Online adaptive radiotherapy for bladder cancer using a simultaneous integrated boost and fiducial markers

Abstract

Purpose

The aim was to assess the feasibility of online adaptive radiotherapy (oART) for bladder cancer using a focal boost by focusing on the quality of the online treatment plan and automatic target delineation, duration of the workflow and performance in the presence of fiducial markers for tumor bed localization.

Methods

Fifteen patients with muscle invasive bladder cancer received daily oART with Cone Beam CT (CBCT), artificial intelligence (AI)-assisted automatic delineation of the daily anatomy and online plan reoptimization. The bladder and pelvic lymph nodes received a total dose of 40Ā Gy in 20 fractions, the tumor received an additional simultaneously integrated boost (SIB) of 15Ā Gy. The dose distribution of the reference plan was calculated for the daily anatomy, i.e. the scheduled plan. Simultaneously, a reoptimization of the plan was performed i.e. the adaptive plan. The target coverage and V95% outside the target were evaluated for both plans. The need for manual adjustments of the GTV delineation, the duration of the workflow and the influence of fiducial markers were assessed.

Results

All 300 adaptive plans met the requirement of the CTV-coverage V95%ā‰„98% for both the boost (55Ā Gy) and elective volume (40Ā Gy). For the scheduled plans the CTV-coverage was 53.5% and 98.5%, respectively. Significantly less tissue outside the targets received 55Ā Gy in case of the adaptive plans as compared to the scheduled plans. Manual corrections of the GTV were performed in 67% of the sessions. In 96% of these corrections the GTV was enlarged and resulted in a median improvement of 1% for the target coverage. The median on-couch time was 22Ā min. A third of the session time consisted of reoptimization of the treatment plan. Fiducial markers were visible on the CBCTs and aided the tumor localization.

Conclusions

AI-driven CBCT-guided oART aided by fiducial markers is feasible for bladder cancer radiotherapy treatment including a SIB. The quality of the adaptive plans met the clinical requirements and fiducial markers were visible enabling consistent daily tumor localization. Improved automatic delineation to lower the need for manual corrections and faster reoptimization would result in shorter session time.

Introduction

Bladder cancer is the 9th most diagnosed cancer worldwide and approximately 1 out of 5 patients develop muscle invasive bladder cancer reducing the 5-year survival rate to 50% [1, 2]. Cystectomy is considered the standard treatment of muscle invasive bladder cancer and combining the surgery with radiotherapy (RT) has shown to improve the outcome [3]. However, the procedure can have significant side effects [4]. Combining RT with chemotherapy and transurethral resection of bladder tumor (TURBT) has shown to be an effective alternative while preserving the bladder [5,6,7,8,9,10].

The main challenge for treating this anatomical site with radiotherapy is the variable bladder size, shape and position. These variations occur both between (interfraction) and during (intrafraction) RT fractions [11]. Even though drinking instructions typically aim to limit the intrafraction variations, the extent remains unpredictable [12,13,14,15]. Interfraction bladder deformation can be accounted for with the application of appropriate (larger) margins, library of plans (LoP) or adaptive reoptimization of the plan. The latter two lead to smaller irradiated volumes than the former [16, 17]. For the LoP approach, different patient-specific treatment plans are made based on different bladder volumes and the best fitting plan is selected for each fraction. An adaptive procedure with plan reoptimization during each treatment session utilizes anatomical information from daily images as acquired with Cone Beam CT (CBCT) or magnetic resonance imaging (MRI) on the linac [18,19,20]. Preclinical studies of these two strategies have shown that daily reoptimization is superior compared to LoP with less integral dose and dose to organs-at-risk (OAR) [21, 22].

Using a focal boost for bladder RT has been shown to be feasible and could further reduce the toxicity [10, 23, 24]. Several studies have shown feasibility of giving such a focal boost as a simultaneous integrated boost (SIB) to the tumor bed [10, 23]. Delivery of a SIB can be complicated by the reduced visibility of the remaining tumor volume when TURBT is performed prior to RT [19]. However, implantation of fiducial markers has shown to facilitate gross tumor volume (GTV) localization and delineation [25, 26]. Liquid fiducial markers are known for their high density which is favorable for (CB)CT scans and the feasibility considering the visibility, stability and safety has been demonstrated [26].

To allow for daily plan reoptimization, different online adaptive radiotherapy (oART) techniques have been developed [19]. The techniques acquire an image of the daily anatomy at the beginning of each treatment fraction. These images are used as input for the reoptimization of the treatment plan. One of these techniques consists of the integration of MRI with a linear accelerator (linac). The modality offers high soft tissue contrast but leads to a relatively long treatment time which is a drawback when considering intrafractional bladder filling, patient comfort and additional work load [18]. A novel concept has been developed, integrating a linac, CBCT and artificial intelligence (AI)-driven software for automatic organ delineation and plan reoptimization [20]. The result of these developments is that on board CBCT and MRI are increasingly used modalities for oART. The median on-couch time (time from first image acquisition to the end of RT delivery) for oART with plan reoptimization using MRI was reported as 39Ā min [18]. For CBCT-guided oART the delineation of the daily anatomy until treatment plan selection took 12Ā min, but the total on-couch time, including RT delivery, has so far not been reported in literature [27, 28]. Several studies have shown bladder oART to be feasible and applicable with CBCT and the image quality to be sufficient to apply automatic bladder segmentation resulting in similar outcomes as manual delineation [29ā€“31]. Recent studies on CBCT-guided oART show the technique to be feasible for whole bladder irradiation without a focal boost [27, 28]. So far no studies have evaluated the performance of CBCT-guided oART for muscle invasive bladder cancer including a focal boost and fiducial markers for tumor localization.

The aim of this study was to prospectively examine the feasibility of a CBCT-guided oART workflow for bladder radiotherapy using fiducial markers and a SIB. We will evaluate the automatic target delineation, quality of the online treatment plan, performance with fiducial markers, the duration of main steps of the workflow and the total on-couch time.

