Skip to main content

Effect of changes in monitor unit rate and energy on dose rate of total marrow irradiation based on Linac volumetric arc therapy

Abstract

Background

This study set out to evaluate the effect of dose rate on normal tissues (the lung, in particular) and the variation in the treatment efficiency as determined by the monitor unit (MU) and energy applied in Linac-based volumetric arc therapy (VMAT) total marrow irradiation (TMI).

Methods

Linac-based VMAT plans were generated for the TMI for six patients. The planning target volume (PTV) was divided into six sub-volumes, each of which had their own isocenter. To examine the effect of the dose rate and energy, a range of MU rates (40, 60, 80, 100, 300, and 600 MU/min) were selected for 6, 10, and 15 MV. All the plans were verified by portal dosimetry.

Results

The dosimetric parameters for the target and normal tissue were consistent in terms of the energy and MU rate. The beam-on time was changed from 59.6 to 6 min for 40 and 600 MU/min. When 40 MU/min was set for the lung, the dose rate delivered to the lung was less than 6 cGy/min (that is, 90%), while the beam-on time was approximately 10 min. The percentage volume of the lung receiving 20 cGy/min was 1.47, 3.94, and 6.22% at 6, 10, and 15 MV, respectively. However, for 600 MU/min, the total lung volume received over 6 cGy/min regardless of the energy, and over 20 cGy/min for 10 and 15 MV (i.e., 54.4% for 6 MV).

Conclusions

In TMI treatment, reducing the dose rate administered to the lung can decrease the incidence of pulmonary toxicity. To reduce the probability of normal tissue complications, the selection of the lowest MU rate is recommended for fields including the lung. To minimize the total treatment time, the maximum MU rate can be applied to other fields.

Background

Total body irradiation (TBI) has long played an important role as a conditioning regimen prior to bone marrow or hematopoietic stem cell transplantation for patients with hematologic malignancies (i.e., leukemia, lymphoma, or multiple myeloma) [1]. The primary reason for the use of TBI is the elimination of residual cancer cells and preventing the immunologic rejection of the transplanted donor stem cells [2]. The TBI treatments are generally performed at an extended source-to-surface distance (SSD) with large open fields to deliver a homogeneous dose to the entire body [3, 4]. In addition, TBI treatments have been performed using techniques such as moving-beam TBI with a sweeping beam, a patient translation technique, multiple-beam TBI, a shrinking-field technique, and the local application of small beams [5,6,7].

However, for patients with advanced acute leukemia, increasing the total TBI dose may improve malignant clone killing, but it is also associated with a potential lethal toxicity to the surrounding normal healthy tissues/organs [8, 9]. The side effects of TBI can be extensive and include acute symptoms such as nausea and a loss of appetite, as well as infertility and secondary malignancies [10]. Pulmonary toxicity is one of the most serious, and potentially life-threatening, complications that may occur following TBI [11]. Therefore, new techniques have concentrated on providing a safer dose escalation [12, 13].

To deliver a therapeutic dose to a target, while not irradiating normal tissue, total marrow irradiation (TMI) has been introduced in combination with helical tomotherapy (HT) or volumetric arc therapy (VMAT) using conventional linear accelerators [12, 14, 15]. TMI, when delivered with an intensity-modulation (IM) technique, may reduce the dose delivered to the organs at risk (OARs), relative to TBI. These techniques have the potential to provide a safer dose escalation for a target [16]. Several studies have shown the potential clinical benefits of IM-TMI [12,13,14,15, 17, 18].

For conventional TBI treatments, the dose rate is a significant parameter. To minimize the dose rate, the patient is generally located at an extended SSD of more than 400 cm [3]. Many studies have reported that, when TBI is performed with a low dose rate, the healthy tissue exhibits less damage [8, 9, 11, 19, 20]. Especially, low dose rates (less than 6 cGy/min) can decrease the risk of pulmonary toxicity [11, 19, 21]. However, TMI treatments with HT or VMAT have conventionally been performed with relatively high dose rates due to limitations on the dose rate selection and the treatment location (i.e., at the isocenter). For example, the lowest dose rate is 100 monitor units (MU)/min and 800 MU/min for the Varian C-series Clinac® (Varian Medical Systems, Palo Alto, CA) and HT, respectively. New delivery platforms that can deliver low dose rates have been released by Varian. Their VitalBeam™ (Varian Medical Systems, Palo Alto, CA) and Truebeam™ systems (Varian Medical Systems, Palo Alto, CA) offer dose rates of as low as 5 MU/min and 20 MU/min for 6–10 MV and 15–20 MV, respectively. In VMAT delivery mode, the dose rate can be varied during the treatment. However, the maximum dose rate can be limited by setting the dose rate in the planning process. If, in the plan optimization, the dose rate is set to less than 100 MU/min, which is supported by the C-series, a safe dose rate can be delivered to the OARs, especially the lungs.

