Twin machines validation for VMAT treatments using electronic portal-imaging device: a multicenter study

Abstract

Purpose

To verify the accuracy of volumetric arc therapy (VMAT) using the RapidArc™ device when switching patients from one single linear accelerator (linac) to a paired energy and mechanics "twin" linac without reoptimization of the original treatment plan.

Patients and Methods

Four centers using 8 linacs were involved in this study. Seventy-four patients previously treated with the 6MV photon RapidArc™ technique were selected for analysis, using 242 measurements. In each institution, all patients were planned on linac A, and their plans were verified both on linac A and on the twin linac B. Verifications were done using the amorphous silicium electronic portal imager (EPID) of the linacs and were analyzed with the EpiQa software (Epidos, Bratislavia, Slovakia). The gamma index formalism was used for validation with a double threshold of 3 % and 3 mm with a measurement resolution of 0.39 mm/pixel, and a smoothed resolution of approximately 2.5 mm.

Results

The number of points passing the gamma criteria between the measured and computed doses was 94.79 ± 2.57 % for linac A and 94.61 ± 2.46 % for linac B. Concerning the smoothed measurement analysis, 98.67 ± 1.26 % and 98.59 ± 1.20 % points passing the threshold were obtained for linacs A and B, respectively.

The difference between the 2 dose matrices acquired on the EPID was very small, with 99.92 ± 0.06 % of the points passing the criteria.

Conclusion

For linacs sharing the same mechanical and energy parameters, this study tends to indicate that patients may be safely switched from treatment with one linac to treatment with its twin linac using the same VMAT plan.

Background

New radiotherapy techniques aiming at improving target coverage and protection of organs-at-risk have emerged in the last decade. Intensity-modulated radiation therapy (IMRT) improved dosimetric results compared with 3-dimensional conformal radiotherapy (3D-CRT) by modulating the beam intensity for static gantry angles [14]. After a development phase followed by a validation process [5, 6], IMRT is now routinely implemented and available for a growing number of patients. New quality assurance procedures had to be designed for IMRT as the complexity of optimization algorithm and technical delivery were increasing. Software tools were developed for quality assurance [79] and for treatment planning systems [10, 11]. Guidelines on the technical and clinical aspects of the validation of IMRT treatments were also established [4, 1215].

Volumetric arc therapy (VMAT), commercialized under the names RapidArc (Varian) or VMAT (Elekta), is a refinement of standard IMRT which uses beam modulation combined with a continuous or discrete arc motion of the linac gantry around the patient. The resulting treatment plan is often more complex than that of the standard IMRT, which raises the question of the feasibility of switching a patient treated with a VMAT plan computed for a specific linac to a “twin” linac e.g. in case of repair or maintenance. While it has been shown possible for 3D-CRT and standard IMRT, no data is available regarding this possibility with VMAT. We thus conducted a study in order to compare the accuracy of the actual patients’ VMAT plans optimized for a specific linac to the same plans when radiotherapy is delivered on a “twin” linac.

Patients and methods

IMRT was implemented in four centers between 2001 and 2008, and more than 10 000 patients were treated using this technique since. Data was collected on the eight linacs (21 EX and 21 iX, Varian Medical System, Palo Alto, CA) used to deliver IMRT treatments in these centers. They were all equipped with 120 leaves multi-leaf collimators (MLC) adjusted and verified by the medical physicists so as to be equivalent regarding the “sliding-window” delivery technique. Between 2008 and 2010, RapidArc options were installed on the linacs and patients started to be treated with this modality. Otto et al. [16] showed that the RapidArc combines dose-rate modulation, gantry-speed modification and leaf position to create a high-dose modulation during a single or few rotations around the patient. Commissioning of the 8 linacs was carried out following the manufacturer’s suggestions [17]. Given the higher number of parameters to verify for the RapidArc technology compared to the static gantry technique, it was important to know if the patients could be switched from a linac for which the treatment plan had been specifically optimized for a “twin” linac. All linacs were equipped with amorphous silicon aS1000 (Varian) electronic portal imaging devices (EPIDs), which allow portal-imaging dosimetry for IMRT verification [18]. In this study, we used EPIDs to validate the linac matching and to investigate the possibility of switching a patient from one linac to another without the need of re-optimizing a new VMAT plan. At the time of the study, the V8 version of ARIA did not allow the RapidArc portal-dose predictions for comparisons with the image measured on the AS1000. We then decided to use the EpiQa software (Epidos, Bratislavia, Slovakia) for data analysis, based on previous publications [19, 20].

