Dosimetric comparison between VMAT plans using the fast-rotating O-ring linac with dual-layer stacked MLC and helical tomotherapy for nasopharyngeal carcinoma

Abstract

Background

To compare the dosimetric profiles of volumetric modulated arc therapy (VMAT) plans using the fast-rotating O-ring linac (the Halcyon system) based on a dual-layer stacked multi-leaf collimator and helical tomotherapy (HT) for nasopharyngeal cancer (NPCa).

Methods

For 30 NPCa patients, three sets of RT plans were generated, under the same policy of contouring and dose constraints: HT plan; Halcyon VMAT plan with two arcs (HL2arc); and Halcyon VMAT plan with four arcs (HL4arc), respectively. The intended dose schedule was to deliver 67.2 Gy to the planning gross target volume (P-GTV) and 56.0 Gy to the planning clinical target volume (P-CTV) in 28 fractions using the simultaneously integrated boost concept. Target volumes and organ at risks dose metrics were evaluated for all plans. Normal tissue complication probabilities (NTCP) for esophagus, parotid glands, spinal cord, and brain stem were compared.

Results

The HT plan achieved the best dose homogeneity index for both P_GTV and P_CTV, followed by the HL4arc and L2arc plans. No significant difference in the dose conformity index (CI) for P_GTV was observed between the HT plan (0.80) and either the HL2arc plan (0.79) or the HL4arc plan (0.83). The HL4arc plan showed the best CI for P_CTV (0.88), followed by the HL2arc plan (0.83) and the HT plan (0.80). The HL4arc plan (median, interquartile rage (Q1, Q3): 25.36 (22.22, 26.89) Gy) showed the lowest Dmean in the parotid glands, followed by the HT (25.88 (23.87, 27.87) Gy) and HL2arc plans (28.00 (23.24, 33.99) Gy). In the oral cavity (OC) dose comparison, the HT (22.03 (19.79, 24.85) Gy) plan showed the lowest Dmean compared to the HL2arc (23.96 (20.84, 28.02) Gy) and HL4arc (24.14 (20.17, 27.53) Gy) plans. Intermediate and low dose regions (40–65% of the prescribed dose) were well fit to the target volume in HL4arc, compared to the HT and HL2arc plans. All plans met the dose constraints for the other OARs with sufficient dose margins. The between-group differences in the median NTCP values for the parotid glands and OC were < 3.47% and < 1.7% points, respectively.

Conclusions

The dosimetric profiles of Halcyon VMAT plans were comparable to that of HT, and HL4arc showed better dosimetric profiles than HL2arc for NPCa.

Background

Radiation therapy (RT) has long been the primary treatment modality in treating the patients with nasopharynx cancer (NPCa) [1,2,3,4]. RT for NPCa is, however, sometimes challenging as the target volumes are frequently large, deep-seated, and close to or even overlapping with the surrounding organs at risk (OARs). Several types of acute and delayed radiation-induced toxicities, including xerostomia, oral mucositis, esophagitis, myelitis, and brain stem necrosis, are inevitably accompanied during and after high dose RT depending on the radiation dose level and anatomical proximity among the radiation targets and surrounding OARs.

In an effort to reduce radiation-induced toxicities, advanced RT techniques which can provide highly conformal dose to the target volume, such as intensity modulated RT (IMRT), have been increasingly applied in treatment of the NPCa patients. In particular, IMRT has contributed to lowering radiation myelitis and brain stem necrosis by reducing the dose to the spinal cord and brain stem, which were rather frequently encountered during the conventional RT technique era [5]. Radiation-induced oral mucositis and xerostomia, however, still remain the annoying prices to be paid by most patients during the dose escalation IMRT for the NPCa patients [4, 6,7,8,9]. Furthermore, most normal tissues in the head and neck (the non-target volume) are sensitive to the radiation damage, so the reduction of the normal tissue dose as much as possible is still critical in reducing radiation-induced toxicity and improving the quality of life during and after RT [4, 6].

To meet these clinical needs, the development of new treatment techniques based on advanced treatment devices has been in progress continuously. The quality of the IMRT plan is closely related to many technical parameters: the characteristics of the multi-leaf collimator (MLC); inverse dose optimization algorithm; dose calculation engine, and beam delivery technique including treatment machine capability, respectively. There have been several technological advances in this regards, and volumetric modulated arc therapy (VMAT) is known to provide better dose distribution than static IMRT technique, and has become the mainstream IMRT beam delivery option in treating various cancer types including NPCa [10, 11].