Methods

Patient characteristics

Fifteen patients with muscle invasive bladder cancer were treated between April 2021 and December 2022 on a ring-based linac integrated with a CBCT and software platform for both treatment planning and delivery (Ethos Therapyā„¢, version 1.1, Varian a Siemens Healthineers Company, USA). The patientsā€™ mean age was 68 years and the number of males and females was 11 and 4, respectively (see Additional file 1). In 20 fractions the bladder, urethra and first pelvic lymph nodes (elective volume) of each patient received a total dose of 40Ā Gy combined with a SIB of an additional 15Ā Gy to the tumor bed resulting in a total of 300 delivered fractions. Radiotherapy was combined with chemotherapy (Mitomycin-c/ Capecitabine).

Pretreatment

Prior to acquiring the planning CT, liquid fiducial markers (BioXmark, Nanovi A/S, Denmark or Lipiodol) were injected by the urologist at the borders of the tumor bed. These markers aid in target localization for both the pretreatment on CT and the online fractions on CBCT. Fiducials were placed in an outpatient setting with rigid cystoscopy in women and flexible cystoscopy in men, after histological confirmation of the bladder tumor by TURBT. The fiducials were placed submucosally with a margin of 0ā€“5Ā mm around the (remnant) tumor or around the resection scar (for simplicity we will refer to both as ā€œGTVā€). Preferably 3ā€“5 dots of 0.1ā€“0.2 cm3 were placed. Within 6 weeks after TURBT, each patient was asked to drink 0.3Ā L of water after voiding the bladder and subsequently refrain from drinking 1.5Ā h prior to CT acquisition (Discovery CT, GE Medical Systems). The CT was made in supine patient position with the arms on the chest and a knee support. To mimic the situation during the online adaptive procedure, after 15Ā min a second CT was acquired to estimate the intrafractional bladder filling. The first acquired planning CT (pCT) was used to delineate the GTV, the clinical target volume (CTV) of both the first pelvic lymph nodes (obturator, internal iliac and hypogastric lymph nodes and external lymph nodes (perivesical until lower part of sacroiliac joint)) and urethra (men: 2Ā cm proximal, women: 1Ā cm proximal) and OARs (bladder, rectum, bowel bag, sigmoid, left and right femur head), see also Additional file 2. If the GTV was in the cranial part of the bladder, the small bowel was also delineated. A GTV-CTVSIB and CTVSIB-PTVSIB margin of 5Ā mm were used. The CTVelective consisted of the urethra, the pelvic lymph nodes and the bladder. The urethra and pelvic lymph nodes had a CTVelective to PTVelective margin of 5ā€“7Ā mm.

The second CT scan was used to generate patient specific margins, to account for intrafraction bladder filling. Firstly, the bladder delineated on the first scan was expanded 5Ā mm in all directions. If the bladder on the second CT was completely encompassed by this expansion a uniform PTVelective margin of 7Ā mm was used. In any direction where the uniform 5Ā mm expansion did not encompass the bladder, the necessary expansion to encompass the bladder was determined. This necessary expansion was then increased by 50% to account for the assumption that on-couch time during the online adaptive procedure is currently longer than the time interval of 15Ā min between the two CTs. With the use of the above described target and OAR volumes a reference plan was generated by using the automatic treatment planning system (TPS) of the Ethos, which utilizes prioritized clinical goals for optimization of the final dose distribution (3 arc VMAT, 6MV FFF). The template used for these clinical goals is given in the additional files (see Additional file 3). All treatment plans were normalized to 98% of the volume of PTVSIB receiving 95% of the prescribed dose of 55Ā Gy.

Online adaptive workflow

The online adaptive treatment sessions were run by a team of two radiation therapists (RTT). A physician and medical physics expert were either present in the operating room or reachable on call. At the start of each fraction a CBCT (CBCT1) was acquired. The bladder and rectum were automatically delineated by the software, using a convolution neural network, allowing for manual correction after presentation [20]. These delineations (so-called influencers) and a deformable registration of the planning CT to the CBCT, resulting in a synthetic CT (sCT), were used to propagate a structure set based on the anatomy of the day. Manual corrections to the target structure and OARs were performed by the RTT/physician if necessary. The dose distribution of the reference plan was calculated on the anatomy of the day using the new structures and the sCT, resulting in the so-called scheduled plan. At the same time, an adaptive plan was generated by running a new optimization. For both the scheduled and adaptive plan, an independent dose calculation was performed for plan QA (Mobius, Varian a Siemens Healthineers Company, USA). Subsequently, one of the two plans was selected for treatment (i.e. adaptive or scheduled plan) based on the clinical goals, regions of high dose (>ā€‰107% of the prescribed dose outside target volume) and isodose lines (target coverage and dose conformity of 95% and 107% of the prescribed dose). Prior to delivery, a second CBCT (CBCT2) was acquired for position verification, which consisted of a bone match and a check of whether the GTV was still encompassed by PTVSIB. If this was not the case, a couch shift was performed if the required shift was larger than 1Ā mm in one or more directions. To evaluate whether the elective and boost targets were inside the PTV margins during the actual irradiation, a post-treatment CBCT (CBCT3) was acquired (see also Fig.Ā 1 for an illustration of the workflow). If the PTV margins were not adequate at the end of the session, the physician was consulted and the margins were adjusted in the TPS for the subsequent sessions.

Fig. 1
figure 1

AI-driven CBCT-guided oART workflow

Workflow evaluation and statistical analysis

To evaluate the CBCT-guided oART workflow, the analysis was focused on the quality and consistency of its online treatment plans, accuracy of its target propagation, duration of the online sessions and the capability of the system with respect to handling fiducial markers. Due to the fact that introducing a technique in the clinic usually comes with a learning curve, the first five patients were considered as a training group. To test this assumption, the workflow performed on the training group was compared with the workflow performed on patient six until patient 15 (the steady group).

