Monte Carlo simulation of linac using PRIMO

Abstract

Background

Monte Carlo simulation is considered as the most accurate method for dose calculation in radiotherapy. PRIMO is a Monte-Carlo program with a user-friendly graphical interface.

Material and method

A VitalBeam with 6MV and 6MV flattening filter free (FFF), equipped with the 120 Millennium multileaf collimator was simulated by PRIMO. We adjusted initial energy, energy full width at half maximum (FWHM), focal spot FWHM, and beam divergence to match the measurements. The water tank and ion-chamber were used in the measurement. Percentage depth dose (PDD) and off axis ratio (OAR) were evaluated with gamma passing rates (GPRs) implemented in PRIMO. PDDs were matched at different widths of standard square fields. OARs were matched at five depths. Transmission factor and dose leaf gap (DLG) were simulated. DLG was measured by electronic portal imaging device using a sweeping gap method.

Result

For the criterion of 2%/2 mm, 1%/2 mm and 1%/1 mm, the GPRs of 6MV PDD were 99.33–100%, 99–100%, and 99–100%, respectively; the GPRs of 6MV FFF PDD were 99.33–100%, 98.99–99.66%, and 97.64–98.99%, respectively; the GPRs of 6MV OAR were 96.4–100%, 90.99–100%, and 85.12–98.62%, respectively; the GPRs of 6MV FFF OAR were 95.15–100%, 89.32–100%, and 87.02–99.74%, respectively. The calculated DLG matched well with the measurement (6MV: 1.36 mm vs. 1.41 mm; 6MV FFF: 1.07 mm vs. 1.03 mm, simulation vs measurement). The transmission factors were similar (6MV: 1.25% vs. 1.32%; 6MV FFF: 0.8% vs. 1.12%, simulation vs measurement).

Conclusion

The calculated PDD, OAR, DLG and transmission factor were all in good agreement with measurements. PRIMO is an independent (with respect to analytical dose calculation algorithm) and accurate Monte Carlo tool.

Background

Monte Carlo (MC) method was considered as the “golden standard” method to perform absorbed dose calculations in external radiotherapy [1,2,3,4,5,6,7,8]. A number of general-purpose MC codes were developed such as FLUKA [9], MCNP [10, 11], EGSnrc [12], PENELOPE [13], GEANT [14, 15], and employed for research and development in medical applications during the past decades. The state-of-the-art Monte Carlo method was available in a few commercial treatment planning systems (TPS) such as Elekta Monaco XVMC (X-ray Voxel MC) [16], Accuray Multiplan [17], Brain Lab IPlan [18], Nomos Corvus [19], which fully takes the electron transport in the medium/water into account and achieves highest accuracy. Despite a continuously increasing interest of MC in radiotherapy treatment planning, the introduction of MC algorithms in clinical practice has been delayed by the excessive calculation time involved.

Typically, the accurate Monte Carlo simulation needed the detailed description of the material properties and geometrical model of the linac head, which was a time-consuming and error-prone task [20]. PRIMO is a program based on PENELOPE 2011 [13, 21, 22], PENEASY [23], dose planning method (DPM) [24], and PENEASYLINAC [23]. It has a graphical user interface and encompasses all the components in a single user-friendly environment. Furthermore, PRIMO is an open access software and everyone can get its license for free.

There were many studies about linac simulation using PRIMO [25,26,27,28,29,30,31,32,33,34,35,36,37,38,39]. Different types of linac machines were simulated using PRIMO, including Varian TrueBeam [25, 29,30,31, 35, 37, 39], Varian 2100 C/D [27], Clinac 2300 [28], Clinac® iX [33, 36], Trilogy [32], Novalis [32], EDGE [34]. Some of the previous studies adopted the phase space files (PSF) from MyVarian website to model the linac head [25, 29, 35, 39]. Some researchers simulated linacs using the default beam parameters given by PRIMO [31, 34], while other studies modified the beam parameters based on the default recommendations in PRIMO [27, 28, 30, 32, 33, 36]. As two critical parameters of linac head model, percentage depth dose (PDD) and off-axis ratio (OAR) were evaluated in many studies. Belosi et al. [25] studied the accuracy of Varian PSF for flattening filter free (FFF) beams, and compared to measured PDD and OAR from ten TrueBeam linacs. Pita et al. [28] simulated a Varian Clinac 2300, compared the PDD and OAR with the measurements, and validated the static field and IMRT field.

