Pretreatment patient-specific quality assurance prediction based on 1D complexity metrics and 3D planning dose: classification, gamma passing rates, and DVH metrics

Abstract

Purpose

Highly modulated radiotherapy plans aim to achieve target conformality and spare organs at risk, but the high complexity of the plan may increase the uncertainty of treatment. Thus, patient-specific quality assurance (PSQA) plays a crucial role in ensuring treatment accuracy and providing clinical guidance. This study aims to propose a prediction model based on complexity metrics and patient planning dose for PSQA results.

Materials and methods

Planning dose, measurement-based reconstructed dose and plan complexity metrics of the 687 radiotherapy plans of patients treated in our institution were collected for model establishing. Global gamma passing rate (GPR, 3%/2mm,10% threshold) of 90% was used as QA criterion. Neural architecture models based on Swin-transformer were adapted to process 3D dose and incorporate 1D metrics to predict QA results. The dataset was divided into training (447), validation (90), and testing (150) sets. Evaluation of predictions was performed using mean absolute error (MAE) for GPR, planning target volume (PTV) HI and PTV CI, mean absolute percentage error (MAPE) for PTV D95, PTV D2 and PTV Dmean, and the area under the receiver operating characteristic (ROC) curve (AUC) for classification. Furthermore, we also compare the prediction results with other models based on either only 1D or 3D inputs.

Results

In this dataset, 72.8% (500/687) plans passed the pretreatment QA under the criterion. On the testing set, our model achieves the highest performance, with the 1D model slightly surpassing the 3D model. The performance results are as follows (combine, 1D, and 3D transformer): The AUCs are 0.92, 0.88 and 0.86 for QA classification. The MAEs of prediction are 0.039, 0.046, and 0.040 for 3D GPR, 0.018, 0.021, and 0.019 for PTV HI, and 0.075, 0.078, and 0.084 for PTV CI. Specifically, for cases with 3D GPRs greater than 90%, the MAE could achieve 0.020 (combine). The MAPE of prediction is 1.23%, 1.52%, and 1.66% for PTV D95, 2.36%, 2.67%, and 2.45% for PTV D2, and 1.46%, 1.70%, and 1.71% for PTV Dmean.

Conclusion

The model based on 1D complexity metrics and 3D planning dose could predict pretreatment PSQA results with high accuracy and the complexity metrics play a leading role in the model. Furthermore, dose-volume metric deviations of PTV could be predicted and more clinically valuable information could be provided.

Introduction

Highly modulated radiation therapy plan offers highly conformal dose distributions to the target volume while maintaining steep dose gradients around the target area. Considering the planning and delivery complexity, patient-specific quality assurance (PSQA) measurements are regularly implemented before the first treatment fraction of intensity modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT) as an important clinical routine [1, 2]. These measurements serve multiple important purposes, including validating dosimetric calculations, confirming accurate data transfer, and ensuring machine deliverability [3]. PSQA typically required the complete delivery of the patient’s treatment plan [4, 5], thus, performing measurements for every patient can be resource-intensive, and the actual benefit of these measurements for certain patients remains unclear [6, 7]. Consequently, in busy healthcare institutions, these measurements are typically conducted outside of regular clinical hours and it may be late and hard to react to failed plans. Therefore, the prediction of pretreatment PSQA results would greatly benefit radiotherapy institutions by quickly identifying treatment plans that may require actual measurements and understanding the potential impact of errors in the delivery.

PSQA commonly employs gamma analysis to quantify the agreement between measured and planned dose distributions [8, 9], and gamma passing rate (GPR) provides information about the overall similarity of the total dose or dose within a region of interests (ROI). The TG 218 report recommends 95% or 90% as tolerance of action limits under the 3%/ 2 mm gamma criterion [2]. Moreover, as clinicians are concerned, dose volume histogram (DVH) metrics can offer a more informative depiction of dose discrepancies for clinical decision [10, 11]. Many studies have shown the relevance of DVH metrics to tumor control and normal tissue sparing [12,13,14], and the ability of DVH to detect clinically relevant dose errors was also revealed [15]. However, there is still a lack of studies that specifically focus on the prediction of DVH metrics.

