Circular collimator arc versus dynamic conformal arc treatment planning for linac-based stereotactic radiosurgery of an intracranial small single lesion: a perspective of lesion asymmetry

Abstract

Background

Although circular collimator arcs (CCA) and dynamic conformal arcs (DCA) are commonly used linear accelerator-based treatment planning techniques for intracranial stereotactic radiosurgery (SRS) of a small single lesion, these two techniques have not been rigorously compared in terms of tumor shape. Therefore, this study compared clinical CCA plans with re-planned DCA plans using conformity index (CI) and V12Gy (volume of normal brain tissue receiving 12 Gy or higher) from a perspective of asymmetry (Asym) of planning target volume (PTV).

Methods

Ninety-five clinical CCA plans delivered for a small single lesion with PTV size < 1.4 cm3 were selected and re-planned using DCA. PTV Asym (%) was defined and calculated from three dimensions of PTV. A pair of the 95 plans was first considered as one group without grouping and then categorized into two groups with respective to either PTV size or PTV Asym, and four groups with respect to PTV size and PTV Asym. For grouping, median values of PTV size and PTV Asym were used. A non-parametric paired test was performed for CI and V12Gy to compare CCA and DCA plans in each group.

Results

Median values of PTV size and PTV Asym were 0.415 cm3 (range: 0.076 cm3–1.369 cm3) and 6.12% (range: 0.52–25.74%), respectively. DCA plans had a lower average CI value than CCA plans for all groups. CCA plans had a smaller average V12Gy value than DCA plans for lesions with PTV Asym ≤6.12%, while CCA and DCA plans had similar average V12Gy values for lesions with PTV Asym > 6.12%.

Conclusions

The DCA technique is recommended when a lesion has PTV Asym > 6.12% regardless of PTV size. For lesions with PTV Asym ≤6.12%, a technique choice would depend on the preference of CI or V12Gy.

Background

One of the treatment modalities for brain metastases is stereotactic radiosurgery (SRS) [1, 2]. SRS is generally intended for treating small lesions and size of lesions is usually limited to ≤4 cm [3,4,5]. A wide array of treatment techniques for intracranial SRS are available and such techniques include the Gamma Knife™ (Elekta AB, Stockholm, Sweden), medical linear accelerator (linac)-based systems, charged particle (proton or other heavy ions) therapy, helical TomoTherapy (Accuray Inc., Sunnyvale, CA) and the robot-assisted linac system, Cyber Knife™ (Accuray Inc., Sunnyvale, CA) [5, 6]. Of these, linac-based systems are capable of multiple delivery techniques such as circular collimator arcs (CCA), dynamic conformal arcs (DCA), intensity modulated radiation therapy and volumetric modulated arc therapy (VMAT) [5]. While CCA uses circular cones, the rest of the techniques require micro multileaf collimators (mMLCs). Treatment of multiple lesions with a single isocenter approach has been employed since VMAT was introduced and its delivery has become highly efficient [7, 8].

Despite the advent of VMAT and its efficiency, CCA and DCA are still commonly used for the treatment of a single lesion. Both techniques are relatively simple and easy to be planned and delivered for a single lesion with one isocenter. The CCA technique uses multiple noncoplanar converging arcs with tertiary circular cones and is a conventional method for linac-based SRS [5, 9]. This technique is ideally suited for a small spherically-shaped tumor [3, 10]. Treatment of an irregularly-shaped tumor is possible using CCA but it often requires multiple isocenters for a good conformity at the expense of treatment time and dose homogeneity [9,10,11,12,13]. On the other hand, the DCA technique uses multiple noncoplanar arcs about a single isocenter with continuously changing field shape using mMLCs [12, 14]. It is usually used for a large and/or irregularly-shaped tumor [3, 9, 10].

SRS plan quality is evaluated with various dosimetric parameters and two common parameters are conformity index (CI) and volumes of normal brain tissue receiving at least 8–12 Gy (e.g., V12Gy). CI is presented as a ratio of the total volume enclosed by the prescription isodose to the target volume [5, 15]. CI indicates a degree of conforming the prescription isodose to shape of the target and for SRS treatment, CI values below 2.0 are recommended by the Radiation Therapy Oncology Group [13]. V12Gy is defined as the volume of normal brain tissue (total brain volume minus planning target volume (PTV)) receiving 12 Gy or higher [1, 16] and is known as the most significant prognostic factor for brain radionecrosis (RN) in SRS patients [1, 17, 18]. It has been reported that the risk of symptomatic RN rapidly increases when V12Gy is greater than 5–10 cm3 [17,18,19].

