Study on the ability of 3D gamma analysis and bio-mathematical model in detecting dose changes caused by dose-calculation-grid-size (DCGS)

Abstract

Objective

To explore the efficacy and sensitivity of 3D gamma analysis and bio-mathematical model for cervical cancer in detecting dose changes caused by dose-calculation-grid-size (DCGS).

Methods

17 patients’ plans for cervical cancer were enrolled (Pinnacle TPS, VMAT), and the DCGS was changed from 2.0 mm to 5.0 mm to calculate the planned dose respectively. The dose distribution calculated by DCGS = 2.0 mm as the “reference” data set (RDS), the dose distribution calculated by the rest DCGS as the“measurement”data set (MDS), the 3D gamma passing rates and the (N) TCPs of the all structures under different DCGS were obtained, and then analyze the ability of 3D gamma analysis and (N) TCP model in detecting dose changes and what factors affect this ability.

Results

The effect of DCGS on planned dose was obvious. When the gamma standard was 1.0 mm, 1.0 and 10.0%, the difference of the results of the DCGS on dose-effect could be detected by 3D gamma analysis (all p value < 0.05). With the decline of the standard, 3D gamma analysis’ ability to detect this difference shows weaker. When the standard was 1.0 mm, 3.0 and 10.0%, the p value of > 0.05 accounted for the majority. With DCGS = 2.0 mm being RDS, ∆gamma-passing-rate presented the same trend with ∆(N) TCPs of all structures except for the femurs only when the 1.0 mm, 1.0 and 10.0% standards were adopted for the 3D gamma analysis.

Conclusions

The 3D gamma analysis and bio-mathematical model can be used to analyze the effect of DCGS on the planned dose. For comparison, the former’s detection ability has a lot to do with the designed standard, and the latter’s capability is related to the parameters and calculated accuracy instrinsically.

Introduction

VMAT technology is widely used for the cervical cancer radiotherapy [1,2,3]. These studies on the application have shown that VMAT technology can be used for cervical cancer radiotherapy, and compared with High Dose Rate brachytherapy, VMAT plan achieves significant dose reduction of rectum, bladder and sigmoid, as well as superior homogeneous target coverage compared to brachytherapy plan [4,5,6].

The dose calculation grid size (DCGS) is a basic parameter setting in the design of the plan. Usually, a commercial treatment planning system (TPS) will provide various DCGS within a certain range for designers to choose for different needs. For example, the commercial Pinnacle TPS provides DCGS ranging from 1.0 mm to 10.0 mm, and the default DCGS is 4.0 mm [7]. Larger DCGS is commonly adopted for calculation in cases with larger target volumes and organ-at-risk (OAR) volumes for better calculation efficiency. However, a smaller DCGS should be chosen for dose calculation in radiotherapy for head and neck tumors to obtain precise doses of small-volume OARs including lens, optic nerve and pituitary, etc., especially important for OARs with strict maximumdose limit [8, 9].

The difference in doses caused by different grid sizes may affect the evaluation of the quality of physical solutions, although the DCGS can not cause the actual absorbed dose when the accelerator’s parameters are certain. (This is why the planned dose values and (N) TCP values calculated under different DCGS are all called “calculated” values in the following sections). Therefore, it is a crucial task to understand the effect of DCGS on the physical and biological doses in radiotherapy for cervical cancer. Gamma analysis is currently the most common and generally accepted method for quantitatively assessing the difference between the two dose-distribution (DD) [10, 11]. It detects the difference between the two DD by a designed gamma standard (e.g. 3.0 mm, 3.0, 10%) and it will provide a report on passing rate [12, 13]. The standard of 3.0 mm, 3.0 and 10% is the most widely used, in which 3.0 mm refers to the consistency of distance, 3.0% refers to the maximum allowable dose difference, and 10% is the threshold (In the following parts of the article we skipped it as it never changed) and when the dose is less than 10% of the reference dose, which does not participate in gamma analysis. Selecting 10% is widely recommended [14, 15]. In intensity modulated radiation therapy (IMRT), gamma analysis is usually used to analyze the difference between the TPS-outputed and the actually measured dose distribution to evaluate the degree of dose deviation caused by various reasons during the execution of the plan, further to determine whether a plan is to execute based on the evaluation. However, previous studies have shown that different dose quality assurance (QA) system (dose QA system refers to the collection of measurement and analysis software and hardware used to ensure that the radiation dose is achieved at the target volume and the OARs as expected) have different abilities to detect errors based on the dose distribution output by TPS. Hussein et al. [16] enrolled pelvis and head & neck IMRT and RapidArc™ plans, and compared the differences in the detecting dose error of five commercial products: PTW Verisoft, Delta4 software, SNC Patient, Varian Portal Dosimetry and IBA OmniPro. The results showed that for the same pass-rate criteria, different devices and software combinations exhibited varying levels of agreement with the predicted gamma analysis. On the other hand, different gamma analysis standards will get different passing rates. Research by Heilemann et al. [17] showed that the 3.0 mm, 3.0% standards were not sufficient to detect the deviation caused by the MLC position uncertainty, and this standard, at least, has to be 2.0 mm, 2.0%.