Previous studies using PRIMO focused on the evaluation of linac head model [25,26,27, 30, 31, 35, 36, 38]. However, the evaluation of multileaf collimator (MLC) model was critical in clinical use [40]. Since the deviation of MLC model in Eclipse™ TPS (transmission factor and DLG) could produce large dosimetric discrepancies, the MLC model should be evaluated comprehensively in beam configuration [40]. Paganini et al. [34] established a series of different MLC patterns to evaluate the MLC model. The difference of DLG values obtained with PRIMO, Acuros, and EBT3 Gafchromic film measurements were within 0.008 cm [34]. Although the film has high spatial resolution, it has the issues of regional non-uniformity and positioning error, which may cause the measurement uncertainty [41]. In this study, we introduced EPID measurement in the evaluation of MLC model, which allows fast measurement, less positioning error, and highly reproducible setup. As far as we know, there were no previous reports about the comparison of simulated DLG and transmission factor with EPID measurements.

In this study, we investigated the linac head and MLC model of Varian VitalBeam™, including PDD, OAR, transmission factor and DLG. Firstly, the linac head with photon beams of 6MV and 6MV FFF was simulated in PRIMO. The initial energy, energy full width at half maximum (FWHM), focal spot FWHM, and beam divergence were tuned in PRIMO to match the PDD and OAR measurement. Secondly, we simulated the transmission factor and DLG in the Millennium 120 MLC model. The simulated results were compared with the EPID measurement and optimized value using clinical treatment plans.

Material and method

Measurement equipment

All measurements were conducted on a Varian VitalBeam™ (Varian Medical Systems, Palo Alto, CA) linac equipped with a Millennium 120 leaf MLC. Photon beams of 6MV and 6MV FFF were used. A PTW BEAMSCAN water phantom (PTW Freiburg, Germany), an ion chamber (Semiflex 3D Type TW31021, PTW, Freiburg, Germany), and an electrometer (PTW UNIDOS E, T10009, PTW, Freiburg, Germany) were used in the measurements. The commercial software MEPHYSTO™ (PTW, Freiburg, Germany) was used to perform the evaluation of dose profiles. PRIMO software (version 0.3.64.1800_x64) was installed on a workstation (Windows 10 Enterprise, Intel Core i7-7820X CPU@3.6 GHz, RAM 32 GB).

Amorphous silicon-based electronic portal imaging device (EPID) was used to measure the transmission factor and DLG. The active area of EPID detector was 30 cm × 40 cm (1024 × 768 matrix) with pixel size of 0.39 mm. The images were acquired in “the integrated image” mode used 4DTC as the acquisition workstation at the source detector distance (SDD) of 105 cm. The frame rate of EPID is 9.6 frames/s. Portal dose image prediction (version 8.0, Varian Medical Systems, Inc., USA) was used to compare the predicted portal dose images with the measured ones. Dose image was calculated using the Portal Dose Image Prediction (PDIP) algorithm in the Eclipse™ version 13.6.23 (Varian Medical Systems Inc., Palo Alto, CA).

The dosimetric leaf gap (DLG) in Eclipse™ TPS was optimized to reduce the discrepancies between the measured dose distributions and the calculated ones of patient treatment plans. The Delta4 phantom system (Scandidos, Uppsala, Sweden) was used in the dose verification for the clinical treatment plans. Gamma-index evaluation method implemented in Delta4 was applied to quantify the discrepancies of dose distributions. Delta4 is a cylindrical polymethyl-methacrylate phantom with 22 cm in diameter and 40 cm in length. The mass density of the phantom was 1.19 g/cm3, and the relative electron density was 1.147. It consists of 1069 p-type Silicon diodes in a crossed array inside the phantom, with 5 mm resolution in central area and 10 mm resolution in the peripheral area. The software ScandiDos Delta4 allowed the users to compare the measured dose distribution with the dose distribution calculated using TPS.

Measurements of the critical parameters for TPS simulation

We measured the PDD and OAR of 6MV and 6MV FFF photon beam for Varian VitalBeam linac at Chinese Academy of Medical Sciences Cancer Hospital, Shenzhen center, respectively. The scan range of the detector in horizontal plane was 50 cm by 50 cm. The measurements were performed in continuous scanning mode at the speed of 10 mm/s. Different square fields (3 cm, 4 cm, 6 cm, 8 cm, 10 cm, 15 cm, 20 cm, 30 cm, and 40 cm) were used with a fixed source-to-phantom distance (SPD) of 100 cm.

Transmission factor and DLG are two crucial parameters to model the rounded MLC leaf in Eclipse™ TPS. The transmission factor is determined by the material and height of the MLC leaf, beam quality, and etc. EPID was used to measure the MLC transmission in two MLC-closed fields: one with MLC bank A closed and the other with MLC bank B closed as shown in Fig. 1. The transmission factor was calculated using the following equations:

$$tr\left( A \right) = R_{A} /R_{open}$$
(1)
$$tr\left( B \right) = R_{B} /R_{open}$$
(2)
$$R_{T} = \left[ {\frac{tr\left( A \right) + tr\left( B \right)}{2}} \right]$$
(3)

where tr(A) and tr(B) are transmission factors of MLC bank A and bank B, respectively. Ropen is the reading of the open field. RA and RB are the readings of closed MLC bank A and closed MLC bank B, respectively. RT is the average MLC leaf transmission for both MLC banks.