In this paper, we describe the plan quality of TMI treatments with a range of MU rates and energies, performed using a Vitalbeam system. The appropriate MU rate setting was determined by evaluating the dose distribution, the dose rate to each OAR, and the duration of the TMI treatment.

Materials and methods

Patient selection and simulation

Six patients being treated by TMI were enrolled in the present study. Their participation in the study was approved by the institutional review board (IRB No. H-1706-007-855). The height and weight of each patient are listed in Table 1. All the patients were scanned using the Brilliance CT Big Bore scanner (Philips, Amsterdam, Netherlands) with a 3.0-mm slice thickness.

Table 1 Patient information (height, weight, and gender)

Treatment plan

A radiation oncologist defined the clinical target volume (CTV) including all the bone marrow in the body from the head to the mid-thighs, but excluding the forearms and hands. The CTVs were segmented for the bone, brain, liver, spleen, and testis. Regarding the lower extremities, one third of the upper femur (active marrow region) was included for CTV definition.

To define the planning target volume (PTV), each CTV was extended by adding a different margin in consideration of the internal motion of each organ. The brain, bone and testis PTV was defined by adding 3, 5, and 7 mm to their CTVs, respectively. For the liver and spleen PTV, a 5-mm margin was set in all except the superior-inferior (SI) directions. For the SI direction, a 15-mm margin was applied.

The total PTV was defined as the sum of each PTV. The total PTV was divided into sub PTVs for each isocenter. A total of six isocenters were defined considering the shape and length of the PTV. The sub PTVs were named as follows: PTV_Brain, PTV_Chest1, PTV_Chest2, PTV_Abdomen, PTV_Pelvic, and PTV_Femur. The lungs, kidneys, eyes, oral cavities, bowels, and heart were designated as being OARs. The prescription dose was 10 Gy, administered in five fractions [12]. The dose constraints of the mean doses administered to the lungs, kidneys, bladder, and bowels were set to 6, 7, 8.5, and 7 Gy, respectively. All the plans were normalized to cover 90% of the PTV volume with 100% of the prescription dose. The patient was treated once a day with a 2-Gy dose, administered by the Vitalbeam system.

A VMAT-TMI treatment plan was generated using an Eclipse™ system (Ver. 13.7, Varian Medical Systems, Palo Alto, CA) with PRO 13.7 (progressive resolution optimization, Ver. 13.7, VarianTM Medical Systems, Palo Alto, CA) as the optimizer algorithm. The dose calculation was performed using Acuros® XB 13.7 with a 2.5-mm grid.

A total of six arc fields corresponding to each isocenter were created with inter-gantry angles of 179° and 181°. Each field was fully opened to cover the entire length of each sub PTV with a collimator rotation of 90° or 180°. For TMI using a multi-isocenter, the verification of the dose accumulation in a junction area should be considered. Previous studies investigated the robustness of the junctions [15, 17, 18]. Those studied have suggested at least 2 cm overlapped region between the previous and following arc on each side to eliminate any hot and cold spots. In the present study, the overlapping regions between neighboring fields were set with at least 2 cm on each side to avoid there being hot or cold regions around the field junctions following literature [15, 17, 18]. Figure 1 shows the arrangement of the field and isocenter corresponding to each PTV.

Fig. 1
figure 1

Field arrangement for VMAT. Pictorial representation of total marrow irradiation (TMI) plan by volumetric arc therapy (VMAT) consisting of six isocenters with a single arc. Field arrangement projected at isocenter in coronal plan (a) and sagittal plane (b)

First, a VMAT-TMI plan was generated by changing the energy (6, 10, and 15 MV) with a dose rate of 600 MU/min, while maintaining the collimator angles, field sizes, and number and location of the isocenters for the total field. Next, new plans were created in accordance with the changing MU rate (40, 60, 80, 100 and 300 MU/min) for each energy. For new created plans, the each VMAT optimization was performed with same optimization parameters.

One MU was calibrated while 1 cGy was delivered to the maximum depth (e.g., 1.5, 2.5, and 2.9 cm depth for 6, 10, and 15 MV) in a water phantom with a 100-cm source to surface distance and a 10 × 10 cm2 field size.