Patients’ selection and study design

Seventy-four patients previously treated with RapidArc for prostate (n = 10), head and neck (n = 30), pelvis (n = 20) or miscellaneous (n = 14) cancers in four French oncology centers (Institut régional du Cancer de Montpellier - ICM, Montpellier, Institut de Cancérologie de la Loire, Saint Etienne, and the ORLAM radiotherapy centers of Mâcon and Villeurbanne) were included in the study between 2008 and 2010, and accounted for 242 measurements. Consecutive patients were selected in each center, independently of the quality assurance results of their pre-treatment plan. Three centers each provided data for 10 prostate or pelvis cases plus for 10 head and neck cancer patients, whereas the fourth center provided the 14 miscellaneous cancer cases. The dosimetric plans were computed using Eclipse treatment planning system for a specific linac (linac A). All the patients’ plans were then verified through a quality assurance process on linacs A and B, the latter having been adjusted by the physics team to be similar during initial commissioning (see “Machine matching”). Rotational treatment plans optimized for linac A were delivered using linac A and linac B on a Silicium detector attached to the gantry. The EPID measured the collapsed plan dose in a 2-dimensional matrix, which was then compared to the one computed in Eclipse on a computer model of the phantom [19, 20]. The gamma index formalism [21] was used for validation with a threshold of 3 % and/or 3 mm, and the number of points reaching these criteria was recorded for comparison. The region of interest corresponded to the entire field with a 1 cm margin. A global gamma comparison method was chosen. Since the anisotropic analytical algorithm calculation grid was 2.5 mm and the measurement resolution was 0.39 mm/pixel, we adopted a smoothing analysis. A Gaussian curve was applied to the measurements in order to decrease the spatial resolution of the matrix to the one computed in Eclipse, resulting in an acquisition matrix of approximately 2.5 mm/pixel. A large range of photon energies (6, 8, 18 and 25 MV) and maximum dose rates (400 to 600 MU/min) were used to analyze if there was specific energy or dose rate dependence of the measurements. A comparison between the dose prediction and dose calculation for linacs A and B was carried out. The results of the plans irradiated with linacs A and B were then compared.

Treatment planning

Optimization and calculations were done at center one (20 patients) using the 8.0.5 version (PRO I) of the Eclipse treatment planning system (Varian, Palo Alto, USA), which only allowed a single arc. Results for the 10 prostate and the 10 head and neck cases were obtained in late 2008 and early 2009. The three other centers (2 to 4) used the 8.2.23 version of the Eclipse system, which includes the second version of the progressive resolution optimizer (PRO II) and allows an optimization for one or two arcs. Prostate treatments were administered with one arc whereas two arcs were used for other pathologies.

RapidArc optimization was performed using the new PRO algorithm that allows iterative changes to the dynamic delivery variables (i.e. MLC, dose rate and gantry angular velocity) via a set of penalty functions. These iterations were separated into five resolution levels. The first represented the full arc with 10 control points (mostly static fields). This number of control points was then doubled plus one for each successive resolution level, with the final arc including up to 177 control points. As each new control point was added, the dynamic variables were interpolated from the two neighboring points. The nature of this process meant that the lower resolution levels were flexible to optimization objective changes but gave a coarse representation of the full arc, while the higher levels were less flexible but gave a much more accurate representation of the full dynamic arc.