The characteristics of MLC have an influence on dosimetric profiles of VMAT plan because they play an important role in both beam shaping and intensity modulation along with dose optimization algorithm. Particularly, the width and transmission of the MLC directly affect the target dose conformity and normal organ dose outside the target, respectively. The recently introduced fast-rotating O-ring linac (FOL), Halcyon system (Varian Medical Systems, Palo Alto, CA, USA) employed the unique staggered DL-MLC by using two MLCs of 1 cm width (actual resolution 0.5 cm) [12]. The DL-MLC can produce low MLC transmission (0.01% of the primary beam), and minimal tongue and groove effect when compared with the traditional C-arm Linac (1.36%) [12,13,14] and helical tomotherapy (HT, 0.53% interleaf leakage) [15]. In contrast, HT has binary MLCs (0.625 cm width), but it can generate fine optimization resolution (of a few millimeters), combined with varying jaw width and pitch [16]. Furthermore, HT employs a helical beam delivery with the dynamic jaw technique to enhance dose conformity to the target volume in the inferior-superior border of the target volume while minimizing the OAR dose [17], which can generate better dosimeric profiles in RT for NPCa [18,19,20,21]. These differences are expected to act differently in the treatment plan for NPCa, but sufficient studies have not been reported.

Understanding the capabilities of the new techniques, which are closely related to the OARs’ dose sparing, is important in selecting the treatment techniques for better clinical outcomes. Several studies have assessed the FOL plan in relation to the C-arm Linac plan in various treatment sites [10,11,12,13,14,15,16,17,18,19,20]. However, there are few plan comparison studies between the FOL and HT [22, 23], and there have been no reports of dose comparison studies for NPCa. Therefore, we performed the first comparative analysis of the dosimeric profiles between VMAT plans using FOL with DL-MLC and HT for NPCa.

Methods and materials

Patient selection and simulation

This dosimetric comparison study did not involve any experiments on humans or animals. With the approval from the institutional review board (IRB SMC 2020-08-120-001), 30 consecutive NPCa patients, who underwent definitive RT based on concurrent chemotherapy between May 2018 and April 2019 with HT (TomoHD™, Accuray®, Sunnyvale, CA, USA) at the authors’ institute were included in this study (Table 1). All patients underwent planning computed tomography (CT) (2.5-mm slice thickness; Discovery RT590, GE Healthcare, Milwaukee, WI) in the supine position with a customized immobilization device (thermoplastic mask, Aquaplast RT™, Q-fix®, USA) and a tongue immobilization device [24].

Table 1 Patients' characteristics

The gross tumor volume (GTV) and clinical target volume (CTV) were delineated based on all available clinical information, including diagnostic images. The planning target volumes (PTVs) were generated by adding 3-mm isotropic expansions from the GTV (P-GTV) and CTV (P-CTV), which were edited considering the actual anatomic boundaries, such as the spinal cord and skin surface. The OARs, including the spinal cord, brainstem, parotid gland, esophagus, optic nerve, optic chiasm, submandibular gland (SMG), cochlea, and OC were delineated according to previously published guidelines [9, 25]. The planning risk volumes were generated by adding 3-mm isotropic expansions for the spinal cord (P-cord) and optic apparatus. Delineation of all contours was performed by one radiation oncologist to ensure consistency.

Treatment planning

The HT plan (Precision™, Version 1.1.1.1, Accuray®, Sunnyvale, USA) and the Halcyon VMAT plans (Eclipse, Version 15.6, Varian Medical systems, Palo Alto, USA) with two arcs (HL2arc) and four arcs (HL4arc) were generated using the same planning CT and contours for the target and OARs with a 6-MV flattening-filter-free (FFF) beam. The typical dose schedule was 67.2 Gy to the P-GTV and 56.0 Gy to the P-CTV in 28 fractions using the simultaneously integrated boost.

The same dose constraints based on the internal plan guideline were applied to all plans (Table 2). Four constraints were set at the highest priority level for the PTVs, P-cord (maximum dose [Dmax] ≤ 45 Gy), optic nerve, and optic chiasm (Dmax ≤ 50 Gy). The second priority level was given to the brain stem (Dmax ≤ 50 Gy), parotid gland (mean dose [Dmean] ≤ 26 Gy), and OC (Dmean ≤ 30 Gy). The third priority level was given to the planning lens (P_lens) (Dmax ≤ 10 Gy), eyeball (Dmax ≤ 50 Gy), and esophagus and cochlea (Dmean ≤ 35 Gy). The lowest level constraint was given to the SMG (Dmean ≤ 30 Gy) and brain (Dmax ≤ 60 Gy).