Previous researches have explored the use of machine learning techniques to predict the outcomes of PSQA in radiotherapy [16,17,18,19]. Several studies have reported QA classification results, achieving a macro area under the curve (AUC) value up to 0.88 [3, 21, 22], These studies highlight the potential of machine learning models in prioritizing treatment plans that would benefit the most from PSQA. Furthermore, for GPR prediction, Valdes et al. demonstrated the potential of machine learning algorithms by utilizing complexity metrics to predict the GPR under a 3%/3mm criterion, with maximum prediction errors smaller than 3%. In a multi-institutional validation study, approximately 86% of the plans had a prediction error smaller than 3.5% [20, 21]. Other studies have investigated the use of deep learning algorithms to predict 2D GPR using flux maps, planar dose distributions, and 3D planning dose distributions [22,23,24,25,26]. Prediction results varied based on the chosen gamma criterion. Generally speaking, under a 3% dose difference criterion, the mean absolute error (MAE) was smaller than 2%. However, there are still some limitations. Firstly, the predictions of GPR were mostly based on 2D GPR, there is a lack of research on more informative 3D GPR. Additionally, the inputs to these models are usually unimodal, only a few researches included partial aperture related plan complexity or volume indices in addition to dose characteristics and get good performance [17, 22]. This suggests that the integration of dose characteristics with 1D metrics is advantageous, however, their study only incorporates a limited set of 1D metrics, indicating the necessity of including more comprehensive measures, such as transfer related plan complexities.

The Swin-Transformer is an innovative deep learning architecture designed to tackle complex vision tasks. It introduces a hierarchical structure where small and efficient transformers are used to process local image patches, and then global transformers are employed to capture long-range dependencies among these patches. In comparison to conventional convolutional neural network, this design allows Swin-Transformer to efficiently handle long-range dependency among various image scales and achieve state-of-the-art performance across a spectrum of medical image analysis tasks [27]. Its attention mechanism also facilitates the capture of intricate dose patterns which may reflect the nuances in dose delivery variability and are well-suited for PSQA prediction.

This is the first study introducing a novel deep learning architecture Swin-transformer to develop a PSQA prediction model that incorporates the planning dose and complexity metrics. In addition to the results of gamma analysis, DVH metrics in target volume are also predicted to provide more comprehensive information for clinical practice. Furthermore, the predictive capabilities of 3D planning dose and 1D complexity metrics were compared.

Materials and methods

Data collection

Patients who received treatment at our institution from March 2022 to April 2023 were included in this study. The patient plans were generated using Eclipse treatment planning system (TPS) (Varian Medical Systems, Palo Alto, CA, version 15.6) or Pinnacle TPS (Philips Radiation Oncology Systems, Fitchburg, WI, version 16.2) and delivered using a Varian IX linear accelerator with a 6MV photon-beam. PSQA was scheduled using the ArcCHECK phantom from Sun Nuclear Corporation (Melbourne, FL, USA). Prior to the study, accelerator commissioning and ArcCHECK phantom configuration were performed following the vendor’s standard procedures. The measurement results obtained from the annular detector matrix of the ArcCHECK phantom were utilized to reconstruct simulated 3D dose distributions on the patient’s CT scans using the planned dose perturbation (PDP) algorithm, which was matched with 3DVH software version 3.0, which is an additional tool to ArcCHECK [28]. The workflow of this research is illustrated in Fig. 1.

Plans with insufficient target coverage resulting from OAR overlap and measurement files with excessive hot or cold points during reconstruction were excluded from the study. Eventually, a total of 687 IMRT and VMAT plan data were successfully reconstructed and used for subsequent model development and verification. Detailed plan information is provided in Table 1. Global GPRs were calculated using a 3%/2mm criterion with a 10% threshold, and a GPR value of 90% was employed to determine QA pass or fail. For DVH metrics, considering the variation in treatment sites among the patients, only metrics of PTV were calculated, which were PTV D95 (minimum dose received by 95% of the planning target volume), PTV D2, PTV Dmean (mean dose received by planning target volume), homogeneity index (HI) and conformity index (CI).

