Dosimetry study of three-dimensional print template for 125I implantation therapy

Abstract

Background

125I seed implantation has been found to show good therapeutic effects on tumors. Recent studies showed that three-dimensional (3D) print template-assisted 125I seed implantation can optimize radiation dose distribution. This study aimed to compare the dose distribution differences in 125I seed implantation among 3D print noncoplanar template- (3DPNCT), 3D print coplanar template- (3DPCT) assisted implantation and traditional free-hand implantation.

Methods

We systematically searched the PubMed, EMbase, Cochrane Library, Wan Fang Med Online, China National Knowledge Infrastructure (CNKI) from the earliest to November 2020 without time or language restrictions. And the references of primary literature were also searched. The outcome measures were dosimetry and operation time. This meta-analysis was carried out using Stata 12.0.

Results

A total of 16 original articles were selected for inclusion. The differences of D90, D100, V90, and V100 values pre- and post-implantation with traditional free-hand implantation showed statistically significant (p < 0.05). The differences of D90, D100, V100, V150, V200, and D2cc of organs at risk (OAR) values pre- and post-implantation with 3D print template showed no statistically significant (p > 0.05). Compared with traditional free-hand implantation without any templates, 3D print template could improve postoperative D90 (Standard mean difference, SMD = 0.67, 95% confidence interval (CI) = 0.35 to 0.98, p < 0.001), D100 (SMD = 0.82, 95%CI = 0.40 to 1.23, p < 0.001), V90 (SMD = 1.48, 95%CI = 0.95 to 2.00, p < 0.001), V100 (SMD = 1.41, 95%CI = 0.96 to 1.86, p < 0.001), and reduce operation time (SMD = − 0.93, 95%CI = − 1.34 to − 0.51, p < 0.001). In three studies, both 3DPNCT and 3DPCT plans were designed for all patients. The prescribed dose and seed activity were same. Pooled analysis of D90, D100, V100, D2cc of OAR, number of seeds and number of needles showed no significant differences between 3DPNCT and 3DPCT groups (p > 0.05). However, in 3DPNCT group, V150 and V200 were increased (SMD = 0.35, 0.49; 95%CI = 0.04 to 0.67, 0.02 to 0.96; p = 0.028, 0.043); the number of through bone needles was reduced (SMD = − 1.03, 95%CI = − 1.43 to − 0.64, p < 0.001).

Conclusions

Compared with traditional free-hand implantation, 3D print template-assisted 125I seeds implantation can optimize dose distribution and reduce the implantation time at the same time. Compared with 3D print coplanar template, 3D print noncoplanar template can increase the volume of high dose within tumor target and is more safer in the respect of puncture route.

Background

Recently, radioactive iodine—125 seeds (RIS) implantation has been widely applied to treat various malignant tumors and has achieved satisfactory therapeutic effects [1, 2]. RIS has the features of a minor trauma surgery, delivering a high local radiation dose to tumor targets and sharply dropping off at adjacent normal tissues. Nevertheless, up to date, there was still no standard procedures for 125I seed implantation for other tumors in the body except prostate cancer. It was challenging to effectively preplan for seed implantation resulted from patients’ body movement, organ movement, and bone structure interference. RIS implantation just relied on individual clinical experience and puncture techniques. Seed location and dose distribution were not the same as the preplan, which could lead to complications of operation and local recurrence of tumor.

In term of high-dose-rate (HDR) brachytherapy, Martinez developed an afterloading applicator that consisted of an template with a predrilled holes which were used as guides for trocars in 1984. And so, trocars could be inserted through the holes and produced optimal dose coverage of the tumor volume, which could reduce the degree of technical difficulty and improve the dose-rate distributions [3]. Aristei designed an 3D template which was confirmed to be a quick, easy, reliable and time-saving method to localize the volume of tumor target for HDR brachytherapy in breast cancer patients [4]. Mahantshetty included 113 patients with gynecologic cancers treated with template-based HDR interstitial brachytherapy boost, which resulted in a satisfactory clinical outcomes without any severe toxicities [5]. Coincidentally, doctors in China tried to implant RIS with the template which was designed individually by 3D printer. The use of 3D print templates including 3D print noncoplanar template (3DPNCT) and 3D print coplanar template (3DPCT) made it more precise to implant RIS, with a highly consistent dose distribution of target volume. Many studies showed that 3D print template-assisted RIS implantation could not only reduce the dosimetric differences between pre-and post-plan but also lower the difficulty of puncture [6,7,8]. However, the number of samples included in the past studies was small, and the quality was uneven. So, we aimed to peform a systematic review and meta-analysis of related researches on the dosimetry after RIS implantation with or without template.

