Dosimetric impact of intrafraction motion on boosts on intraprostatic lesions: a simulation based on actual motion data from real time ultrasound tracking

A Data Descriptor to this article was published on 11 November 2019

Abstract

Background

Intrafraction motion is particularly problematic in case of small target volumes and narrow margins. Here we simulate the dose coverage of intraprostatic lesions (IPL) by simultaneous integrated boosts (SIB). For this purpose, we use a large sample of actual intrafraction motion data.

Methods

Fifty-three h of intra-fraction motion of the prostate were recorded in real-time by 4D ultrasound (4DUS) during 720 fractions in 28 patients. We simulate spherical IPLs with 3, 5, and 7 mm radius and matching spherical SIBs with 0, 2, and 5 mm safety margins. The volumetric overlap between IPLs and SIBs is calculated. Dose volume histograms (DVH) are estimated by Monte Carlo simulation.

Results

On average, the distance of the prostate was 1.3 mm from its initial position over all fractions and patients. Average volumetric overlap was 73, 82, and 87% of IPL volume in case of 3, 5, and 7 mm IPLs and SIBs without safety margins. These improved to 95% or more in case of 2 mm safety margins and 98% or more in case of 5 mm safety margins. DVHs showed that 80% of the IPL volume received 60, 72, and 79% of maximum dose in case of 3, 5, and 7 mm IPLs and SIBs without safety margins. These improved to 94% or more given moderately sized safety margins of 2 mm.

Conclusions

On average over all fractions and patients, the dose coverage would have been acceptable even for small target volumes such as IPLs of radius 3 to 7 mm and narrow fields. Moderate safety margins of 2 mm could have ensured a delivery of 90% or more of the SIB dose to the IPL. In this case, SIB volume would have been considerably larger than IPL volume, but still considerably smaller than the overall PTV of the prostate.

Background

Recurrence after radiotherapy of prostate carcinoma often originates from a macroscopic intraprostatic lesion (IPL). A simultaneous integrated boost (SIB) to the dominant IPL is a strategy to fight such recurrences [1,2,3,4]. Toxicity seems to be less of a problem [5,6,7,8], however it has been suggested that intrafraction motion requires either intrafraction motion tracking in case of small margins [9] or substantial margins [10].

Generally, the recording of the prostate position by 4D ultrasound [11] or kV-imaging [12, 13] during a fraction has received some interest in recent years and is a natural extension of these techniques from inter-fraction position verification and correction to online intra-fraction monitoring and adaption. The hope is, e.g. to account for prostate intra-fraction motion by gating the beam or tracking the prostate by continually adjusting the multi-leaf collimator [ 14,15,16,17].

In this simulation study we calculate the impact of intrafraction motion on the volumetric coverage between IPL and SIB and on dose volume histograms (DVH). We use motion data from real time ultrasound tracking during actual fractions and apply this data to idealized spherical IPL volumes of 3, 5, and 7 mm radius and to idealized spherical SIB volumes without and with safety margins of 2 mm resp. 5 mm.

Our main question is, if and what size of margins is required to counterbalance intrafraction motion and ensure an acceptable dose coverage of the IPL.

Methods

Intrafraction motion data

The analysis is based on the same set of patients and raw data as reported in [18]. The study is based on 28 patients with adenocarcinoma of the prostate who received a definitive external beam radiotherapy between June 2014 and March 2017 at our department. The 28 patients were of a typical demographics for prostate cancer, as were the treatment regimes, see Table 1.

Table 1 Patient demographics and treatment regimes

Up to the calculation of prostate trajectories, the raw data and methods were identical to those detailed in [18].

Vertical (anterior-posterior), longitudinal (cranial-caudal), lateral (left-right), and radial (‘vector length’ or Euclidean ‘3D distance’) prostate displacements were calculated relative to the starting position of the prostate (as recorded by 4DUS) after image guide repositioning (by kV-CBCT) at the beginning of the respective fraction.

Averages, standard deviations, quantiles (5, 25%, median, 75, and 95%) and root mean squares (r.m.s.) of vertical, longitudinal, lateral, and radial displacements were calculated.

Volumetric coverage

The IPL was idealized as a sphere of radius r. The dose distribution was idealized as constant within a spherical region of radius r + m and zero outside, where m denotes the safety margin. These idealizations were assumed for simplicity, and because they are not uncommon, e.g. in optimal margin recipes.