The quality and consistency of the scheduled and adaptive plans were analyzed with help of home built software (Matlab R2021a, Mathworks) by assessing the target coverage, conformity index (CI), homogeneity index (HI) and volume of healthy tissue outside the target receiving high dose (55Ā Gy/40 Gy) [32, 33]. To make the conformity comparable with previous studies a conversion of the CI was performed (CIRTOG) [34]. The target coverage is given by the percentage of the volume of the target receiving at least 95% of the prescribed dose (V95%). The clinical requirement for the target coverage of the PTV and CTV was a minimum of 98% for V95%.

The healthy tissue outside the target receiving high dose (>ā€‰55Ā Gy for tissue outside the GTV and >ā€‰40Ā Gy for tissue outside the elective volume) is given by

$${V}_{95\%, out}= {V}_{95\%, Body}- {V}_{95\%}$$

,

where V95%,Body is the volume of the body receiving 95% of the prescribed dose. To evaluate the differences between the scheduled and adaptive plans, a paired Wilcoxon signed-rank test was performed. A Bonferroni corrected significance level of 0.4% was used (5%/12). To assess the magnitude of interfraction bladder variability, we determined the bladder volume on the pCTs and CBCT1s for all patients.

To evaluate the automatic target delineation, manual corrections of the GTV delineation during the online sessions (GTVclin) were first geometrically assessed. This was done by comparing GTVclin with the AI-supported automatically propagated GTV delineations (GTVAI). GTVAI was generated by reproducing the workflow in an online Ethos test environment (Emulator, version 1.1, Varian a Siemens Healthineers Company, USA) with the online acquired CBCTs and the same bladder structure as was used (but without manual corrections to the GTV). The differences between GTVclin and GTVAI were evaluated by calculating the Dice Similarity Coefficient (DSC), Hausdorff Distance (HD) and difference in volume [35, 36]. The difference in volume was also determined with respect to the GTV delineated on the reference plan (GTVref). The next step was to evaluate whether manual corrections of the GTV caused a significant difference in dosimetry. This was done by first exporting the treatment plan from the Emulator (planAI) including GTVAI. For each simulated session the same treatment plan was exported as selected during the online session (i.e. adaptive plan or scheduled plan). The V95% of the PTV and CTV and V95%, out as described in the previous section were determined (for PTVSIB, CTVSIB, PTVelective and CTVelective). To compare these parameters with the clinical plan, the clinical delineations (including GTVclin) were first propagated to the dose distribution of planAI. The same parameters, i.e. V95% of the PTV and CTV and V95%, out, were then determined for these clinical structures. To leave out the possible effect of the learning curve on manual corrections, the training group of the first 5 patients was excluded from the evaluation of the GTV propagation.

The on-couch time was defined as the time from image acquisition (CBCT1) until the end of RT (excluding time spend on patient setup and acquisition of CBCT3). The session time, defined by the time between the patient entering and leaving the treatment room, was manually registered by the RTTs. The duration of different steps in the adaptive procedure was extracted from file records: Patient setup & CBCT1, AI supported structure set propagation & evaluation, Plan reoptimization, Plan evaluation & CBCT2, Position verification, RT & CBCT3 and Patient leaving (see also Fig.Ā 1). The total session time for the training group was compared with the steady group to determine the presence of a learning curve. The comparison was tested for statistical significance by performing a Mann-Whitney test.

Fiducial markers were visually used in the online workflow for delineation of the GTV. We examined the role of these fiducial markers during the structure-guided deformable registration. The position of these fiducial markers on the sCT with respect to the position on the daily CBCT1 was determined. The evaluation was done by first manually delineating the fiducial markers on the pCT, which were subsequently used as input for the simulated sessions in the Emulator. The same steps of the oART workflow as described in the previous section were performed. The marker delineations propagated by the system to the CBCT (Markerauto) were compared with manually delineated markers on the CBCT, i.e. Markerman (in the clinical workflow it is these visible markers that are used to determine the location of the tumor by the RTT and physician). The Euclidean Distance between the center of mass of both marker delineations (Ī”CoM) were compared (Markerauto vs. Markerman). We assumed that an evaluation on the whole data set would not be necessary. Therefore, a representative sample of 40 sessions was created for two groups: sessions where the GTV was manually corrected and sessions where the proposed GTV was accepted (each group half of the sample size). The selection was done by randomly selecting one session per week for all patients in the two different groups. To compare these two groups, a Mann-Whitney test was performed. As also described in the previous section, the training group was excluded here.

Results

In three patients the patient specific PTV was adjusted based on the early post-treatment evaluations. These adjustments were made after session three for the first patient and after session five and six for the other patients (see also Additional file 1 for more detail). On the post-treatment CBCT (CBCT3), it was observed that the bladder was inside the PTV in 90% of all sessions for all patients (for the initial and interfraction bladder volume see Additional file 4).

Plan quality and consistency

The adaptive plan was selected in 99.7% of the sessions and the scheduled plan was selected once. All 300 adaptive plans met the requirement of the CTV and PTV coverage (V95%>98%) for both the boost (55Ā Gy) and elective (40Ā Gy) volume (Fig.Ā 2A). For the scheduled plans this requirement was met by 49% and 96% of the treatment plans for CTVSIB and CTVelective, respectively. For the PTVSIB this was achieved by 8% and for the PTVelective this was 63% for the scheduled plans. Even though the adaptive plans showed a significant different target coverage compared to the reference treatment plans, it was not a clinically relevant difference as the median target coverage and range were still within the clinical requirements (Fig.Ā 2A). For the scheduled plans, the median target coverage was lower and the range was higher meaning there was less consistency when considering the target coverage. Furthermore, significantly (pā€‰<ā€‰0.001) less external tissue received high dose (40 and 55Ā Gy) with the adaptive plans compared to the scheduled and reference plans. An exception was the healthy tissue outside the PTVelective of the adaptive plans, which was comparable to the reference plan (Fig.Ā 2B). The adaptive plans outperformed the scheduled plans with respect to the CI (for more details see also Additional file 5), the HI and Dmean. Moreover, compared to the scheduled plans, the adaptive plans were closer to the values extracted from the reference plans and showed less variation, i.e. more consistency (Fig.Ā 2C).