Methods

Study selection

The meta-analysis was carried out according to the PRISMA (Preferred Reporting Items for Systematic Reviews and Meta-Analyses) statement for reporting reviews and meta-analysis [9].

Major electronic literature databases were systematically searched, which included EMbase, PubMed, Cochrane Library, Wan Fang Med Online, CNKI. The search used various combination of subject words and free words, which included brachytherapy, iodine radioisotopes, iodine-125, 3-dimensional printing, 3D print. And the search strategy was determined after multiple presearches. Articles published before November 2020 were found in the search without publication and any language restriction. In addition, The researchers review the full texts of the included literature and carefully checked the list of references of the selected literature so as to avoid missing any other relevant researches on this topic.

Inclusion criteria

In order to be included in this meta-analysis, the study had to meet all the criteria as follows: (1) randomized controlled trials (RCTs) or non-RCTs or retrospective study; (2) 2-arm studies in which patients received 3D print template-assisted RIS implantation in the treatment group and traditional free-hand implantation in the control group or a single-arm study reporting OAR with 3D print template-assisted RIS implantation; (3) Studies had outcomes of dosimetry or operation time.

Exclusion criteria

The exclusion criteria include the following: (1) abstract, letter, case report, editorial, animal experiments, review, and other irrelevant studies; (2) no outcome measures.

Data extraction

Two researchers (E.C. and Y.Z.) searched and reviewed related studies and carried out the data extraction independently. When there was any controversy, articles would be sent to a third reviewer (H.Z) for assessment until they achieved a agreement. We extract information for the following items: study characteristics (author, publication year, study design), demographic data (tumor site, sample size), treatment characteristics (with or without 3D print template), and outcome data including D90(the dose of 90% of the target volume), D100, V90(the percent of the tumor target receiving 90% of the prescribed dose), V100, V150, V200, D2cc (the dose received by 2 cm3 of normal tissue) and operation time.

Quality assessment

For RCTs, the methodological quality were assessed by Cochrane risk of bias tool. Non-RCTs were assessed by the Newcastle–Ottawa Scale (NOS) [10]. Two researchers (E.C. and J.W.) carried out the scoring independently, and debated until a full agreement was reached. Studies with a score more than 7 were considered high quality, 4–6 moderate and below than 4 low quality.

Statistical analysis

Standard mean difference (SMD) was adopted as the effect indicator for dosimetry and operation time. Pooled SMD and 95% confidence interval (CI) were calculated. We used I2 statistics to evaluate statistical heterogeneity. An I2 value of 0–40% indicates low heterogeneity; 30–60%, moderate heterogeneity; 50–90%, substantial heterogeneity; 75–100%, considerable heterogeneity. The Mantel–Haenszel fixed effect model [11] was applied for p > 0.1, I2 < 50%; data were pooled with the random-effects model when the I2 > 50%. p < 0.05 was considered to indicate that the difference was statistically significant. Publication bias was assessed with Egger’s regression. The statistical analysis were carried out using Stata 12.0 software.

Results

Literature search results

In total, 145 candidate publications were retrieved. Ultimately, 16 studies [12,13,14,15,16,17,18,19,20,21,22,23,24,25,26,27] which fullfilled the eligibility criteria were included in the final analysis (see Fig. 1 for further details). All of the researches originated from China., including 16 non-RCTs. Eight original articles [12,13,14,15,16,17,18,19] with 280 patients compared the dose distribution differences between 3D print template-assisted implantation and traditional free-hand implantation. Among them, 128 patients received 3D print template-assisted RIS implantation and 152 without template. Five studies [20,21,22,23,24] with 76 patients compared the differences of D90, D100, V100, V150, V200, and D2cc of organs at risk (OAR) values pre- and post-implantation with 3D print template. In three studies [25,26,27], both 3DPNCT and 3DPCT plans were designed for all patients. The prescribed dose and seed activity were same. The data including D90, D100, V100, V150, V200, D2cc of OAR, number of seeds, number of needles and number of through bone needles in the two plans were compared. Basic information for inclusion in the study is presented in Table 1.