The 3D-distance or Euclidean distance d of the prostate from the origin determined the distance of the centers of these two spheres. The volume of the overlap of the two spheres in case m < d < r + m was.

$$ {\mathrm{V}}_{\mathrm{overlap}}=\uppi /\left(12\mathrm{d}\right)\ast {\left(2\mathrm{r}+\mathrm{m}-\mathrm{d}\right)}^2\ast \left({\mathrm{d}}^2+2\mathrm{d}\left(2\mathrm{r}+\mathrm{m}\right)-3\;{\mathrm{m}}^2\right) $$

See Fig. 1.

Fig. 1
figure 1

Calculation of the overlap between SIB dose and IPL volume

This volume was divided by the volume of the IPL, VIPL = 4π/3 * r2 to arrive at the coverage.

$$ {\displaystyle \begin{array}{cc}{\mathrm{c}}_{\mathrm{IPL}}=100\%& \mathrm{in}\ \mathrm{case}\ \mathrm{d}<=\mathrm{m}\\ {}{\mathrm{c}}_{\mathrm{IPL}}=0\%& \mathrm{in}\ \mathrm{case}\ \mathrm{d}>=2\mathrm{r}+\mathrm{m}\\ {}{\mathrm{c}}_{\mathrm{IPL}}={\mathrm{V}}_{\mathrm{overlap}}/{\mathrm{V}}_{\mathrm{IPL}}& \mathrm{in}\ \mathrm{case}\ \mathrm{m}<\mathrm{d}<2\mathrm{r}+\mathrm{m}\end{array}} $$

This coverage cIPL,i was calculated for every single one of the 39,295 measured 3D-distances di and for every combination of IPL radius r (3 mm, 5 mm, 7 mm) and safety margin m (0 mm, 2 mm, 5 mm). Finally, the respective averages, standard deviations, quantiles (5, 25%, median, 75, and 95%) over all measurements were calculated.

The aforementioned coverage of the IPL volume by the SIB volume, in % of IPL volume, concerns the question how much dose was received by the IPL. The other interesting question is how much of the overall SIB dose was received by the IPL. To this end, the coverage cSIB of the larger SIB volume by the smaller IPL volume, in % of SIB volume, was also calculated. This ‘reverse’ calculation yields:

$$ {\displaystyle \begin{array}{cc}{\mathrm{c}}_{\mathrm{SIB}}={\mathrm{r}}^3/{\left(\mathrm{r}+\mathrm{m}\right)}^3& \mathrm{in}\ \mathrm{case}\ \mathrm{d}<=\mathrm{m}\\ {}{\mathrm{c}}_{\mathrm{SIB}}=0\%& \mathrm{in}\ \mathrm{case}\ \mathrm{d}>=2\mathrm{r}+\mathrm{m}\\ {}{\mathrm{c}}_{\mathrm{SIB}}={\mathrm{V}}_{\mathrm{overlap}}/{\mathrm{V}}_{\mathrm{SIB}}& \mathrm{in}\ \mathrm{case}\ \mathrm{m}<\mathrm{d}<2\mathrm{r}+\mathrm{m}\\ {}\mathrm{Generally},{\mathrm{c}}_{\mathrm{SIB}}={\mathrm{r}}^3/{\left(\mathrm{r}+\mathrm{m}\right)}^3{\mathrm{c}}_{\mathrm{IPL}.}& \end{array}} $$

Dose volume histograms

For the calculation of DVHs, 5000 points within the IPL sphere of radius r were randomly and uniformely sampled (Monte Carlo). The points were parametrized as a triple (a, b, c). For each measured position (xi,yi,zi) of the prostate, it was checked if (a + xi)2 + (b + yi)2 + (c + zi)2 < = (r + m)^2 lay within the sphere covered by the SIB. If so, a unit of dose was added, and the received units of dose were divided by the number of samples, 39,925. Each of these dose percentages was then added to the histogram of 5000 such dose percentages which was plotted as the DVH. Again, this procedure was done for every combination of IPL radius r (3 mm, 5 mm, 7 mm) and safety margin m (0 mm, 2 mm, 5 mm).

Data handling, statistical analysis and plots were performed with Microsoft Excel and VBA.

Results

Intrafraction motion data

On average over all patients and fractions, the prostate was displaced by about 1.3 mm radially from its initial position. In some cases, the prostate drifted off by 1 cm or more, but during 95% of the time, it remained within 4.3 mm of its initial position, see Table 2.

