Evaluation of the tumor movement and the reproducibility of two different immobilization setups for image-guided stereotactic body radiotherapy of liver tumors

Abstract

Background

The purpose of this study is to evaluate the tumor movement and accuracy of patient immobilization in stereotactic body radiotherapy of liver tumors with low pressure foil or abdominal compression.

Methods

Fifty-four liver tumors treated with stereotactic body radiotherapy were included in this study. Forty patients were immobilized by a vacuum couch with low pressure foil, 14 patients by abdominal compression. We evaluated the ratio of gross tumor volume/internal target volume, the tumor movement in 4D-computed tomography scans and daily online adjustments after cone beam computed tomography scans.

Results

The ratio of gross tumor volume/internal target volume was smaller with low pressure foil. The tumor movement in 4D-computed tomography scans was smaller with abdominal compression, the cranial movement even significantly different (p = 0.02). The mean online adjustments and their mean absolute values in the vertical, lateral and longitudinal axis were smaller with abdominal compression. The online adjustments were significantly different (p < 0.013), their absolute values in case of the longitudinal axis (p = 0.043). There was no significant difference of the adjustments’ 3D vectors.

Conclusions

In comparison to low pressure foil, abdominal compression leads to a reduction of the tumor movement. Online adjustments decreased significantly, thus leading to higher accuracy in patient positioning.

Background

Stereotactic body radiotherapy (SBRT) is able both to deliver a high radiation dose to the tumor, and to preserve sufficient liver function by little dose exposure to the surrounding liver tissue. SBRT is a therapeutic option for patients with primary and secondary liver tumors [1,2,3,4,5].

The liver in general is believed to have a relatively low radiation tolerance and this is a major limitation for the use of SBRT or conventionally fractionated external beam radiotherapy [6, 7]. On top of that, most patients with primary liver cancer have hepatitis B, C or alcoholic cirrhosis, resulting in an impaired liver function and even lower radiation tolerance already prior to radiotherapy. Radiation to the liver in general may result in radiotherapy-induced liver disease (RILD), possibly leading to liver failure and death and especially in case of reirradiation [8,9,10,11]. In order to guard against these events, high accuracy in patient repositioning and reduction of tumor movement due to breathing are necessary. This is of utmost importance especially in case of SBRT with high dose exposure to the liver. Due to intracorporal tumor and liver movement and difficult patient immobilization, SBRT of liver tumors is still a challenge. Intrafractional and interfractional robustness have to be improved as far as possible. Therefore, advantages and disadvantages of the following ways of patient immobilization setups have to be reevaluated: By the use of breath hold, radiotherapy mainly relies on the patients’ individual ability to control their way of breathing and can be regarded as a special way of gating. Gating in general ensures dose application only during the planned breathing cycles. There is only a small number of institutions, which can also offer tracking the target volume e.g. by the use of fiducials. Nevertheless, the implementation of fiducials is attended by a risk of complications, which is why in our institution abdominal compression is a well-established way of intracorporal immobilization. Our institution used low pressure foil until 2014 and changed patient immobilization in radiotherapy of liver tumors to abdominal compression afterwards. Both setups are used in clinical routine, but the true impact on robustness and treatment planning has not yet been evaluated. So, the present study aims to elucidate the impact of patient immobilization with abdominal compression, with regard to reducing tumor movement and increasing the accuracy in patient positioning.

Methods

Patient characteristics

Fifty-four patients with liver tumors, treated with SBRT at our institution between 2010 and 2016 were included in this analysis. Twenty-three and Thirty-one patients had primary and secondary liver cancer. Only patients with one lesion being irradiated were included. Fiducial markers were not implanted.

Stereotactic body radiotherapy

Forty patients were immobilized using a vacuum couch with low pressure foil (Medical Intelligence Medizintechnik GmbH, Schwabmünchen, Germany), 14 patients by the use of abdominal compression (Fa. ITV, Innsbruck, Österreich) (Fig. 1). The abdominal compression was performed by a rectangle plate with a triangle peak, being reproducibly placed directly under the xiphoid bone.

Fig. 1
figure 1

Immobilization setups. a Immobilization with a vacuum couch with low pressure foil. b Immobilization with abdominal compression

Computed tomography (CT) scans for treatment planning were performed with no breath-hold by a Somatom Emotion computed tomography scanner (Siemens medical solutions, Erlangen, Germany). The thickness of one treatment planning CT slice was 3 mm. Furthermore, a 4D-CT scan was carried out with the same CT scanner to evaluate the maximum extension of the tumor movement by covering the whole breathing cycle (measured by respiratory position management system (RPM, Varian Medical Systems, Palo Alto, CA)). One CT slice of the 4D-CT had the thickness of 2.1 mm. Treatment planning itself was carried out by use of the treatment planning system (TPS) Eclipse™ (Varian Medical Systems, Palo Alto, CA).