Fig. 1
figure 1

Flow diagram of the study selection process. CNKI China Knowledge Resource Integrated Database

Table 1 Characteristics of the included studies

Quality evaluation

All included studies were assessed using Newcastle Ottawa Scale. All 2-arm studies [12,13,14,15,16,17,18,19] achieved a score of ≥ 7. Five studies [20,21,22,23,24] comparing dosimetry values pre- and post-implantation with 3D print template achieved a score of 6. Three studies [25,26,27] comparing 3DPNCTwith 3DPCT plans achieved a score of 9 (as shown in Table 2).

Table 2 Quality assessment of included studies

Dosimetry

Eight studies [12,13,14,15,16,17,18,19] compared post-implantation dosimetry data between 3D print template and traditional free-hand implantation. For D90, D100, V90, V100, there occurred no significant heterogeneities among results (I2 = 0, 0, 0, and 37%, respectively; p = 0.498, 0.315, 0.831, and 0.174, respectively). The pooled effect was therefore evaluated using a fixed-effects model. The result showed that all of these dosimetric parameters in 3D print template group were higher than those in traditional group with statistically significant (SMD = 0.67, 0.82, 1.48, and 1.41, respectively; 95%CI = 0.35 to 0.98, 0.40 to 1.23, 0.95 to 2.00, 0.96 to 1.86, respectively; p < 0.001) (as shown in Figs. 2 and 3). Six studies [12, 14,15,16, 18, 19] reported dosimetry pre- and post-implantation with traditional free-hand implantation. The result showed that all dosimetric parameters including D90, D100, V90, and V100 values showed significant differences between pre- and post-implantation (SMD = 0.87, 0.73, 1.89, and 1.61, respectively; 95%CI = 0.21 to 1.53, 0.18 to 1.28, 1.28 to 2.49, and 1.20 to 2.02 respectively; p = 0.010, 0.010, < 0.001, and < 0.001, respectively)(as shown in Fig. 4). Five studies [20,21,22,23,24] reported dosimetry pre- and post-implantation with 3DPNCT. The result showed that all dosimetric parameters including D90, D100, V100, V150, V200 and D2cc of OARs showed no significant differences between pre- and post-implantation(SMD = 0.11, − 0.26, 0.30, 0.13, − 0.20, and 0.01, respectively; 95%CI = − 0.21 to 0.43, − 0.62 to 0.10, − 0.20 to 0.80, − 0.24 to 0.49, − 0.58 to 0.17, and − 0.20 to 0.21 respectively; p = 0.489, 0.151, 0.243, 0.494, 0.289, and 0.954, respectively) (as shown in Figs. 5 and 6). In three studies [25,26,27], both 3DPNCT and 3DPCT plans were designed for all patients. Pooled analysis of D90, D100, V100, D2cc of OAR, number of seeds and number of needles showed no significant differences between 3DPNCT and 3DPCT groups (p = 0.930, 0.215, 0.766, 0.863, 0.904, and 0.575, respectively). V150, V200 increased (SMD = 0.35, 0.49; 95%CI = 0.04 to 0.67, 0.02 to 0.96; p = 0.028, 0.043) and number of through bone needles decreased (SMD = − 1.03, 95%CI = − 1.43 to − 0.64, p < 0.001) with 3DPNCT (as shown in Figs. 7, 8 and 9).

Fig. 2
figure 2

Forest plot of studies comparing post-implantation D90 and D100 between 3D print template and traditional free-hand implantation

Fig. 3
figure 3

Forest plot of studies comparing post-implantation V90 and V100 between 3D print template and traditional free-hand implantation

Fig. 4
figure 4

Forest plot of studies comparing D90, D100, V90, and V100 between pre- and post-implantation with traditional free-hand implantation

Fig. 5
figure 5

Forest plot of studies comparing D90, D100, V100, V150, and V200 between pre- and post-implantation with 3D print template

Fig. 6
figure 6

Forest plot of studies comparing D2cc of OARs between pre- and post-implantation with 3D print template