Fig. 1
figure 1

Workflow of this study. The metrics in the orange boxes are utilized as 1D predictive input, while the planning dose is utilized as 3D input to establish 3D only predictive models. And both of them are utilized as multimodal inputs for combined model. ACPDP: ArcCHECK planned dose perturbation algorithm

Table 1 Plan characteristic distributions

A total of 687 plans were utilized to calculate the complexity metrics, following the methodology described in previous studies [29, 30]. These complexity metrics encompassed information regarding the machine unit, leaf aperture, and leaf movement. Regular QA for linear accelerators (linacs) is performed as reported in TG142 [31], and the linac QA metrics on or closest to the measuring day were also recorded, such as absolute dose variation, flatness and symmetry. Additionally, certain dosimetric parameters of the plan were considered in the model, including the HI and CI of the target volume, the volume of the PTV, and the prescription dose. Thus, a total of 71 one-dimensional metrics were incorporated into the model, and the specific metrics are outlined in Supplementary Materials Table S1.

Combined model establishing

Given the diverse treatment sites of the patients, the sizes of the planning dose grid varied within the range of 63 × 68 × 62 to 336 × 259 × 263. Additionally, there were two sizes of grid spacing: 2.5 × 2.5 × 2.5 mm3 and 3 × 3 × 3 mm3. To ensure consistency and facilitate analysis, the dose data underwent preprocessing and scaling, resulting in a standardized size of 192 × 192 × 192 before entering the model. As for the 71 1D metrics, normalization was applied using the Z-score method, enabling effective concatenation within the model. In this study, we develop a novel combined architecture based on the Swin-transformer to effectively fuse multimodal inputs [32]. We employed 2 Swin-Transformer blocks, with 4 and 8 heads incorporated in each block. The 3D Swin-transformer block was employed to process the planning doses of patients. Subsequently, the extracted features from the 3D dose were subjected to an average pooling layer, resulting in a one-dimensional feature vector with 256 elements. On the other side, 71 1D inputs were processed through a multi-layer perceptron (MLP) with a hidden layer of 128 units to obtain another 256 1D features. These two types of features were then combined and directly used for multi-task prediction. In the Swin-Transformer block, the MLP was configured with 2 hidden layers, each having 2 times the input dimension units. Dropout layers were incorporated into the model in both 1D metrics processing and Swin-Transformer blocks to prevent overfitting. The architecture of the network employed in this study is illustrated in Fig. 2.

Fig. 2
figure 2

Network architecture utilized in this study. (a) The overall architecture of the combined model. (b) Reprocess block of the input image, including rescale and a convolution layer for all backbones. (c) The simplified Swin-transformer block of 3D version, consisting of two successive layers. (d), (e) The 3D residual blocks and U-net Encoder utilized in this study for comparison. The number of n or 2n means output channels of the convolution layer and n is the input channel of the block, the s means stride of the convolution kernel or pooling kernel. Norm: Normalization layer, in (b), (d) and (e) means Batch Normalization, in (c) means Layer Normalization. MSA: Multi-head Self-attention. MLP: Multilayer Perceptron

The dataset was split into train, validation and testing. The validation dataset was utilized to fine-tune the hyperparameters of the model during the training phase, and the independent test set was used only once for testing after the model was developed. To assess the regression model’s performance in predicting DVH metrics and GPR, the mean square error (MSE) loss function was employed. For QA classification task, the binary cross-entropy (BCE) loss function was utilized. The overall loss function was defined as the sum of these three individual losses, as depicted in Eq. 1.

$$\begin{array}{l}Total\,loss = \frac{1}{N}\sum\limits_{i = 1}^N {{{(\gamma - \hat \gamma )}^2}} + \frac{1}{N}\sum\limits_{i = 1}^N {\sum\limits_{j = 1}^n {{{(D - \hat D)}^2}} } \\\,\,\,\,\,\,\, - \frac{1}{N}\sum\limits_{i = 1}^N ( {y^i}{\rm{log}}{{\hat y}^i} + (1 - {y^i}){\rm{log}}(1 - {{\hat y}^i}))\end{array}$$
(1)

Note: N: total number of samples, n: number of DVH metrics, γ: gamma passing rate, D: DVH metrics, y: the probability of passing the QA criteria, means the label of the variable.

