogeneity levels of target doses in lung SBRT plans have been investigated. We hope that the automatic planning program could make our results more objective and more stable. Moreover, the adjustment precision level of the automatic plan (2 cGy) was more accurate than that of the manual plan (more than 50 cGy).

Unlike conventional radiotherapy, the PTV in SBRT plans mainly involves tumor tissue with its necessary margins arising from tumor motion. Accordingly, it is not necessary for SBRT plans to obtain a higher level of target dose homogeneity [30]. The advantage of the heterogeneous target dose has been demonstrated in the SBRT plans to abdominal malignancies [31]. In the study of Widder et al. on the prescription strategy of SBRT for lung tumors [30], they found that the lower the percentage of the prescription isodose level to the isocenter dose, the faster the dose falloff would be, which was beneficial to improve the normal tissue dose sparing. Our study indirectly confirmed their findings. We found that PTV conformity, OARs sparing, and dose falloff beyond targets improved significantly in MHE and HE plans compared to those in HO plans, while dose improvements in MHE and HE plans were very similar. In addition, the dose sparing of rib and rings derived from the heterogeneous target were associated with the distance from the tumor to normal tissue.

Most planning studies of SBRT have previously been based on manual plans [23,24,25,26]. However, the quality of manual IMRT plans has often been dependent on the experience of each planner [22], as well as on the clinical preferences of planners and centers [23,24,25]. In the present study, all SBRT plans were based on the automated planning software, which guaranteed that the prescription and optimization strategies for each plan were identical. These measures could thereby avoid or minimize the impact caused by the experience or clinical preferences of different planners or centers and guarantee unbiased dose comparison.

At the cost of homogeneity, the CIs calculated from MHE and HE plans were superior to those calculated from HO plans, which is inconsistent with Miao et al., who performed a study for the conventional fraction radiotherapy of NSCLC and found that there was no significant difference in CI between inhomogeneous and homogeneous plans [19]. A possible explanation for this might be that the SBRT technique pays more attention to conformity than the conventional fraction radiotherapy technique. In the present study, ten rings outside the PTV were used to limit dose to normal tissue and improve PTV conformity. These results suggested that target conformity of SBRT plans could be improved by loosening or abandoning dose homogeneity. Other possible reasons were the PTV volumes in their study was much bigger than those in ours and they also used a Step-and-Shoot technique, not VMAT. In the study by Widder et al. [30], the dose was prescribed to include the PTV with the prescription isodose level specified in a range between 50 and 80% of the isocenter dose. They found that poor CIs was showed in homogeneous plans, not in inhomogeneous plans. Their results were consistent with ours, suggesting that it is not necessary for SBRT plans to obtain a higher level of target dose homogeneity.

Radiation-induced lung toxicity (RILT) is one of the major factors limiting the maximal radiation dose that can be safely delivered to thoracic tumors [32, 33]. Dosimetric parameters are associated with the development of RILT [8]. Chang et al. demonstrated that MLD > 6 Gy, V20 > 12%, and ipsilateral V30 > 15% were significantly associated with grade 2–3 radiation pneumonitis (RP) [8]. In our study, both MLD and V20 in MHE and HE plans were lower than those in HO plans, as well as the number of patients whose MLD and V20 exceeded 6 Gy and 12%, which could be beneficial to reducing the risk of RILT. However, the improvement of lung doses in our study was less than those of Miao et al. who found that mean lung volume received 0–60 Gy and MLD was reduced by up to 5.5% (approximately 1.4 Gy). This result may be attributed to fewer limitations on dose falloff in conventional fraction radiotherapy than in SBRT. In their study, only two rings were used to limit dose outside the target. Nevertheless, we still believe lung dose and the risk of RILT can be reduced by loosening or abandoning the dose homogeneity of the target.

Chest wall toxicities are usually associated with peripheral lung tumor treated with SBRT [9,10,11]. A recent pooled analysis indicated that dosimetric parameters including Dmax of 0.5–5 cc and V30 for chest wall or ribs were significantly associated with chest wall pain and rib fracture [9]. Various maximum cutoff doses ranging from 21 to 60 Gy in 2–5 fractions with 0% to 55.7% associated risk of chest wall toxicity were reported [10, 11, 34]. In the present study, the results showed that median rib D2cc in MHE and HE plans were reduced by approximately 4 Gy compared to those in HO plans, and the number of patients with D2cc above 30.8 Gy in MHE and HE plans was 3 and 4 less than those in HO plans. Similarly, the median rib V30 was reduced by 2.2%. In addition, the dose improvements of V30 in MHE and HE plans were associated with the distance from tumor to rib. These results support the previous finding that tumor-to-chest wall distance was significantly associated with chest wall pain and rib fracture [9]. For skin, the dosimetric parameters related to skin toxicity included skin Dmax [12] and chest wall V30 [35]. In our study, the median dose reduction of skin Dmax in MHE and HE plans was approximately 2 Gy compared to HO plans. Therefore, inhomogeneous plans could be beneficial to reducing the risk of chest wall and skin toxicities.