Fig. 2
figure 2

Comparison between the reference plan, scheduled plan and adaptive plan of all sessions from all 15 patients (nā€‰=ā€‰300 sessions). The boxplots represent the 1st and 3rd quartile, the line inside the median and the whiskers represent the range. A) Target coverage of the three plans from left to right: the CTVSIB, PTVSIB, CTVelective and PTVelective. The dotted green line represents the clinical goal. B) Volume of external tissue outside the CTVSIB and PTVSIB receiving a dose of 55Ā Gy and outside the CTVelective and PTVelective receiving a dose of 40Ā Gy. C) The median CI for both the boost and elective PTVs (left axis), the HI for the boost PTV (left axis) and mean dose for the boost area (right axis)

Accuracy and target propagation

Manual adjustment of the GTV was performed in 67% of the sessions of all patients. For the training group (patient 1ā€“5) manual correction of the GTV was done in 80% of the sessions and reduced to 60% for the subsequent 10 patients, i.e. the steady group. 96% of the corrections were performed to significantly (pā€‰<ā€‰0.001) enlarge the GTVAI delineations with a median volume of 2.5 cm3 (Fig.Ā 3a and Additional file 6). The median volume of the resulting GTVclin was 0.9 cm3 smaller compared to the reference GTV (pā€‰<ā€‰0.001). GTVAI was 3.9 cm3 smaller than the median volume of GTVref (pā€‰<ā€‰0.001). Comparing GTVclin with GTVAI from sessions where manual GTV corrections were performed resulted in a median Dice Similarity Coefficient of 0.70 and a median Hausdorff Distance of 9Ā mm (Fig.Ā 3b). Performing the GTV corrections, i.e. GTVclin, resulted in a significant (pā€‰<ā€‰0.001) median increase in target coverage of 1% for the V95% of the CTVSIB compared to that same parameter for the GTVAI (Fig.Ā 3c). Manual adjustments of the GTV never resulted in a decrease of the CTVSIB coverage. For the PTVSIB, manual corrections of the GTV caused a significant (pā€‰<ā€‰0.001) increase in target coverage of 19%.

Fig. 3
figure 3

Evaluation of the GTV contour propagation by the Ethos software in sessions where the GTV was manually adjusted (Nā€‰=ā€‰10 patients, nā€‰=ā€‰134 sessions). The boxplots represent the 1st and 3rd quartile, the line inside the median and the whiskers represent the range. A) Difference between the volume of GTVclin, GTVref and GTVAI. B) The Dice similarity coefficient and Hausdorff distance (mm) between the GTVclin and GTVAI. C) Difference in target coverage (volume receiving minimum of 95% of the prescribed dose of 55Ā Gy) between the CTV and PTV of GTVAI (i.e. generated by the Ethos software) and the CTV and PTV of GTVclin (i.e. manually corrected), respectively

On-couch time and duration of oART

The median session time was 30Ā min (range: 19ā€“56) for all patients and consisted for about a third of reoptimization of the treatment plan (Fig.Ā 4). Patients setup and acquiring CBCT1 took a median of 4Ā min, for propagation and evaluation of the structure set this was 7Ā min and reoptimization of the treatment plan consisted of a median duration of 8Ā min. For plan evaluation a median of 4Ā min was needed, for position verification 1ā€“2Ā min and for dose delivery 2ā€“3Ā min. The median on-couch time was 22Ā min for the steady group (range: 14ā€“51) and reduced significantly compared to the training group (pā€‰<ā€‰0.001). The steady group also showed a shortened plan evaluation time of 1.5Ā min (see also Additional file 7). Considering sessions in which the GTV delineation was corrected, the AI supported delineation propagation/evaluation reduced with 1.5Ā min for the steady group in comparison to the training group. For the training group the median on-couch time was 26Ā min and for the total patient group the median on-couch time was 23Ā min. Manual correction of the GTV resulted in a median addition of 5Ā min to the session time compared to session where GTVAI was accepted.

Fig. 4
figure 4

Median session time with the median duration of each step in the oART workflow and median on-couch time for: all patients (Nā€‰=ā€‰15), first five patients, subsequent ten patients, sessions where GTVAI was accepted and sessions where the GTVAI was corrected (Nā€‰=ā€‰number of patients and nā€‰=ā€‰number of sessions)

Fiducial markers

Fiducial markers used as aid during online target evaluation were visible and could be distinguished on the CBCT (Additional file 8). For the representative sample of 40 sessions the center of mass position, CoM, of Markerauto (the fiducial marker delineation propagated by the Ethos system) differed with a median of 7.8Ā mm from the CoM position given by Markerman, i.e. the manually delineated marker (Fig.Ā 5). The median āˆ†CoM of sessions where GTVAI was immediately accepted was 2.6Ā mm lower than for the sessions where manual GTV adjustment was performed.

Fig. 5
figure 5

Distance between the center of mass of marker contours propagated by the Ethos software, i.e. Markerauto and manually delineated, i.e. Markerman (Nā€‰=ā€‰10 patients; nā€‰=ā€‰number of sessions). The boxplots represent the 1st and 3rd quartile, the line inside the median and the whiskers represent the range. ā€œGTV correctedā€ represents the sessions where the GTV was online adjusted. ā€œGTV acceptedā€ represents the sessions where the propagated GTV was accepted. No significant difference was found between these two groups (pā€‰<ā€‰0.001)

Discussion

In this study we are the first to demonstrate the feasibility of the CBCT-guided oART workflow for bladder cancer using a focal boost and fiducial markers. With the adaptive procedure online treatment plans with excellent conformity were generated with a median on-couch time, after a training period, of 22Ā min. These adaptive plans always met the clinical requirements considering the target coverage and reduced the high dose to healthy tissue outside the boost and elective targets compared to the reference plan.