Fig. 1
figure 1

The BEV projection of the completely blocked MLC field with MLC bank A closed (RA) and bank B closed (RB), respectively

DLG was measured using sweeping gap technique [42]. The sweeping gap test plans were designed in Eclipse™ TPS (Varian Medical Systems, Inc., USA), including dynamic sweeping gaps of 2 mm, 4 mm, 6 mm, 10 mm, 14 mm, 16 mm, and 20 mm (shown in Fig. 2), two closed-MLC fields, and one open-MLC field. The test plans were generated with gantry angle of 0 degree, collimator setting of 90 degree, SSD at 100 cm, MLC moving from − 6 to + 6 cm in dynamic mode at the speed of 2.5 cm/s. For each treatment plan, the jaws were set to 10 cm by 10 cm as the reference field. Each sweeping gap traveled across the reference field, and the delivered dose was 100 MU.

Fig. 2
figure 2

DLG test plan with gap width of 20 mm

In order to evaluate the contribution of sweeping gap field, the MLC transmission reading RgT was subtracted from the initial ionization reading \(\cdot R_{g}\). The corrected gap reading (\({R}_{{g}^{^{\prime}}}\)) and RgT for each gap (g) were defined by Varian guideline as:

$$R_{gT} = R_{T} \cdot \left[ {1 - \frac{g}{120}} \right]$$
(4)
$$R_{{g^{\prime}}} = R_{g} - R_{gT}$$
(5)

where g is the nominal gap width (unit: mm), the sweeping gap movement range is 120 mm. Rg is the initial sweeping gap field reading (CU values for EPID). The corrected gap readings were fitted using linear regression method. DLG was the value at the intersection between the extrapolated extension of the fitted line and y-axis (Fig. 6). R-squared was used to measure the goodness-of-fit for the linear regression model.

PRIMO simulation

The PRIMO simulation setup consists of three segments [31, 32, 43] as shown in Fig. 3. In segment 1 (s1), the upper part of the linac was simulated. The user could adjust the primary beam parameters including nominal energy, initial energy, energy FWHM, focal spot FWHM, and beam divergence. In s1, the four parameters were tuned to match the PDDs in different field size and OARs in different depths. In the segment 2 (s2), the field parameters were edited, including the treatment technique, beam weights, gantry start and end angles, collimator angle, couch angle, MLC type, aperture size, applicator, and isocenter location. In segment 3 (s3), the linac model was applied in patient and phantom geometry. The three segments can be grouped in a user-defined way [43].

Fig. 3
figure 3

The three segments in PRIMO simulation

In PRIMO, Varian 2100 linac model was recommended to simulate TrueBeam for 6MV, while FakeBeam linac model is an experimentally based geometry of TrueBeam for 6MV FFF and 10MV FFF. In segment 1, we selected Varian 2100 and Fake Beam from PRIMO linac model list, as the initial models of 6MV and 6MV FFF Varian VitalBeam linac, respectively. The linac head geometry is shown in Fig. 4. The transport parameters of linac head components are shown in the Additional file 1: Table 1. To determine the beam characteristics, there were four crucial parameters in PRIMO, including initial electron energy, energy FWHM, FWHM of the focal spot size, and beam divergence. We fine-tuned the four crucial parameters to match the simulated PDD and OAR with our measurements. Among the four parameters, the initial electron energy was firstly determined based on the depth of the maximum dose on the depth dose curves based on field sizes of 10 cm × 10 cm. The other three parameters were adjusted until both simulated PDD and OAR had the highest GPRs compared with our measurements in all the considered field sizes (For field sizes smaller than 10 cm × 10 cm (included), the GPRs (1%/1 mm) were higher than 90%, other field sizes were higher than 85%).

Fig. 4
figure 4

The illustration of head geometry for Varian VitalBeam™ operating in photon mode (From PRIMO user’s manual software version 0.3.(32–64).1880)

In our study, the search range of initial electron energy was 5.4–6.6 MeV, and the interval was 0.1 MeV. The energy FWHM varied from 0 to 0.1 MeV with an interval of 0.01 MeV. The FWHM of the focal spot size varied from 0 to 0.5 cm with an interval of 0.05 cm. The divergence of the parallel beams was considered to be zero degree.

In segment (s1), the PENELOPE computation engine was used to obtain the phase-space file. Splitting roulette was adopted as the variance reduction technique to improve simulating efficiency. In segment 2 (s2) and segment 3 (s3), the dose planning method (DPM) was employed to calculate dose distribution. The splitting factor in DPM was 300. Particle histories increased from 109 in s1 to 1010 in s2 and s3. The number of particle histories increased along with the increase of the field size to reduce the uncertainty.