The focus on the variation of planned dose (PD) is due to the fact that it could cause changes in the biological effects. Specifically, in the clinical practice of radiotherapy, the alteration in physical dose will bring about changes in tumor control probability (TCP) and nomal tissue complication probability (NTCP). Therefore, the current project, under the condition of DCGS changes, used the dose distribution calculated by DCGS = 2.0 mm as the reference data set (RDS) to explore the efficacy and sensitivity of the 3D gamma analysis and the bio-mathematical model on dose change detection by analyzing the 3D gamma passing rates of all structures and the relationship between gamma passing rate and (N)TCP.

Materials and methods

Patient meterials

A retrospective study was performed on the physical plan of 17 patients with cervical cancer who were treated in the Department of Radiation Therapy of our hospital from December 2017 to November 2018. And the 17 patients’ plans were intially designed and evaluated with DCGS = 4.0 mm. The patient’s planning gross target volume (PGTV) volume was 20.0–395.0cm3, and the planning target volume (PTV) volume was 880.0–2587.0cm3. The average volumes of the two target volume were 128.9 ± 110.2 cm3, 1752.9 ± 460.1 cm3, respectively. The rectum’s mean volume was 59.3 ± 25.4 cm3, and the bladder’s was 257.5 ± 165.6 cm, and the L-femur’ s and R-femur’s were 107.2 ± 19.1 cm3, 108.2 ± 19.6 cm3. These patients were in a supine position with both hands surrounding their heads, and the patient was fixed with a thermoplastic mesh. The Siemens Somatom Sensation Open 24 CT (Siemens Co., Munich, Germany) was used as the data acquisition system. The range of scanning was from the head of the diaphragm to lower 1.0 cm of the bottom pubic symphysis. And the CT data of each patient was reconstructed with a 3.0 mm layer thickness, was transmitted to Pinnacle TPS 9.10.

Design of volumetric modulated arc therapy (VMAT) radiotherapy plan

17 patients were treated with a Versa HD linear accelerator (Elekta Medical Systems Co., Stockholm, Sweden) of 6 MV photon beams. The VMAT plan of a 360o full bow with 2 arcs was designed for every patient based on Smart Arc inverse optimization. The objective functions were shown in Table 1. The doses were calculated with the Collapsed Cone Convolution (CCC) algorithm [18]. The planning prescription setting was as follows: the PTV prescription being 45.0–50.0Gy/25 fractions, and the PGTV prescription being 60.0–62.5Gy/25 fractions.

Table 1 Dose-volume criteria used in the cervix cancer VMAT plans

All VMAT physical schemes were designed with Pinnacle TPS (version 9.10). When the default value was DCGS = 4.0 mm, the planners optimized and adjusted the treatment plans for cervical cancer patients based on their own previous experience. After all the indicators of the plans met the clinical requirements, changed the DCGS (from 2.0 mm to 5.0 mm) and recalculated dose in the target volumes and OARs.

D gamma standard and passing rate

After the emergence of IMRT technology, verification of radiation dose before the implementation of treatment has become a very important part of the radiotherapy process. Dose verification can be divided into point dosimetry verification, plane dosimetry verification and gel dosimetry verification. The point dosimetry verification and Gel dosimetry verification [19, 20] have not been widely accepted because of various reasons, and the plane dosimetry verification has become a popular method. The commercially available PTW-ARRAYs [21] and IAB-ARRAYs [22] are the most popular tools for plane dose verification. These ARRAYs only respond correctly to beams perpendicular to their matrix plane, and it is necessary to combine the scattered beams into one direction when using these ARRAYs. This feature makes these ARRAYs less suitable for dose verification of VMAT with rotating beams. So, Delta 4, ArcCheck and Octavius 4D for dose verification of VMAT are more common in daily work [23,24,25].

Although the tools advanced, there has been no fundamental change in analytical methods. The gamma analysis has been used throughout IMRT dose verification. The commonly recommended gamma analysis standards are 3.0 mm, 3.0% [14, 15], but studies have shown that, depending on the technology and the disease, we should adopt stricter standards or other supplement analysis to analyze errors [26, 27].

The 3D gamma analysis is also a standard gamma analysis. The 3D gamma analysis is that during the dose of reconstruction before the analysis of gamma analysis, the software will be used the planned dosed Perturbation (PDP) algorithm corresponding the hardware. For example, if the hardware is ArcCHECK, the corresponding algorithm is ArcCHECK-planned-dosed-perturbation (AC-PDP) [28].