Fig. 7
figure 7

Forest plot of studies comparing D90, D100, V100, V150 and V200 between 3DPNCT and 3DPCT groups

Fig. 8
figure 8

Forest plot of studies comparing D2cc of OAR between 3DPNCT and 3DPCT groups

Fig. 9
figure 9

Forest plot of studies comparing number of seeds, number of needles and through bone needles between 3DPNCT and 3DPCT groups

Operation time

Three studies reported operation time. The heterogeneity result showed no significant heterogeneity (I2 = 35.4%, p = 0.213). The fixed effect model was utilized. The result showed that 3D print template assisted RIS implantation could reduce operation time with statistically significant compared to taditional free-hand implantation (SMD = − 0.93; 95%CI = − 1.3 to − 0.51; p < 0.001) (as shown in Additional file 1: Fig. 1).

Publication bias

For studies comparing 3D print template-assisted implantation with traditional free-hand implantation, publication bias was evaluated by a funnel plot. The regression test of the funnel plot symmetry confirmed that no publication bias was found (p = 0.999, 0.373, 0.903, 0.833, and 0.964, respectively) for D90, D100, V90, V100, and operation time (as shown in Additional file 2: Fig. 2, Additional file 3: Fig. 3).

Discussion

RIS implantation has the dosimetric advantage of sharp dose gradients between tumor target area and adjacent normal tissues, which allows more sparing of the surrounding organs at risk. It has been widely used in China for the following tumors: head and neck, thorax, breast, abdomen, and pelvic cavity; and it has a good effect on relieving pain, reducing tumor burden, improving life quality and prolonging survival time of patients [28,29,30,31,32]. But except for prostate cancer, there is still no standard method to treat other tumors. In the past, 125I seeds were implanted just by doctors’ experience. However, it is challenging for doctors to insert many needles at 1 time into the targer in line with the preplan, which leads to a large difference in location of seeds and dose distribution between pre- and post- plan, finally leading to tumor local recurrence and complications [12,13,14,15,16,17,18,19]. Also, traditional free-hand 125I seed implantation is very complicated and time consuming; the operators usually spend a lot of time learning these special skills [16, 17]. Therefore, how to achieve a precise distribution of RIS in the tumor target area and to ensure that RIS implantation treatment is strictly followed up as preplanning are the research focuses.

3D print template is a personalized template which contains preset implant channel information and body surface information of patients' treatment area. Through accurate intraoperative reset of template and real-time planning of TPS, the consistency between actual implant channel and planned needle channel can be improved. With the assistence of a 3D print template, it is easy and efficient to insert needles at any arbitrary angle into the tumor target, meanwhile accurately reproducing the needle positions according to the preplan. This method has been shown to improve the accuracy of seed location and dose distribution in many studies [9,10,11,12,13], but as far as we know, there is still a lack of evidence-based medical data regarding its effectiveness.

Our meta-analysis result firstly showed that there was no significant difference between pre- and post-implantation for all the parameters including D90, D100, V100, V150, V200 and D2cc of OARs with 3D print template and that the D90, D100, V90 and V100 in the template group were higher than those in the freehand group, indicating that 3D print template could provide good accuracy for RIS implantation. And through the template guidance, the operation time was also reduced. Liang et al. [7] treated 15 patients with cervical lymph node metastasis by 3D print template assisted RIS implantation, and found that the dose distribution in preplan can be achieved easily and satisfactorily by 3D print template. Zhang et al. [8] used 3D print template to assist RIS implantation in 14 patients, and the differences of D90, V90, V100 and V150 values pre- and post-implantation showed no statistically significant. Zhang et al. [12] analyzed the clinical data of 27 patients with RIS implantation (13 patients with template, 14 with freehand); the result showed that V90 (92.76% ± 1.89%) in the template group was significantly higher than that in the freehand group (84.59% ± 7.56%), the difference was statistically significant (p = 0.001). Huang et al. [33] treated 25 patients with head and neck tumors by RIS implantation with the guidance of 3D print template. According to the insertion site of the needle, the patients were divided into four groups: parotid gland and masseter area group (9 cases); maxillary and paranasal area group (8 cases); submandibular and upper neck area group (5 cases); posterior area group (6 cases). All the needles were inserted at their predetermined positions once. The average insertion time was 7.5 s for each needle, and no complications were observed. These studies all have indicated that 3D print template could not only improve postoperative dose distribution but also lower the difficulty of puncture and reduce the operation time.