Table 2 Recorded displacements of the prostate from its initial position

Volumetric coverage

Even without safety margins, a same-sized SIB would have covered 73, 82% resp. 87% of the volume of an IPL of radius 3 mm, 5 mm resp. 7 mm, on average. 95% of the time, it would have covered at least 11, 40% resp. 56% of the IPL volume. These numbers improve substantially for a moderately sized safety margin of 2 mm. In this case, average coverage would have been at least 95, and 95% of the time (and hence, dose) would have been applied to a majority of the IPL volume. Furthermore, with 5 mm safety margin, average coverage would be around 99, and 95% of the time the SIB would have completely covered the IPL volume. See Table 3.

Table 3 Volumetric coverage cIPL of the IPL by the SIB, in % of IPL volume

Safety margins, however, would also have had a downside in terms of dose efficiency resp. potentially toxicity. Due to the cubic power in the ratio between IPL volume (proportional to r^3) and SIB volume (proportional to (r + m)^3) the ratio would have fallen off quickly even in the best case of completely overlapping spheres. For example, in case of a 2 mm margin for a 5 mm radius lesion, the volume of the IPL would have been only 36% of the SIB, with two thirds of the boost dose going to surrounding non-lesion prostate tissue or even non-prostate tissue. The percentage of SIB dose that would have gone to the IPL is shown in Table 4 for the different situations.

Table 4 Volumetric coverage cSIB of the SIB by the IPL, in % of SIB volume

Dose volume histograms

The dose volume histograms draw a slightly more differentiated picture, see Fig. 2.

Fig. 2
figure 2

Dose volume histograms. Cumulative histogram of the dose volume relationship with dose on the horizontal axis (in percentage of maximum dose) and volume on the vertical axis (in percentage of IPL volume). Result of Monte Carlo simulations for different combinations of IPL radius (3 5 and 7 mm) and SIB safety margin (0 2 and 5 mm)

In all situations, 30% of the dose would have been received by 100% of the IPL volume, with our without safety margins.

Without safety margins, 50% of the dose would have been received by 91, 96% resp. 98% of the IPL volume of radius 3 mm, 5 mm resp. 7 mm. With safety margins of either 2 mm or 5 mm, that 50% dose would have been received by 100% of the IPL volume.

At 95% of dose, the size of the safety margin begins to matter. An IPL volume of 3 mm receives 0% (no safety margins), 42% (2 mm safety margin) resp. 100% (5 mm safety margin) of the dose. Similarly, the dose received by an IPL volume of 5 mm increases from 9% (no safety margin) to 83% (2 mm) to 100%. In case of an IPL volume of 7 mm, this receives already 31% of dose without any safety margins, and this value increases to 90% in case of 2 mm margins. See in the Additional file 1: Table S5.

Conversely, e.g. 95% of the IPL volume of radius 3 mm, 5 mm resp. 7 mm receive 44, 53% resp. 58% of the nominal SIB dose in the absence of safety margins. This percentage rises to over 90% in case of 2 mm margins and to 98% in case of 5 mm margins. These figures can be recovered by scaling the earlier results by (r + m)3/r3, as explained above.

Patient variability

The motility of the prostate differs in patients. While the average radial displacement of the prostate was 1.28 mm across all patients and fractions, it was as small as 0.50 mm in one patient (patient #13) and as large as 2.53 mm in another (patient #11). The median was 1.06 mm with quartiles 0.75 mm and 1.39 mm. Similarly, the maximum radial displacement of the prostate was only 2.68 mm in one patient (patient #27) and as large as 21.20 mm in another (patient #10). The median was 7.82 mm with quartiles 4.96 mm and 10.07 mm. These differences translate into the DHVs. To illustrate the two extreme cases, Fig. 3 shows the DVHs for patients #13 and #11 side by side:

Fig. 3
figure 3

Dose volume histograms for individual patients (extreme cases of minimal and maximal motility). As Fig. 1 but for individual patients. a The top plot shows the DVHs for patient #13 who featured a particular small average radial displacement of the prostate of 0.50 mm. b By contrast the bottom plot shows the DVHs for patient #11 with a particularly large average radial displacement of the prostate of 2.53 mm

Discussion

A recent review on IPL boosts [3] reports a mean of 2.4 cm3 for the mean/median IPL volumes reported by multiple studies. Due to the nonlinear relationship between radius and volume, this corresponds to a mean IPL radius probably well below 8 mm. In our simulation we use a cautious assumption of 3, 5, and 7 mm in order to highlight the impact of intrafraction motion on smaller targets.