For mediastinal OARs, some dose–response models developed by different SBRT teams show that the probability of radiotherapy toxicities increases with dose, such as Dmax of esophagus [13], great vessels [14], trachea, and bronchus [15]. In our study, the median dose of D2% or Dmax reductions of esophagus, great vessels, trachea, and bronchus were less than 2 Gy, which were lower than those for rib. This discrepancy may be due to the fact that peripheral lung tumors are farther from mediastinal OARs than from ribs. Nevertheless, the results still indicated the dose of mediastinal OARs can be reduced by loosening or abandoning homogeneity, which may help reduce the risk of radiation-induced toxicities. Especially for patients treated with radiotherapy combined with chemotherapy or immunotherapy, which has been found to increase the risk of toxicity [36,37,38].

Steep dose falloff gradients beyond the target are very critical for SBRT, because tissues exposed to high fraction doses are prone to significant dysfunction [7]. Additionally, normal tissue dose tolerances in the context of SBRT are still evolving and only have limited experimentation. For 3D-CRT, beam penumbra at the target edge can be set manually to produce a steep dose falloff outside the PTV with small to no margins. Unlike 3D-CRT, the aperture in VMAT plans was always optimized automatically. Because of this, we used ten rings outside the PTV in this study to limit the dose around the target. We found that Dmax and Dmean for all rings (excluding the Dmax of Ring1) in MHE and HE plans were reduced by 0.71–8.46 Gy compared to those in HO plans. Analogously, D2cm, R50%, and R100% in MHE and HE plans were superior to those in HO plans. Therefore, we believe that a steeper dose gradient can be obtained by loosening or abandoning homogeneity of target dose. In addition, the median Dmean improvements of inhomogeneous plans were negatively correlated with the distance from tumor to rings. Combining the results of the ribs and rings, we speculated that normal tissue dose sparing derived from the heterogeneous target could decrease as the distance from the tumor increases.

Comparing MHE and HE plans, there was little to no difference in dose parameters for all OARs or in dose falloff, although different levels of heterogeneity for them were used. A possible explanation for this might be that too flat of a dose curve for PTV in the HE plans results in OARs doses raised by the prescription dose normalization. When D95% and D99% were fulfilled by a normalizing prescription dose, a flatter PTV dose line on the DVH led to a dose increase of OARs and other normal tissue, which offset dose sparing obtained by abandoning homogeneity. Therefore, we speculate that 90–125% of the prescription dose may be a tradeoff of heterogeneity levels between target coverage and OARs sparing. There seems to be no further advantage of dose sparing for higher Dmax.

Several limitations should be addressed here. Firstly, this is a single center study and the sample size is relatively small. Secondly, we have not validated clinical results of different heterogeneity plans. A larger cohort and multi-center study may be necessary to identify dose and clinical outcome differences among SBRT plans with different levels of heterogeneity.

Conclusions

For SBRT plans of peripheral lung tumor, the conformity of the target, the dose of OARs, and the dose falloff around the target could be improved by loosening or abandoning dose homogeneity. While there was little further dose benefit for a maximum target dose above 125% of the prescription. In addition, the dose sparing of normal tissue derived from a heterogeneous target tended to decrease as distance from the tumor increased.

Availability of data and materials

The datasets generated and/or analyzed during the current study are not publicly available because it contains personal information, but are available from the corresponding author on reasonable request.