Both PDD and OAR were simulated in the water phantom under the following condition: gantry angle, 0 degree; collimator angle, 0 degree; SSD, 100 cm; fixed bin size in the tallying volume, 0.2 × 0.2 × 0.2cm3. In our study, there were nine square field sizes considered, including 3 cm, 4 cm, 6 cm, 8 cm, 10 cm, 15 cm, 20 cm, 30 cm, and 40 cm. All the square fields were collimated by the jaws. The tallying volumes increased with the increase of the square field size. The corresponding tallying volumes in PRIMO were 9 × 9 × 35cm3, 12 × 12 × 35cm3, 18 × 18 × 35cm3, 24 × 24 × 35cm3, 30 × 30 × 35cm3, 40 × 40 × 35cm3, 50 × 50 × 35cm3, 60 × 60 × 35cm3, and 70 × 70 × 35cm3, respectively.

The model of Millennium 120 MLC was selected in PRIMO to evaluate the MLC model. The model consists of 40 central leaf pairs with a 5 mm projection width at the isocenter (SSD = 100 cm) and 20 outer leaf pairs with a width of 10 mm. The MLC transmission factor were obtained using Eqs. (1)–(3). The designed sweeping gap test plans were imported into PRIMO. The dose tallying volume in simulation was 15 × 15 × 10cm3. The point dose (Rg) was acquired at 5 cm below the dose tallying volume. The Rg readings of different sweeping gap test plans were fitted using linear regression method. DLG was acquired as the value at the intersection of the extrapolated fitting line and y-axis (Fig. 5).

Fig. 5
figure 5

Comparison between simulations and measurements of PDD and OAR (The dose distribution was normalized to the maximum dose point). a PDD 6MV, 10 × 10 cm2, b PDD 6MV FFF, 10 × 10 cm2, c the gamma index for 6MV PDD, d the gamma index for 6MV FFF PDD, e OAR 6MV, 10 × 10 cm2 depth at 1.5 cm, f OAR 6MV FFF, 10 × 10 cm2 depth at 1.5 cm, g the gamma index for 6MV OAR, h the gamma index for 6MV FFF OAR

Results

PDDs and OARs of different field sizes

For PDD and OAR, we modified the critical beam parameters to match the simulation with measurement results. The critical parameters of 6MV photon beam were listed as follows: initial electron energy, 5.9 MeV; FWHM, 0.03 MeV; focal spot FWHM, 0.1 cm; beam divergence, 0 degree. The critical parameters of 6MV FFF photon beam were listed as follows: initial electron energy, 5.8 MeV; FWHM, 0.04 MeV; focal spot FWHM, 0.20 cm; beam divergence, 0 degree. The average statistical dose uncertainty in the PRIMO simulation was below 2%. All uncertainties reported by PRIMO are given at 2 standard deviations (2σ).

The comparison of measured and simulated PDD and OAR were shown in Fig. 5. The GPRs for PDD and OAR were shown in the Additional file 1: Tables 2, 3, 4 and 5, respectively. For 6MV photon beam PDD, the GPRs were 100% in all the considered fields as shown in the Additional file 1: Table 2, when the 3%/3 mm gamma criterion was used. The GPRs of PDD in all the nine field sizes were over 99.33%, 99% and 99% with the criteria of 2%/2 mm, 1%/2 mm and 1%/1 mm, respectively.

For 6MV FFF photon beam PDD, the GPRs were 100% in all the considered fields using the criteria of 3%/3 mm. As shown in the Additional file 1: Table 3, the GPRs of PDD in the nine field sizes were over 99.33%, 98.99% and 97.64% with the criteria of 2%/2 mm, 1%/2 mm and 1%/1 mm, respectively.

For 6MV photon beam OAR, the GPRs were 100% in all the nine fields with the criteria of 3%/3 mm. As shown in the Additional file 1: Table 4, the GPRs of OAR in the nine field sizes were over 96.4%, 90.6%, and 85.12% with the criteria of 2%/2 mm, 1%/2 mm and 1%/1 mm, respectively.

For 6MV FFF photon beam OAR, the GPRs were 100% in all the nine fields with the criteria of 3%/3 mm. As shown in the Additional file 1: Table 5, the GPRs of OAR in the nine field sizes were over 95.15%, 89.32%, and 86.55% with the criteria of 2%/2 mm, 1%/2 mm and 1%/1 mm, respectively.

Transmission factor and DLG

The transmission factor for 6MV photon beam was 1.25% versus 1.32%, EPID measured vs PRIMO simulated. For 6MV FFF photon beam, the transmission factor was 0.8% versus 1.12%, EPID measured vs PRIMO simulated (Table 1).