Due to the increased time needed for optimization using PRO I and PRO II, we adopted a strategy where the set of constraints for a specific patient was searched with the dose volume optimizer (DVO). The algorithm, also used for sliding-window IMRT, is based on fluence determination and any change in the constraints has a direct impact on the dose-volume histograms during the optimization. After a few iterations with the DVO, a personalized set of constraints for the specific patients was obtained, which was used to facilitate RapidArc optimization.

As opposed to the DVO algorithm, machine parameters were directly used during optimization with the PRO algorithm. This could have an impact on the final results. Indeed, the DVO algorithm generates a theoretical fluence map at the end of its optimization process, which is converted in an actual fluence by a sequencer including both mechanical and dosimetric linac parameters. The actual dose distribution was then computed on patient data using the predicted fluence and this often resulted in differences between the dose-volume histograms shown at the end of the optimization process and those obtained after the sequencer generated the fluence. With the PRO algorithm, the dose-volume histogram results at the end of the optimization process are usually very close to those obtained after dose calculation. Dose calculation on Eclipse used the anisotropic analytical algorithm with a calculation grid of 2.5 mm.

Machine matching

Many parameters were checked to consider that two linacs are matched together or form a “twin”-pair. The first concerns relative dosimetric data. Depth dose curves were identical (we applied a threshold of 0.5 %) and special attention was given to small fields, especially when the linacs were used for IMRT treatments. The dose profiles were also checked to be as similar as possible, and a focus on larger fields was necessary since tuning the beam to obtain identical profiles was more difficult for large fields. Dosimetric parameters were primordial factors for the matching procedure. Machine calibration was done in the same conditions and output factors did not differ by more than 0.5 %. After these goals were achieved and beam matching was obtained for 3D-CRT, matching for IMRT also needed a fine MLC tuning and leaf calibration to obtain equivalent results for both linacs. Sharing the same MLC leakage transmission, the calibration of the dosimetric leave gaps was mechanically adjusted to obtain less than 0.5 % difference for the same plan delivered on two different linacs. This method has previously been described in the literature [4, 22, 23]. Furthermore, leaf transmission and dosimetric leave gaps parameters were adjusted in the treatment planning system to obtain similar measurements for the same treatment plan, as described by Chauvet et al. [24].

Machine tests and patient quality assurance

Delivery measurements were acquired with the EPID of the linacs using amorphous silicium (AS1000) and analysed with the EpiQa software (Epidos, Bratislavia, Slovakia). This program allows the conversion of a dosimetric image acquired by an EPID into a dose map and to compare the dose map with a reference dose distribution. It is possible to use Epiqa for the verification of static and intensity modulated fields.

The portal dosimetry image conversion to dose map was based on the GLAaS algorithm as described by Nicolini et al. [19].

Technical quality assurance for linacs A and B was accomplished using the DICOM files provided by Varian and following the process defined by Ling et al. [17]. The accuracy of the leave movements was tested using the “sliding-window” and “picket fence” tests on different gantry positions [22, 23, 25]. The results obtained with linacs A and B were used to evaluate the quality of the matching (Fig. 1). These tests provided a more qualitative than quantitative analysis, and were similar to the film dosimetry quality assurance, which has been used for many years. A garden fence methodology was followed to check the quality of these results [22]. Specific tests were done using the RapidArc to verify the dose-rate, the gantry speed and the leaf speed variations during rotation as to be assured of the consistency of the two linacs.