Abbreviations

SBRT:

Stereotactic body radiation therapy

VMAT:

Volumetric-modulated arc therapy

HO, MHE, and HE:

Homogeneous, moderate heterogeneous, and heterogeneous

MLD:

Mean lung dose

PTV:

Planning target volume

D max :

Maximum dose

OARs:

Organs at risk

NSCLC:

Non-small cell lung cancer

HI:

Heterogeneity index

CI:

Conformity index

R 50%  and  R 100% :

Ratio of 50% and 100% prescription isodose volume to the PTV volume

MUs:

Monitor units

D V :

The absorbed dose that covers a specified volume V

TPBT:

Trachea and proximal bronchial tree

D mean :

Mean dose

RILT:

Radiation-induced lung toxicity

D min :

Minimum dose

D V :

Absorbed dose that covers a specified fractional volume V

V D :

Volume that receives at least the absorbed dose D Gy

D 2cm :

Maximum dose (in % of dose prescribed) 2 cm from PTV in any direction

D 2cc :

Minimuhcfy-anchorm absorbed dose that covers 2 cc of the volumhcfy-anchore

References

  1. Bray F, Ferlay J, Soerjomataram I, Siegel RL, Torre LA, Jemal A. Global cancer statistics 2018: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries. CA Cancer J Clin. 2018;68:394–424.

    Article  Google Scholar 

  2. Qiao X, Tullgren O, Lax I, Sirzen F, Lewensohn R. The role of radiotherapy in treatment of stage I non-small cell lung cancer. Lung Cancer. 2003;41:1–11.

    Article  Google Scholar 

  3. Guckenberger M, Andratschke N, Dieckmann K, Hoogeman MS, Hoyer M, Hurkmans C, et al. ESTRO ACROP consensus guideline on implementation and practice of stereotactic body radiotherapy for peripherally located early stage non-small cell lung cancer. Radiother Oncol. 2017;124:11–7.

    Article  Google Scholar 

  4. Baumann P, Nyman J, Hoyer M, Wennberg B, Gagliardi G, Lax I, et al. Outcome in a prospective phase II trial of medically inoperable stage I non-small-cell lung cancer patients treated with stereotactic body radiotherapy. J Clin Oncol. 2009;27:3290–6.

    Article  Google Scholar 

  5. Rusthoven KE, Kavanagh BD, Burri SH, Chen C, Cardenes H, Chidel MA, et al. Multi-institutional phase I/II trial of stereotactic body radiation therapy for lung metastases. J Clin Oncol. 2009;27:1579–84.

    Article  Google Scholar 

  6. Chang JY, Senan S, Paul MA, Mehran RJ, Louie AV, Balter P, et al. Stereotactic ablative radiotherapy versus lobectomy for operable stage I non-small-cell lung cancer: a pooled analysis of two randomised trials. Lancet Oncol. 2015;16:630–7.

    Article  Google Scholar 

  7. Timmerman RD. An overview of hypofractionation and introduction to this issue of seminars in radiation oncology. Semin Radiat Oncol. 2008;18:215–22.

    Article  Google Scholar 

  8. Chang JY, Li QQ, Xu QY, Allen PK, Rebueno N, Gomez DR, et al. Stereotactic ablative radiation therapy for centrally located early stage or isolated parenchymal recurrences of non-small cell lung cancer: how to fly in a “no fly zone.” Int J Radiat Oncol Biol Phys. 2014;88:1120–8.

    Article  Google Scholar 

  9. Ma JT, Liu Y, Sun L, Milano MT, Zhang SL, Huang LT, et al. Chest wall toxicity after stereotactic body radiation therapy: a pooled analysis of 57 studies. Int J Radiat Oncol Biol Phys. 2019;103:843–50.

    Article  Google Scholar 

  10. Andolino DL, Forquer JA, Henderson MA, Barriger RB, Shapiro RH, Brabham JG, et al. Chest wall toxicity after stereotactic body radiotherapy for malignant lesions of the lung and liver. Int J Radiat Oncol Biol Phys. 2011;80:692–7.

    Article  Google Scholar 

  11. Pettersson N, Nyman J, Johansson KA. Radiation-induced rib fractures after hypofractionated stereotactic body radiation therapy of non-small cell lung cancer: a dose- and volume-response analysis. Radiother Oncol. 2009;91:360–8.

    Article  Google Scholar 

  12. Hoppe BS, Laser B, Kowalski AV, Fontenla SC, Pena-Greenberg E, Yorke ED, et al. Acute skin toxicity following stereotactic body radiation therapy for stage I non-small-cell lung cancer: who’s at risk? Int J Radiat Oncol Biol Phys. 2008;72:1283–6.

    Article  Google Scholar 

  13. Nuyttens JJ, Moiseenko V, McLaughlin M, Jain S, Herbert S, Grimm J. Esophageal dose tolerance in patients treated with stereotactic body radiation therapy. Semin Radiat Oncol. 2016;26:120–8.