Table 1 The transmission factor and DLG values for 6MV and 6MV FFF photon beams

For 6MV photon beam, the linear regression equation based on PRIMO simulation was y = 269.34x − 1.36 with R square of 0.99. The slope of the fitted line is 269.34, and the DLG value was 1.36. For 6MV FFF photon beam, the linear regression equation based on PRIMO simulation was y = 268.35x − 1.07 with R square of 0.99 (Fig. 6). The slope of the fitted line is 268.35, and the DLG value was 1.07.

Fig. 6
figure 6

DLG obtained from PRIMO simulation. a 6MV photon beam, b DLG obtained using PRIMO for 6MV FFF photon beam (R2 = 0.999)

For 6MV photon beam, the linear regression equation based on EPID measurements was y = 53.79x − 1.41 with R square of 0.99. The slope of the fitted line is 53.79, and the DLG value was 1.41. For 6MV FFF photon beam, the linear regression equation based on EPID measurements was y = 53.29x − 1.03 with R square of 0.99 (Fig. 7). The slope of the fitted line is 53.29, and the DLG value was 1.03.

Fig. 7
figure 7

DLG obtained from EPID measurements. a 6MV photon beam, b 6MV FFF photon beam (R2 = 0.999)

Discussion

In this study, we tuned the four crucial parameters in PRIMO, including initial energy, energy FWHM, focal spot FWHM, and beam divergence. The fine-tuned Monte Carlo program was used to verify linac head model and MLC model. PDD and OAR are two vital parameters of linac head model. Transmission factor and DLG are the important parameters of MLC model in Eclipse™ TPS. Nine field sizes were considered, including 3 cm, 4 cm, 6 cm, 8 cm, 10 cm, 15 cm, 20 cm, 30 cm, and 40 cm. The results demonstrated that the simulated PDD, OAR, transmission factor, and DLG matched well with the measurements.

Although no programming was required when using PRIMO, the PDD and OAR match in beam commissioning of the linac head model was intensive and laborious. Some studies used the default parameters in the primary beam setup segment [31, 34]. Some research groups imported the external phase space files offered by Varian into PRIMO to reduce the workload [25, 29, 35, 39]. In this study, we tuned the parameters in the primary beam setup segment to match our measurements of the linac machines. Many studies about beam configuration for linac model only compared simulated PDD and OAR with measurements [25, 26, 29, 30, 38, 39]. It could lead to dose deviation due to the lack of MLC model configuration when clinical treatment plans were compared with simulation results. In this study, MLC modeling combined with PDD and OAR matching covered a more integrated and accurate work. The fine-tuned linac models using PRIMO have potential to be used as an independent tool in quality assurance of clinical treatment plans.

There are mainly three methods to measure the transmission factor and DLG, including ion-chamber, film and EPID [44]. Due to the independence of the energy spectrum and linear increase with dose, the ion chamber measurements were considered as the gold standard [45]. However, the positioning error and the volume effect may result in measurement error in the condition of small MLC gaps. Paganini studied the DLG and transmission factor of 10MV FFF photon beam using PRIMO and compared the results with the microdiamond detector and Gafchromic film measurements [34]. Film has the advantage of high spatial resolution and reproducibility, but the non-uniform response and data analysis may increase the measurement uncertainty. Compared with ion chamber and film, EPID has the advantage of fast measurement, easy operation, less positioning error, and highly reproducible set-up [45]. In this study, EPID was used to measure the transmission factor and DLG.

Some studies found it could generate large difference between calculated and measured dose distributions when using the DLG obtained by sweeping gap method. It is recommended to introduce the optimized DLG using clinical treatment plans [46, 47]. In this study, we optimized the DLG value using Delta4 phantom and 100 clinical treatment plans. The optimized DLG produced the highest GPRs between the calculated and measured dose distributions in the clinical treatment plans. We found the optimized DLGs were similar with the measured ones, and larger than the simulated ones, which is consistent with the previous study [48]. Our result demonstrated that the initial values of transmission factor and DLG can be obtained using PRIMO. In the future, we can generate transmission factor and DLG based on the calculation using PRIMO and fine-tune the two parameters using patient specific quality assurance measurements. This will help the future commissioning of Eclipse in new linac site.

As an independent dose verification software, PRIMO has potential to be used in clinical application, such as beam commission, treatment planning verification, and quality assurance. Our future work will focus on the clinical applications of the commissioned PRIMO program, such as the treatment planning calculations of skin doses with bolus, and the verification of dose distributions in the inhomogeneous medium.