The gross tumor volume (GTV) was defined by the visible tumor on an arterial phase contrast enhanced treatment planning CT scan for primary liver cancer and on a venous phase contrast enhanced treatment planning CT scan for secondary liver cancer. The internal target volume (ITV) was defined by the maximum extension of the GTV movement in every direction in the 4D-CT scan. The planning target volume (PTV) was defined by addition of 5 mm margin in all directions.

Twenty-six patients were treated by Volumetric modulated arc therapy (VMAT) with 1 to 7 arcs. Twenty-eight patients were treated by 3D conformal radiotherapy with 6 to 11 coplanar beams. The patients’ dose prescription is summarized in Table 1.

Table 1 Dose prescription for 54 patients with liver tumors treated with Stereotactic Body Radiotherapy

Two patients receiving two to three fractions were scheduled to receive five fractions (5 × 7Gy = 35Gy), but the treatment was interrupted due to clinical complications. Generally, at least 700ccm of the liver should receive a radiation dose below 15Gy and the mean dose to the liver was kept as low as possible.

Treatment was performed on a Varian Clinac Trilogy linear accelerator (Varian Medical Systems, Palo Alto, CA).

Prior to each fraction image guidance was performed by a free breath cone beam computed tomography (CBCT) control scan of the liver in the immobilization setup as described before. On the basis of the manual and automatic registration of the treatment planning CT and CBCT scan the online adjustments are conducted respectively. Adjustments according to daily CBCT control scans were based on the total liver tissue together with the vertebral bones.

Data analysis

The treatment planning volumes GTV and ITV were retrieved from the TPS Eclipse™ (Varian Medical Systems, Palo Alto, CA) and were compared for both patient groups, in order to analyze the immobilization impact on the general intracorporal movement of the liver. The mean ratio GTV/ITV was calculated to indicate the relative movement of the GTV in the ITV. Aiming at specifying the impact of the immobilization, the tumor movement in the 4D-CT scan has been analyzed. The movement has been defined as the distance between ITV and GTV on the height of the center of the GTV in the 6 directions “right lateral, left lateral, anterior, posterior, cranial, and caudal”.

Adjustments according to CBCT scan prior to each fraction of radiotherapy were analyzed, in order to evaluate the accuracy and reproducibility of immobilization: online adjustments, the absolute values of online adjustments and the summarizing 3D vector of online adjustments were analyzed. Vector calculation for each online adjustment was done by the following Eq. 1.

Equation 1. Vector equation with \( \overrightarrow{\mathrm{v}} \) representing the vector and x, y and z representing its online adjustments (in the vertical, lateral and longitudinal axis), according to the cone beam computed tomography scan [cm].

$$ \overrightarrow{\mathrm{v}}=\sqrt{{\mathrm{x}}^2+{\mathrm{y}}^2+{\mathrm{z}}^2} $$

The non-parametric Mann-Whitney test was used for differences between two independent groups. Significance was defined by a level of the p-value below 0.05. The software SPSS (Statistical package for the Social Sciences) version 23 (SPSS Inc., Chicago, USA) was used for statistical analysis.

Results

The GTV in the group with low pressure foil and with abdominal compression has a mean volume of 73.9ccm and 37.6ccm. The ITV in both groups has a mean volume of 94.1ccm and 67.9ccm. The mean ratio of GTV/ITV is 0.54 and 0.61 in case of low pressure foil and abdominal compression. There was no statistically significant difference between both groups (p > 0.05). The mean tumor movements at the height of the center of the GTV during 4D-CT are listed in Table 2. Only the cranial movement showed a significant difference between both groups (p = 0.02).

Table 2 Mean tumor movement [cm] at the height of the center of the gross tumor volume during 4D-computed tomography scan

The online adjustments were evaluated in general and with regard to its absolute values, thus leading to a comparison of the true amount of movements. The mean online adjustments in the vertical, lateral and longitudinal axis account for − 0.11 cm (±0.40 cm), 0.16 cm (±0.95 cm) and − 0.28 cm (±0.87 cm) with low pressure foil and 0.09 cm (±0.43 cm), 0.16 cm (±0.29 cm) and 0.08 cm (±0.44 cm) with abdominal compression (Figs. 2 and 3).

Fig. 2
figure 2

Online adjustments after cone beam computed tomography scans with low pressure foil [cm]

Fig. 3
figure 3

Online adjustments after cone beam computed tomography scans with abdominal compression [cm]

The absolute mean value of the online adjustments in the vertical, lateral and longitudinal axis account for 0.29 cm (±0.50 cm), 0.48 cm (±1.50 cm) and 0.57 cm (±0.89 cm) with low pressure foil and 0.28 cm (±0.28 cm), 0.29 cm (±0.43 cm) and 0.33 cm (±0.32 cm) with abdominal compression, respectively. The vertical, lateral and longitudinal online adjustments were statistically significant different between both groups (p < 0.013). The absolute values of the online adjustments were statistically different in case of the longitudinal axis (p = 0.043).