    Article  Google Scholar 

  14. Xue J, Kubicek G, Patel A, Goldsmith B, Asbell SO, LaCouture TA. Validity of current stereotactic body radiation therapy dose constraints for aorta and major vessels. Semin Radiat Oncol. 2016;26:135–9.

    Article  Google Scholar 

  15. Duijm M, Schillemans W, Aerts JG, Heijmen B, Nuyttens JJ. Dose and volume of the irradiated main bronchi and related side effects in the treatment of central lung tumors with stereotactic radiotherapy. Semin Radiat Oncol. 2016;26:140–8.

    Article  Google Scholar 

  16. Zhao J, Yorke ED, Li L, Kavanagh BD, Li XA, Das S, et al. Simple factors associated with radiation-induced lung toxicity after stereotactic body radiation therapy of the thorax: a pooled analysis of 88 studies. Int J Radiat Oncol Biol Phys. 2016;95:1357–66.

    Article  Google Scholar 

  17. Potters L, Kavanagh B, Galvin JM, Hevezi JM, Janjan NA, Larson DA, et al. American Society for Therapeutic Radiology and Oncology (ASTRO) and American College of Radiology (ACR) practice guideline for the performance of stereotactic body radiation therapy. Int J Radiat Oncol Biol Phys. 2010;76:326–32.

    Article  Google Scholar 

  18. Chen GP, Ahunbay E, Schultz C, Li XA. Development of an inverse optimization package to plan nonuniform dose distributions based on spatially inhomogeneous radiosensitivity extracted from biological images. Med Phys. 2007;34:1198–205.

    Article  Google Scholar 

  19. Miao J, Yan H, Tian Y, Ma P, Liu Z, Li M, et al. Reducing dose to the lungs through loosing target dose homogeneity requirement for radiotherapy of non small cell lung cancer. J Appl Clin Med Phys. 2017;18:169–76.

    Article  Google Scholar 

  20. Benedict SH, Yenice KM, Followill D, Galvin JM, Hinson W, Kavanagh B, et al. Stereotactic body radiation therapy: the report of AAPM Task Group 101. Med Phys. 2010;37:4078–101.

    Article  Google Scholar 

  21. Videtic GM, Paulus R, Singh AK, Chang JY, Parker W, Olivier KR, et al. Long-term follow-up on NRG oncology RTOG 0915 (NCCTG N0927): a randomized phase 2 study comparing 2 stereotactic body radiation therapy schedules for medically inoperable patients with stage I peripheral non-small cell lung cancer. Int J Radiat Oncol Biol Phys. 2019;103:1077–84.

    Article  Google Scholar 

  22. Bohsung J, Gillis S, Arrans R, Bakai A, De Wagter C, Knoos T, et al. IMRT treatment planning:- a comparative inter-system and inter-centre planning exercise of the ESTRO QUASIMODO group. Radiother Oncol. 2005;76:354–61.

    Article  Google Scholar 

  23. Giglioli FR, Strigari L, Ragona R, Borzi GR, Cagni E, Carbonini C, et al. Lung stereotactic ablative body radiotherapy: a large scale multi-institutional planning comparison for interpreting results of multi-institutional studies. Phys Med. 2016;32:600–6.

    Article  Google Scholar 

  24. Esposito M, Maggi G, Marino C, Bottalico L, Cagni E, Carbonini C, et al. Multicentre treatment planning inter-comparison in a national context: the liver stereotactic ablative radiotherapy case. Phys Med. 2016;32:277–83.

    Article  Google Scholar 

  25. Marino C, Villaggi E, Maggi G, Esposito M, Strigari L, Bonanno E, et al. A feasibility dosimetric study on prostate cancer: are we ready for a multicenter clinical trial on SBRT? Strahlenther Onkol. 2015;191:573–81.

    Article  Google Scholar 

  26. Xiao J, Li Y, Jiang Q, Sun L, Henderson F Jr, Wang Y, et al. Hepatic arterial phase and portal venous phase computed tomography for dose calculation of stereotactic body radiation therapy plans in liver cancer: a dosimetric comparison study. Radiat Oncol. 2013;8:264.

    Article  Google Scholar 

  27. Xiao J, Li Y, Shi H, Chang T, Luo Y, Wang X, et al. Multi-criteria optimization achieves superior normal tissue sparing in intensity-modulated radiation therapy for oropharyngeal cancer patients. Oral Oncol. 2018;80:74–81.

    Article  Google Scholar 

  28. Hodapp N. The ICRU Report 83: prescribing, recording and reporting photon-beam intensity-modulated radiation therapy (IMRT). Strahlenther und Onkol. 2012;188:97–9.