Although it is known that CCA is usually suited for a small spherically-shaped tumor and DCA is for a large and/or irregularly-shaped tumor [3, 10], a comparison of these two techniques in terms of tumor shape has not been rigorously discussed in the literature. In this study, herein, CCA and DCA techniques were compared from a perspective of lesion asymmetry using the two dosimetric parameters, CI and V12Gy, for SRS of a small (volumes < 1.4 cm3) single lesion with a single isocenter approach.

Methods

Patient cohort

A total of 95 patients who received intracranial SRS treatment for a small (volume < 1.4 cm3) single lesion were selected for the current retrospective study. Treatments for all the selected patients were planned and delivered using CCA. For the same patient cohort, re-planning was performed using DCA. This study was approved by our institutional review board.

Clinical circular collimator arc treatment planning

For each patient, CCA treatment planning was performed in the iPlan RT (ver. 4.5.5, Brainlab AG, Munich, Germany) treatment planning system (TPS) which supports a pencil beam algorithm with tissue inhomogeneity corrections. Before treatment planning, PTV was constructed by a 1-mm expansion from gross tumor volume delineated with assistance of magnetic resonance images. Five to six noncoplanar partial (90–160 degrees) arcs with circular cones (from 7.5 mm to 17.5 mm in diameters in 2.5 mm increments) were configured with a single isocenter. Various beam parameters such as gantry angles, couch angles, arc length, cone size and beam weighting for each arc were adjusted such that the plan met our institutional SRS planning criteria. The planning criteria include PTV coverage (COV) ≥98.3%, CI ≤1.5 if possible, V12Gy ≤3.3 cm3 and prescribed isodose line (IDL) between 50 and 80%. For 95 plans, a prescribed dose ranged from 16 Gy to 24 Gy with a median value of 21 Gy. Sixteen plans with a prescribed dose other than 21 Gy were retrospectively re-scaled to 21 Gy. Dose grid size was 1 mm × 1 mm × 1.5 mm.

Dynamic conformal arc treatment planning

Ninety-five DCA plans were generated from the 95 clinical CCA plans in the same TPS. Circular cones in each CCA plan were replaced with mMLCs to create a DCA plan. Based on our institutional practice, an MLC aperture margin was set to 0 mm to achieve similar prescribed IDL to that for CCA plans. Isocenter position, gantry and couch angles, arc length for each arc and beam weighting were kept the same as for the CCA plan. Each DCA plan was adjusted by changing prescribed IDL such that PTV COV was within ±1.0% from that for the corresponding CCA plan.

Asymmetry calculation for each PTV

Asymmetry (Asym) was defined and calculated for each PTV. Following response evaluation criteria in solid tumors 1.1 [20], the longest diameter of PTV and its perpendicular diameter on the transverse plane were measured. Then the longest dimension perpendicular to the transverse plane was measured. From the measurements, PTV Asym was computed using the following equation:

$$ \mathrm{Asym}\ \left(\%\right)=\frac{\left|\mathrm{a}-\mathrm{m}\right|+\left|\mathrm{b}-\mathrm{m}\right|+\left|\mathrm{c}-\mathrm{m}\right|}{4\mathrm{m}}\times 100 $$
(1)

where a, b and c are three dimensions of PTV and m is a mean value of a, b and c. The value “4” in the denominator is a normalization factor to scale Asym (%) ranging from 0% (most symmetric) to 100% (most asymmetric).