The superior target coverage, healthy tissue sparing and plan quality of the adaptive plans compared to the scheduled plans demonstrate the benefit of implementing online reoptimization of the treatment plan. These aspects could be maintained while allowing for smaller PTVs which is in line with previous studies [19, 21, 22, 28]. The target coverage of the clinically used online plans always met the clinical requirement which is also observed by previous CBCT- and MR-guided oART studies [18, 28]. In terms of the HI and CI, the plan quality was also comparable with a previous CBCT-guided oART study for bladder cancer [28]. These similarities between our CI values (using VMAT) and the CI values reported by ƅstrƶm et al. (using IMRT), imply that the delivery method does not need to have an effect on the conformity [28]. Foroudi et al. published about the differences between the delivery methods and found no significant difference in CI between IMRT and VMAT for bladder cancer patients [37]. However, for MR-guided oART the conformity, CIRTOG, is reported to be around 2.5 indicating considerable dose to healthy tissue [18]. In our study the conversion of the CI to CIRTOG was between 1.0 and 1.3 for all adaptive plans, showing an improved conformity (with the ideal CIRTOG value equal to 1). In view of the observed similarity of IMRT and VMAT, as mentioned above [37] it seems more likely that this is due to larger treatment times (and corresponding larger margins to deal with intrafraction motion). Furthermore, a previous study on employing LoP for the treatment of bladder cancer, including a SIB, reported an even higher value for CIRTOG of 3.5 [38], which might be explained by interfraction variability of the bladder shape as compared to the available library plans. These findings suggests the potential gain of dose reduction to healthy tissue by reoptimized CBCT-guided oART for bladder cancer as compared to MR-guided oART and LoP.

When no manual corrections were needed, the session time was reduced with 5Ā min compared to sessions where the GTV was adjusted. The GTV volume appeared to be smaller when delineated by the Ethos software as compared to the clinically manually adapted GTV. The geometric differences between GTVclin and GTVAI were large enough to result in significant dosimetric differences, i.e. lower dose in the tumor. The manual adjustments of the GTV performed in 6 out of 10 sessions resulted in longer session times. More accurate GTV propagation by the software is needed to allow for a shorter workflow and thus less intrafraction bladder filling and potentially smaller PTV margins. The effect of interobserver variation during correction of manual corrections was not considered in this study.

A detailed time schedule of the oART workflow was analyzed giving departments a more detailed insight of what to expect during clinical implementation. The on-couch time was about 10Ā min lower than reported by Hunt et al. for MR-guided oART for bladder cancer [18]. Plan reoptimization took about 2Ā min longer in our study and the dose delivery time was 6Ā min shorter. Their findings show that the recontouring part of the workflow took 7Ā min which is the same as the time for our total structure set propagation and performance of manual correction. Reducing the need for manual correction would result in a structure set delineation time of 2Ā min as shown by the fully automatic sessions. The on-couch time would be 18% shorter, meaning a reduction of the total session time of 5Ā min. Therefore, improving the software for accurate GTV propagation would be beneficial for bladder treatments.

Compared to a previous CBCT-guided whole bladder oART study without a focal boost to the tumor, the review of the automatically generated structure set was about 2Ā min shorter in our case [28]. The plan reoptimization and selection took approximately 6Ā min longer resulting in a longer adaptive procedure of 4Ā min [28]. The main difference between these studies, including a SIB (thus two dose levels) and employing VMAT instead of IMRT, apparently leads to a longer reoptimization time in our case. However, VMAT has shown to deliver treatment within a shorter time compared to IMRT [37]. Compared to Library of Plans, as reported by a study including a SIB, the on-couch time was about 10Ā min longer in our study [38].

The on-couch time decreased for patient 6 to 15, as compared to the first 5 patients. This is indicative of a learning curve caused by experience with evaluating the treatment plan. The learning curve regarding the duration of the manual corrections yielded an overall speed up of the work flow. This might be due to the staffā€™s increasing skill in applying the manual corrections. Additionally, during the pilot phase manual corrections could be made that afterwards would not be considered clinically relevant. As the staff gaining experience, the frequency of performing manual corrections decreased. As 96% of the corrections were made to increase the GTV, we do not expect this to have an effect on the target coverage of the first 5 patients compared to the latter 10 patients, but we do expect it to affect the volume of healthy tissue receiving high dose (i.e. 40/55 Gy). The decrease in frequency of manual corrections also gives another explanation for the reduction in session time. Longer session times would result in more intrafraction bladder filling necessitating larger PTV margins. Nevertheless, our study shows that the bladder was included by the PTV at CBCT3 in 90% of the sessions. A limitation of this study is the assumption of linear bladder filling with a fixed filling rate between the 2 pCTs. Additionally, the different session times are currently not taken into account. However, the CBCT acquired after dose delivery allows for monitoring and gives the possibility to modify the PTV margins in between treatment fractions.

To be superior to LoP with respect to the irradiated volume of healthy tissue, an MR-guided study suggested that the session time should be around 15Ā min [39]. Another preclinical study reported the requirement of a PTV margin of 5Ā mm [21]. These requirements are currently not met by the oART workflow on the Ethos. It should be noted that these requirements were for whole bladder treatments without a SIB and requirements for treatments including a SIB are to our knowledge currently lacking. An advantage of the oART workflow over LoP, is taking into account the OARs [40]. This is a potential benefit especially for patients with the small bowel close to the tumor. Besides the difference in conformity mentioned earlier, this also illustrates the potential for the reoptimization oART workflow to reduce gastrointestinal toxicity. Furthermore, during LoP it might occur that no suitable treatment plan fits the daily anatomy, which is not the case for oART [41].

To allow for accurate boost dose to the tumor, fiducial markers were used for tumor localization. The markers could be clearly distinguished on the online CBCT images and was an essential part of the target evaluation and adaptation process. However, the system was not able to localize the markers accurately enough on the daily anatomy with aid of the structure-guided deformable registration. Sessions where manual correction of the GTV was performed showed a larger displacement of the markers. Allowing the structure-guided deformable registration to account for fiducial markers, as they are used for tumor localization, might increase the accuracy and lower the treatment time when using a focal boost.