Sing et al. [1] found that four IPLs in three patients were successfully targeted with a 3 mm margin. In our simulation, using a margin of 2 mm or more ensures at least 90% of prescribed dose to the entire IPL volume.

A large and recent study [7] reports that 284 patients with an additional SIB of 95 Gy to the IPL did not differ in toxicity from the base group of 287 patients receiving only 77 Gy to the entire prostate. Other studies show similar levels of tolerance [5, 6, 8]. Our data shows that the average displacement of the prostate is 1.3 mm. 95% of the time, the prostate remains within 4.3 mm of its initial position, notably without any mitigation (such as online correction of the table position or the multi leaf collimator). While this means that the SIB is not always centered on the IPL, the SIB will hit prostate tissue predominantly, depending on the location of the IPL within the prostate. Our simulation indicates that the IPL would receive the majority of the SIB in case of an IPL radius of at least 8 mm and a safety margin of at most 2 mm. Without safety margin (reducing the coverage of the IPL by the SIB), the coverage of the SIB by the IPL rises to 73% resp. 87% in case of an IPL radius of 3 mm resp. 7 mm.

In a planning study with and without simulated intrafraction motion [10] a safety margin of 7 mm was found to make both sequential or simultaneous boosts feasible. This is in agreement with our simulation showing that already a safety margin of 5 mm will yield 97% of the dose to 100% of the IPL volume.

Another planning study [9] concludes that even a hypofractionated SIB is feasible but ‘requires small margins needing intrafraction motion tracking’. This is in agreement with our simulation indicating that a safety margin of only around 2 mm might be a good compromise in terms of dose delivery to the IPL and toxicity. From other work [19, 20], we agree that online tracking and correction is the preferred approach to counter intrafraction motion.

However, there was considerable variability between patients. On the one hand, in a patient with little motility, an average displacement of the prostate of only 0.5 mm meant that 2 mm margins would have covered the complete lesion volume with at least 98% of the dose. On the other hand, in a patient with especially high motility, an average displacement of the prostate of 2.5 mm meant that 2 mm margins would have ensured only 72% of the total dose to 100% of the lesion volume. Here even a margin of 5 mm would ensure coverage with 90% of the prescribed dose to 69% of the volume of a lesion of radius 7 mm.

There are a couple of limitations to our study. Foremost, our simulation takes the recorded trajectories as ground truth and does not consider noise on top of them. Realistic margins will take all sources of error into account. The simplest way to do so (albeit not technically correct) would be to treat the intra-fraction motion as one of several sources of variance and simply add those up. The caveat lies in the fact that while other sources of error have constant variance, the variance of intra-fraction increases linearly in time [18, 19]. This leads to non-Gaussian kernels of dose convolution [21]. This also means that standard optimal margin recipes for inter-fractional motion, in so far as they assume a Gaussian distribution of displacements [22], are not perfectly valid for intra-fraction motion.

Also, our simulation does not take any interference between the motion of the apparatus (e.g. collimator leaves) and the motion of the prostate into account. The simulation simply assumes a constant and perfectly spherical dose distribution. We also did not consider any effects of target rotation or deformation. The IPL is assumed to be a constant and perfectly spherical volume. In comparison to more realistic 4D dose reconstruction techniques [14, 23] our approach is a simplification dictated by available data, i.e. tracking of the prostate center only.

Furthermore, an interesting question which is out of the scope of this simulation is the question how much dose a neighboring organ at risk receives due to intra-fraction motion. On the one hand, one could use our model for a simple calculation by looking at two spheres at a certain initial distance. However, a sphere is already an idealization of an IPL and is even less so a suitable proxy for the shape of organs at risk such as e.g. the rectum.

Conclusion

While sometimes the prostate strayed far from its initial position, on average over all fractions and patients, the dose coverage would have been acceptable even for small target volumes such as IPLs of radius 3 to 7 mm and narrow fields. Moderate safety margins of 2 mm could have ensured a delivery of 90% or more of the SIB dose to the IPL in the average patient. In this case, SIB volume would have been considerably larger than IPL volume, but still considerably smaller than the overall PTV of the prostate. However, there is considerable variability between patients, with some extreme cases requiring larger margins still for a satisfactory dose coverage.