Table 2 Dose constraints for inverse planning

For the HT plan, fine plan conditions were used, including a field width of 2.5 cm, modulation factor of 2.0, and pitch of 0.287 to avoid the thread effect [26]. Dynamic jaw mode (TomoEDGE™, Accuray) was employed to improve the longitudinal dose conformity by reducing the penumbra at the inferior and superior borders of the PTV [27]. The final dose was calculated using the collapsed-cone convolution algorithm with a fine dose calculation resolution (0.98 mm in the x–y plane and 2.5 mm in z).

HL2arc and HL4arc plans were created using two and four full dynamic arcs with DL-MLC, respectively. One isocenter with an automatic collimator angle option was employed for both plans. The final dose was calculated using the anisotropic analytical algorithm with fine-dose resolution (0.25 cm) [13].

HT and VMAT plans were generated by dosimetrists specializing in each plan under blind conditions. The plans shared the planning CT image, contour, and plan constraints with the optimization strategy. The same dose optimization strategy was applied based on the same plan conditions as follows. For all plans, the same order of the dose optimization priority was applied by controlling “importance (tomotherapy plan)” and “priority (Eclipse plan)” based on internal guideline (Table 2). Once the PTV dose met the goal, optimization was continued to reduce the doses to the OARs as much as possible while maintaining the PTV dose coverage.

Dosimetric comparison of the HT, HL2arc, and HL4arc plans

All planning data including calculated dose and contour information with CT image set of three VMAT plans on each patient were transferred to MIM Maestro® (MIM Software Inc., USA) using the Digital Imaging and Communications in Medicine protocol, and quantitative analysis of the dose and volume parameters was performed. To evaluate the target dose coverage, the Dmax received by 2% (D2) and the minimum dose received by 98% (D98) of the P_GTV and P_CTV, respectively, were compared among the three plans. The homogeneity index (HI = D5/D98) [28] and conformity index (CI) [29] were also compared.

To evaluate the dose to normal tissues, the following OAR-related dosimetric parameters were compared (Table 3): the Dmean to the parotid, esophagus, cochlea, eyeball, and OC; the Dmax to the brain stem, esophagus, P-cord, optic chiasm, optic nerve, SMG, eyeball, P-lens, and OC; and the volume of the OC that received doses of 15 Gy (V15), 30 Gy (V30), 45 Gy (V45), or more.

Table 3 Comparison of dosimetric characteristics

Furthermore, dose-spillage volumes (DSVs) were calculated to assess the rapid dose fall-off ability near the PTV according to the treatment technique [30,31,32]. The DSV was calculated as follows:

$${\text{DSV}} = \frac{{V_{X\% } - PTV}}{PTV}$$
(1)

where VX% is the volume covered by the X% isodose surface. High DSV (VHS) for the P_GTV and P_CTV and intermediate DSV (VIS) and low DSV (VLS) for the P_CTV were calculated by taking into account the volumes that received ≥ 90%, ≥ 50%, and ≥ 25% of the prescribed dose, respectively [30,31,32]. The ideal value of the DSV is close to zero, which indicates that the volume that received the prescribed dose fit the shape of the target volume well.

To estimate the impact of dose differences on clinical complications, normal tissue complications (NTCPs), including clinical stricture/perforation of esophagus, xerostomia (parotid), oral mucositis (OC), myelitis (spinal cord), and brain stem necrosis, were calculated using the Poisson-LQ model [33, 34] for all plans. The parameters used for NTCP calculation were taken from a previous study (Additional file 1: Table 1).

For statistical analysis of the dosimetric and NTCP comparisons between the three plans, pairwise comparisons were performed using the Wilcoxon signed-rank test based on the Bonferroni correction (SPSS statistics, version 27, IBM®). A probability level with a p value < 0.05 was considered significant.