Finally, we employed the determined optimal hyperparameters to train the model using a combined dataset consisting of the 447 training samples and 90 validation samples and the performance of the model was evaluated on the independent test set comprising 150 cases. The evaluation of predictions was based on the MAE and the AUC of ROC curve. The proposed deep network architecture was implemented using PyTorch [33] and executed on a NVIDIA GeForce RTX 3090Ti GPU with 24GB memory.

Other models

Additionally, we also calculated and compared the results obtained using either the 3D dose or the 1D metrics alone (shown in Fig. 1) to assess the individual predictive capabilities of these components as well as the combined model. For 3D model, we adopted commonly used network models in medical dose process, such as ResNet and U-Net encoder, the architectures are shown in Fig. 2. And for 1D model based on 1D metrics, we adopted a three-layer MLP with 2 hidden layers of 128 and 256 units. These training processes were on the same dataset as before. For clinical practice, the results of predicting QA classification directly or determining by predicting GPR were also compared using sensitivity and specificity.

Results

Hyperparameters determination

After conducting a search for optimal hyperparameters on a dedicated validation set, we determined that the initial learning rate for the 3D Swin-transformer parameters should be set to 3 × 10− 7 and RMSProp (root mean square propagation) optimizer was used. For the remaining parameters of linear layers, the initial learning rate was set to 3 × 10− 6. As the training progressed, both learning rates decayed with a dropping rate of 0.98 per 4 epochs. The weight decays of the parameters were set to 10− 6 for the 3D Swin-transformer and 10− 5 for the other parameters. The Swin-transformer and MLP were trained simultaneously. To train the prediction model, we utilized a batch size of 4, a dropout rate of 0.2, and an epoch number of 200. Subsequently, these hyperparameters were employed to build the prediction model using the 537 cases from the training and validation sets.

GPR and DVH metrics prediction

Figure 3 illustrates the distribution of absolute errors or absolute percentage errors in the predicted 3D GPRs and the DVH metrics across different ranges of TPS deviations of combined model. The regions characterized by higher GPRs and smaller absolute deviations in DVH metrics display higher prediction accuracy. 60% (90/150) of the plans exhibited predicted percent deviations of PTV D95 within 1%, and 49% (73/150) of the plans demonstrated predicted GPRs with errors within 2%. In the supplementary materials, Figure S.1-S.3 showed the result distributions of other models. Furthermore, Table 2 presents the comparison the performance of various models. The 1D metrics exhibited superior predictive capabilities for GPR and DVH metrics compared to using 3D dose alone. Notably, the model that combined multiple modalities, including 3D dose and plan complexity metrics, achieved the highest prediction accuracy.

Fig. 3
figure 3

The distribution of absolute errors or absolute percentage errors between the predicted metrics by combined model and the ground truth values for (a) GPRs, (b) PTV HI, (c)PTV CI, (d) PTV D95, (e) PTV D2 and (f) PTV Dmean across different ranges of comparisons between TPS calculation results and ground truth values

Table 2 Evaluation indicators of various predictive models with various inputs

QA classification

Figure 4 illustrates the ROC curves of the QA classification results obtained using various models. The corresponding AUC values were 0.92, 0.88, and 0.86 for combined model, 1D model and 3D transformer model respectively. For the QA classification performed using the combined model, the sensitivity was 0.93, and the specificity was 0.92, indicating a high classification accuracy. On the other hand, when classifying based on the predicted GPRs (also 90% as QA criterion), the sensitivity was 0.83, and the specificity was 0.94. The confusion matrices of the classification results in both ways are listed in Supplementary Table S2.