Fig. 1
figure 1

Comparison tests for MLC and EPID calibrations. a Comparison of sliding window tests for verification of the dosimetric leave separation; (b) picket-fence tests for leave positioning accuracy. The profiles were taken perpendicularly following the cross-hair axes display

Regarding the treatment plans, once the dosimetry was finished, a quality control plan was realized using the GlAas algorithm formalism [19], commercially available in the Epiqa software (Epidos, Bratislava, Slovakia). A dose distribution on a virtual water phantom at the depth of the maximum of the depth dose curve and a distance of 100 cm to the source was calculated in Eclipse. This verification plan, using the original parameters (MLC and dose rate) was “collapsed” on an infinitesimal gantry rotation, generating a dose distribution of the whole arc on a single plane, perpendicular to the beam central axis. This dose distribution, calculated with a resolution of 2.5 mm/pixel, was exported and compared with the data acquired on the EPID (AS1000, Varian Medical System, Palo Alto, CA, USA) during the delivery of the real patients’ plan using the Epiqa software. The GlAas algorithm was configured to convert the images acquired without any build-up on the EPID cassette into a dose at the depth of the maximum of the depth dose curve (e.g. 1.4 cm for X6). The measurement consisted of an acquisition of the real treatment plan during the RapidArc delivery using the integrated mode of the EPID. The detector was positioned at the isocenter distance without any patient, phantom or couch inside the beam. [20] Spatial resolution was an important factor for the evaluation of the results. As the detector provided a resolution of 0.39 mm/pixel, lower than the 2.5 mm/pixel grid calculation in Eclipse, we performed a second analysis using a Gaussian smoothing method of 2 mm provided by the Epiqa software, in order to have a comparison with similar spatial resolution. Our tolerance level for acceptance of a treatment plan was a 3 %-3 mm gamma index threshold with an acceptance value of 95 % of the point passing these criteria. For cases below the 95 % acceptance level, qualitative analysis could still lead to a quality assurance validation. Using this data, we first compared linacs A and B of each center calculating the difference between dose-calculation and dose-measurement for the same treatment plan. Second, we compared the treatment delivery using the measurements made with linac A to those made with linac B. This evaluation excluded all treatment planning system algorithm configurations and was purely an indicator of the linac beam, MLC and EPID parameters calibrations.

Results

Analysis of the “sliding window” and the “picket fence” tests were made qualitatively by looking at the curve correspondence on portal measurements. It showed a good agreement between linacs A and B (Fig. 1). We took a series of mechanical and imaging tests of the MLC using the same DICOM file on the twin linacs, showing equivalent results. First, the sliding window IMRT files were used to measure absolute dose with an ionisation chamber in a PMMA phantom to ensure proper calibration of the leaves on each MLC. Results were within 1%. We then performed portal IMRT tests (e.g. chair, step wedges or pyramid) and the difference for a gamma 3 %-3 mm between the 2 linacs was less than 1 %. The specific RapidArc quality assurance test results were concordant with those described by Jorgensen et al. [26].

Concerning the patients treatment plans, the comparison between measured and computed dose showed 94.79 ± 2.57 % of points passing the gamma criteria for linac A and 94.61 ± 2.46 % for linac B (Fig. 2a). For the smooth analysis, we reached 98.67 ± 1.26 % and 98.59 ± 1.20 % for linacs A and B, respectively (Fig. 2b). Regarding the ten head and neck patients, the results were 94.03 ± 2.08 % and 94.20 ± 2.10 % with a measurement grid of 0.39 mm for linacs A and B, respectively. Once again, a smoothed analysis improved the results: 98.51 ± 1.39 % (A) and 98.59 ± 1.12 % (B) of points passing a gamma ≤1 for 3 % or 3 mm.