Results

Treatment plan comparison

The comparisons of dosimetric parameters using the three VMAT techniques are summarized in Table 3. In comparison of target dose coverage, D2 (maximum dose) of the P_GTV and P_CTV was closest to the prescribed dose in the HT plan, followed by the HL4arc and HL2arc plans (p < 0.05). For the D98 (minimum dose) for P_GTV, no significant differences were observed between the HT and HL4arc, and HL2arc and HL4arc in the pairwise comparisons (p > 0.05), however, there was significant difference between the HT and HL2arc plans (p < 0.05). There was no significant difference in the D98 (minimum dose) for P_CTV between the three plans (p > 0.05). Therefore, the HT plan achieved the best HI for both P_GTV and P_CTV, followed by the HL4arc and HL2arc plans (p < 0.05).

No significant difference in the CI for P_GTV was observed between the HT plan (0.80) and either the HL2arc plan (0.79) or the HL4arc plan (0.83) (p > 0.05), although the HL4arc plan showed superiority over the HL2arc plan in the pairwise comparison (p < 0.05). However, the HL4arc plan showed the best CI for P_CTV (0.88), followed by the HL2arc plan (0.83) and the HT plan (0.80) (p < 0.05).

Although the HL4arc plan showed the lowest Dmax in the P_cord, optic nerve, and eyeball, followed by the HL2arc and HT plans, all plans met the dose constraints with sufficient dose margins. Furthermore, both HL2arc and HL4arc plans showed lower Dmax in the brain stem and optic chiasm compared to the HL plan (p < 0.05). In contrast, the HT plan showed a lower Dmax in the lens and a lower Dmean in the SMG, cochlea and esophagus than the HL2arc and HL4arc plans. However, all values were within the dose constraints. The HL4arc (median, interquartile range [IQR] (Q1, Q3): 25.36 (22.22, 26.89) Gy) and HT (25.88 (IQR: 23.87, 27.87) Gy) plans showed lower Dmean in the parotid gland compared to the HL2arc plan (28.00 (IQR: 23.24, 33.99) Gy) (p < 0.05). In OC dose comparison, the HT (22.03 (IQR: 19.79, 24.85) Gy and 63.57 (IQR: 56.46, 75.79)%) plan showed the lowest Dmean and V15 compared to the HL2arc (23.96 (IQR: 20.84, 28.02) Gy and 86.46 (IQR: 68.99, 93.65)%) and HL4arc (24.14 (IQR: 20.17, 27.53) Gy and 82.62 (IQR: 60.65, 97.29)%) plans (p < 0.05), whereas there was no significant difference in V30 between the plans (p > 0.05). However, the HL4arc (6.01 (IQR: 3.08, 8.30)%) plan had a lower V45 than the HT (6.29 (IQR: 4.34, 9.73)%) plan.

For DSV analysis, the HT (1.68 (IQR: 1.35, 2.12) plan had the lowest median VHS for P_GTV, followed by the HL4arc (1.81 (IQR: 1.22, 2.64) and HL2arc (2.03 (IQR:1.40, 3.01) plans. No significant difference was observed between the HT and HL4arc plans in pairwise comparison (p > 0.05), although the VHS were significantly different between the HT and HL2arc, and HL2arc and HL4arc plans (p < 0.05). However, the HL4arc (0.40 (IQR: 0.36, 0.42) and 2.62 (IQR: 2.46, 2.85)) plan had the lowest median VHS and VIS for the P_CTV, followed by the HL2arc (0.43 (IQR: 0.39, 0.55) and 2.79 (IQR: 2.56, 3.10)) and HT (0.61 (IQR: 0.56, 0.71) and 3.75 (IQR: 3.40, 4.24)) plans, and significant differences were observed in pairwise comparisons (p < 0.05). The median VLS for the P_CTV showed a more pronounced pattern between HT and the HL4arc and HL2arc (p < 0.05), except no significant difference was observed between the HL4arc and HL2arc plans in a pairwise comparison (p > 0.05). This was well represented in the dose distribution for a patient (Fig. 1). Intermediate and low dose regions (40–65% of the prescribed dose) were well fit to the target volume in HL4arc, compared to the HT and HL2arc plans.