Plan analysis

The beam-on time was calculated to evaluate the dose rate for each field. To evaluate the non-linear correlation between the dose rate and beam-on time, the non-linear correlation factor was defined as the ratio of MUs to beam-on times. The dosimetric quality was evaluated with respect to the target coverage and OAR dose. For the PTV, the dose parameters were analyzed using the near-minimum (D98%), near-maximum, (D2%), and mean absorbed doses, the latter being designated D50%. For the OAR dose, the mean dose and V5Gy (cc) were evaluated.

The average dose rate delivered to the lungs was calculated by dividing the mean dose by the beam-on time. According to the guidelines published by the ACR and the ASTRO, a dose rate to the lungs of less than 20 cGy/min has been suggested to decrease the incidence of pulmonary toxicity. Therefore, the average volume irradiated at a dose rate of more than 20 cGy/min was evaluated for the lungs. The PTV_Chest1 and PTV_Chest2 isocenters were set to irradiate the chest region including the lungs and ribs. The sum of the beam-on time for PTV_Chest1 and PTV_Chest2 was used as the beam-on time for calculating the average dose rate for the entire lung.

Plan verification

The plans were verified using a Portal dosimetry system with an electronic portal imaging device. The verification plans were created using a portal dose image prediction (PDIP, Ver 13.7, Varian Medical Systems, Inc., Palo Alto, CA) algorithm. For gamma index evaluation, the criteria was set to a 2% dose difference and a 2-mm distance to agreement. The threshold dose was set to 10%.

Results

The beam-on times and MU for each field, energy, and dose rate are listed in Table 2. Figure 2 shows the non-linear correlation factor between the MUs and beam-on times. The beam-on time decreases as the energy increases. The beam-on time for field 1 (PTV_Brain) was the shortest. The non-linear correlation factor increases dramatically with the MU rate.

Table 2 Beam-on time and MU of each isocenter for energy and MU rate
Fig. 2
figure 2

Non-linear correlation factor. The factors were defined as a ratio of total MU to beam-on time. 6 MV (black solid line), 10 MV (red solid line) and 15 MV (blue solid line)

Dose-volumetric parameters for each MU rate of TMI by VMAT using 6, 10 and 15 MV was provided in supplementary material (Additional file 1: Table S1).

The TMI isodose distributions in colorwash are illustrated in Fig. 3. The average values of the dose-volumetric parameters of PTV and normal tissue for the energy and dose rate of each patient are provided as supplementary material. The different dose-volumetric parameters for the energy and dose rates are not shown.

Fig. 3
figure 3

Isodose distributions. Isodose distributions in colorwash of TMI by VMAT for 6 (a), 10 (b), and 15 MV (c). Dose range: 20% (blue) to 120% (red)

Figure 4 shows the dose-rate volumetric histogram for the lung volume of a representative case (Patient 1). When the 40-MU rate was used, the average volume of the lung receiving a dose of less than 6 cGy/min was 92.45 ± 8.88%, 93.47 ± 6.53% and 94.04 ± 6.11% at 6, 10, and 15 MV, respectively. The average lung volumes irradiated at a dose rate of more than 20 cGy/min are listed in Table 3.

Fig. 4
figure 4

Dose-rate versus lung volume histogram. Dose-rate versus lung volume histogram of representative case for 6 MV (solid line), 10 MV (dotted line), and 15 MV (dashed line) according to the monitor unit (MU) rate. Black: 40 MU/min, red: 60 MU/min, green: 80 MU/min, blue: 100 MU/min, cyan: 300 MU/min and magenta: 600 MU/min

Table 3 Average volume (%) of lung irradiated at dose rate of more than 20 cGy/min

For all the plans, the gamma passing rate was more than 90%. For those plans with 300 and 600 MU/min at every energy, the gamma passing rate was 99%. For those plans with 40, 60, and 80 MU/min at every energy, the range of the gamma passing rate was 90–94%.

Discussion

In the present study, we evaluated the dose rate and dosimetric parameter based on the MU rate of the machine and the energy.

TMI treatments using the VMAT technique are clinically favorable in that they can overcome the limitations of TBI since TMI can reduce the dose administered to an OAR. Additionally, the accuracy of the TMI treatment can be improved using an image guidance system. However, there remain issues with the dose rate when using TMI delivery with a C-series or HT device. However, this problem has been solved by the release of new machines with a low MU rate.