Abbreviations

4DUS:

4D ultrasound; ultrasound imaging resolving both volume (3D) and time (1D)

DVH:

Dose volume histogram

IPL:

Intraprostatic lesion

SIB:

Simultaneous integrated boost

References

  1. Singh AK, Guion P, Sears-Crouse N, Ullman K, Smith S, Albert PS, et al. Simultaneous integrated boost of biopsy proven, MRI defined dominant intra-prostatic lesions to 95 gray with IMRT: early results of a phase I NCI study. Radiat Oncol. 2007;2:36.

    Article  Google Scholar 

  2. Rischke HC, Nestle U, Fechter T, Doll C, Volegova-Neher N, Henne K, et al. 3 tesla multiparametric MRI for GTV-definition of dominant Intraprostatic lesions in patients with prostate Cancer--an interobserver variability study. Radiat Oncol. 2013;8:183.

    Article  Google Scholar 

  3. von Eyben FE, Kiljunen T, Kangasmaki A, Kairemo K, von Eyben R, Joensuu T. Radiotherapy boost for the dominant Intraprostatic Cancer lesion-a systematic review and meta-analysis. Clin Genitourin Cancer. 2016;14:189–97.

    Article  Google Scholar 

  4. Zamboglou C, Klein CM, Thomann B, Fassbender TF, Rischke HC, Kirste S, et al. The dose distribution in dominant intraprostatic tumour lesions defined by multiparametric MRI and PSMA PET/CT correlates with the outcome in patients treated with primary radiation therapy for prostate cancer. Radiat Oncol. 2018;13:65.

    Article  Google Scholar 

  5. Pinkawa M, Piroth MD, Holy R, Klotz J, Djukic V, Corral NE, et al. Dose-escalation using intensity-modulated radiotherapy for prostate cancer - evaluation of quality of life with and without (18) F-choline PET-CT detected simultaneous integrated boost. Radiat Oncol. 2012;7:14.

    Article  CAS  Google Scholar 

  6. Sundahl N, De Meerleer G, Villeirs G, Ost P, De Neve W, Lumen N, et al. Combining high dose external beam radiotherapy with a simultaneous integrated boost to the dominant intraprostatic lesion: analysis of genito-urinary and rectal toxicity. Radiother Oncol. 2016;119:398–404.

    Article  Google Scholar 

  7. Monninkhof EM, van Loon JWL, van Vulpen M, Kerkmeijer LGW, Pos FJ, Haustermans K, et al. Standard whole prostate gland radiotherapy with and without lesion boost in prostate cancer: toxicity in the FLAME randomized controlled trial. Radiother Oncol. 2018;127:74–80.

    Article  Google Scholar 

  8. Timon G, Ciardo D, Bazani A, Marvaso G, Riva G, Volpe S, et al. Short-term high precision radiotherapy for early prostate cancer with concomitant boost to the dominant lesion: ad interim analysis and preliminary results of phase II trial AIRC-IG-13218. Br J Radiol. 2018;91:20160725.

    Article  Google Scholar 

  9. Tree A, Jones C, Sohaib A, Khoo V, van As N. Prostate stereotactic body radiotherapy with simultaneous integrated boost: which is the best planning method? Radiat Oncol. 2013;8:228.

    Article  Google Scholar 

  10. Abdellatif A, Craig J, Jensen M, Mulligan M, Mosalaei H, Bauman G, et al. Experimental assessments of intrafractional prostate motion on sequential and simultaneous boost to a dominant intraprostatic lesion. Med Phys. 2012;39:1505–17.

    Article  Google Scholar 

  11. Sihono DSK, Ehmann M, Heitmann S, von Swietochowski S, Grimm M, Boda-Heggemann J, Lohr F, Wenz F, Wertz H. Determination of Intrafraction prostate motion during external beam radiation therapy with a Transperineal 4-dimensional ultrasound real-time tracking system. Int J Radiat Oncol Biol Phys. 2018;101:136–63.

    Article  Google Scholar 

  12. Keall PJ, Aun Ng J, O'Brien R, Colvill E, Huang CY, Rugaard Poulsen P, Fledelius W, Juneja P, Simpson E, Bell L, Alfieri F, Eade T, Kneebone A, Booth JT. The first clinical treatment with kilovoltage intrafraction monitoring (KIM): a real-time image guidance method. Med Phys. 2015;42:354–8.