Conclusion

This study used the PRIMO software to simulate the photon beam 6MV and 6MV FFF of a Varian VitalBeam linac. Our study showed that PRIMO is an accurate, self-contained, Monte Carlo-based linac simulator and dose calculator with a user-friendly graphical interface. The commissioned PRIMO can be used as an effective tool to provide a potential independent quality assurance.

Availability of data and materials

The datasets used during the current study are available from the corresponding author on reasonable request.

Abbreviations

MC:

Monte Carlo

TPS:

Treatment planning system

DPM:

Dose planning method

PSF:

Phase space file

FFF:

Flattening filter free

PDD:

Percentage depth dose

OAR:

Off-axis ratio

MLC:

Multileaf collimator

DLG:

Dose leaf gap

FWHM:

Energy full width at half maximum

EPID:

Electronic portal imaging device

SDD:

Source detector distance

SPD:

Source-to-phantom distance

SBRT:

Stereotactic body radiation therapy

PO:

Photon optimizer

AAA:

Anisotropic analytical algorithm

DD:

Dose-difference

DTA:

Distance-to-agreement

GPR:

Gamma passing rate

IMRT:

Intensity modulated radiation therapy

References

  1. Vilches M, Garcia-Pareja S, Guerrero R, et al. Monte Carlo simulation of the electron transport through thin slabs: A comparative study of PENELOPE, GEANT3, GEANT4, EGSnrc and MCNPX. Nucl Instrum Methods Phys Res Sect B. 2007;254(2):219–30.

    Article  CAS  Google Scholar 

  2. Vilches M, Garcia-Pareja S, Guerrero R, et al. Monte Carlo simulation of the electron transport through air slabs: a comparative study of PENELOPE, GEANT3, Geant4 and EGSnrc Monte Carlo codes. IEEE Trans Nucl Sci. 2008;55(2):710–6.

    Article  Google Scholar 

  3. Vilches M, García-Pareja S, Guerrero R, et al. Multiple scattering of 13 and 20 MeV electrons by thin foils: a Monte Carlo study with GEANT, Geant4, and PENELOPE. Med Phys. 2009;36(9Part1):3964–70.

    Article  CAS  PubMed  Google Scholar 

  4. Faddegon BA, Kawrakow I, Kubyshin Y, et al. The accuracy of EGSnrc, Geant4 and PENELOPE Monte Carlo systems for the simulation of electron scatter in external beam radiotherapy. Phys Med Biol. 2009;54(20):6151.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  5. Reynaert N, Van der Marck SC, Schaart DR, et al. Monte Carlo treatment planning for photon and electron beams. Radiat Phys Chem. 2007;76(4):643–86.

    Article  CAS  Google Scholar 

  6. Nahum A. Monte Carlo techniques in radiation therapy. Med Phys. 2015;42(8):4999–5000.

    Article  Google Scholar 

  7. Verhaegen F, Seuntjens J. Monte Carlo modelling of external radiotherapy photon beams. Phys Med Biol. 2003;48(21):R107-164.

    Article  CAS  PubMed  Google Scholar 

  8. Reynaert N, De Smedt B, Coghe M, et al. MCDE: a new Monte Carlo dose engine for IMRT. Phys Med Biol. 2004;49(14):N235-241.

    Article  CAS  PubMed  Google Scholar 

  9. Ferrari A, Sala PR, Fasso A, et al. FLUKA: a multi-particle transport code. Stanford Linear Accelerator Center (SLAC). 2005.

  10. Brown FB, Barrett RF, Booth TE, et al. MCNP version 5. Trans Am Nucl Soc. 2002;87(273):02–3935.

    Google Scholar 

  11. Pietrzak R, Konefał A. Determination of energy spectra in water for 6 MV X rays from a medical linac. Acta Phys Polonica Ser B. 2016;47(3):783.

    Article  Google Scholar 

  12. Kawrakow I. The EGSnrc code system, Monte Carlo simulation of electron and photon transport. NRCC Report Pirs-701. 2001.

  13. Baro J, Sempau J, Fernández-Varea JM, et al. PENELOPE: an algorithm for Monte Carlo simulation of the penetration and energy loss of electrons and positrons in matter. Nucl Instrum Methods Phys Res B. 1995;100(1):31–46.

    Article  CAS  Google Scholar 

  14. Agostinelli S, Allison J, Amako KA, et al. GEANT4-a simulation toolkit. Nucl Instrum Methods Phys Res A: Accel Spectrom Detect Assoc Equip. 2003;506(3):250–303.

    Article  CAS  Google Scholar 

  15. Konefał A, Bakoniak M, Orlef A, et al. Energy spectra in water for the 6 MV X-ray therapeutic beam generated by Clinac-2300 linac. Radiat Meas. 2015;72:12–22.

    Article  Google Scholar 

  16. Fippel M. Fast Monte Carlo dose calculation for photon beams based on the VMC electron algorithm. Med Phys. 1999;26(8):1466–75.