In this meta-analysis, we also found that both 3DPNCT and 3DPCT plans could achieve prescription dose. Pooled analysis of D90, D100, V100, and D2cc of OAR showed no significant differences between 3DPNCT and 3DPCT groups. But compared with 3DPCT, in 3DPNCT group, both V150 and V200 were increased, indicating that 3DPNCT could increase the volume of high dose within tumor target. And the larger volume of high dose might produce more beneficial effects on local control. On the other hand, the number of through bone needles was reduced in 3DPNCT group, which showed that 3DPNCT was more safer in the respect of puncture route. However, there are several challenges in the broader use of 3D print template, which include good preoperative implantation designing, accurate 3D print template calibration, and the confidence of doctors in carrying out the procedures. The template is advantageous in challenging clinical cases, which include tumors close to important organs and tissues, blocked by the bones, or next to the important blood vessels. On the other hand, it is time-consuming to design and prepare 3D print template, though the needle could be inserted into the target more accurately and efficiently, leading to better dose consistency.

In this study, the advantages of 3D print template in RIS implantation were revealed by evidence-based medicine study for the first time. Our meta-analysis result showed that both 3DPNCT and 3DPCT assisted RIS implantation can realize the accurate distribution of RIS, and make the post implantation dosimetry more predictable, which provided evidences for clinical practice. However, there are shortcomings in our study: as the relevant studies included were all from China, it may lead to the selection bias of the literature. Also, we used SMD as the effect indicator for dosimetry and operation time to perform meta-analysis due to the differences in preplan among studies. The advantage of 3D print template in clinical efficacy such as local tumor control and long-term survival of patients still needs a larger sample and high-quality randomized controlled trial to verify in the future.

Conclusion

Our meta-analysis result showed that 3D print template assisted RIS implantation can realize the accurate distribution of RIS, optimize dose distribution and reduce the operation time at the same time. Compared with 3D print coplanar template, 3D print noncoplanar template could increase the volume of high dose within tumor target and was more safer in the respect of puncture route, which provided evidences for clinical practice.

Availability of data and materials

The datasets analysed for this study are available from J.W. on reasonable request.

Abbreviations

3D:

Three-dimensional

3DPNCT:

3D print noncoplanar template

3DPCT:

3D print coplanar template

CNKI:

China National Knowledge Infrastructure

CI:

Confidence interval

SMD:

Standard mean difference

RIS:

Radioactive iodine-125 seeds

PRISMA:

Preferred Reporting Items for Systematic Reviews and Meta-Analyses

RCTs:

Randomized controlled trials

D90:

The dose of 90% of the target volume

D100:

The dose of 10% of the target volume

V90:

The percent of the tumor target receiving 90% of the prescribed dose

V100:

The percent of the tumor target receiving 100% of the prescribed dose

V150:

The percent of the tumor target receiving150% of the prescribed dose

V200:

The percent of the tumor target receiving 200% of the prescribed dose

D2cc:

The dose received by 2 cm3 of normal tissue

References

  1. Wang L, Wang H, Jiang Y, et al. The efficacy and dosimetry analysis of CT-guided 125I seed implantation assisted with 3D-printing non-co-planar template in locally recurrent rectal cancer. Radiat Oncol. 2020;15:179.

    Article  CAS  Google Scholar 

  2. Huo XD, Wang HX, Yang JK, et al. Effectiveness and safety of CT-guided (125) I seed brachytherapy for postoperative locoregional recurrence in patients with non-small cell lung cancer. Brachytherapy. 2016;15:370–80.

    Article  Google Scholar 

  3. Martinez A, Cox RS, Edmundson GK. A multiple-site perineal applicator (MUPIT) for treatment of prostatic, anorectal, and gynecologic malignancies. Int J Radiat Oncol Biol Phys. 1984;10:297–305.

    Article  CAS  Google Scholar 

  4. Aristei C, Lancellotta V, Piergentini M, et al. Individualized 3D-printed templates for high-dose-rate interstitial multicathether brachytherapy in patients with breast cancer. Brachytherapy. 2019;18:57–62.