Fig. 1
figure 1

Comparison of axial dose distribution for a nasopharyngeal cancer patient, helical tomotherapy (HT, upper), two-arc halcyon (HL2arc, middle), and four-arc halcyon (HL4arc, low) plans. Intermediate- and low dose regions (40–65% of the prescribed dose) were well fit to target volume in HL4arc compared to HT and HL2arc plans

No significant differences in NTCPs for the esophagus, spinal cord, and brain stem were observed between the three plans (Table 4). The HL4arc (22.45 (IQR: 18.87, 27.87)) plan had the lowest median NTCP in the parotid glands, followed by the HT (24.82 (IQR: 20.15, 29.66)) and HL2arc (25.92 (IQR: 21.11, 33.71)) plans. In pairwise comparison, the HL4arc plan showed a significant difference median NTCP for parotid glands compared with the HT and HL2arc plan (p < 0.05), but no significant difference was observed between the HT and HL2arc plans (p > 0.05). The HL2arc (4.48 (IQR: 1.63, 12.80) plan had the lowest median NTCP in the OC, followed by the HL4arc (5.06 (IQR: 1.74, 9.06) and HT (6.18 (IQR: 2.81, 11.15) plans. Furthermore, the HL4arc plan was significantly different compared to the HT plan in the median incidence of NTCP for the OC in pairwise comparison (p < 0.05), but no significant difference was observed between the HL4arc and HL2arc, and HL2arc and HT plans (p > 0.05).

Table 4 Comparison of normal tissue complication

Discussion

Efforts have been made to reduce the dose to OARs near the target volume as much as possible to reduce radiation-induced toxicity and provide a better quality of life during and after RT for NPCa. HT, which was introduced relatively earlier, showed favorable outcomes [18], and FOL was implemented relatively recently for the same purposes in clinical practice. FOL and HT provide similar VMATs based on the same 6 megavoltage photon beam with the FFF design [12, 16] but have distinctly different beam delivery techniques. Various dosimetric characteristics, including dose optimization and calculation algorithm, characteristics of the MLC [32], beam model condition, beam delivery technique etc., interact simultaneously to meet the dose constraints during the invers-dose optimization [35]. Therefore, it is important to evaluate the dosimetric profile of the treatment plan by a new treatment machine in selecting the optimal treatment techniques in order to achieve better clinical outcomes. It is, however, not easy to distinguish which parameters influenced a certain dosimetric profile particularly. This is a plan comparison study, comparing the overall plan quality, based on a typical clinical environment.

Low transmissions of the MLC and interleaf leakage can be effective in reducing normal organ dose outside the target volume during dose optimization based on intensity modulation. It can be considered as one of the many possibilities that contributed to lowering the VIS and VLS for P_CTV in both HL2arc and HL4arc plans, compared to HT in our study. In contrast, Li et al. reported that DL-MLC of the Halcyon™ had no significant impact on plan quality of the head and neck VMAT compared to conventional MLC of Truebeam® (Varian Medical Systems, Palo Alto, CA, USA). It was only effective in IMRT [14].

For OARs sparing, the Dmean of the parotid glands was the lowest in the following order—HL4arc, HT, and HL2arc. However, it doesn’t seem to have clinical significant because both HL4arc and HT met the dose threshold (\(\approx 25\mathrm{ Gy}\) for both glands) for severe xerostomia (long-term salivary function < 25%) [7]. High OC dose leads to radiation-induced acute and late toxicity associated with mucositis include oral pain, dysphagia, weight loss, and secondary infections [36, 37]. All plans met the dose threshold for grade \(\ge\) 3 acute toxicity (V30 > 71.8%) [9] and grade \(\ge\) 2 (Dmean \(\ge\) 50 Gy) [38]. Although the HT plan showed lower Dmean, V15, and V30 than both HL2arc and HL4arc, NTCP was slightly higher than that of HL2arc and HL4arc plans. This is because the volume receiving high dose (V45), which contributes more to NTCP calculation, is relatively larger than these plans. As a result, the differences in NTCP values for parotid glands and OC between the three plans were within 3.47% and 1.7% points, respectively. Furthermore, most of the OARs met the dose criteria in all plans. Although there were slight dose differences depending on the specific plan technique, it does not seem to be clinically meaningful.

When evaluating a new VMAT plan technique, the number of arcs used is an important concern because it has a close relationship with plan quality as well as planning time and beam delivery efficiency. Multiple-arc VMAT plans are likely to provide better dosimetric profile than single-arc VMAT plans for complex targets, with increased delivery times and spread of low doses [10]. However, HL4arc showed better dosimetric profiles than HL2arc but no significant MU increase (median total MU: 836.30 (IQR: 733.15, 918.90) (HL2arc) vs. 856.35 (IQR: 750.00, 963.83) (HL4arc)) in our study (p > 0.05). Also, four arc beams were sufficient to meet the dose constraints. Michiels et al. reported similar results. Triple-arc VMAT plan on a Halcyon provides better dosimeric profile compare with double-arc VMAT plan, and beam delivery time was slightly increased, but significantly decreased compared to double-arc VMAT on a Truebeam [22]. Although more research is needed to determine how many arcs are appropriate for NPCa, FOL provides a greater opportunity to use the multiple-arc VMAT technique because of extremely short beam delivery time, because of fast gantry rotation (four vs. one revolution per minute compared with C-arm Linac) [14].