    Article  CAS  PubMed  Google Scholar 

  17. Sharma S, Ott J, Williams J, et al. Dose calculation accuracy of the Monte Carlo algorithm for CyberKnife compared with other commercially available dose calculation algorithms. Med Dosim. 2011;36(4):347–50.

    Article  PubMed  Google Scholar 

  18. Fragoso M, Wen N, Kumar S, et al. Dosimetric verification and clinical evaluation of a new commercially available Monte Carlo-based dose algorithm for application in stereotactic body radiation therapy (SBRT) treatment planning. Phys Med Biol. 2010;55(16):4445–64.

    Article  PubMed  Google Scholar 

  19. Redmond KJ, Lo SS, Dagan R, et al. A multinational report of technical factors on stereotactic body radiotherapy for oligometastases. Future Oncol. 2017;13(12):1081–9.

    Article  CAS  PubMed  Google Scholar 

  20. Theis C, Buchegger KH, Brugger M, et al. Interactive three-dimensional visualization and creation of geometries for Monte Carlo calculations. Nucl Instrum Methods Phys Res Sect A. 2006;562(2):827–9.

    Article  CAS  Google Scholar 

  21. Salvat F, Fernández-Varea J M, Sempau J. PENELOPE-2006: a code system for Monte Carlo simulation of electron and photon transport. Workshop proceedings. Nuclear Energy Agency, Organization for Economic Cooperation and Development, Barcelona, Spain 2006; 4 (6222): 7.

  22. Sempau J, Acosta E, Baro J, et al. An algorithm for Monte Carlo simulation of coupled electron-photon transport. Nucl Instrum Methods Phys Res B Beam Interact Mater At. 1997;132(3):377–90.

    Article  CAS  Google Scholar 

  23. Sempau J, Badal A, Brualla L. A PENELOPE-based system for the automated Monte Carlo simulation of clinacs and voxelized geometries—application to far-from-axis fields. Med Phys. 2011;38(11):5887–95.

    Article  PubMed  Google Scholar 

  24. Sempau J, Wilderman SJ, Bielajew AF. DPM, a fast, accurate Monte Carlo code optimized for photon and electron radiotherapy treatment planning dose calculations. Phys Med Biol. 2000;45(8):2263–91.

    Article  CAS  PubMed  Google Scholar 

  25. Belosi MF, Rodriguez M, Fogliata A, et al. Monte Carlo simulation of TrueBeam flattening-filter-free beams using Varian phase-space files: comparison with experimental data. Med Phys. 2014;41(5):051707.

    Article  PubMed  Google Scholar 

  26. Omer MOME. Monte Carlo simulation of medical linear accelerator using primo code. Thesis/Dissertation, Sudan Academy of Sciences (SAS), Khartoum (Sudan), 2014.

  27. Maskani R, Tahmasebibirgani MJ, Hoseini-Ghahfarokhi M, et al. Determination of initial beam parameters of varian 2100 CD linac for various therapeutic electrons using PRIMO. Asian Pac J Cancer Prev. 2015;16(17):7795–801.

    Article  PubMed  Google Scholar 

  28. Pita VCI. Monte Carlo LINAC Simulations using PRIMO for IMRT Treatment Verification. Thesis/Dissertation, UNIVERSIDADE DE LISBOA, Lisboa, 2016.

  29. Lloyd S, Gagne IM, Bazalova-Carter M, et al. Validation of Varian TrueBeam electron phase–spaces for Monte Carlo simulation of MLC-shaped fields. Med Phys. 2016;43(6):2894–903.

    Article  PubMed  Google Scholar 

  30. Efendi MA, Funsian A, Chittrakarn T, et al. Monte carlo simulation of 6 MV flattening filter free photon beam of TrueBeam STx LINAC at Songklanagarind hospital. Sains Malays. 2017;46(9):1407–11.

    Article  CAS  Google Scholar 

  31. Hermida-López M, Sánchez-Artuñedo D, Calvo-Ortega JF. PRIMO Monte Carlo software benchmarked against a reference dosimetry dataset for 6 MV photon beams from Varian linacs. Radiat Oncol. 2018;13(1):1–10.

    Article  Google Scholar 

  32. Esposito A, Silva S, Oliveira J, et al. Primo software as a tool for Monte Carlo simulations of intensity modulated radiotherapy: a feasibility study. Radiat Oncol. 2018;13(1):1–13.

    Article  CAS  Google Scholar 

  33. Rodriguez M, Brualla L. Treatment verification using Varian’s dynalog files in the Monte Carlo system PRIMO. Radiat Oncol. 2019;14(1):1–7.