    Article  Google Scholar 

  5. Mahantshetty U, Shrivastava S, Kalyani N, et al. Template-based high-dose-rate interstitial brachytherapy in gynecologic cancers: a single institutional experience. Brachytherapy. 2014;13:337–42.

    Article  Google Scholar 

  6. Ji Z, Jiang YL, Guo FX, et al. Dosimetry verification of radioactive seed implantation for malignant tumors assisted by 3D printing individual templates and CT guidance. Appl Radiat Isot. 2017;124:68–74.

    Article  CAS  Google Scholar 

  7. Liang Y, Wang Z, Zhang H, et al. Three-dimensional-printed individual template-guided 125I seed implantation for the cervical lymph node metastasis: a dosimetric and security study. J Cancer Res Ther. 2018;14:30.

    Article  Google Scholar 

  8. Zhang H, Dev D, Yu H, et al. Feasibility of three-dimensional-printed template-guided 125I seed brachytherapy and dosimetric evaluation in patients with malignant tumor. J Cancer Res Ther. 2019;15:793–800.

    Article  Google Scholar 

  9. Moher D, Liberati A, Tetzlaff J, et al. Preferred reporting items for systematic reviews and meta-analyses: The PRISMA statement. Int J Surg. 2011;8:336–41.

    Article  Google Scholar 

  10. Wells GA, Shea B, O’Connell D. The Newcastle-Ottawa Scale(NOS) for assessing the quality of nonrandomised studies in meta-analyses. 2016. Available at: http://www.ohri.ca/programs/clinical_epidemiology/nos_manual.pdf.

  11. Mantel N, Haenszel W. Statistical aspects of the analysis of data from retrospective studies of disease. J Natl Cancer Inst. 1959;22:719–48.

    CAS  PubMed  Google Scholar 

  12. Hongtao Z, Xuemin D, Huimin Y, et al. Dosimetry study of three dimensional print template guided precision 125 I seed implantation. J Cancer Res Ther. 2016;12:159–65.

    Article  Google Scholar 

  13. Han T, Yang X, Xu Y, et al. Therapeutic value of 3-D printing template-assisted 125I-seed implantation in the treatment of malignant liver tumors. Onco Targets Ther. 2017;10:3277–83.

    Article  Google Scholar 

  14. Huang W, Lu J, Chen KM, et al. Preliminary application of 3D-printed coplanar template for iodine-125 seed implantation therapy in patients with advanced pancreatic cancer. World J Gastroenterol. 2018;24:96–103.

    Article  Google Scholar 

  15. Cao Q, Huo B, Huo XD, et al. Dosimetric study of 3D-printing coplanar template combined with CT-guided 125I seed implantation for treating non-smaIl cell lung cancer. Chin J Radiol Med Prot. 2017;37:528–32.

    Google Scholar 

  16. Pan TF. Clinical research of 3D printing template-assisted 125I radioactive seed interstitial brachytherapy for malignant tumors. Southeast University 2018:1–40.

  17. Shen YX. Clinical Study of 3D Printing individualized template Assisted 125I Particles Implantation in the Therapy of Head and Neck. Zhengzhou University 2018:1–47.

  18. Zheng YF, Gong J, Xia N. Clinical application of 3D printing coplanar template assist CT-guided radioactive seed implantation for mediastinal lymph node metastasis. Chin Comput Med Imag. 2019;25:79–84.

    Google Scholar 

  19. Di XM. Clinical research on 125I seed implantation in treatment of para-aortic lymph node metastasis. Hebei Medical University 2017:1–44.

  20. Ji Z, Jiang Y, Su L, et al. Dosimetry verification of 125I seeds implantation with three-dimensional printing noncoplanar templates and CT guidance for paravertebral/retroperitoneal malignant tumors. Technol Cancer Res Treat. 2017;16(6):1044–50.

    Article  CAS  Google Scholar 

  21. Hao W, Junjie W, Yuliang J, et al. CT guidance 125I seed implantation for pelvic reeurrent rectal cancer assisted by 3D printing individual non-coplanar template. Natl Med J Chin. 2016;96(47):3782–6.