Fig. 2
figure 2

Global comparison of the predicted and effective dose distribution for the twin linacs with (a) and without (b) smoothing

For the prostate cancer patients, for whom the modulation was much simpler, the results were 99.00 ± 0.48 % for linac A, and 97.98 ± 0.59 % for linac B without smoothing even for the plans optimized using the PRO version (center 1). The results using the smoothing method were 99.77 ± 0.31 % and 99.84 ± 0.16 % for the linacs A and B, respectively. For patients with pelvic cancer, results were 94.13 ± 2.44 % and 98.26 ± 1.19 % for linac A, and 93.99 ± 231 % and 98.12 ± 1.23 % for linac B, before and after smoothing, respectively. For centers 2, 3 and 4, the quality assurance results on linac A compared to those on linac B were good with a maximum difference of 1.9 % in the values obtained (Fig. 3b, c, and d). For center 1, the maximal difference was 6.4 % for a head and neck treatment delivered with a single arc. This difference can be explained by the specificity of verification using collapsed plans on 2D views and because of an over modulation measured by the portal imager (peaks on the blue line on Fig. 4), especially for this complex case with a high modulation level. Adding a second arc decreased the modulation level by arc, and consequently allowed an improved result.

Fig. 3
figure 3

Comparison of the predicted and effective dose distribution for the twin linacs by center. The X- and Y-axes represent the gamma index result for a 3 %-3 mm analysis for linacs A and B, respectively, for a) center 1, b) center 2, c) center 3 and d) center 4

Fig. 4
figure 4

Over-modulation on measurement before smoothing. Profiles were taken perpendicularly following the cross-hair axes display

Fig. 5
figure 5

Comparison of the dose measurements between the twin machines: global (a) and per center (b) results. Each dot on a ray represents a value for patient measurement. The percentages of points passing a gamma index of 3 %-3 mm are represented by the different circle levels

All plans but one passed the criteria of 3 % - 3 mm for 95 % of the points analyzed with the smooth evaluation for the twin machines in the four institutes (Fig. 2b).

Regarding the dose distribution comparison test for the same treatment plans acquired on different portal devices, the differences between linacs A and B were very small, 99.92 ± 0.06 % of points passing the criteria (Fig. 5a). In this analysis, raw data with a 0.39 mm/pixel resolution was used. This test is the most robust to compare two matched linacs and showed excellent results for all the participating centers (Fig. 5b).

Discussion

IMRT has the ability to deliver highly conformal dose distributions to complex targets. In both static and rotational IMRT, many parameters such as the leaf position, dose rate and gantry speed have to be accurately controlled by the treatment planning system in order to achieve such coverage. Consequently, IMRT quality assurance procedures have to be specifically developed, using ionisation chambers, films or a 2D array [12, 27]. Although VMAT represents an evolution of IMRT, the quality assurance process remains similar. Films and 2D arrays are mostly used to compare the predicted and measured doses, with results reaching the acceptation criteria usually admitted [2831]. More recently, the use of EPIDs has made the quality assurance easier to set up by taking out the need to use phantoms [32]. This methodology, concerning IMRT quality assurance, was adopted by many institutions [20, 3235] Validation of RapidArc treatments with EPIDs on 3 different matched beam using Varian linacs was previously described by Fredh et al. [36]. The effect of the gantry movement was not found to significantly affect the results leading to the use of EPIDs as a detector even in the rotational mode. The results were quite similar even after segmenting the 360° arc into multiple sub-arcs of 6° and 12° to “collapse” the irradiation into a 2D measurement [20]. The consistency and reproducibility of VMAT plan delivery were shown with 3 different quality assurance systems by Chandraraj et al. [37].

Our study focused on the capacity of twin linacs to deliver the same dose distribution from the same VMAT plan. The algorithm version of the optimiser was not included in this comparison; only dosimetric and mechanical tunings of the linacs were compared. Our results showed good and equivalent results for all the centers, allowing the possibility for a patient to be switched from one linac to another without the need of repositioning a new plan, provided that the beams and MLC are perfectly matched.

The percentage of points passing the gamma criteria of 3 %-3 mm were similar to other studies using EPID or films for quality assurance [38, 39], even if another study has reported better results [40]. The value of the gamma index highly depends on the spatial resolution used for the analysis. Most of the publications showing overall gamma values better than 99 % have used 2D matrixes with a spatial resolution ranging from 5 to 10 mm, whereas EPIDs have a sub-millimeter resolution of 0.39 mm/pixel. We performed a smoothed analysis to render the resolution closer to a calculated planar dose. Doing so, the results of the gamma index were significantly improved. Another method could have been to increase the spatial resolution of the calculation grid in Eclipse to the lowest value proposed of 1 mm, but the commercial hardware actually available could not carry out this calculation on a real patient volume.