Although both are gamma analysis, there are differences between 2D gamma analysis and 3D gamma analysis in the range of detection analysis (2D dose is extracted from the 3D dose at the isocenter position) and sensitivity of the error detection [15].

This article was to investigate errors among planned dose caused by DCGS with dose distribution calculated by DCGS = 2.0 mm as RDS and the dose distribution calculated by DCGS = 3.0 mm, 4.0 mm, 5.0 mm as MDS, respectively. And we thought the location uncertainty was scarcely influential element under the situation. Therefore, when setting the gamma analysis standard, we setted the following 3 standards: 1.0 mm, 1.0%; 1.0 mm, 2.0%; 1.0 mm, 3.0%.

TCP and NTCP calculation

The link between physical dose (change) and biological effect (change) has always been our focus. Changes in DCGS will certainly bring about changes in the physical dose, as well as the biological effects. It is well known that changes in biological effects have more direct clinical significance [29], so this study ignored the physical dose and directly calculated the changes in biological effects caused by changes in DCGS by using biological model.

Some biologically related models for plan optimization and/or evaluation have been introduced into treatment planning tools for clinical use. A variety of dose- response models with a series of organ-specific model parameters were reported in the literatures, and were widely accepted as follows [30, 31]:

$$ \mathrm{TCP}=\frac{1}{1+{\left(\frac{{\mathrm{TCD}}_{50}}{\mathrm{EUD}}\right)}^{4\gamma 50}} $$
(1)
$$ \mathrm{NTCP}=1/\sqrt{2\pi }{\int}_{-\infty}^u{e}^{-{t}^{\raisebox{1ex}{$2$}\!\left/ \!\raisebox{-1ex}{$2$}\right.}} dt $$
(2.1)
$$ u=\left(\mathrm{EUD}\hbox{-} {\mathrm{TD}}_{50}\right)/\left({\mathrm{mTD}}_{50}\right) $$
(2.2)
$$ {\mathrm{TD}}_{50}(v)={\mathrm{TD}}_{50}(1)\ast {v}^{-n} $$
(2.3)
$$ \mathrm{EUD}={\left({\sum}_i{v}_i{\mathrm{D}}_i^a\right)}^{1/a} $$
(2.4)

Where, a is an organ-specific constant, and its corresponding value is in the literatures [32, 33]. vi is the fractional volume of the organ receiving Di. m and n are unique organ-specific constants [32, 33]. TD50 is an uniform dose that is absorbed dose at a 50% complication probability, and TCD50 is an uniform dose that is absorbed dose at a 50% control probability.

Statistical analysis

Origin 8.0 was used for drawing and SPSS 20.0 was used for statistical analysis. Paired t test was used for statistical analysis, and p > 0.05 indicates no significant difference, and 0.01 < p < 0.05 indicates significant difference, and p < 0.01 indicates very significant difference.

Results

(N) TCP and absolute dose (AD) changes with DCGS

The TPS software takes the cube formed by DCGS as the unit to collect the dose, and then generates the DVH. DCGS’s change indicated the change of “collection-dose-unit”. Key dose points on DVH of each structure with DCGS’s change were counted and analysed. As we all know, for target, minimum dose (MinD), maximum dose (MaxD) and mean dose (MeanD) were important for efficacy [34]. And for parallel organ-at-risk, MeanD was important for toxicity [35]. So, We counted on PGTV’s MinD, MaxD and MeanD, and PTV’s MinD and MeanD (because PGTV is included in PTV, the maximum dose in PTV is located in PGTV), and MeanD of Bladder, Rectum and Femurs (shown in Figs. 1, 2, 3 and 4).

Fig. 1
figure 1

Display of MinD, MaxD and MeanD of PGTV change casused by DCGS

Fig. 2
figure 2

Display of MinD and MeanD of PTV change casused by DCGS

Fig. 3
figure 3

Display of MeanDs of Rectum and Bladder change casused by DCGS

Fig. 4
figure 4

Display of MeanDs of R-Femur and L-Femur change casused by DCGS

The 68 differential DVH of 17 patients’ radiotherapy plans (17*4 = 68) were exported as .txt data files. The data in the .txt file were read by MATLAB program and put into formulas (1) and (2.4) to calculate the EUD of each OAR and each target. Then TCP and NTCP were calculated using the formulas (1, 2.1). Figures drawn by Origin8.0 software were shown in Fig. 5. and Fig. 6. As shown in the figures, the influence of DCGS on the calculated value of (N) TCP was obvious, and the calculated value of (N) TCP decreased with the increase of DCGS.