    Google Scholar 

  22. Yuliang J, Hao W, Zhe Ji, et al. Computed tomography image-guided and personalized 3D printed template-assisted 125-iodine seed implantation for recurrent pelvic tumor: a dosimetric study. Chin J Radiat Oncol. 2016;25(9):959–64.

    Google Scholar 

  23. Jiang P, Guo F, Jiang Y, et al. 3D-printing non-coplanar template assisted CT-guided 125I seed implantation on pelvic recurrent cervical cancer. Chin J Radiol Med Prot. 2017;37(7):490–4.

    Google Scholar 

  24. Zhe J, Yuliang J, Fuxin G, et al. Dosimetric assessment of CT-guided radioactive seed implantation assisted by 3D printing non-coplanar template in treatment of chest malignant tumor. Chin J Radiat Oncol. 2017;26(7):754–8.

    Google Scholar 

  25. Ji Z, Sun H, Jiang Y, et al. Comparative study for CT-guided 125I seed implantation assisted by 3D printing coplanar and non-coplanar template in peripheral lung cancer. J Contemp Brachytherapy. 2019;11(2):169–73.

    Article  Google Scholar 

  26. Ang Q, Junjie W, Yuliang J, et al. Comparison of preoperative planning of radioactive seed implantation for pelvic wall recurrent gynecological malignant tumors between 3D-printing non-coplanar template and 3D-printing coplanar template. Natl Med J Chin. 2019;99(11):841–3.

    Google Scholar 

  27. Xuemin L, Ran P, Yuliang J, et al. Dosimetric analysis of 3D-printing template assisted and CT-guided 125I seed implantation for treatment of soft tissue sarcoma. Chin J Radiol Med Prot. 2018;38(5):350–4.

    Google Scholar 

  28. Li Q, Tian Y, Yang D, et al. Permanent iodine-125 seed implantation for the treatment of nonresectable retroperitoneal malignant tumors. Technol Cancer Res Treat. 2019. https://doi.org/10.1177/1533033819825845.

    Article  PubMed  PubMed Central  Google Scholar 

  29. Qu A, Jiang P, Sun H, et al. Efficacy and dosimetry analysis of image-guided radioactive 125I seed implantation as salvage treatment for pelvic recurrent cervical cancer after external beam radiotherapy. J Gynecol Oncol. 2019;30:e9.

    Article  CAS  Google Scholar 

  30. Yu YH, Wei CY, Qin QH, et al. Efficacy of iodine-125 seed implantation in locoregionally recurrent and unresectable breast cancer: a retrospective study. Pathol Oncol Res. 2019;25:327–32.

    Article  CAS  Google Scholar 

  31. Huang MW, Liu SM, Zheng L, et al. A digital model individual template and CT guided 125I seed implants for malignant tumors of the head and neck. J Radiat Res. 2012;53:973–7.

    Article  Google Scholar 

  32. Lin L, Wang J, Jiang Y, et al. Interstitial 125I seed implantation for cervical lymph node recurrence after multimodal treatment of thoracic esophageal squamous cell carcinoma. Technol Cancer Res Treat. 2015;14:201–7.

    Article  CAS  Google Scholar 

  33. Huang MW, Zhang JG, Zheng L, et al. Accuracy evaluation of a 3D-printed individual template for needle guidance in head and neck brachytherapy. J Radiat Res. 2016;57:662–7.

    Article  CAS  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

Not applicable.

Author information

Authors and Affiliations

Authors

Contributions

JW, EC, YZ, HZ, CJ, and YL are the study sponsors. JW and EC designed research process; EC, YZ and HZ screened and reviewed all literature that conforms to the inclusion criteria and complete statistical analysis of the data; EC, YZ, CJ, and YL interpreted the outcomes; JW made the decision to submit the article for publication. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Juan Wang.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1

. Forest plot of studies evaluating operation time.

Additional file 2

. Funnel plot of studies evaluating D90, D100.

Additional file 3

. Funnel plot of studies evaluating V90, V100.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Chen, E., Zhang, Y., Zhang, H. et al. Dosimetry study of three-dimensional print template for 125I implantation therapy. Radiat Oncol 16, 115 (2021). https://doi.org/10.1186/s13014-021-01845-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13014-021-01845-y

Keywords