    Article  CAS  Google Scholar 

  29. Paddick I. A simple scoring ratio to index the conformity of radiosurgical treatment plans. Technical note. J Neurosurg. 2000;93(Suppl 3):219–22.

    Article  Google Scholar 

  30. Widder J, Hollander M, Ubbels JF, Bolt RA, Langendijk JA. Optimizing dose prescription in stereotactic body radiotherapy for lung tumours using Monte Carlo dose calculation. Radiother Oncol. 2010;94:42–6.

    Article  Google Scholar 

  31. Lax I, Blomgren H, Naslund I, Svanstrom R. Stereotactic radiotherapy of malignancies in the abdomen. Methodological aspects. Acta Oncol. 1994;33:677–83.

    Article  CAS  Google Scholar 

  32. Pollom EL, Chin AL, Diehn M, Loo BW, Chang DT. Normal tissue constraints for abdominal and thoracic stereotactic body radiotherapy. Semin Radiat Oncol. 2017;27:197–208.

    Article  Google Scholar 

  33. Hawkins PG, Boonstra PS, Hobson ST, Hearn JWD, Hayman JA, Ten Haken RK, et al. Radiation-induced lung toxicity in non-small-cell lung cancer: understanding the interactions of clinical factors and cytokines with the dose–toxicity relationship. Radiother Oncol. 2017;125:66–72.

    Article  CAS  Google Scholar 

  34. Murray L, Karakaya E, Hinsley S, Naisbitt M, Lilley J, Snee M, et al. Lung stereotactic ablative radiotherapy (SABR): dosimetric considerations for chest wall toxicity. Br J Radiol. 2016;89:20150628.

    Article  Google Scholar 

  35. Welsh J, Thomas J, Shah D, Allen PK, Wei X, Mitchell K, et al. Obesity increases the risk of chest wall pain from thoracic stereotactic body radiation therapy. Int J Radiat Oncol Biol Phys. 2011;81:91–6.

    Article  Google Scholar 

  36. Rakovitch E, Tsao M, Ung Y, Pignol JP, Cheung P, Chow E. Comparison of the efficacy and acute toxicity of weekly versus daily chemoradiotherapy for non-small-cell lung cancer: a meta-analysis. Int J Radiat Oncol Biol Phys. 2004;58:196–203.

    Article  Google Scholar 

  37. Rowell NP, O'Rourke NP. Concurrent chemoradiotherapy in non-small cell lung cancer. Cochrane Database Syst Rev. 2004; CD002140.

  38. Tian S, Switchenko JM, Buchwald ZS, Patel PR, Shelton JW, Kahn SE, et al. Lung stereotactic body radiation therapy and concurrent immunotherapy: a multicenter safety and toxicity analysis. Int J Radiat Oncol Biol Phys. 2020;108:304–13.

    Article  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

This research supported by grants from the Sichuan Science and Technology Program (No: 2020YFG0079) and the National Natural Science Foundation of China (No: 62071314).

Author information

Authors and Affiliations

Authors

Contributions

Conceptualization, J.H.X., Z.G.W., X.C.P.; methodology, Z.G.W., X.C.P., Y.W.; validation, Z.G.W., X.C.P.; formal analysis, Z.G.W., X.C.P., Y.W., L.L.Y., L.H.; investigation, Y.W., L.L.Y., L.H., J.J.W.; resources, Z.R.L., X.L.M.; data curation, Z.R.L., X.L.M., R.D.L.; writing—original draft preparation, Z.G.W., X.C.P. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Jianghong Xiao.

Ethics declarations

Ethics approval and consent to participate

Institutional Review Board of West China hospital, Sichuan university approved the protocol, and the need for written informed consent was waived.

Consent for publication

Not applicable.

Competing interests

The authors declare no conflict of interest.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1: Figure 1.

Total score of three kinds of plan. Horizontal bars indicate median values. HO, MHE, and HE plans = homogeneous, moderate heterogeneous, and heterogeneous plans.

Additional file 2: Table 1.

PTV dose, monitor units, and statistical comparisons. Table 2. OARs dose and statistical comparisons. Table 3. Dose parameters of dose falloff beyond PTV and statistical comparisons.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Wei, Z., Peng, X., Wang, Y. et al. Influence of target dose heterogeneity on dose sparing of normal tissue in peripheral lung tumor stereotactic body radiation therapy. Radiat Oncol 16, 167 (2021). https://doi.org/10.1186/s13014-021-01891-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13014-021-01891-6

Keywords