Conclusion

This study carried out on 8 linacs in 4 centers and for different cancer types tends to indicate that it is possible to treat a patient with the same VMAT plan on twin linacs, provided that the linacs are perfectly matched.

References

  1. Boehmer D, Bohsung J, Eichwurzel I, Moys A, Budach V. Clinical and physical quality assurance for intensity modulated radiotherapy of prostate cancer. Radiother Oncol. 2004;71:319–25.

    Article  PubMed  Google Scholar 

  2. McNair H, Adams E, Clark C, Miles E, Nutting C. Implementation of IMRT in the radiotherapy department. Brit J Radiol. 2003;76:850.

    Article  PubMed  CAS  Google Scholar 

  3. Miles E, Clark C, Urbano M, Bidmead M, Dearnaley DP, Harrington KJ, et al. The impact of introducing intensity modulated radiotherapy into routine clinical practice. Radiother Oncol. 2005;77:241–6.

    Article  PubMed  Google Scholar 

  4. Van Esch A, Bohsung J, Sorvari P, Tenhunen M, Paiusco M, Iori M, et al. Acceptance tests and quality control (QC) procedures for the clinical implementation of intensity modulated radiotherapy (IMRT) using inverse planning and the sliding window technique: experience from five radiotherapy departments. Radiother Oncol. 2002;65:53–70.

    Article  PubMed  Google Scholar 

  5. Adams E, Convery D, Cosgrove V, McNair HA, Staffurth JN, Vaarkamp J, et al. Clinical implementation of dynamic and step-and-shoot IMRT to treat prostate cancer with high risk of pelvic lymph node involvement. Radiother Oncol. 2004;70:1–10.

    Article  PubMed  Google Scholar 

  6. Clark C, Mubata C, Meehan C, Bidmead AM, Staffurth J, Humphreys ME, et al. IMRT clinical implementation: Prostate and pelvic node irradiation using Helios and a 120-leaf multileaf collimator. J Appl Clin Med Phys. 2002;3:273.

    Article  PubMed  CAS  Google Scholar 

  7. Kutcher G, Coia L, Gillin M, Hanson WF, Leibel S, Morton RJ, et al. Comprehensive QA for radiation oncology: report of AAPM radiation therapy committee task group 40. Med Phys. 1994;21:581–618.

    Article  PubMed  CAS  Google Scholar 

  8. Klein E, Hanley J, Bayouth J, Yin FF, Simon W, Dresser S, et al. Task Group 142 report: Quality assurance of medical accelerators. Med Phys. 2009;36:4197.

    Article  PubMed  Google Scholar 

  9. Bayouth J, Wendt D, Morrill S. MLC quality assurance techniques for IMRT applications. Med Phys. 2003;30:743.

    Article  PubMed  CAS  Google Scholar 

  10. Van Dyk J. Quality assurance of radiation therapy planning systems: current status and remaining challenges. Int J Radiat Oncol Biol Phys. 2008;71:23–7.

    Article  Google Scholar 

  11. Fraass B, Doppke K, Hunt M, Kutcher G, Starkschall G, Stern R, et al. American Association of Physicists in Medicine Radiation Therapy Committee Task Group 53: quality assurance for clinical radiotherapy treatment planning. Med Phys. 1998;25:1773.

    Article  PubMed  CAS  Google Scholar 

  12. Alber M. Guidelines for the verification of IMRT. ESTRO Booklet;7. 2008 - Estro Brussels, Belgium.

  13. Ezzell G, Galvin J, Low D, Palta JR, Rosen I, Sharpe MB, et al. Guidance document on delivery, treatment planning, and clinical implementation of IMRT: Report of the IMRT subcommittee of the AAPM radiation therapy committee. Med Phys. 2003;30:2089.