The radiation toxicity of a normal organ, especially the incidence of pulmonary toxicity, is a major factor determining the prescription dose for patients with hematolymphoid malignancies [8, 11, 19, 21]. Pulmonary toxicity can significantly affect the overall survival rate. The dose rate administered to the lung is a key factor, as well as the total dose. The published guidelines recommend that TBI treatments not exceed a dose rate of 20 cGy/min [21]. TBI with a dose rate of less than 6 cGy/min can reduce pulmonary toxicity [11, 19]. When 40 MU/min was set for the PTV_Chest1 and 2 fields, the lung volume receiving less than 6 cGy/min was 90% while the beam-on time of two field (PTV_Chest1 and 2) was 10 min, approximately. However, if MU/min was set for all the fields, the total treatment time would be 60 min. This long treatment time can result in increasing the intra-fraction error (e.g., breathing movement or patient motion. Moreover, the patient can feel discomfort due to lying down on the couch with an immobilization device. To avoid such an adverse effect, we can select the proper MU rate for a specific field which includes an OAR such as the lung. If the maximum MU rate (i.e., 600 MU/min) is applied to the PTV_Brain, Pelvic and Femur fields with 40MU/min for PTV_Chest1 and 2 only, the beam-on time was approximately 24 min. Even if the imaging acquisition time is included, the total treatment can be finished within 1 h. For imaging guide, cone beam computed tomography (CBCT) was taken for each single isocenter. The reference CT image set and digitally reconstructed radiograph was compare to CBCT and pair of orthogonal kV radiographs.

However, in terms of dose voxel, each voxel in OAR has a different dose rate. Each beam of field beam is subdivided into hundreds of beamlets. The voxel was partially irradiated by beamlet which have an individual intensity level [22]. In this study, the dose rate was calculated under the assumption that the dose to OAR is sum total of dose during beam on time.

The total dose administered to the lung did not exhibit any dependency on the energy. However, the beam-on time decreased slightly as the energy increased, except when the dose rate was 600 MU/min. A higher energy can result in the delivery of a similar dose in less time with a limited MU rate. However, when the maximum MU rate (600 MU/min) is allowed, the machine can deliver a sufficiently high dose rate at the maximum gantry speed. The beam-on time with an MU rate of 600 MU/min was constant since the beam-on time is predominantly determined by the maximum gantry speed.

The PTV should include the ribs since the bone marrow is primarily located in the ribs. The lungs, which are the most important OARs, are enclosed by the ribs. In planning the processing, we tried to reduce the dose administered to the lungs as far as possible. However, it was a difficult to reduce this dose due to the PTV coverage including the ribs. The DVH of the lung had a steep dose gradient with a range of 3–10 Gy.

The beam-on time and dose slightly decreased with an increase in the energy. With conventional TBI treatment, high-energy beams (10–18 MV) have historically been used, particularly for patients for which the thickness is greater than 35 cm, and therefore considered to be homogenous. However, when high-energy photon beams of more than 10 MV were used, the photoneutrons can lead to an increase in the undesired dose to the patient [23]. TMI techniques using 10 and 15 MV offer no advantages in consideration of the higher integral dose, larger penumbra and similar beam-on time and dose rate.

The variation in the MU rate was extremely limited relative to the general capacity of the machine. The optimization was processed with the extra constraints for a treatment plan with a lower dose rate. Therefore, plan verification was performed to check the delivery accuracy. The gamma passing rates of the plan with the lower MU rate were lower than those with the higher dose rate. The response of the Si detector used to perform portal dosimetry can be affected by the dose rate. The relatively low gamma passing rate was caused by this effect [24]. However, in every case, the gamma passing rate was more than 90%. All the plans could be delivered clinically.

In this study, the part of the lower extremity (i.e. lower legs from mid-femurs) was not included in VMAT optimization. The current treatment table on the LINAC has a maximum travel length of approximately 150 cm in longitudinal direction [25]. Therefore, two different plans are required for TMI which includes lower leg. In literature, reverse patient position approach has been introduced [18]. However, the field junction problem should be considered carefully for the approach [17]. VMAT plan could be generated for this part in feet-first position, keeping an overlapping region with the upper sector to prevent the risk of cold or hot spots [25].

Conclusion

We evaluated the beam-on time and dose volume parameter for given energies and MU rates for VMAT TMI. The dose volume parameter was found to be not dependent on the energy and MU rate. However, the received PTV dose rate for normal tissue was found to differ significantly with the MU rate. In the case of TMI, the dose rate delivered to the lung is critical. To reduce the probability of normal tissue complications, the selection of the lowest MU rate was recommended for fields included the lung. Considering the total treatment time, maximum dose rate can be selected for other fields.