Fig. 5
figure 5

Display of TCP change casused by DCGS

Fig. 6
figure 6

Display of NTCP change casused by DCGS

p-value analysis

The change of DCGS would bring about the change of each structure’s absorption dose. In order to quantify the dose change, we selected gamma analysis to analyze the difference between the dose corresponding to DCGS = 2.0 mm acting as RDS and the dose corresponding to DCGS = 3.0 mm, 4.0 mm, 5.0 mm acting as MDS, respectively.

The paired t-test results of each structure were shown in Fig. 7. The gamma-standard had a significant impact on the gamma-analysis’ sensitivity. When the gamma-standard was 1.0 mm, 1.0%, the difference of the results of the DCGS on dose-effect could be detected by 3Dgamma analysis (all p value< 0.05). With the decline of the standard, 3Dgamma analysis’ ability to detect this difference was also declining. When the standard was 1.0 mm, 3.0%, the p value of > 0.05 accounted for the majority. It was a high probability event that the dose difference between DCGS = 5.0 mm and DCGS = 3.0 mm (or the other two DCGSs) could not be detected by this analysis.

Fig. 7
figure 7

The map of every structure’s paired t test under varied gamma analysis standard. When the gamma analysis standard setting was 1.0 mm, 1.0% and the dose distributions of DCGS = 3.0, 4.0, 5.0 mm were compared with DCGS = 2.0 mm respectively and the results were grouped as Aa, Ba, Ca. When the gamma analysis standard setting was 1.0 mm, 2.0% and the dose distributions of DCGS = 3.0, 4.0, 5.0 mm were compared with DCGS = 2.0 mm respectively and the results were grouped as Ab, Bb, Cb. When the gamma analysis standard setting was 1.0 mm, 3.0% and the dose distributions of DCGS = 3.0, 4.0, 5.0 mm were compared with DCGS = 2.0 mm respectively and the results were grouped as Ac, Bc, Cc

Correlation of ∆(N) TCP and ∆gamma

The (N)TCPs’ and gamma values both changed with DCGS. To explore whether there was a certain correlation between these two changes, we investigated (N) TCP andgamma separately. (N) TCP was defined as the (N) TCP value when DCGS = 2.0 mm minus the (N) TCP value when DCGS was the other value. gamma was defined as 100.0% minus the gamma passing rate when the dose distribution of DCGS = 3.0 mm, or 4.0 mm, or 5.0 mm compared that of DCGS = 2.0 mm with gamma analysis. Because gamma analysis was carried out with three different standards in this paper, the gamma were divided into gamma1, gamma2 and gamma3 corresponding to 1.0 mm, 1.0%; 1.0 mm, 2.0% and 1.0 mm, 3.0%.

In order to simplify the following writing, we describe the corresponding relationship as follows.

“M” corresponding to ∆(N) TCP = (N) TCP (DCGS = 2.0 mm) -(N) TCP (DCGS = 3.0 mm) and.

∆gamma = 100.0% - gamma passing rate when DCGS = 3.0 mm compared DCGS = 2.0 mm.

“N”corresponding to ∆(N)TCP = (N) TCP (DCGS = 2.0 mm) -(N) TCP (DCGS = 4.0 mm) and.

∆gamma = 100.0% - gamma passing rate when DCGS = 4.0 mm compared DCGS = 2.0 mm.

“L”corresponding to ∆ (N)TCP = (N) TCP (DCGS = 2.0 mm) -(N) TCP (DCGS = 5.0 mm) and.

∆gamma = 100.0% - gamma passing rate when DCGS = 5.0 mm compared DCGS = 2.0 mm.

The ∆(N) TCP and ∆gamma were shown in Table 2, Table 3 and Table 4. As shown in the tables, when the calculated values of (N) TCP of the targets, the bladder and the rectum decreased with DCGS increasing, the gamma passing rate also decreased when the standard was 1.0 mm, 1.0%.

Table 2 Display of TCP and γ
Table 3 Display of NTCP and γ for the bladder and the rectum
Table 4 Display of NTCP and γ for the femurs

Discussion and conclusion

Compared with conformal radiotherapy, intensity-modulated radiotherapy can improve the conformal degree of the target area, reduce the dose of organs at risk, and reduce the acute and late toxicity of organs [36, 37]. VMAT is a higher form of modulated radiation therapy, VMAT and IMRT have been compared in many studies [38, 39]. The publications relating to planning [40], commissioning [13], QA [41] and clinical implementation [42] have been published, which made VMAT technology spread quikly around the world. Therefore, the cervical cancer’s VMAT plans were selected as the research object in this study. The gantry and dose rate change simultaneously when a VMAT plan was executed, which makes VMAT’s QA are more complicated than IMRT [43]. It was therefore difficult to define whether the failure of a VMAT QA was possibly caused by dose calculation from TPS, dose delivery from linac, detectors of phatom, or other aspects.