    Article  PubMed  Google Scholar 

  14. Ailleres N, Azria D, Thezenas S, Barbier N, Fenoglietto P, Delard R, et al. Pilot study of conformal intensity modulated radiation therapy for localized prostate cancer. Cancer Radiother. 2004;8:59.

    Article  PubMed  CAS  Google Scholar 

  15. Fenoglietto P, Laliberté B, Aillères N, Riou O, Dubois J-B, Azria D. Eight years of IMRT quality assurance with ionization chambers and film dosimetry: experience of the montpellier comprehensive cancer center. Radiat Oncol. 2011;6:1–11.

  16. Otto K. Volumetric modulated arc therapy: IMRT in a single gantry arc. Med Phys. 2008;35:310.

    Article  PubMed  Google Scholar 

  17. Ling C, Zhang P, Archambault Y, Bocanek J, Tang G, LoSasso T. Commissioning and quality assurance of RapidArc radiotherapy delivery system. Int J Radiat Oncol Biol Phys. 2008;72:575–81.

    Article  PubMed  Google Scholar 

  18. Van Esch A, Depuydt T, Huyskens D. The use of an aSi-based EPID for routine absolute dosimetric pre-treatment verification of dynamic IMRT fields. Radiother Oncol. 2004;71:223–34.

    Article  PubMed  Google Scholar 

  19. Nicolini G, Fogliata A, Vanetti E, Clivio A, Cozzi L. GLAaS: An absolute dose calibration algorithm for an amorphous silicon portal imager. Applications to IMRT verifications. Med Phys. 2006;33:2839.

    Article  PubMed  Google Scholar 

  20. Nicolini G, Vanetti E, Clivio A, Fogliata A, Korreman S, Bocanek J, et al. The GLAaS algorithm for portal dosimetry and quality assurance of RapidArc, an intensity modulated rotational therapy. Radiat Oncol. 2008;3:24.

    Article  PubMed  PubMed Central  Google Scholar 

  21. Low DA, Harms WB, Mutic S, Purdy JA. A technique for the quantitative evaluation of dose distributions. Med Phys. 1998;25:656.

    Article  PubMed  CAS  Google Scholar 

  22. LoSasso T, Chui CS, Ling CC. Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy. Med Phys. 1998;25:1919.

    Article  PubMed  CAS  Google Scholar 

  23. LoSasso T, Chui CS, Ling CC. Comprehensive quality assurance for the delivery of intensity modulated radiotherapy with a multileaf collimator used in the dynamic mode. Med Phys. 2001;28:2209.

    Article  PubMed  CAS  Google Scholar 

  24. Chauvet I, Petitfils A, Lehobey C, Kristner JY, Brunet Y, Lembrez R, et al. The sliding slit test for dynamic IMRT: a useful tool for adjustment of MLC related parameters. Phys Med Biol. 2005;50:563.

    Article  PubMed  CAS  Google Scholar 

  25. Essers M, de Langen M, Dirkx MLP, Heijmen BJM. Commissioning of a commercially available system for intensity-modulated radiotherapy dose delivery with dynamic multileaf collimation. Radiother Oncol. 2001;60:215–24.

    Article  PubMed  CAS  Google Scholar 

  26. Jo-rgensen M, Hoffmann L, Petersen Jr B, Praestegaard L, Hansen R, Muren L. Tolerance levels of EPID-based quality control for volumetric modulated arc therapy. Med Phys. 2011;38:1425.

  27. Alber M, Broggi S, De Wagter C, et al. Guidelines for the verification of IMRT. ESTRO Booklet;7. 2008 - Estro Brussels, Belgium.

  28. Korreman S, Medin J, Kjaer-Kristoffersen F. Dosimetric verification of RapidArc treatment delivery. Acta Oncol. 2009;48:185–91.