Grouping and statistical analysis

To compare CCA and DCA plans, statistical analyses were performed for three different cases. First, without grouping, 95 clinical CCA plans were compared with re-planned DCA plans using four dosimetric parameters (COV, IDL, CI and V12Gy) (one-group analysis). Second, a pair of the 95 plans was categorized into two groups with respect to either PTV size or PTV Asym and then CCA and DCA plans in each group were compared using two dosimetric parameters (CI and V12Gy) (two-group analysis). Third, a pair of the 95 plans was categorized into four groups with respect to both PTV size and PTV Asym, and then CCA and DCA plans in each group were compared using CI and V12Gy (four-group analysis). For each group analysis, a non-parametric test (i.e., Wilcoxon matched pairs test) was run. A difference in a dosimetric parameter between CCA and DCA plans was considered statistically significant when the p-value was < 0.05. It is noted that CI and V12Gy were used to compare CCA and DCA plans (one-group, two-group and four-group analyses) while COV and IDL were used to generate DCA plans with similar plan quality to CCA plans.

Results

Two independent variables for grouping: PTV size and PTV Asym

Statistics of PTV size and PTV Asym for 95 lesions is as follows. Mean ± standard deviation of PTV size is 0.505 cm3 ± 0.333 cm3 and its range (minimum - maximum) is 0.076 cm3–1.369 cm3 with a median value of 0.415 cm3. Those for PTV Asym are 6.64% ± 4.10 and 0.52% - 25.74% with a median value of 6.12%. As shown in Fig. 1a and b, PTV size and PTV Asym did not follow normal distributions and were skewed to low values. PTV Asym as a function of PTV size is displayed in Fig. 1c. There was no distinct trend (R2 = 0.011 from linear regression line) observed between PTV size and PTV Asym. The median values of PTV size and PTV Asym (dotted lines in Fig. 1c) were used to categorize a pair of the 95 plans for two-group analysis and four-group analysis.

Fig. 1
figure 1

Distributions of (a) planning target volume (PTV) size and (b) PTV asymmetry (Asym), and (c) the plot of PTV Asym as a function of PTV size for 95 lesions. Dotted lines represent median values of PTV size (0.415 cm3) and PTV Asym (6.12%)

One-group analysis

Statistics of COV and IDL for 95 CCA and DCA plans is summarized in Table 1. COV for CCA was slightly higher than that for DCA (mean: 98.9% vs. 98.8%) and the COV difference between the two techniques was statistically significant (p-value = 0.0023). Mean of IDL (78.9%) for CCA was also higher than that (78.2%) for DCA and the difference in IDL was statistically significant (p-value = 0.0144).

Table 1 Comparison between circular collimator arc (CCA) and dynamic conformal arc (DCA) treatment planning for 95 lesions. Planning target volume (PTV) coverage (COV) (%), prescribed isodose line (IDL) (%), conformity index (CI), and V12Gy (cm3) were compared between CCA and DCA

CI comparison between CCA and DCA is shown in Table 1 and Fig. 2. More conformal plan was feasible with DCA than with CCA (mean: 1.30 vs. 1.46), resulting in a statistically significant difference (p-value < 0.0001) between the two techniques (Table 1). CI distributions as a function of PTV size showed that there was no noticeable trend (R2 = 0.2778 for DCA and R2 = 0.0256 for CCA from linear regression lines) observed between CI and PTV size (Fig. 2a). Similarly, CI distributions as a function of PTV Asym revealed that there was no dependence of CI on PTV Asym (R2 = 0.0334 for DCA and R2 = 0.3436 for CCA from linear regression lines) (Fig. 2b). As shown in Fig. 2, for most lesions, CI values were < 1.5 with DCA, satisfying our institutional SRS planning criteria better than with CCA. The upper 95% confidence interval (i.e., 97.5 percentile value) of CI for DCA was 1.49 in comparison with that (1.97) for CCA. In CCA plans, there were three lesions with CI values > 2.0. Those lesions had fairly small PTV sizes (0.144 cm3 - 0.244 cm3) and high PTV Asym values (12.1–25.7%).

Fig. 2
figure 2

Conformity index (CI) as a function of (a) planning target volume (PTV) size and (b) PTV asymmetry (Asym) for 95 lesions. Dotted lines in (a) and (b) represent median values of PTV size (0.415 cm3) and PTV Asym (6.12%), respectively