It is an important basis for us to set up DCGS in the planning design to consider the efficiency of calculation under the precise of satisfying the accuracy of dose calculation. Many studies from radiation oncology deparments recommend DCGS = 2.0 mm be for the clinical requirements [44, 45]. This conclusion was the main reason why in this paper we chose DCGS = 2.0 mm as the research basis. Secondly, the low computational efficiency of DCGS = 1.0 mm makes it difficult in clinical practice, when was DCGS = 1.0 mm, the Pinnacle TPS would spend about two 2.0 h to calculate once a dose distribution for a patient.

Gamma analysis is a commonly used method to compare differences between two dose distributions, but the ability of gamma analysis to detect errors is closely related to the criteria set. Figure 7 of the “p-value analysis” in this paper showed that when 1.0 mm, 1.0% was used as the standard in the gamma analysis, there was a statistical difference (all p value< 0.05) between any two results of gamma analysis when DGCS was 3.0 mm, 4.0 mm or 5.0 mmvs 2.0 mm, indicating that the gamma analysis was sensitive to changes of DCGS. However, when 1.0 mm, 3.0% was used as the standard in gamma analysis, there was mostly no significant difference (all p value> 0.05) between any two results of gamma analysis when DGCS was 3.0 mm, 4.0 mm or 5.0 mmvs 2.0 mm, and at this time the gamma analysis was not sensitive to changes of DCGS. Many studies’ results all reflected the similar conclusion of gamma-standard and gamma analysis’sensitivity [46, 47]. The difference was that these studies had set change of the standard in two dimensions (distance and dose) at the same time, for example, change the gamma3.0%, 3.0mm to gamma2.0%, 2.0mm. When the dose standard of gamma analysis was relaxed from 1.0 to 3.0%, and those “points” that fail to pass the 1.0 mm, 1.0% standard were passed under the 1.0 mm, 3.0% standard, which meaned that the new standard lost the ability to detect dose errors in the range of 1.0–3.0%. At the same time, we could also get from the results that the majority of the dose “calculated value” changes as DCGS from 2.0 mm to 5.0 mm were < 3.0%.

However, the situation was different when the (N) TCP biological mathematical model was used to detect these dose changes, and theoretically any dose change caused by DCGS could be represented in the value of (N) TCP as long as the value of (N) TCP was accurate enough. So, when we changed DCGS we got the trend of (N) TCP and ∆ (N) TCP, even though ∆(N) TCP was not a big value. In the study, ∆NTCP of the Femurs was always 0.00 in Table 4. It was not because NTCP did not change, but because the value was omitted because of too small, which was cause by parameters’ value of NTCP model.

The TCP and NTCP were expected to be obtained by studying the targets’ and OARs’ physical dose, because the two formers were of greater clinical significance. So this paper studied the relationship between ∆(N) TCP and ∆gamma. With the DCGS becoming larger, the relatively-lower-dose in the normal tissues located around the target was more calculated into the target, so the overall dose of the target decreased and the TCP decreased. As shown in Fig. 8, in most cases, DM (absorbed dose at M point) > DN (absorbed dose at N point) > DL (absorbed dose at L point), When DCGS was small (corresponding to the green box), it collected DM and DN. As.

Fig. 8
figure 8

Schematic diagram of dose collection by different DCGS

the DCGS became larger (corresponding to the purple box), it collected a lower dose of DL than the DN dose. The focused irradiation mode of radiotherapy determines the general trend of dose decrease from the target area to the periphery, with the highest dose in the target. In this paper, we investigated the OARs, bladder, rectum, and femurs were the organs of adjacent to the targets, as the DCGS became larger, their relatively-high-dose was “deprived” by the targets, and the dose “deprived” by them from their surrounding was relatively-high-dose, so the DCGS became larger, their overall doses were falling, and their NTCPs were falling.

Some previous studies reported the relationship between DCGS and gamma passing rate, and the results showed that DCGS = 1.0–4.0 mm could meet most clinical needs [48, 49]. However, from the perspective of plan evaluation in this paper, large DCGS made us make a “low evaluation” for the patient’s radiation dose, and the patient’s actual radiation dose was larger than the dose we “saw”.

The 3Dgamma analysis and bio-mathematical model can be used to analyze the effect of DCGS on the planned dose, and the former’s detection ability has a lot to do with the designed standard, and the latter’s capability is related to the parameters and calculated accuracy of the latter.

Availability of data and materials

The datasets generated and/or analysed during the current study are not publicly available due to the General Data Protection Regulation (GDPR) but are available from the corresponding author on reasonable request.