    Article  PubMed  CAS  Google Scholar 

  29. Lagerwaard FJ, Meijer OWM, Van der Hoorn EAP, Verbakel WFAR, Slotman BJ, Senan S. Volumetric modulated arc radiotherapy for vestibular schwannomas. Int J Radiat Oncol Biol Phys. 2009;74:610–15.

  30. Van Esch A, Clermont C, Devillers M, Iori M, Huyskens DP. On-line quality assurance of rotational radiotherapy treatment delivery by means of a 2D ion chamber array and the Octavius phantom. Med Phys. 2007;34:3825.

    Article  PubMed  Google Scholar 

  31. Feygelman V, Opp D, Javedan K, Saini A, Zhang G. Evaluation of a 3D diode array dosimeter for helical tomotherapy delivery QA. Med Dosim. 2010;35:324–9

  32. Bakhtiari M, Kumaraswamy L, Bailey D, de Boer S, Malhotra H, Podgorsak M. Using an EPID for patient-specific VMAT quality assurance. Med Phys. 2011;38:1366–73.

  33. van Elmpt W, McDermott L, Nijsten S, Wendling M, Lambin P, Mijnheer B. A literature review of electronic portal imaging for radiotherapy dosimetry. Radiother Oncol. 2008;88:289–309.

    Article  PubMed  Google Scholar 

  34. Vieillot S, Azria D, Lemanski C, Moscardo CL, Gourgou S, Dubois JB. et al. Plan comparison of volumetric-modulated arc therapy (RapidArc) and conventional intensity-modulated radiation therapy (IMRT) in anal canal cancer. Radiat Oncol. 2010;5:92.

  35. Llacer-Moscardo C, Quenet F, Azria D, Fenoglietto P. Feasibility study of volumetric modulated arc therapy for the treatment of retroperitoneal sarcomas. Radiat Oncol. 2010;5:83.

  36. Fredh A, Korreman S, Rosenschöld PM. Automated analysis of images acquired with electronic portal imaging device during delivery of quality assurance plans for inversely optimized arc therapy. Radiot Oncol. 2010;94:195–98.

  37. Chandraraj V, Stathakis S, Manickam R, Esquivel C, Supe SS, Papanikolaou N. Consistency and reproducibility of the VMAT plan delivery using three independent validation methods. J Appl Clin Med Phys. 2010;12:3373

  38. Pesce GA, Clivio A, Cozzi L. et al. Early clinical experience of radiotherapy of prostate cancer with volumetric modulated arc therapy. Radiat Oncol. 2010;5:54.

  39. Hsu F, Carolan H, Nichol A, Cao F, Nuraney N. Lee R. et al. Whole brain radiotherapy with hippocampal avoidance and simultaneous integrated boost for 1–3 brain metastases: a feasibility study using volumetric modulated arc therapy. Int J Radiat Oncol Biol Phys.2010;76:1480–85.

  40. Wagner D, Vorwerk H. Two years experience with quality assurance protocol for patient related Rapid Arc treatment plan verification using a two dimensional ionization chamber array. Radiat Oncol. 2010;6:21.

Download references

Acknowledgements

The authors thank Dr Benoit Laliberté and Dr Hélène de Forges for editorial help.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to P. Fenoglietto.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

PF designed and coordinated the study. Patient accrual and clinical data collection were done by PF, MK, DN, FJ. Data analysis, physics data and treatment planning data collection were done by PF, who also drafted the manuscript. NA and PF revised the manuscript critically for important intellectual content. All authors read and approved the final manuscript.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Fenoglietto, P., Khodri, M., Nguyen, D. et al. Twin machines validation for VMAT treatments using electronic portal-imaging device: a multicenter study. Radiat Oncol 11, 2 (2016). https://doi.org/10.1186/s13014-015-0577-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13014-015-0577-3

Keywords