Abbreviations

CTV:

Clinical target volume

DVH:

Dose volume histogram

IM:

Intensity modulation

MU:

Monitor unit

OAR:

Organs at risk

PDIP:

Portal dose image prediction

PTV:

Planning target volume

TMI:

Total marrow irradiation

VMAT:

Volumetric modulated arc therapy

References

  1. Thomas ED, Buckner CD, Banaji M, Clift RA, Fefer A, Flournoy N, et al. One hundred patients with acute leukemia treated by chemotherapy, total body irradiation, and allogeneic marrow transplantation. Blood. 1977;49(4):511–33.

    CAS  PubMed  Google Scholar 

  2. Sorror ML, Storer BE, Maloney DG, Sandmaier BM, Martin PJ, Storb RJB. Outcomes after allogeneic hematopoietic cell transplantation with nonmyeloablative or myeloablative conditioning regimens for treatment of lymphoma and chronic lymphocytic leukemia. Blood. 2008;111:446–52.

    Article  CAS  Google Scholar 

  3. Choi CH, Park JM, Park S-Y, Chun M, Han JH, Cho JD, et al. Prediction of midline dose from entrance and exit dose using OSLD measurements for Total body irradiation. J Radiat Protect Res. 2017;42:77–82.

    Article  Google Scholar 

  4. Wong JY, Filippi AR, Dabaja BS, Yahalom J, Specht L. Total body irradiation: guidelines from the international lymphoma radiation oncology group (ILROG). Int J Radiat Oncol Biol Phys. 2018;101:521–9.

    Article  Google Scholar 

  5. Sarfaraz M, Yu C, Chen D, Der L. A translational couch technique for total body irradiation. J Appl Clinic Medic Phys. 2001;2:201–9.

    Article  CAS  Google Scholar 

  6. Pla M, Chenery SG, Podgorsak EB. Total body irradiation with a sweeping beam. Int J Radiat Oncol Biol Phys. 1983;9:83–9.

    Article  CAS  Google Scholar 

  7. Quast U. Total body irradiation—review of treatment techniquesin Europe. Radiother Oncol. 1987;9:91–106.

    Article  CAS  Google Scholar 

  8. Barrett A, Depledge MH, Powles RL. Interstitial pneumonitis following bone marrow transplantation after low dose rate total body irradiation. Int J Radiat Oncol Biol Phys. 1983;9:1029–33.

    Article  CAS  Google Scholar 

  9. Ozsahin M, Belkacemi Y, Pens F, Dominique C, Schwartz LH, Uzal C, et al. Total-body irradiation and cataract incidence: a randomized comparison of two instantaneous dose rates. Int J Radiat Oncol Biol Phys. 1994;28:343–7.

    Article  CAS  Google Scholar 

  10. Kawaguchi H, Ishibashi N, Soejima T, Oguchi M, Hasegawa M, Isobe K, et al. National Survey on Total body irradiation before reduced-intensity stem cell transplantation in Japan. Int J Radiat Oncol Biol Phys. 2017;99(2):E430–E1.

    Article  Google Scholar 

  11. Kim D-Y, Kim IH, Yoon S-S, Kang HJ, Shin HY, Kang H-C. Effect of dose rate on pulmonary toxicity in patients with hematolymphoid malignancies undergoing total body irradiation. Radiat Oncol. 2018;13:180.

    Article  Google Scholar 

  12. Wong JY, Liu A, Schultheiss T, Popplewell L, Stein A, Rosenthal J, et al. Targeted total marrow irradiation using three-dimensional image-guided tomographic intensity-modulated radiation therapy: an alternative to standard total body irradiation. Biol Blood Marrow Transplant. 2006;12:306–15.

    Article  Google Scholar 

  13. Yeginer M, Roeske JC, Radosevich JA, Aydogan B. Linear accelerator-based intensity-modulated total marrow irradiation technique for treatment of hematologic malignancies: a dosimetric feasibility study. Int J Radiat Oncol Biol Phys. 2011;79:1256–65.

    Article  Google Scholar 

  14. Corvò R, Zeverino M, Vagge S, Agostinelli S, Barra S, Taccini G, et al. Helical tomotherapy targeting total bone marrow after total body irradiation for patients with relapsed acute leukemia undergoing an allogeneic stem cell transplant. Radiother Oncol. 2011;98:382–6.