V12Gy comparison between CCA and DCA is presented in Table 1 and Fig. 3. Average V12Gy (1.37 cm3) for CCA was smaller than that (1.43 cm3) for DCA and the difference was statistically significant (p-value < 0.0001). V12Gy had a linear relationship with PTV size for each of CCA and DCA plans (Fig. 3a). Linear regression lines are V12Gy = 2.0713 × PTV + 0.3226 for CCA plans and V12Gy = 2.0792 × PTV + 0.3846 for DCA plans, where V12Gy is the volume receiving 12 Gy or higher in cm3 for the prescribed dose of 21 Gy and PTV is the target volume in cm3. DCA plans had a better fit to the linear regression line than CCA plans (R2: 0.9487 vs. 0.9115). Figure 3b shows V12Gy distributions as a function of PTV Asym for 95 lesions. PTV Asym did not affect V12Gy distributions (R2 = 0.0088 for DCA and R2 = 0.0013 for CCA from linear regression lines). As shown in Fig. 3, for most lesions, CCA plans had smaller V12Gy values than DCA plans.

Fig. 3
figure 3

V12Gy as a function of (a) planning target volume (PTV) size and (b) PTV asymmetry (Asym) for 95 lesions. Dotted lines in (a) and (b) represent median values of PTV size (0.415 cm3) and PTV Asym (6.12%), respectively

Two-group analysis

Analysis with respect to PTV size

Table 2 shows two-group analysis with respect to PTV size. A pair of the 95 plans was divided into two groups (Group A: PTV size ≤0.415 cm3; Group B: PTV size > 0.415 cm3) using the median value (0.415 cm3) of PTV size. Sample sizes were 48 (50.5%) and 47 (49.5%) for Groups A and B, respectively.

Table 2 Two-group analysis in terms of planning target volume (PTV) size. For each group, CI and V12Gy were compared between circular collimator arc (CCA) and dynamic conformal arc (DCA) treatment planning

For Group A, DCA plans had a lower average CI value than CCA plans (1.34 vs. 1.49). The CI difference between the two techniques was statistically significant (p-value < 0.0001). For Group B, similar results were observed. The average CI value (1.26) for DCA plans was lower than that (1.43) for CCA plans and the difference was statistically significant (p-value < 0.0001). Therefore, CI is lower in DCA plans than in CCA plans regardless of PTV size.

For Group A, CCA plans had a smaller average V12Gy value than DCA plans (0.82 cm3 vs. 0.87 cm3). There was a statistically significant difference in V12Gy between the two techniques (p-value = 0.0017). For Group B, average V12Gy (1.93 cm3) for CCA plans was also smaller than that (2.01 cm3) for DCA plans and the difference was statistically significant (p-value = 0.0051). Therefore, V12Gy for CCA plans is smaller than for DCA plans regardless of PTV size.

Analysis with respect to PTV Asym

Table 3 summarizes two-group analysis with respect to PTV Asym. A pair of the 95 plans was divided into two groups (Group C: PTV Asym ≤6.12%; Group D: PTV Asym > 6.12%) using the median value (6.12%) of PTV Asym. Sample sizes were 48 (50.5%) and 47 (49.5%) for Groups C and D, respectively.

Table 3 Two-group analysis in terms of planning target volume (PTV) asymmetry (Asym). For each group, CI and V12Gy were compared between circular collimator arc (CCA) and dynamic conformal arc (DCA) treatment planning

For Group C, DCA plans had a lower average CI value than CCA plans (1.30 vs. 1.41). The CI difference was statistically significant (p-value < 0.0001). For Group D, the average CI value (1.30) for DCA plans was also lower than that (1.52) for CCA plans and the difference was statistically significant (p-value < 0.0001). Therefore, CI is lower in DCA plans than in CCA plans regardless of PTV Asym.

For Group C, CCA plans had a smaller average V12Gy value than DCA plans (1.36 cm3 vs. 1.48 cm3). The V12Gy difference was statistically significant (p-value < 0.0001). For Group D, average V12Gy (1.38 cm3) for CCA plans was almost the same as that (1.39 cm3) for DCA plans and the difference was not statistically significant (p-value = 0.4002). Therefore, V12Gy is smaller in CCA plans than in DCA plans for lesions with PTV Asym ≤6.12%, while there is no difference in V12Gy between the two techniques when PTV Asym is > 6.12%.

Four-group analysis

CI and V12Gy comparisons among four groups (Groups #1-#4) divided using median values of PTV size (0.415 cm3) and PTV Asym (6.12%) are presented in Table 4. Sample sizes of the four groups are 23 (24.2%), 25 (26.3%), 25 (26.3%) and 22 (23.2%) for Groups #1-#4 in order.