Fig. 4
figure 4

The ROC curves of the QA classification results obtained from various methods. (a) Comparison of 1D and 3D inputs. (b) Comparison of various models based on 3D input

Discussion

In this study, we developed a transformer-based deep learning prediction model for PSQA by utilizing patient multimodal input including 3D planning dose and 1D complexity metrics parameters. The model demonstrated high accuracy in predicting DVH metrics of PTV, QA classification, and 3D GPR on an independent test dataset. Furthermore, the model showed promising applicability across various treatment sites and treatment technologies. The predictive capabilities of 3D and 1D inputs were also compared.

Measurement-based PSQA remains a crucial step in the clinical workflow of radiotherapy, serving to validate data transfer and ensure the consistency between linac delivery and TPS calculations. Typically, the widely adopted evaluation method for PSQA is gamma analysis, this prediction of PSQA is feasible, but the clinical value is still controversial [34]. And in this context, additional dosimetric analysis based on DVH metrics or dose distributions [35] holds the potential to offer enhanced guidance for clinical practice. In our study, our combined model accurately predicts several crucial DVH metrics of the PTV. However, we did not propose specific action levels for PSQA based on deviations of DVH metrics due to variations in PTV across different treatment sites. While a methodology based on patient populations may be a feasible solution for threshold establishment [36], and it may be more suitable for single-site treatment scenarios.

The complexity of the plan is considered to be associated with delivery accuracy, where lower plan complexity indicates a higher likelihood of accurate delivery [37]. Our study also shows that plan complexity metrics have the ability to predict QA results. Moreover, the predictive performance of complexity metrics surpasses that of using patient planning dose alone, with the combined model incorporating both dose and complexity metrics achieving the highest accuracy. To assess the significance of each 1D metrics, we employed all 1D metrics to train a random forest classification model. The importance of each metric was then calculated using the SHAP (SHapley Additive exPlanations) method [38] The results are presented in Supplementary Figure S4. Among these metrics, the top five important ones were identified as follows: CI, LSmean (Mean value of leaf speed), APLSS2.0−2.4 (The average proportion of leaf speed from a given range 2.0-2.4), LAmean (Mean value of leaf acceleration) and SYMY (Symmetry in Y direction). This indicates that the planning conformity and the complexity of MLC motion significantly impact QA outcomes. These findings can provide guidance of input choice for the development of future QA prediction models.

An important objective of a QA predictive model is to determine whether a treatment plan will pass or fail the QA procedure, thereby assisting the clinical workflow, reducing physicists’ workload. For QA classification, studies showed that direct classification models exhibit higher sensitivity than models classified based on GPR prediction, it is consistent with our results. In clinical PSQA procedure, sensitivity is often prioritized over specificity, as predicting the passing of a heavily biased plan can result in higher clinical costs. Li et al.‘s study achieved 100% (19/19) sensitivity and 87.71% (207/236) specificity in their random forest model during cross-validation [39], the distributions of positive and negative samples imbalanced greatly. While Granville et al.‘s multi-classification model based on mean dose error achieved a macro-AUC of 0.88 using a self-defined classification method [3]. Our study adopts the standard recommended by TG218, which is consistent with our institution’s QA process, and achieves a sensitivity of 0.93 (39/42) in an independent 150 test dataset, considering failures constitute approximately 30% of the total samples.

One limitation of our study is that we focused on predicting only 5 critical DVH metrics of the PTV. However, different institutions may prioritize different DVH metrics for the PTV in different treatment sites. Additionally, the prediction of DVH metrics for OARs is also important in clinical practice. However, due to the inclusion of patients with multiple treatment sites in our study, obtaining the same OAR group is challenging. In the future, our single-site studies will try to address the prediction of DVH metrics of OARs. Another limitation is that we solely focused on the 3%/2mm gamma criterion for evaluating the results. This choice was based on our clinical routine and the distribution of classification results in our dataset, but other criteria were also reported useful and might be considered.