Abbreviations

DCGS:

Dose calculation grid size

PD:

Planned dose

DVH:

Dose-volume histogram

RDS:

Reference data set

MDS:

Measurement data set

TCP:

Tumor control probability

NTCP:

Normal tissue complication probability

TPS:

Treatment planning system

OAR:

Organ-at-risk

DD:

Dose distribution

IMRT:

Intensity modulated radiation therapy

VMAT:

Volumetric modulated arc therapy

PTV:

Planning target volume

PGTV:

Planning gross tumor volume

QA:

Quality assurance

References

  1. Yoshihiro U, Shingo O, Masaru I, et al. Strategies for reducing ovarian dose in volumetric modulated arc therapy (VMAT) for postoperative uterine cervical cancer [J]. Br J Radiol. 2018;91(1081):20160777.

    PubMed  Google Scholar 

  2. Sharfo, AW ; Breedveld, S ; Voet, PW ; et al. Validation of Fully Automated VMAT Plan Generation for Library-Based Plan-of-the-Day Cervical Cancer Radiotherapy [J].PLoS One, 2016, 11(12) :e0169202.

  3. Sharfo AW, Voet PW, Breedveld S, et al. Comparison of VMAT and IMRT strategies for cervical cancer patients using automated planning [J]. Radiother Oncol. 2015;114(3):395–401.

    PubMed  Google Scholar 

  4. Tinoco M, Waga E, Tran K, et al. RapidPlan development of VMAT plans for cervical cancer patients in low- and middle-income countries [J]. Med Dosim. 2020;45(2):172–8.

    PubMed  Google Scholar 

  5. Wali L, Helal A, Darwesh R, et al. A dosimetric comparison of volumetric modulated arc therapy (VMAT) and high dose rate (HDR) brachytherapy in localized cervical cancer radiotherapy [J]. J Xray Sci Technol. 2019;27(3):473–83.

    CAS  PubMed  Google Scholar 

  6. Wang Y, Chen L, Zhu F, et al. A study of minimum segment width parameter on VMAT plan quality, delivery accuracy, and efficiency for cervical cancer using Monaco TPS [J]. J Appl Clin Med Phys. 2018 Sep;19(5):609–15.

    CAS  PubMed  PubMed Central  Google Scholar 

  7. pinnacle3 classic plan (user manual) [M], 2012, 183.

  8. Park JM, Park SY, Kim JI, et al. The influence of the dose calculation resolution of VMAT plans on the calculated dose for eye lens and optic pathway [J]. Australas Phys Eng Sci Med. 2017;40(1):209–17.

    PubMed  Google Scholar 

  9. SS, Gou; RW, Jiang, QE, Ding, et al. Effects of mesh accuracy on radiotherapy plan for different cancers [J]. Chin J Cancer Prev Treat, 2017, 24(20): 1465–1468.

  10. Lam D, Zhang X, Li H, et al. Predicting gamma passing rates for portal dosimetry-based IMRT QA using machine learning [J]. Med Phys. 2019;46(10):4666–75.

    PubMed  Google Scholar 

  11. Han C, Yu W, Zheng X, et al. Composite QA for intensity-modulated radiation therapy using individual volume-based 3D gamma indices [J]. J Radiat Res. 2018;59(5):669–76.

    CAS  PubMed  PubMed Central  Google Scholar 

  12. Low DA, Harms WB, Mutic S, et al. A technique for the quantitative evaluation of dose distributions [J]. Med Phys. 1998;25(5):656–61.

    CAS  PubMed  Google Scholar 

  13. Ezzell GA, Burmeister JW, Dogan N, et al. IMRT commissioning: multiple institution planning and dosimetry comparisons, a report from AAPM task group 119[J]. Med Phys. 2009;36(11):5359–73.

    PubMed  Google Scholar 

  14. Shiba E, Saito A, Furumi M, et al. Predictive gamma passing rate by dose uncertainty potential accumulation model [J]. Med Phys. 2019;46(2):999–1005.

    CAS  PubMed  Google Scholar 

  15. Zhang D, Wang B, Zhang G, et al. Comparison of 3D and 2D gamma passing rate criteria for detection sensitivity to IMRT delivery errors [J]. J Appl Clin Med Phys. 2018;19(4):230–8.

    PubMed  PubMed Central  Google Scholar 

  16. Hussein M, Rowshanfarzad P, Ebert MA, et al. A comparison of the gamma index analysis in various commercial IMRT/VMAT QA systems [J]. Radiother Oncol. 2013;109(3):370–6.

    PubMed  Google Scholar 

  17. Heilemann G, Poppe B, Laub W. On the sensitivity of common gamma-index evaluation methods to MLC misalignments in Rapidarc quality assurance [J]. Med Phys. 2013;40(3):031702 -1-13.

    CAS  PubMed  Google Scholar 

  18. Zhen H, Hrycushko B, Lee H, et al. Dosimetric comparison of Acuros XB with collapsed cone convolution/superposition and anisotropic analytic algorithm for stereotactic ablative radiotherapy of thoracic spinal metastases [J]. J Appl Clin Med Phys. 2015;16(4):181–92.