A limitation of our study was that we evaluated the dose on CBCT1. An evaluation of the dose on CBCT2 or CBCT3 would probably better reflect the actually delivered dose. In our paper we choose to focus on the data that is used for the actual clinical decision making, as reflected on CBCT1. This is in line with earlier studies with a LoP, also using the initial (and only) CBCT for dose evaluation [38]. Future investigation should reveal if there is significant difference between the dose evaluated on CBCT1 and CBCT2 or CBCT3 which is not unlikely due the time between the scans. Additionally, using a larger number of patients for the study would make the data more representative considering the generalizability.

Conclusion

We have shown the clinical feasibility of delivering a focal boost with aid of fiducial markers during daily online adaptive radiotherapy for muscle-invasive bladder cancer. Employing an AI-driven CBCT-guided RT technique yields consistent plan quality for all treatments, comparable to the pretreatment reference plan with a median on-couch time of 22Ā min.

Data Availability

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

AI:

Artificial intelligence

CBCT:

Cone Beam CT

CI:

Conformity index

CoM:

Center of mass

CT:

Computed tomography

CTV:

Clinical target volume

DSC:

Dice Similarity Coefficient

FFF:

Flattening-Filter-Free

Gy:

Gray

GTV:

Gross tumor volume

HD:

Hausdorff Distance

HI:

Homogeneity index

LoP:

Library of plans

MRI:

Magnetic resonance imaging

MV:

Megavolt

OAR:

Organ-at-risk

oART:

Online adaptive radiotherapy

pCT:

Planning computed tomography

PTV:

Planning target volume

QA:

Quality assurance

RT:

Radiotherapy

RTT:

Radiation therapist

SIB:

Simultaneously integrated boost

sCT:

Synthetic computed tomography

TPS:

Treatment planning system

TURBT:

Transurethral resection of bladder tumor

TV:

Target volume

V95% :

Volume receiving a minimum of 95% of the prescribed dose

VMAT:

Volumetric modulated arc therapy

References

  1. Sung H, Ferlay J, Siegel RL, Laversanne M, Soerjomataram I, Jemal A, et al. Global Cancer Statistics 2020: GLOBOCAN estimates of incidence and Mortality Worldwide for 36 cancers in 185 countries. CA Cancer J Clin. 2021;71(3):209ā€“49.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  2. Knowles MA, Hurst CD. Molecular biology of bladder cancer: new insights into pathogenesis and clinical diversity. Nat Rev Cancer. 2015;15(1):25ā€“41.

    ArticleĀ  CASĀ  PubMedĀ  Google ScholarĀ 

  3. Zohar WHH, Machele D. Defining optimal therapy for muscle invasive bladder Cancer. J Urol. 2007;177(2):437ā€“43.

    ArticleĀ  Google ScholarĀ 

  4. Shabsigh A, Korets R, Vora KC, Brooks CM, Cronin AM, Savage C, et al. Defining early morbidity of radical cystectomy for patients with bladder cancer using a standardized reporting methodology. Eur Urol. 2009;55(1):164ā€“74.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  5. Zlotta AR, Ballas LK, Niemierko A, Lajkosz K, Kuk C, Miranda G, et al. Radical cystectomy versus trimodality therapy for muscle-invasive bladder cancer: a multi-institutional propensity score matched and weighted analysis. Lancet Oncol. 2023;24(6):669ā€“81.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  6. Arcangeli G, Arcangeli S, Strigari L. A systematic review and meta-analysis of clinical trials of bladder-sparing trimodality treatment for muscle-invasive bladder cancer (MIBC). Crit Rev Oncol Hematol. 2015;94(1):105ā€“15.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  7. Kotwal S, Choudhury A, Johnston C, Paul AB, Whelan P, Kiltie AE. Similar treatment outcomes for Radical Cystectomy and Radical Radiotherapy in invasive bladder Cancer treated at a United Kingdom specialist treatment Center. Int J Radiat Oncol. 2008;70(2):456ā€“63.

    ArticleĀ  Google ScholarĀ 

  8. Giacalone NJ, Shipley WU, Clayman RH, Niemierko A, Drumm M, Heney NM, et al. Long-term outcomes after bladder-preserving tri-modality therapy for patients with muscle-invasive bladder Cancer: an updated analysis of the Massachusetts General Hospital Experience. Eur Urol. 2017;71(6):952ā€“60.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  9. Vashistha V, Wang H, Mazzone A, Liss MA, Svatek RS, Schleicher M, et al. Radical cystectomy compared to combined modality treatment for muscle-invasive bladder Cancer: a systematic review and Meta-analysis. Int J Radiat Oncol. 2017;97(5):1002ā€“20.

    ArticleĀ  Google ScholarĀ 

  10. Lutkenhaus LJ, van Os RM, Bel A, Hulshof MCCM. Clinical results of conformal versus intensity-modulated radiotherapy using a focal simultaneous boost for muscle-invasive bladder cancer in elderly or medically unfit patients. Radiat Oncol. 2016;11(1):45.

    ArticleĀ  PubMedĀ  PubMed CentralĀ  Google ScholarĀ 

  11. Pos FJ, Koedooder K, Hulshof MCCM, van Tienhoven G, GonzĆ”lez GonzĆ”lez D. Influence of bladder and rectal volume on spatial variability of a bladder tumor during radical radiotherapy. Int J Radiat Oncol. 2003;55(3):835ā€“41.

    ArticleĀ  Google ScholarĀ 

  12. Dees-Ribbers HM, Betgen A, Pos FJ, Witteveen T, Remeijer P, van Herk M. Inter- and intra-fractional bladder motion during radiotherapy for bladder cancer: a comparison of full and empty bladders. Radiother Oncol. 2014;113(2):254ā€“9.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  13. Biancia C, Della, Yorke E, Kollmeier MA. Image guided radiation therapy for bladder cancer: Assessment of bladder motion using implanted fiducial markers. Pract Radiat Oncol. 2014;4(2):108ā€“15.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  14. Miralbell R, Nouet P, Rouzaud M, Bardina A, Hejira N, Schneider D. Radiotherapy of bladder cancer: relevance of bladder volume changes in planning boost treatment. Int J Radiat Oncol. 1998;41(4):741ā€“6.