Conclusions

In this study, we developed a transformer-based deep learning model for patient-specific QA using complexity metrics and 3D dose distributions. The accurate prediction of QA classification and the PTV DVH metrics provides valuable guidance for clinical practice. Our findings demonstrate that complexity metrics play a leading role in the predictive capability of the model, surpassing the predictive performance of 3D dose in this task.

Data Availability

The data that support the findings of this study are available from the corresponding author, upon reasonable request.

Abbreviations

AUC:

Area under the curve

BCE:

Binary cross-entropy

CI:

Conformity index

DVH:

Dose volume histogram

GPR:

Gamma passing rate

HI:

Homogeneity index

IMRT:

Intensity modulated radiation therapy

Linac:

Linear accelerator

MAE:

Mean absolute error

MAPE:

Mean absolute percentage error

MLP:

Multi-layer perceptron

MSE:

Mean square error

PDP:

Planned dose perturbation

PSQA:

Patient specific quality assurance

PTV:

Planning target volume

ROC:

Receiver operating characteristic curve

ROI:

Region of interests

TPS:

Treatment planning system

VMAT:

Volumetric modulated arc therapy

References

  1. Ezzell GA, Galvin JM, Low D, et al. Guidance document on delivery, treatment planning, and clinical implementation of imrt: report of the imrt subcommittee of the aapm radiation therapy committee. Med Phys. 2003;30:2089–115.

    Article  PubMed  Google Scholar 

  2. Miften M, Olch A, Mihailidis D, et al. Tolerance limits and methodologies for imrt measurement-based verification qa: recommendations of aapm task group no. 218. Med Phys. 2018;45:e53–e83.

    Article  PubMed  Google Scholar 

  3. Granville DA, Sutherland JG, Belec JG, et al. Predicting vmat patient-specific qa results using a support vector classifier trained on treatment plan characteristics and linac qc metrics. Phys Med Biol. 2019;64:095017.

    Article  CAS  PubMed  Google Scholar 

  4. Van Esch A, Huyskens DP, Behrens CF, et al. Implementing rapidarc into clinical routine: a comprehensive program from machine qa to tps validation and patient qa. Med Phys. 2011;38:5146–66.

    Article  PubMed  Google Scholar 

  5. Smilowitz JB, Das IJ, Feygelman V, et al. Aapm medical physics practice guideline 5.A.: commissioning and qa of treatment planning dose calculations - megavoltage photon and electron beams. J Appl Clin Med Phys. 2015;16:14–34.

    Article  PubMed  PubMed Central  Google Scholar 

  6. McKenzie EM, Balter PA, Stingo FC, et al. Toward optimizing patient-specific imrt qa techniques in the accurate detection of dosimetrically acceptable and unacceptable patient plans. Med Phys. 2014;41:121702.

    Article  PubMed  PubMed Central  Google Scholar 

  7. McKenzie EM, Balter PA, Stingo FC, et al. Reproducibility in patient-specific imrt qa. J Appl Clin Med Phys. 2014;15:4741.

    Article  PubMed  Google Scholar 

  8. Low DA, Dempsey JF. Evaluation of the gamma dose distribution comparison method. Med Phys. 2003;30:2455–64.

    Article  PubMed  Google Scholar 

  9. Xu Y, Zhang K, Liu Z, et al. Treatment plan prescreening for patient-specific quality assurance measurements using Independent monte carlo dose calculations. Front Oncol. 2022;12:1051110.

    Article  PubMed  PubMed Central  Google Scholar 

  10. Zhen H, Nelms BE, Tome WA. Moving from gamma passing rates to patient dvh-based qa metrics in pretreatment dose qa. Med Phys. 2011;38:5477–89.

    Article  PubMed  Google Scholar 

  11. Stasi M, Bresciani S, Miranti A, et al. Pretreatment patient-specific imrt quality assurance: a correlation study between gamma index and patient clinical dose volume histogram. Med Phys. 2012;39:7626–34.