    PubMed  PubMed Central  Google Scholar 

  19. De Deene Y. Gel dosimetry for the dose verification of intensity modulated radiotherapy treatments [J]. Z Med Phys. 2002;12(2):77–88.

    PubMed  Google Scholar 

  20. Ibbott GS, Maryanski MJ, Eastman P, et al. Three-dimensional visualization and measurement of conformal dose distributions using magnetic resonance imaging of BANG polymer gel dosimeters [J]. Int J Radiat Oncol Biol Phys. 1997;38(5):1097–103.

    CAS  PubMed  Google Scholar 

  21. Jursinic, PA ; Sharma, R ; Reuter, J. MapCHECK used for rotational IMRT measurements: Step-and-shoot, Tomotherapy, RapidArc [J]. Med Phys, 2010,37(6Part1) :2837–2846.

  22. Varasteh Anvar M, Attili A, Ciocca M, et al. Quality assurance of carbon ion and proton beams: a feasibility study for using the 2D MatriXX detector [J]. Phys Med. 2016 Jun;32(6):831–7.

    CAS  PubMed  Google Scholar 

  23. Bedford JL, Fast MF, Nill S, et al. Effect of MLC tracking latency on conformal volumetric modulated arc therapy (VMAT) plans in 4D stereotactic lung treatment [J]. Radiother Oncol. 2015;117(3):491–5.

    PubMed  PubMed Central  Google Scholar 

  24. Morrison CT, Symons KL, Woodings SJ, et al. Verification of junction dose between VMAT arcs of total body irradiation using a sun Nuclear ArcCHECK phantom [J]. J Appl Clin Med Phys. 2017;18(6):177–82.

    PubMed  PubMed Central  Google Scholar 

  25. Yang B, Geng H, Ding Y, et al. Development of a novel methodology for QA of respiratory-gated and VMAT beam delivery using Octavius 4D phantom [J]. Med Dosim. 2019;44(1):83–90.

    PubMed  Google Scholar 

  26. Alharthi T, Pogson EM, Arumugam S, et al. Pre-treatment verification of lung SBRT VMAT plans with delivery errors: toward a better understanding of the gamma index analysis [J]. Phys Med. 2018;49(5):119–28.

    PubMed  Google Scholar 

  27. Kim JI, Park SY, Kim HJ, et al. The sensitivity of gamma-index method to the positioning errors of high-definition MLC in patient-specific VMAT QA for SBRT [J]. Radiat Oncol. 2014;28(9):167–76.

    Google Scholar 

  28. 3DVH Reference Guide –Complete 3D Patient Dose Analysis [M]. 2018,93–93.

  29. Leibel,SA;Zelefsky,MJ;Kutcher,GJ; et al. The biological basis and clinical application of three-dimensional conformal external beam radiation therapy in carcinoma of the prostate [J]. Semin Oncol, 1994, 21(5):580–597.

  30. Chow JL, Jiang R. Dose-volume and radiobiological dependence on the calculation grid size in prostate VMAT planning [J]. Med Dosim. 2018;43(4):383–9.

    PubMed  Google Scholar 

  31. Gulliford SL, Partridge M, Sydes MR, et al. Parameters for the Lyman Kutcher Burman (LKB) model of Normal tissue complication probability (NTCP) for specific rectal complications observed in clinical practise [J]. Radiother Oncol. 2012;102(3):347–51.

    PubMed  Google Scholar 

  32. Luxton G, Keall PJ. King; CR. A new formula for normal tissue complication probability (NTCP) as a function of equivalent uniform dose (EUD) [J]. Phys Med Biol. 2008;53(1):23–36.

    PubMed  Google Scholar 

  33. Fogliata A, Thompson S, Stravato A, et al. On the gEUD biological optimization objective for organs at risk in photon optimizer of eclipse treatment planning system [J]. J Appl Clin Med Phys. 2018;19(1):106–11.

    PubMed  Google Scholar 

  34. Aras S, İkizceli T, Aktan M. Dosimetric comparison of three-dimensional conformal radiotherapy (3D-CRT) and intensity modulated radiotherapy techniques (IMRT) with radiotherapy dose simulations for left-sided mastectomy patients [J]. Eur J Breast Health. 2019;15(2):85–9.

    PubMed  PubMed Central  Google Scholar 

  35. Johansen S, Cozzi L, Olsen DR. A planning comparison of dose patterns in organs at risk and predicted risk for radiation induced malignancy in the contralateral breast following radiation therapy of primary breast using conventional, IMRT and volumetric modulated arc treatment techniques [J]. Acta Oncol. 2009;48(4):495–503.