Table 4 Four-group analysis in terms of planning target volume (PTV) size and PTV asymmetry (Asym). For each group, CI and V12Gy were compared between circular collimator arc (CCA) and dynamic conformal arc (DCA) treatment planning

DCA plans had lower average CI than CCA plans for all four groups. Two groups with PTV size > 0.415 cm3 (Groups #2 and #4) had the lowest average CI value (1.26) when DCA was used. Group #3 (PTV size ≤0.415 cm3 and PTV Asym > 6.12%) had the highest (1.54) when CCA was used.

For two groups with PTV Asym ≤6.12% (Group #1 and #2), CCA plans had a smaller average V12Gy value than DCA plans. On the other hand, for two groups with PTV Asym > 6.12% (Groups #3 and #4), the two techniques had similar V12Gy values (p-value > 0.05). In Group #3 (PTV size ≤0.415 cm3 and PTV Asym > 6.12%), V12Gy for CCA was slightly smaller than for DCA (0.87 cm3 vs. 0.90 cm3), whereas in Group #4 (PTV size > 0.415 cm3 and PTV Asym > 6.12%), V12Gy for CCA was slightly larger than for DCA (1.97 cm3 vs. 1.95 cm3). Group #1 (PTV size ≤0.415 cm3 and PTV Asym ≤6.12%) had the smallest average V12Gy value (0.77 cm3) with CCA. Group #2 (PTV size ≤0.415 cm3 and PTV Asym > 6.12%) had the largest (2.07 cm3) with DCA.

Discussion

This study limited PTV size to < 1.4 cm3 in comparing CCA and DCA plans for SRS treatment. Minniti et al.’s outcome study for 310 brain metastases reported that the actuarial risk at 1 year for the development of brain RN was 0% in the first quartile (V12Gy < 3.3 cm3) [18]. To be conservative, our institution follows this V12Gy constraint. Our experience has shown that lesions with PTV size > 1.4 cm3 usually do not meet this V12Gy constraint in CCA plans. Hence, all subjects included in this study had PTV size < 1.4 cm3.

As shown in the Results, PTV size and PTV Asym do not affect CI in CCA and DCA plans (Fig. 2) and CI is always lower with DCA (Tables 2, 3 and 4). This would be attributed to field shaping capability of mMLCs in DCA plans. Several studies also reported that mMLCs conform to large and/or irregularly-shaped tumors better than circular cones [3, 9, 10].

In this study, linear relationships between V12Gy and PTV size for CCA and DCA techniques were established (Fig. 3a). Bohoudi et al. derived a linear prediction model of V12Gy for a given prescribed dose from 30 single brain metastases with PTV size ranging from 0.14 cm3 to 43.4 cm3 treated using DCA: V12Gy = (0.12 × prescribed dose – 1.44) × PTV + (0.12 × prescribed dose – 0.96) [1]. This prediction model for a prescribed dose of 21 Gy leads to V12Gy = 1.08 × PTV + 1.56 (Fig. 4). The comparison of this model with our linear model (V12Gy = 2.08 × PTV + 0.38 for DCA) shows that Bohoudi et al.’s V12Gy values are larger than ours for PTV size < 1.2 cm3 (Fig. 4). This comparison is supported by Bohoudi et al.’s finding: their model slightly overestimates V12Gy for small tumors [1]. Zhao et al. also derived a linear prediction model of V12Gy (V12Gy = 1.90 × PTV + 2.11) from 22 clinical targets with volumes from 0.5 cm3 to 41.7 cm3 treated using DCA [14]. The slope (1.90) of their model is close to ours (2.08) (Fig. 4). However, their model was generated for an MLC aperture margin of 1 mm, whereas ours was obtained without a margin. As a result, their V12Gy values are larger than ours by a range of 1.4 cm3 to 1.7 cm3. Except for these models, there has been no published data on a V12Gy prediction model especially for small tumors. Therefore, the linear relationships between V12Gy and PTV size for CCA and DCA techniques obtained in this study would allow for estimation of V12Gy for PTV size < 1.4 cm3 and for the prescribed dose of 21 Gy.