    Article  CAS  PubMed  Google Scholar 

  12. Wang H, Cooper BT, Schiff P, et al. Dosimetric assessment of Tumor control probability in intensity and volumetric modulated radiotherapy plans. Br J Radiol. 2019;92:20180471.

    Article  PubMed  Google Scholar 

  13. Dimopoulos JC, Lang S, Kirisits C, et al. Dose-volume histogram parameters and local Tumor control in magnetic resonance image-guided Cervical cancer brachytherapy. Int J Radiat Oncol Biol Phys. 2009;75:56–63.

    Article  PubMed  Google Scholar 

  14. Milano MT, Grimm J, Niemierko A, et al. Single- and multifraction stereotactic radiosurgery dose/volume tolerances of the brain. Int J Radiat Oncol Biol Phys. 2021;110:68–86.

    Article  PubMed  Google Scholar 

  15. Fredh A, Scherman JB, Fog LS, et al. Patient qa systems for rotational radiation therapy: a comparative experimental study with intentional errors. Med Phys. 2013;40:031716.

    Article  PubMed  Google Scholar 

  16. Chan MF, Witztum A, Valdes G. Integration of Ai and machine learning in radiotherapy qa. Front Artif Intell. 2020;3:577620.

    Article  PubMed  PubMed Central  Google Scholar 

  17. Hirashima H, Ono T, Nakamura M, et al. Improvement of prediction and classification performance for gamma passing rate by using plan complexity and dosiomics features. Radiother Oncol. 2020;153:250–7.

    Article  CAS  PubMed  Google Scholar 

  18. Ono T, Hirashima H, Iramina H, et al. Prediction of dosimetric accuracy for vmat plans using plan complexity parameters via machine learning. Med Phys. 2019;46:3823–32.

    Article  PubMed  Google Scholar 

  19. Wall PDH, Hirata E, Morin O, et al. Prospective clinical validation of virtual patient-specific quality assurance of volumetric modulated arc therapy radiation therapy plans. Int J Radiat Oncol Biol Phys. 2022;113:1091–102.

    Article  PubMed  Google Scholar 

  20. Valdes G, Scheuermann R, Hung CY, et al. A mathematical framework for virtual imrt qa using machine learning. Med Phys. 2016;43:4323.

    Article  CAS  PubMed  Google Scholar 

  21. Valdes G, Chan MF, Lim SB, et al. Imrt Qa using machine learning: a multi-institutional validation. J Appl Clin Med Phys. 2017;18:279–84.

    Article  PubMed  PubMed Central  Google Scholar 

  22. Tomori S, Kadoya N, Takayama Y et al. A deep learning-based prediction model for gamma evaluation in patient-specific quality assurance. Med Phys 2018.

  23. Interian Y, Rideout V, Kearney VP, et al. Deep nets vs expert designed features in medical physics: an imrt qa case study. Med Phys. 2018;45:2672–80.

    Article  PubMed  Google Scholar 

  24. Huang Y, Pi Y, Ma K, et al. Virtual patient-specific quality assurance of imrt using unet++: classification, gamma passing rates prediction, and dose difference prediction. Front Oncol. 2021;11:700343.

    Article  PubMed  PubMed Central  Google Scholar 

  25. Hao Y, Zhang X, Wang J, et al. Improvement of imrt qa prediction using imaging-based neural architecture search. Med Phys. 2022;49:5236–43.

    Article  CAS  PubMed  Google Scholar 

  26. Wang L, Li J, Zhang S, et al. Multi-task autoencoder based classification-regression model for patient-specific vmat qa. Phys Med Biol. 2020;65:235023.

    Article  PubMed  PubMed Central  Google Scholar 

  27. Liu Z, Lv Q, Yang Z, et al. Recent progress in transformer-based medical image analysis. Comput Biol Med. 2023;164:107268.

    Article  PubMed  Google Scholar 

  28. Olch AJ. Evaluation of the accuracy of 3dvh software estimates of dose to virtual ion chamber and film in composite imrt qa. Med Phys. 2012;39:81–6.