Although we tried to make a fair and quantitative plan comparison, our study has some limitations. The results of this study may reflect not only the differences in the intrinsic performance of the treatment machine and beam delivery technique, but also the differences in dose calculation grid resolution, dose calculation error, and dose optimization capability of the treatment planning system and planning skill of the dosimetrists between the two comparison groups. It seems necessary to consider them when referring to the results of this study.

Conclusions

With the advancement of RT technology, new treatment machines and beam delivery techniques are being introduced steadily to meet clinical demands for the improvement of clinical outcomes and reduction of radiation-induced complications. It is timely and important to have a clear understanding of the capabilities and limitations of a new device prior to patient assignment in clinical practice.

We performed the first dosimetric comparison study between VMAT plans using the FOL with DL-MLC and HT with dynamic jaw for NPCa. Although HT and FOL plans showed different advantages and disadvantages in dosimetric characteristics, overall, they showed comparable dosimetric profiles. Our results, combined with the facility’s availability of treatment machines and resource, may contribute to wise selection of appropriate RT techniques for NPCa.

Availability of data and materials

The datasets analyzed during the current study are available from the corresponding author on reasonable request.

References

  1. Kubicek GL, Machtay M. New advances in high-technology radiotherapy for head and neck cancer. Hematol Oncol Clin N Am. 2008;22:1165.

    Article  Google Scholar 

  2. Bhide SA, Nutting CM. Advances in radiotherapy for head and neck cancer. Oral Oncol. 2010;46:439–41.

    Article  CAS  Google Scholar 

  3. Gregoire V, Langendijk JA, Nuyts S. Advances in radiotherapy for head and neck cancer. J Clin Oncol. 2015;33:3277–84.

    Article  CAS  Google Scholar 

  4. Sun XS, Li XY, Chen QY, Tang LQ, Mai HQ. Future of radiotherapy in nasopharyngeal carcinoma. Br J Radiol. 2019;92:20190209.

    Article  Google Scholar 

  5. Li YC, Chen FP, Zhou GQ, Zhu JH, Hu J, Kang DH, et al. Incidence and dosimetric parameters for brainstem necrosis following intensity modulated radiation therapy in nasopharyngeal carcinoma. Oral Oncol. 2017;73:97–104.

    Article  Google Scholar 

  6. Little M, Schipper M, Feng FY, Vineberg K, Cornwall C, Murdoch-Kinch CA, et al. Reducing xerostomia after chemo-IMRT for head-and-neck cancer: beyond sparing the parotid glands. Int J Radiat Oncol Biol Phys. 2012;83:1007–14.

    Article  Google Scholar 

  7. Deasy JO, Moiseenko V, Marks L, Chao KSC, Nam J, Eisbruch A. Radiotherapy dose-volume effects on salivary gland function. Int J Radiat Oncol Biol Phys. 2010;76:S58–63.

    Article  Google Scholar 

  8. Eisbruch A, Kim HM, Terrell JE, Marsh LH, Dawson LA, Ship JA. Xerostomia and its predictors following parotid-sparing irradiation of head-and-neck cancer. Int J Radiat Oncol Biol Phys. 2001;50:695–704.

    Article  CAS  Google Scholar 

  9. Li K, Yang L, Hu QY, Chen XZ, Chen M, Chen Y. Oral mucosa dose parameters predicting grade ≥ 3 acute toxicity in locally advanced nasopharyngeal carcinoma patients treated with concurrent intensity-modulated radiation therapy and chemotherapy: an independent validation study comparing oral cavity versus mucosal surface contouring techniques. Transl Oncol. 2017;10:752–9.

    Article  Google Scholar 

  10. Guckenberger M, Richter A, Krieger T, Wilbert J, Baier K, Flentje M. Is a single arc sufficient in volumetric-modulated arc therapy (VMAT) for complex-shaped target volumes? Radiother Oncol. 2009;93:259–65.

    Article  Google Scholar 

  11. Stieler F, Wolff D, Schmid H, Welzel G, Wenz F, Lohr F. A comparison of several modulated radiotherapy techniques for head and neck cancer and dosimetric validation of VMAT. Radiother Oncol. 2011;101:388–93.