    ArticleĀ  CASĀ  Google ScholarĀ 

  15. Turner SL, Swindell R, Bowl N, Marrs J, Brookes B, Read G, et al. Bladder movement during radiation therapy for bladder cancer: implications for treatment planning. Int J Radiat Oncol. 1997;39(2):355ā€“60.

    ArticleĀ  CASĀ  Google ScholarĀ 

  16. Foroudi F, Wong J, Kron T, Rolfo A, Haworth A, Roxby P, et al. Online adaptive radiotherapy for muscle-invasive bladder Cancer: results of a pilot study. Int J Radiat Oncol. 2011;81(3):765ā€“71.

    ArticleĀ  Google ScholarĀ 

  17. Lutkenhaus LJ, Visser J, de Jong R, Hulshof MCCM, Bel A. Evaluation of delivered dose for a clinical daily adaptive plan selection strategy for bladder cancer radiotherapy. Radiother Oncol. 2015;116(1):51ā€“6.

    ArticleĀ  Google ScholarĀ 

  18. Hunt A, Hanson I, Dunlop A, Barnes H, Bower L, Chick J, et al. Feasibility of magnetic resonance guided radiotherapy for the treatment of bladder cancer. Clin Transl Radiat Oncol. 2020;25:46ā€“51.

    CASĀ  PubMedĀ  PubMed CentralĀ  Google ScholarĀ 

  19. Kong V, Hansen VN, Hafeez S. Image-guided adaptive radiotherapy for bladder Cancer. Clin Oncol. 2021;33(6):350ā€“68.

    ArticleĀ  CASĀ  Google ScholarĀ 

  20. Archambault Y, Boylan C, Drew B, Morgas T, Peltola J, Ruokokoski E et al. Making on-line adaptive radiotherapy possible using artificial intelligence and machine learning for efficient daily re-planning. Med Phys Int J. 2020;8(2).

  21. Kong VC, Taylor A, Chung P, Craig T, Rosewall T. Comparison of 3 image-guided adaptive strategies for bladder locoregional radiotherapy. Med Dosim. 2019;44(2):111ā€“6.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  22. Vestergaard A, Muren LP, SĆøndergaard J, ElstrĆøm UV, HĆøyer M, Petersen JB. Adaptive plan selection vs. re-optimisation in radiotherapy for bladder cancer: a dose accumulation comparison. Radiother Oncol. 2013;109(3):457ā€“62.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  23. Piet AHM, Hulshof MCCM, Pieters BR, Pos FJ, de Reijke TM, Koning CCE. Clinical results of a concomitant Boost Radiotherapy technique for muscle-invasive bladder Cancer. Strahlentherapie und Onkol. 2008;184(6):313ā€“8.

    ArticleĀ  Google ScholarĀ 

  24. Pos FJ, van Tienhoven G, Hulshof MCCM, Koedooder K, GonzĆ”lez GonzĆ”lez D. Concomitant boost radiotherapy for muscle invasive bladder cancer. Radiother Oncol. 2003;68(1):75ā€“80.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  25. Chai X, van Herk M, van de Kamer JB, Remeijer P, Bex A, Betgen A, et al. Behavior of lipiodol markers during image guided radiotherapy of bladder Cancer. Int J Radiat Oncol. 2010;77(1):309ā€“14.

    ArticleĀ  CASĀ  Google ScholarĀ 

  26. de Ridder M, Gerbrandy LC, de Reijke TM, Hinnen KA, Hulshof MCCM. BioXmarkĀ® liquid fiducial markers for image-guided radiotherapy in muscle invasive bladder cancer: a safety and performance trial. Br J Radiol. 2020;93(1111):20200241.

    ArticleĀ  PubMedĀ  PubMed CentralĀ  Google ScholarĀ 

  27. Sibolt P, Andersson LM, Calmels L, Sjƶstrƶm D, Bjelkengren U, Geertsen P, et al. Clinical implementation of artificial intelligence-driven cone-beam computed tomography-guided online adaptive radiotherapy in the pelvic region. Phys Imaging Radiat Oncol. 2021;17:1ā€“7.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  28. ƅstrƶm LM, Behrens CP, Calmels L, Sjƶstrƶm D, Geertsen P, Mouritsen LS, et al. Online adaptive radiotherapy of urinary bladder cancer with full re-optimization to the anatomy of the day: initial experience and dosimetric benefits. Radiother Oncol. 2022;171:37ā€“42.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  29. Henry AM, Stratford J, McCarthy C, Davies J, Sykes JR, Amer A, et al. X-ray volume imaging in bladder radiotherapy verification. Int J Radiat Oncol. 2006;64(4):1174ā€“8.

    ArticleĀ  Google ScholarĀ 

  30. Nishioka K, Shimizu S, Kinoshita R, Inoue T, Onodera S, Yasuda K, et al. Evaluation of inter-observer variability of bladder boundary delineation on cone-beam CT. Radiat Oncol. 2013;8(1):185.

    ArticleĀ  PubMedĀ  PubMed CentralĀ  Google ScholarĀ 

  31. van de Schoot AJAJ, Schooneveldt G, Wognum S, Hoogeman MS, Chai X, Stalpers LJA, et al. Generic method for automatic bladder segmentation on cone beam CT using a patient-specific bladder shape model. Med Phys. 2014;41(3):31707.

    ArticleĀ  Google ScholarĀ 

  32. Paddick I. A simple scoring ratio to index the conformity of radiosurgical treatment plans: technical note. J Neurosurg. 2000;93(supplement3):219ā€“22.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  33. International Commission on Radiation Units. Measurements prescribing, recording, and reporting photon-beam intensity-modulated radiation therapy (IMRT). ICRU Rep 83 J ICRU. 2010;10:1ā€“106.