    Article  Google Scholar 

  34. Paganini L, Reggiori G, Stravato A, et al. MLC parameters from static fields to VMAT plans: an evaluation in a RT-dedicated MC environment (PRIMO). Radiat Oncol. 2019;14(1):1–13.

    Article  Google Scholar 

  35. Mamballikalam G, Senthilkumar S, Jayadevan PM, et al. Evaluation of dosimetric parameters of small fields of 6 MV flattening filter free photon beam measured using various detectors against Monte Carlo simulation. J Radiother Pract. 2020;20(3):369–369.

    Article  Google Scholar 

  36. Sarin B, Bindhu B, Saju B, et al. Validation of PRIMO Monte Carlo model of ClinaciX 6MV photon beam. J Med Phys. 2020;45(1):24–35.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  37. Aamri H, Fielding A, Aamry A, et al. Comparison between PRIMO and EGSnrc Monte Carlo models of the varian true beam linear accelerator. Radiat Phys Chem. 2021;178:109013.

    Article  CAS  Google Scholar 

  38. Bacala AM. Linac photon beam fine-tuning in PRIMO using the gamma-index analysis toolkit. Radiat Oncol. 2020;15(1):1–11.

    Article  Google Scholar 

  39. Efendi MA, Funsian A, Chittrakarn T, et al. Monte Carlo simulation using PRIMO code as a tool for checking the credibility of commissioning and quality assurance of 6 MV TrueBeam STx varian LINAC. Rep Pract Oncol Radiother. 2020;25(1):125–32.

    Article  PubMed  Google Scholar 

  40. Wasbø E, Valen H. Dosimetric discrepancies caused by differing MLC parameters for dynamic IMRT. Phys Med Biol. 2008;53(2):405–15.

    Article  PubMed  Google Scholar 

  41. Kumaraswamy LK, Schmitt JD, Bailey DW, et al. Spatial variation of dosimetric leaf gap and its impact on dose delivery. Med Phys. 2014;41(11):111711.

    Article  PubMed  Google Scholar 

  42. Eclipse Algorithms Reference Guide. Varian Medical Systems, Inc., Palo Alto. 2010.

  43. Rodriguez M, Sempau J, Brualla L. PRIMO: a graphical environment for the Monte Carlo simulation of Varian and Elekta linacs. Strahlenther Onkol. 2013;189(10):881–6.

    Article  CAS  PubMed  Google Scholar 

  44. Mei X, Nygren I, Villarreal-Barajas JE. On the use of the MLC dosimetric leaf gap as a quality control tool for accurate dynamic IMRT delivery. Med Phys. 2011;38(4):2246–55.

    Article  PubMed  Google Scholar 

  45. Lorenz F, Nalichowski A, Rosca F, et al. Spatial dependence of MLC transmission in IMRT delivery. Phys Med Biol. 2007;52(19):5985–99.

    Article  PubMed  Google Scholar 

  46. Szpala S, Cao F, Kohli K. On using the dosimetric leaf gap to model the rounded leaf ends in VMAT/RapidArc plans. J Appl Clin Med Phys. 2014;15(2):67–84.

    Article  PubMed Central  Google Scholar 

  47. Kielar KN, Mok E, Hsu A, et al. Verification of dosimetric accuracy on the TrueBeam STx: rounded leaf effect of the high definition MLC. Med Phys. 2012;39(10):6360–71.

    Article  PubMed  Google Scholar 

  48. Lin CY, Shiau AC, Ji JH, et al. A simple method for determining dosimetric leaf gap with cross-field dose width for rounded leaf-end multileaf collimator systems. Radiat Oncol. 2018;13(1):1–7.

    Article  Google Scholar 

Download references

Acknowledgements

We would like to thank Dr. Xiaokun Liang for his helpful advice on various technical issues examined in this study.

Funding

Supported by Shenzhen High-level Hospital Construction Fund, Shenzhen Key Medical Discipline Construction Fund (No. SZXK013), Sanming Project of Medicine in Shenzhen (No. SZSM201612063), Shenzhen High-level Hospital Construction Fund, National Natural Science Foundation of China (No. 12005301), Hospital Research Project (No. SZ2020MS002), Guangdong Basic and Applied Basic Research Foundation (No. 2022A1515012456).

Author information

Authors and Affiliations

Authors

Contributions

Y and XS performed the simulations and carried out to the analysis of the Monte Carlo simulated results and provided editing assistance. Y and YH AC helped develop the idea of the study. CL guided the direction of research and edited the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Chenbin Liu.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1

. The transport parameters configuration in PRIMO and gamma passing rates for PDD and OAR.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Li, Y., Sun, X., Liang, Y. et al. Monte Carlo simulation of linac using PRIMO. Radiat Oncol 17, 185 (2022). https://doi.org/10.1186/s13014-022-02149-5

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13014-022-02149-5

Keywords