    Article  Google Scholar 

  13. Rosario T, van der Weide L, Admiraal M, Piet M, Slotman B, Cuijpers J. Toward planning target volume margin reduction for the prostate using intrafraction motion correction with online kV imaging and automatic detection of implanted gold seeds. Pract Radiat Oncol. 2018;8(6):422–8.

    Article  Google Scholar 

  14. Colvill E, Booth JT, O'Brien RT, Eade TN, Kneebone AB, Poulsen PR, Keall PJ. Multileaf collimator tracking improves dose delivery for prostate Cancer radiation therapy: results of the first clinical trial. Int J Radiat Oncol Biol Phys. 2015;92(5):1141–7.

    Article  Google Scholar 

  15. Murtaza G, Toftegaard J, Khan EU, Poulsen PR. Volumetric modulated arc therapy with dynamic collimator rotation for improved multileaf collimator tracking of the prostate. Radiother Oncol. 2017;122:109–15.

    Article  Google Scholar 

  16. Fast MF, O'Shea TP, Nill S, Oelfke U, Harris EJ. First evaluation of the feasibility of MLC tracking using ultrasound motion estimation. Med Phys. 2016;43:4628–33.

    Article  Google Scholar 

  17. Ipsen S, Bruder R, O’Brien R, Keall PJ, Schweikard A, Poulsen PR. Online 4D ultrasound guidance for real-time motion compensation by MLC tracking. Med Phys. 2016;43:5695–704.

    Article  Google Scholar 

  18. Ballhausen H, Li M, Ganswindt U, Belka C. Shorter treatment times reduce the impact of intra-fractional motion : a real-time 4DUS study comparing VMAT vs. step-and-shoot IMRT for prostate cancer. Strahlenther Onkol. 2018;194:664–74.

    Article  Google Scholar 

  19. Ballhausen H, Li M, Hegemann NS, Ganswindt U, Belka C. Intra-fraction motion of the prostate is a random walk. Phys Med Biol. 2015;60:549–63.

    Article  CAS  Google Scholar 

  20. Ballhausen H, Ganswindt U, Belka C, Li M. Intra-fraction motion of the prostate is not increased by patient couch shifts. Radiat Oncol. 2016;11:49.

    Article  Google Scholar 

  21. Ballhausen H, Reiner M, Kantz S, Belka C, Söhn M. The random walk model of intrafraction movement. Phys Med Biol. 2013;58(7):2413–27.

    Article  CAS  Google Scholar 

  22. van Herk M, Remeijer P, Lebesque JV. Inclusion of geometric uncertainties in treatment plan evaluation. Int J Rad Onc Biol Phys. 2002;52(5):1407–22.

    Article  Google Scholar 

  23. Poulsen PR, Schmidt ML, Keall P, Worm ES, Fledelius W, Hoffmann L. A method of dose reconstruction for moving targets compatible with dynamic treatments. Med Phys. 2012;39(10):6237–46.

    Article  Google Scholar 

Download references

Acknowledgements

We thank Patrick Dominik Thum, Andrea Beisel, Anne Kolberg, Gabriela Danilkiewicz and Anja Weber for their excellent work as radiation therapist.

Funding

Not applicable. No third party had any role in the design of the study and collection, analysis, and interpretation of data and in writing the manuscript.

Availability of data and materials

The datasets generated and/or analysed during the current study are not publicly available due to patient privacy concerns.

Author information

Authors and Affiliations

Authors

Contributions

CB suggested and designed the study. MR took care of the physics department and the measurement apparatus. ML took care of patients and physicians and gathered the data. HB performed data handling and calculations and wrote the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Hendrik Ballhausen.

Ethics declarations

Ethics approval and consent to participate

All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards. For this type of study formal consent is not required.

Consent for publication

Not applicable.

Competing interests

CB has a research cooperation with Elekta AB, Stockholm, Sweden. The other authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Additional file

Additional file 1:

Table S5. DVH, volume in % of IPL volume as a function of dose in % of nominal SIB dose. (DOCX 26 kb)

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Ballhausen, H., Li, M., Reiner, M. et al. Dosimetric impact of intrafraction motion on boosts on intraprostatic lesions: a simulation based on actual motion data from real time ultrasound tracking. Radiat Oncol 14, 81 (2019). https://doi.org/10.1186/s13014-019-1285-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13014-019-1285-1

Keywords