    Google ScholarĀ 

  34. Shaw E, Kline R, Gillin M, Souhami L, Hirschfeld A, Dinapoli R, et al. Radiation therapy oncology group: Radiosurgery quality assurance guidelines. Int J Radiat Oncol. 1993;27(5):1231ā€“9.

    ArticleĀ  CASĀ  Google ScholarĀ 

  35. Crum WR, Camara O, Hill DLG. Generalized overlap measures for evaluation and validation in Medical Image Analysis. IEEE Trans Med Imaging. 2006;25(11):1451ā€“61.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  36. Taha AA, Hanbury A. Metrics for evaluating 3D medical image segmentation: analysis, selection, and tool. BMC Med Imaging. 2015;15(1):29.

    ArticleĀ  PubMedĀ  PubMed CentralĀ  Google ScholarĀ 

  37. Foroudi F, Wilson L, Bressel M, Haworth A, Hornby C, Pham D et al. A dosimetric comparison of 3D conformal vs intensity modulated vs volumetric arc radiation therapy for muscle invasive bladder cancer. Radiat Oncol. 2012;7(1).

  38. Hafeez S, Warren-Oseni K, McNair HA, Hansen VN, Jones K, Tan M, et al. Prospective study delivering Simultaneous Integrated high-dose Tumor Boost (ā‰¤ 70 gy) with image guided adaptive Radiation Therapy for Radical Treatment of localized muscle-invasive bladder Cancer. Int J Radiat Oncol. 2016;94(5):1022ā€“30.

    ArticleĀ  Google ScholarĀ 

  39. den Boer D, den Hartogh MD, Kotte ANTJ, van der Voort JRN, Noteboom JL, Bol GH, et al. Comparison of Library of plans with two daily adaptive strategies for whole bladder radiotherapy. Phys Imaging Radiat Oncol. 2021;20:82ā€“7.

    ArticleĀ  Google ScholarĀ 

  40. Winkel D, Kroon PS, Werensteijn-Honingh AM, Bol GH, Raaymakers BW, JĆ¼rgenliemk-Schulz IM. Simulated dosimetric impact of online replanning for stereotactic body radiation therapy of lymph node oligometastases on the 1.5T MR-linac. Acta Oncol. 2018;57(12):1705ā€“12.

    ArticleĀ  CASĀ  PubMedĀ  Google ScholarĀ 

  41. Collins SD, Leech MM. A review of plan library approaches in adaptive radiotherapy of bladder cancer. Acta Oncol. 2018;57(5):566ā€“73.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

Download references

Acknowledgements

The authors would like to thank all involved radiation therapists, medical physics experts and physicians for the clinical support. Special thanks to Anna Dinkla, the project group led by her and Amy de la Fuente. We would also like to thank Varian a Siemens Healthineers Company for providing a test environment and supporting this work. Our research was embedded within Cancer Center Amsterdam.

Funding

This study was sponsored by a research grant from Varian a Siemens Healthineers Company.

Author information

Authors and Affiliations

Authors

Contributions

All authors significantly contributed to the study design, concept and interpretation of data. All authors were involved in revising the article and all approved the final manuscript.

Corresponding author

Correspondence to Sana Azzarouali.

Ethics declarations

Ethics approval and consent to participate

All procedures performed in this study were in accordance with the ethical standards of the institutional research committee. Written informed consent was provided by all patients.

Consent for publication

Written informed consent was provided by all patients of which individual personā€™s data was used.

Competing interests

The department had received grants from Varian a Siemens Healthineers Company. Varian a Siemens Healthineers Company had no influence on the study design, data analysis and revisions of the article manuscripts.

Additional information

Publisherā€™s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Electronic supplementary material

Below is the link to the electronic supplementary material.

Supplementary Material 1. Additional file 1 (.pdf) : Patient characteristics including sex, age, tumor stage and other.

13014_2023_2348_MOESM2_ESM.pdf

Supplementary Material 2. Additional file 2 (.pdf) : Representation of a planning CT used to calculate the reference plan.

13014_2023_2348_MOESM3_ESM.pdf

Supplementary Material 3. Additional file 3 (.pdf) : Representation of the template with prioritized planning directives used to constraint treatment plans. The exact values of the clinical goals were patient-specific.

13014_2023_2348_MOESM4_ESM.pdf

Supplementary Material 4. Additional file 4 (.pdf) : The initial bladder volume from the pCT and the interfraction bladder volume from CBCT1.

13014_2023_2348_MOESM5_ESM.pdf

Supplementary Material 5. Additional file 5 (.pdf) : The conformity index (CIRTOGā€‰=ā€‰V95% / PTV volume) of the adaptive plans for the tumor and elective area (Nā€‰=ā€‰15 patientsā€‰=ā€‰300 sessions). Different ways of calculating the CI are used in literature. To be able to compare our work with previous studies we added the conformity index as proposed by the Radiation Therapy Oncology Group (RTOG) [34].

13014_2023_2348_MOESM6_ESM.pdf

Supplementary Material 6. Additional file 6 (.pdf) : A representation of the automatic propagated GTV (GTVAI), the manually corrected GTV delineation on the online CBCT (GTVclin) and the reference GTV on the planning CT.

13014_2023_2348_MOESM7_ESM.pdf

Supplementary Material 7. Additional file 7 (.pdf) : Comparison between the training group (P1-5) and steady group (P6-15) considering the duration of different steps from the oART workflow. The number of sessions is indicated by n.

13014_2023_2348_MOESM8_ESM.pdf

Supplementary Material 8. Additional file 8 (.pdf) : Representation of a GTV delineation (yellow) on an online CBCT (from up to down: axial, sagittal and coronal). Fiducial markers (white spots) indicated with red arrows are used as aid for the delineation.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the articleā€™s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the articleā€™s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Azzarouali, S., Goudschaal, K., Visser, J. et al. Online adaptive radiotherapy for bladder cancer using a simultaneous integrated boost and fiducial markers. Radiat Oncol 18, 165 (2023). https://doi.org/10.1186/s13014-023-02348-8

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13014-023-02348-8

Keywords