    Article  PubMed  Google Scholar 

  29. Chiavassa S, Bessieres I, Edouard M, et al. Complexity metrics for imrt and vmat plans: a review of current literature and applications. Br J Radiol. 2019;92:20190270.

    Article  PubMed  PubMed Central  Google Scholar 

  30. Antoine M, Ralite F, Soustiel C, et al. Use of metrics to quantify imrt and vmat treatment plan complexity: a systematic review and perspectives. Phys Med. 2019;64:98–108.

    Article  PubMed  Google Scholar 

  31. Klein EE, Hanley J, Bayouth J, et al. Task group 142 report: quality assurance of medical accelerators. Med Phys. 2009;36:4197–212.

    Article  PubMed  Google Scholar 

  32. Liu Z, Lin YT, Cao Y et al. Swin transformer: Hierarchical vision transformer using shifted windows. 2021 Ieee/Cvf International Conference on Computer Vision (Iccv 2021) 2021:9992–10002.

  33. Paszke A, Gross S, Massa F et al. Pytorch: an imperative style, high-performance deep learning library. Adv Neur in 2019;32.

  34. Valdes G, Adamson J, Cai J. Artificial intelligence for prediction of measurement-based patient-specific quality assurance is ready for prime time. Med Phys. 2021;48:2701–4.

    Article  PubMed  Google Scholar 

  35. Potter NJ, Mund K, Andreozzi JM, et al. Error detection and classification in patient-specific imrt qa with dual neural networks. Med Phys. 2020;47:4711–20.

    Article  PubMed  Google Scholar 

  36. Chen L, Zhang Z, Yu L et al. A clinically relevant online patient qa solution with daily ct scans and epid-basedin vivodosimetry: a feasibility study on rectal cancer. Phys Med Biol 2022;67.

  37. Viola P, Romano C, Craus M et al. Prediction of vmat delivery accuracy using plan modulation complexity score and log-files analysis. Biomed Phys Eng Express 2022;8.

  38. Ribeiro MT, Singh S, Guestrin C. “Why should i trust you?“ Explaining the predictions of any classifier. Kdd’16: Proceedings of the 22nd Acm Sigkdd International Conference on Knowledge Discovery and Data Mining 2016:1135–1144.

  39. Li J, Wang L, Zhang X, et al. Machine learning for patient-specific quality assurance of vmat: prediction and classification accuracy. Int J Radiat Oncol Biol Phys. 2019;105:893–902.

    Article  PubMed  PubMed Central  Google Scholar 

Download references

Funding

This work was supported by Natural Science Foundation Project of Chongqing [grant number cstc2021jcyjmsxmX0441], Overall rationing Program of Chongqing Talents Program (Basic research and Frontier exploration) [grant number cstc2021ycjh-bgzxm0023] and Science-Health Joint Project of Chongqing [grant number 2022DBXM005].

Author information

Authors and Affiliations

Authors

Contributions

L.C. and H.L. proposed the study idea, wrote the main manuscript text, and participate the model trainning. S.L. and X.T. collected the patient data. B.F. and X.Y wrote the code of training network. Y.W and F.J. provided provide technical support and study schedule guidance. All authors reviewed the manuscript.

Corresponding author

Correspondence to Fu Jin.

Ethics declarations

Ethical approval and consent to participate

This retrospective study was reviewed and approved by the Institutional Review Board of Chongqing University Cancer hospital (CZLS2023164-A) and the requirement for individual informed consent was waived.

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interests.

Additional information

Declarations.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Electronic supplementary material

Below is the link to the electronic supplementary material.

Supplementary Material 1

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Chen, L., Luo, H., Li, S. et al. Pretreatment patient-specific quality assurance prediction based on 1D complexity metrics and 3D planning dose: classification, gamma passing rates, and DVH metrics. Radiat Oncol 18, 192 (2023). https://doi.org/10.1186/s13014-023-02376-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13014-023-02376-4

Keywords