Fig. 4
figure 4

Comparison of three V12Gy prediction models for the dynamic conformal arc (DCA) technique

The V12Gy comparisons between CCA and DCA shown in Tables 2, 3 and 4 demonstrated that V12Gy is smaller with CCA regardless of PTV size. However, V12Gy is smaller with CCA for lesions with PTV Asym ≤ 6.12%, while there is no difference in V12Gy between CCA and DCA techniques for lesions with PTV Asym > 6.12%. Tertiary cones used for the CCA technique increase a source to diaphragm distance in a linac, resulting in a decrease of geometric penumbra [21]. Sharper penumbra can reduce a V12Gy value in CCA plans. However, when PTV Asym becomes larger (> 6.12%), in CCA plans, both CI and V12Gy values are increased due to less flexibility of field shaping with cones. In contrast, in DCA plans, field shaping capability of mMLCs compensates for an increase of V12Gy. As a result, V12Gy becomes similar in the two techniques (1.38 cm3 for CCA vs. 1.39 cm3 for DCA) (Table 3). Therefore, lesions with PTV Asym > 6.12% can benefit from the DCA technique in maintaining similar V12Gy.

A dose gradient index (GI) is another tool to measure dose fall off outside the target and can be a useful predictor of adverse outcomes in the same way as V12Gy [22, 23]. The GI, originating from the Gamma Knife treatment of benign lesions, is defined as the ratio of the volume of half the prescription isodose to the volume of the prescription isodose [22, 23]. In our study, V12Gy was preferred to the GI for plan comparison based on our institutional practice. For 95 patients (one-group analysis), GI (mean ± standard deviation: 3.22 ± 0.40; range: 2.51–4.23) for CCA plans is lower than that (3.82 ± 0.47; 3.03–5.00) for DCA plans and the difference is statistically significant (p-value < 0.0001) from the Wilcoxon matched pairs test. For two-group (A vs. B or C vs. D) and four-group (1, 2, 3 vs. 4) analyses, the results are similar: GI for CCA plans is lower and the differences are statistically significant (p-value < 0.05).

Several studies [23,24,25] in the literature showed that the MLC margin between − 1.5 mm and − 0.5 mm yields the minimum GI. A margin < 0 mm would reduce V12Gy (and GI) but prescribed IDL will become lower to achieve the PTV coverage goal. Our institutional practice shows that most clinical CCA plans had prescribed IDL between 75 and 80% even though our planning criteria allow for prescribed IDL between 50 and 80%. Likewise, most clinical DCA plans had prescribed IDL between 75 and 80% and a MLC aperture margin of 0 mm was used. Therefore, based on our institutional practice, in the current study, 0 mm margin was selected to generate DCA plans which have similar PTV coverage and prescribed IDL to those for CCA plans.

Group analyses presented in this study would be useful in choosing an SRS planning technique between CCA and DCA for a small single lesion (Tables 2, 3 and 4). From the information on PTV size and PTV Asym, a better technique can be selected in terms of CI and V12Gy. For lesions with PTV Asym > 6.12% (Groups #3 and #4), the DCA technique would be better to achieve lower CI without compromising V12Gy as discussed above. For those lesions, using DCA, CI values can be improved by 0.21 for Group #3 (PTV Asym > 6.12% and PTV size ≤0.415 cm3) and by 0.23 for Group #4 (PTV Asym > 6.12% and PTV size > 0.415 cm3). For lesions with PTV Asym ≤6.12% (Groups #1 and #2), the choice of a technique would depend on the dosimetric parameter (CI or V12Gy) to which the clinician pays more attention. If lower CI is preferred, DCA would be a better option with compromised V12Gy. If smaller V12Gy is preferred, CCA should be used but CI would be increased.

Conclusions

DCA plans have lower CI than CCA plans regardless of PTV size or PTV Asym. On the other hand, CCA plans have smaller average V12Gy for lesions with PTV Asym ≤6.12% but the two techniques have similar V12Gy when PTV Asym is larger than 6.12%. Therefore, for lesions with PTV Asym > 6.12%, regardless of PTV size, the DCA technique would be more beneficial to achieve lower CI and simultaneously, to maintain similar V12Gy compared with the CCA technique.

Availability of data and materials

Not applicable.