    Article  Google Scholar 

  12. Lim TY, Dragojevic I, Hoffman D, Flores-Martinez E, Kim GY. Characterization of the Halcyon(TM) multileaf collimator system. J Appl Clin Med Phys. 2019;20:106–14.

    Article  Google Scholar 

  13. Netherton T, Li Y, Gao S, Klopp A, Balter P, Court LE, et al. Experience in commissioning the halcyon linac. Med Phys. 2019;46:4304–13.

    Article  CAS  Google Scholar 

  14. Li T, Scheuermann R, Lin A, Teo BK, Zou W, Swisher-McClure S, et al. Impact of multi-leaf collimator parameters on head and neck plan quality and delivery: a comparison between halcyon and Truebeam(R) treatment delivery systems. Cureus. 2018;10:e3648.

    PubMed  PubMed Central  Google Scholar 

  15. Sarkar V, Lin L, Shi C, Papanikolaou N. Quality assurance of the multileaf collimator with helical tomotherapy: design and implementation. Med Phys. 2007;34:2949–56.

    Article  Google Scholar 

  16. Langen KM, Papanikolaou N, Balog J, Crilly R, Followill D, Goddu SM, et al. QA for helical tomotherapy: report of the AAPM Task Group 148. Med Phys. 2010;37:4817–53.

    Article  Google Scholar 

  17. Rong Y, Chen Y, Shang L, Zuo L, Lu W, Chen Q. Helical tomotherapy with dynamic running-start-stop delivery compared to conventional tomotherapy delivery. Med Phys. 2014;41:051709.

    Article  Google Scholar 

  18. Bibault JE, Dussart S, Pommier P, Morelle M, Huguet M, Boisselier P, et al. Clinical outcomes of several IMRT techniques for patients with head and neck cancer: a propensity score-weighted analysis. Int J Radiat Oncol Biol Phys. 2017;99:929–37.

    Article  Google Scholar 

  19. Murthy V, Master Z, Gupta T, Ghosh-Laskar S, Budrukkar A, Phurailatpam R, et al. Helical tomotherapy for head and neck squamous cell carcinoma: dosimetric comparison with linear accelerator-based step-and-shoot IMRT. J Cancer Res Ther. 2010;6:194–8.

    Article  Google Scholar 

  20. Jacob V, Bayer W, Astner ST, Busch R, Kneschaurek P. A planning comparison of dynamic IMRT for different collimator leaf thicknesses with helical tomotherapy and RapidArc for prostate and head and neck tumors. Strahlenther Onkol. 2010;186:502–10.

    Article  Google Scholar 

  21. Wiezorek T, Brachwitz T, Georg D, Blank E, Fotina I, Habl G, et al. Rotational IMRT techniques compared to fixed gantry IMRT and tomotherapy: multi-institutional planning study for head-and-neck cases. Radiat Oncol. 2011;6:20.

    Article  Google Scholar 

  22. Michiels S, Poels K, Crijns W, Delombaerde L, De Roover R, Vanstraelen B, et al. Volumetric modulated arc therapy of head-and-neck cancer on a fast-rotating O-ring linac: plan quality and delivery time comparison with a C-arm linac. Radiother Oncol. 2018;128:479–84.

    Article  Google Scholar 

  23. Panda S, Swamidas J, Chopra S, Mangaj A, Fogliata A, Kupelian P, et al. Treatment planning comparison of volumetric modulated arc therapy employing a dual-layer stacked multi-leaf collimator and helical tomotherapy for cervix uteri. Radiat Oncol. 2020;15:22.

    Article  CAS  Google Scholar 

  24. Ju SG, Ahn YC, Kim YB, Park SG, Choi YM, Na CH, et al. Development of a tongue immobilization device using a 3D printer for intensity modulated radiation therapy of nasopharyngeal cancer patients. Cancer Res Treat. 2021;53:45–54.

    Article  CAS  Google Scholar 

  25. Brouwer CL, Steenbakkers RJ, Bourhis J, Budach W, Grau C, Gregoire V, et al. CT-based delineation of organs at risk in the head and neck region: DAHANCA, EORTC, GORTEC, HKNPCSG, NCIC CTG, NCRI, NRG Oncology and TROG consensus guidelines. Radiother Oncol. 2015;117:83–90.

    Article  Google Scholar 

  26. Chen M, Chen Y, Chen Q, Lu W. Theoretical analysis of the thread effect in helical TomoTherapy. Med Phys. 2011;38:5945–60.