    CAS  PubMed  Google Scholar 

  36. Cho KH, Kim JY, Lee SH, et al. Simultaneous integrated boost intensity-modulated radiotherapy in patients with high-grade gliomas [J]. Int J Radiat Oncol Biol Phys. 2010;78(2):390–7.

    PubMed  Google Scholar 

  37. Daly-Schveitzer N, Juliéron M, Tao YG, et al. Intensity-modulated radiation therapy (IMRT): toward a new standard for radiation therapy of head and neck cancer?[J]. Eur Ann Otorhinolaryngol Head Neck Dis. 2011;128(5):241–7.

    CAS  PubMed  Google Scholar 

  38. Ouyang Z, Liu Shen Z, Murray E, et al. Evaluation of auto-planning in IMRT and VMAT for head and neck cancer [J]. J Appl Clin Med Phys. 2019;20(7):39–47.

    PubMed  PubMed Central  Google Scholar 

  39. Wiehle R, Knippen S, Grosu AL, et al. VMAT and step-and-shoot IMRT in head and neck cancer: a comparative plan analysis [J]. Strahlenther Onkol. 2011;187(12):820–5.

    PubMed  Google Scholar 

  40. Ezzell GA, Galvin JM, Low D, et al. Guidance document on delivery, treatment planning, and clinical implementation of IMRT: report of the IMRT subcommittee of the AAPM radiation therapy committee [J]. Med Phys. 2003;30(8):2089–115.

    PubMed  Google Scholar 

  41. Basran PS, Woo MK, et al. An analysis of tolerance levels in IMRT quality assurance procedures [J]. Med Phys. 2008;35(6):2300–7.

    PubMed  Google Scholar 

  42. McNair HA, Adams EJ, Clark CH, et al. Implementation of IMRT in the radiotherapy department [J]. Br J Radiol. 2003;76(912):850–6.

    CAS  PubMed  Google Scholar 

  43. Sanghangthum T, Suriyapee S, Srisatit S, et al. Statistical process control analysis for patient-specific IMRT and VMAT QA [J]. J Radiat Res. 2013;54(3):546–52.

    PubMed  Google Scholar 

  44. Park JY, Kim S, Park HJ, et al. Optimal set of grid size and angular increment for practical dose calculation using the dynamic conformal arc technique: a systematic evaluation of the dosimetric effects in lung stereotactic body radiation therapy [J]. Radiat Oncol. 2014;9(4):5–12.

    PubMed  PubMed Central  Google Scholar 

  45. Gros, S ; Descovich, M ; Barani, I ;et al. SU-E-T-446: Effect of Dose Calculation Grid Size Variability on the Specification of Spinal Cord Dose Tolerance for Spinal Stereotactic Body Radiation Therapy [J]. Med Phys, 2012 Jun ;39(6Part17) :3807.

  46. Bresciani S, Di Dia A, Maggio A, et al. Tomotherapy treatment plan quality assurance: the impact of applied criteria on passing rate in gamma index method [J]. Med Phys. 2013;40(12):121711–20.

    PubMed  Google Scholar 

  47. Templeton AK, Chu JC, Turian JV. The sensitivity of ArcCHECK-based gamma analysis to manufactured errors in helical tomotherapy radiation delivery [J]. J Appl Clin Med Phys. 2015;16(1):4814–22.

    PubMed  Google Scholar 

  48. Park JY, Kim S, Park HJ, et al. Optimal set of grid size and angular increment for practical dose calculation using the dynamic conformal arc technique: a systematic evaluation of the dosimetric effects in lung stereotactic body radiation therapy [J]. Radiat Oncol. 2014;9(4):5.

    PubMed  PubMed Central  Google Scholar 

  49. Huang B, Wu L, Lin P, et al. Dose calculation of Acuros XB and anisotropic analytical algorithm in lung stereotactic body radiotherapy treatment with flattening filter free beams and the potential role of calculation grid size [J]. Radiat Oncol. 2015;10(6):53.

    PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

The authors have received the financial support from the planned project of science and technology department of yunnan province of China. Item number is 2017FE468(− 221).

Author information

Authors and Affiliations

Authors

Contributions

HB conducted all digital analyses and performed thestatistical analysis. HB and XRW made substantial contributions to theconception and design of the work. HB, SJZ and XHL were majorcontributors in drafting and writing the manuscript. FHC, JWY, SJZ wereresponsible for radiation treatment planning. The author(s) read and approved the final manuscript.

Corresponding author

Correspondence to Xingrao Wu.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

The authors agreed to publish this paper in Radiation Oncology.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Bai, H., Zhu, S., Wu, X. et al. Study on the ability of 3D gamma analysis and bio-mathematical model in detecting dose changes caused by dose-calculation-grid-size (DCGS). Radiat Oncol 15, 161 (2020). https://doi.org/10.1186/s13014-020-01603-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13014-020-01603-6

Keywords