Abbreviations

Asym:

Asymmetry

CCA:

Circular collimator arcs

CI:

Conformity index

COV:

Coverage

DCA:

Dynamic conformal arcs

IDL:

Isodose line

linac:

Linear accelerator

mMLCs:

Micro multileaf collimators

PTV:

Planning target volume

RN:

Radionecrosis

SRS:

Stereotactic radiosurgery

TPS:

Treatment planning system

V12Gy:

Volume of normal brain tissue receiving 12 Gy or higher

VMAT:

Volumetric modulated arc therapy

References

  1. Bohoudi O, Bruynzeel AM, Lagerwaard FJ, Cuijpers JP, Slotman BJ, Palacios MA. Isotoxic radiosurgery planning for brain metastases. Radiother Oncol. 2016;120(2):253–7.

    Article  Google Scholar 

  2. Solberg TD, Boedeker KL, Fogg R, Selch MT, DeSalles AA. Dynamic arc radiosurgery field shaping: a comparison with static field conformal and noncoplanar circular arcs. Int J Radiat Oncol Biol Phys. 2001;49(5):1481–91.

    Article  CAS  Google Scholar 

  3. Cardinale RM, Benedict SH, Wu Q, Zwicker RD, Gaballa HE, Mohan R. A comparison of three stereotactic radiotherapy techniques; ARCS vs. noncoplanar fixed fields vs. intensity modulation. Int J Radiat Oncol Biol Phys. 1998;42(2):431–6.

    Article  CAS  Google Scholar 

  4. Chen CC, Chapman PH, Kooy H, Loeffler JS. Neuroimaging in radiosurgery treatment planning and follow-up evaluation. In: Chin LS, Regine WF, editors. Principles and practice of stereotactic radiosurgery. New York: Springer; 2008. p. 9–23.

    Chapter  Google Scholar 

  5. Yu C, Shepard D. Treatment planning for stereotactic radiosurgery with photon beams. Technol Cancer Res Treat. 2003;2(2):93–104.

    Article  Google Scholar 

  6. Heller C, Yu C, Apuzzo MLJ. Techniques of stereotactic radiosurgery. In: Chin LS, Regine WF, editors. Principles and practice of stereotactic radiosurgery. New York: Springer; 2008. p. 25–30.

    Chapter  Google Scholar 

  7. Ruggieri R, Naccarato S, Mazzola R, Ricchetti F, Corradini S, Fiorentino A, et al. Linac-based VMAT radiosurgery for multiple brain lesions: comparison between a conventional multi-isocenter approach and a new dedicated mono-isocenter technique. Radiat Oncol. 2018;13(1):38.

    Article  Google Scholar 

  8. Yuan Y, Thomas EM, Clark GA, Markert JM, Fiveash JB, Popple RA. Evaluation of multiple factors affecting normal brain dose in single-isocenter multiple target radiosurgery. J Radiosurg SBRT. 2018;5(2):131–44.

    PubMed  PubMed Central  Google Scholar 

  9. Shiu AS, Kooy HM, Ewton JR, Tung SS, Wong J, Antes K, et al. Comparison of miniature multileaf collimation (MMLC) with circular collimation for stereotactic treatment. Int J Radiat Oncol Biol Phys. 1997;37(3):679–88.

    Article  CAS  Google Scholar 

  10. Verhey LJ, Smith V, Serago CF. Comparison of radiosurgery treatment modalities based on physical dose distributions. Int J Radiat Oncol Biol Phys. 1998;40(2):497–505.

    Article  CAS  Google Scholar 

  11. Dellaretti M, Barbosa Pereira JL, Tagawa E, Pedrini M. Stereotactic radiosurgery of intracranial tumors: a comparison of intensity-modulated radiosurgery and dynamic conformational arc. J Radiosurg SBRT. 2012;1(4):273–80.

    PubMed  PubMed Central  Google Scholar 

  12. Hazard LJ, Wang B, Skidmore TB, Chern SS, Salter BJ, Jensen RL, et al. Conformity of LINAC-based stereotactic radiosurgery using dynamic conformal arcs and micro-multileaf collimator. Int J Radiat Oncol Biol Phys. 2009;73(2):562–70.