    Article  Google Scholar 

  15. Aydogan B, Yeginer M, Kavak GO, Fan J, Radosevich JA, Gwe-Ya K. Total marrow irradiation with RapidArc volumetric arc therapy. Int J Radiat Oncol Biol Phys. 2011;81:592–9.

    Article  Google Scholar 

  16. Hui S, Brunstein C, Takahashi Y, DeFor T, Holtan SG, Bachanova V, et al. Dose escalation of total marrow irradiation in high-risk patients undergoing allogeneic hematopoietic stem cell transplantation. Biol Blood Marrow Transplant. 2017;23:1110–6.

    Article  Google Scholar 

  17. Fogliata A, Cozzi L, Clivio A, Ibatici A, Mancosu P, Navarria P, et al. Preclinical assessment of volumetric modulated arc therapy for total marrow irradiation. Int J Radiat Oncol Biol Phys. 2011;80:628–36.

    Article  Google Scholar 

  18. Mancosu P, Navarria P, Castagna L, Roggio A, Pellegrini C, Reggiori G, et al. Anatomy driven optimization strategy for total marrow irradiation with a volumetric modulated arc therapy technique. J Appl Clin Medic Phys. 2012;13:138–47.

    Article  Google Scholar 

  19. Travis E, Peters L, McNeill J, Thames H Jr, Karolis C. Effect of dose-rate on total body irradiation: lethality and pathologic findings. Radiother Oncol. 1985;4:341–51.

    Article  CAS  Google Scholar 

  20. Glass TJ, Hui SK, Blazar BR, Lund TC. Effect of radiation dose-rate on hematopoietic cell engraftment in adult zebrafish. PLoS One. 2013;8:e73745.

    Article  CAS  Google Scholar 

  21. Wolden SL, Rabinovitch RA, Bittner NH, Galvin JM, Giap HB, Schomberg PJ, et al. American College of Radiology (ACR) and American Society for Radiation Oncology (ASTRO) practice guideline for the performance of total body irradiation (TBI). Am J Clinic Oncol. 2013;36:97–101.

    Article  Google Scholar 

  22. Taylor A, Powell MEB. Intensity-modulated radiotherapy—what is it? Cancer Imaging. 2004;4(2):68–73.

    Article  CAS  Google Scholar 

  23. Cherpak AJ, Monajemi T, Chytyk-Praznik K, Mulroy L. Energy-dependent OAR sparing and dose conformity for total marrow irradiation of obese patients. J Appl Clinic Medic Phys. 2018;19:532–8.

    Article  Google Scholar 

  24. Esch AV, Huyskens DP, Hirschi L, Scheib S, Baltes C. Optimized Varian aSi portal dosimetry: development of datasets for collective use. J Appl Clinic Medic Phys. 2013;14(6):82–99.

    Article  Google Scholar 

  25. Mancosu P, Navarria P, Castagna L, Reggiori G, Stravato A, Gaudino A, et al. Plan robustness in field junction region from arcs with different patient orientation in total marrow irradiation with VMAT. Phy Med. 2015;31(7):677–82.

    Article  Google Scholar 

Download references

Acknowledgments

Not applicable.

Funding

This work was supported by the National Research Foundation of Korea (NRF) grant funded by the Korean government (MSIP) (no. 2017M2A2A7A02020641) and the National R&D Program for Cancer Control, Ministry of Health & Welfare, Republic of Korea (no. HA 16C0025).

Availability of data and materials

Data sharing is not applicable to this article because no datasets were generated or analyzed during the current study.

Author information

Authors and Affiliations

Authors

Contributions

CHC and HCK conceived the study concept, compiled and analyzed the data, drafted the manuscript, and participated in all aspects of the study. NC helped measure the data. JMP and JIK analyzed the data from a clinical perspective and engaged in discussions to improve the significance of this study. HGW reviewed and advised the clinical aspect of this study. All authors read and approved the final manuscript.

Corresponding authors

Correspondence to Hyun-Cheol Kang or Chang Heon Choi.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Additional file

Additional file 1:

Table S1. Dose-volumetric parameters for each MU rate of TMI by VMAT using 6, 10 and 15 MV. (XLSX 17 kb)

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Son, J., Choi, N., Kim, Ji. et al. Effect of changes in monitor unit rate and energy on dose rate of total marrow irradiation based on Linac volumetric arc therapy. Radiat Oncol 14, 87 (2019). https://doi.org/10.1186/s13014-019-1296-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13014-019-1296-y

Keywords