    Article  Google Scholar 

  27. Chen Y, Chen Q, Chen M, Lu W. Dynamic tomotherapy delivery. Med Phys. 2011;38:3013–24.

    Article  Google Scholar 

  28. Hong CS, Oh D, Ju SG, Ahn YC, Na CH, Kwon DY, et al. Development of a semi-customized tongue displacement device using a 3D printer for head and neck IMRT. Radiat Oncol. 2019;14:79.

    Article  Google Scholar 

  29. Paddick I. A simple scoring ratio to index the conformity of radiosurgical treatment plans. Technical note. J Neurosurg. 2000;93(Suppl 3):219–22.

    Article  Google Scholar 

  30. Timmerman RD, Kavanagh BD, Cho LC, Papiez L, Xing L. Stereotactic body radiation therapy in multiple organ sites. J Clin Oncol. 2007;25:947–52.

    Article  Google Scholar 

  31. Tanyi JA, Summers PA, McCracken CL, Chen Y, Ku LC, Fuss M. Implications of a high-definition multileaf collimator (HD-MLC) on treatment planning techniques for stereotactic body radiation therapy (SBRT): a planning study. Radiat Oncol. 2009;4:22.

    Article  Google Scholar 

  32. Hong CS, Ju SG, Kim M, Kim JI, Kim JM, Suh TS, et al. Dosimetric effects of multileaf collimator leaf width on intensity-modulated radiotherapy for head and neck cancer. Med Phys. 2014;41:021712.

    Article  Google Scholar 

  33. Fogliata A, Reggiori G, Stravato A, Lobefalo F, Franzese C, Franceschini D, et al. RapidPlan head and neck model: the objectives and possible clinical benefit. Radiat Oncol. 2017;12:1–12.

    Article  Google Scholar 

  34. Kallman P, Agren A, Brahme A. Tumour and normal tissue responses to fractionated non-uniform dose delivery. Int J Radiat Biol. 1992;62:249–62.

    Article  CAS  Google Scholar 

  35. Chiavassa S, Bessieres I, Edouard M, Mathot M, Moignier A. Complexity metrics for IMRT and VMAT plans: a review of current literature and applications. Br J Radiol. 2019;92:20190270.

    Article  Google Scholar 

  36. Schwartz DL, Hutcheson K, Barringer D, Tucker SL, Kies M, Holsinger FC, et al. Candidate dosimetric predictors of long-term swallowing dysfunction after oropharyngeal intensity-modulated radiotherapy. Int J Radiat Oncol Biol Phys. 2010;78:1356–65.

    Article  Google Scholar 

  37. Sanguineti G, Gunn GB, Parker BC, Endres EJ, Zeng J, Fiorino C. Weekly dose-volume parameters of mucosa and constrictor muscles predict the use of percutaneous endoscopic gastrostomy during exclusive intensity-modulated radiotherapy for oropharyngeal cancer. Int J Radiat Oncol Biol Phys. 2011;79:52–9.

    Article  Google Scholar 

  38. Mazzola R, Ricchetti F, Fersino S, Fiorentino A, Giaj Levra N, Di Paola G, et al. Predictors of mucositis in oropharyngeal and oral cavity cancer in patients treated with volumetric modulated radiation treatment: a dose-volume analysis. Head Neck. 2016;38:E815–9.

    Article  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

This work was supported by the research program, NRF-2021R1F1A1047378 (Republic of Korea).

Author information

Authors and Affiliations

Authors

Contributions

Conception, design, and drafting the manuscript were performed by SGJ, and YCA. Data collection and interpreting were performed by YK, JMK, DYK, BSP, KY. All authors read and approved the manuscript.

Corresponding author

Correspondence to Yong Chan Ahn.

Ethics declarations

Ethics approval and consent to participate

The institutional review board at Samsung Medical Center (file no. SMC 2020-08-120-001, August 28, 2021).

Consent for publication

Not applicable.

Competing interests

Conflict of interest relevant to this article was not reported.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1.

The parameters used for NTCP evaluation.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Ju, S.G., Ahn, Y.C., Kim, Yb. et al. Dosimetric comparison between VMAT plans using the fast-rotating O-ring linac with dual-layer stacked MLC and helical tomotherapy for nasopharyngeal carcinoma. Radiat Oncol 17, 155 (2022). https://doi.org/10.1186/s13014-022-02124-0