    Article  Google Scholar 

  13. Soisson ET, Mehta MP, Tome WA. A comparison of helical tomotherapy to circular collimator-based linear-accelerator radiosurgery for the treatment of brain metastases. Am J Clin Oncol. 2011;34(4):388–94.

    Article  Google Scholar 

  14. Zhao B, Wen N, Chetty IJ, Huang Y, Brown SL, Snyder KC, et al. A prediction model of radiation-induced necrosis for intracranial radiosurgery based on target volume. Med Phys. 2017;44(8):4360–7.

    Article  CAS  Google Scholar 

  15. Feuvret L, Noel G, Mazeron JJ, Bey P. Conformity index: a review. Int J Radiat Oncol Biol Phys. 2006;64(2):333–42.

    Article  Google Scholar 

  16. Limon D, McSherry F, Herndon J, Sampson J, Fecci P, Adamson J, et al. Single fraction stereotactic radiosurgery for multiple brain metastases. Adv Radiat Oncol. 2017;2(4):555–63.

    Article  Google Scholar 

  17. Blonigen BJ, Steinmetz RD, Levin L, Lamba MA, Warnick RE, Breneman JC. Irradiated volume as a predictor of brain radionecrosis after linear accelerator stereotactic radiosurgery. Int J Radiat Oncol Biol Phys. 2010;77(4):996–1001.

    Article  Google Scholar 

  18. Minniti G, Clarke E, Lanzetta G, Osti MF, Trasimeni G, Bozzao A, et al. Stereotactic radiosurgery for brain metastases: analysis of outcome and risk of brain radionecrosis. Radiat Oncol. 2011;6:48.

    Article  Google Scholar 

  19. Lawrence YR, Li XA, el Naqa I, Hahn CA, Marks LB, Merchant TE, et al. Radiation dose-volume effects in the brain. Int J Radiat Oncol Biol Phys. 2010;76(3 Suppl):S20–7.

    Article  Google Scholar 

  20. Eisenhauer EA, Therasse P, Bogaerts J, Schwartz LH, Sargent D, Ford R, et al. New response evaluation criteria in solid tumours: revised RECIST guideline (version 1.1). Eur J Cancer. 2009;45(2):228–47.

    Article  CAS  Google Scholar 

  21. Khan FM, Gibbons JP Jr. Clinical radiation generators. The physics of radiation therapy. 5th ed. Philadelphia: Lippincott Williams & Wilkins; 2014. p. 39–57.

    Google Scholar 

  22. Paddick I, Lippitz B. A simple dose gradient measurement tool to complement the conformity index. J Neurosurg. 2006;105 Suppl:194–201.

    Article  Google Scholar 

  23. Zhao B, Jin JY, Wen N, Huang Y, Siddiqui MS, Chetty IJ, et al. Prescription to 50-75% isodose line may be optimum for linear accelerator based radiosurgery of cranial lesions. J Radiosurg SBRT. 2014;3(2):139–47.

    PubMed  PubMed Central  Google Scholar 

  24. Cardinale RM, Wu Q, Benedict SH, Kavanagh BD, Bump E, Mohan R. Determining the optimal block margin on the planning target volume for extracranial stereotactic radiotherapy. Int J Radiat Oncol Biol Phys. 1999;45(2):515–20.

    Article  CAS  Google Scholar 

  25. Ding C, Solberg TD, Hrycushko B, Xing L, Heinzerling J, Timmerman RD. Optimization of normalized prescription isodose selection for stereotactic body radiation therapy: conventional vs robotic linac. Med Phys. 2013;40(5):051705.

    Article  Google Scholar 

Download references

Acknowledgements

The authors are grateful to Jason Wei-Yeong Huynh for data collection.

Funding

None.

Author information

Authors and Affiliations

Authors

Contributions

YCL collected data, analyzed data and wrote a manuscript. YK designed the study and analyzed data. Both authors read and approved the final manuscript.

Corresponding author

Correspondence to Yongbok Kim.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Lee, Y.C., Kim, Y. Circular collimator arc versus dynamic conformal arc treatment planning for linac-based stereotactic radiosurgery of an intracranial small single lesion: a perspective of lesion asymmetry. Radiat Oncol 14, 91 (2019). https://doi.org/10.1186/s13014-019-1307-z

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13014-019-1307-z

Keywords