Summarizing the adjustments in vectors results in a mean vector of 0.99 cm (±1.59 cm) with low pressure foil and 0.72 cm (±0.46 cm) with abdominal compression. There was no statistically significant difference of the vectors neither between the two groups, nor between the vectors of two fractions within one group.

Discussion

Robust treatment planning is based on contouring an ITV, which takes into account the maximum extent of tumor motion. Liver motion is due to the breathing amplitude, which should be minimized by an appropriate patient immobilization. There are a lot of possible ways of patient immobilization in SBRT, but low pressure foil and abdominal compression are highly promising, easy to use, with little interfractional displacements and characterized by good reproducibility [12,13,14,15]. In our results, there was not any significant difference in case of the GTV/ITV ratio in both groups. This may be explained on the one hand by the small collective of this study, the heterogeneous body mass index and on the other hand by the tumor movement itself. On top of that, this is a retrospective study and the group with abdominal compression was smaller than the group with low pressure foil. Another explanation might be the big difference of the gross tumor volumes between both groups. With the GTV being bigger in the group with low pressure foil, the ratio of GTV/ITV might possibly be even smaller in reality (with similar volumes in both groups), thus leading to a significant difference between both groups. This assumption is most likely true as the real tumor movement in the group of abdominal compression compared to the group of low pressure foil was smaller in every direction. We could even show a significant difference in case of cranial movement.

However, the extent of online corrections showed a statistically significant difference between both groups. Abdominal compression leads to a shift in overall liver motion which is especially due to a decreased motion in the longitudinal axis and a compensating motion away from the abdominal compression axis. The fact that a change in patient immobilization mainly leads to a displacement shift with one dominant direction is represented in our evaluation of the 4D-CTs and could also be shown in a similar study about SBRT of lung tumors [16]. Nevertheless, the significant change in one main direction is probably compensated by the other directions and leads to a compensation in case of an ITV-generation. This can also explain the significant differences in each direction of online adjustments, whereas the mean values of online adjustments are only statistically significantly different in the longitudinal axis but not in the other dimensions. Online adjustments, that is to say the precision of patient positioning can be improved by the usage of abdominal compression but not in every setting. The values with but not without abdominal compression match with the results from a study from Herfarth et al., showing that this is a feasible way of patient immobilization, providing high accuracy [17]. Together with the studies from Herfarth et al. and Navarro-Martin et al. the impact of immobilization on interfractional robustness and setup reproducibility in general could be shown [16, 17]. Nevertheless, these studies about immobilization in SBRT are based on rigid immobilization setups. Promising results about frameless SBRT could be shown for lung cancer [18,19,20]. In case of liver cancer, rigid immobilization setups, such as abdominal compression, are still the most established safety devices to ensure robust and accurate SBRT. This gain in accuracy is mainly dependent on adjustments according to CBCT. But, online matching by the use of CBCT has to be critically evaluated, as matching is based on the analysis of the liver’s soft tissue. CBCT does not allow safe identification of the target volume. Online adjustments are therefore based on the total liver volume but not the target volume. But Eccles et al. could show that the interfractional change or variability of the liver volume is small and therefore reliable [21]. However, fiducial markers such as lipiodol or gold fiducials are highly promising and effective in case of image-guided radiotherapy [22, 23]. Nevertheless, these markers can only be used by invasive interventions prior to radiotherapy, coming along with distinct complications, which is why other principles of guiding are the topic of present studies [24].

With regard to robust treatment planning, breathing is summarized in the treatment planning CT scan by both inspiration and expiration – no breath-hold was performed during CT scanning. The robustness of this CT scan is ensured by an ITV increasing the accuracy of treatment application. With the CBCT scan again being performed with both breathing cycles, the registration of the CBCT on the treatment planning CT scan can be done. The robustness of this kind of registration could be increased by gating, that is to say irradiating only during expiration or inspiration, thus reducing the impact of tumor motion [25, 26].

Tracking, that is to say following the tumor motion is another method of robust irradiation with highly precise dose application in the tumor volume. Unfortunately, there is still little knowledge on the best conditions of tracking, which is why future studies have to show its feasibility [27].

Conclusion

In summary, clinical practice in radiotherapy of liver tumors is still mainly based on SBRT with distinct methods of patient positioning and reducing tumor motion. Generation of an ITV and patient immobilization in our study could show its feasibility and the potential advantage of abdominal compression in contrast to immobilization by low pressure foil. By increasing robustness the general underuse of SBRT in the therapy of primary and secondary liver cancer can possibly be overcome and SBRT will not only be used in case of a big tumor volume or mainly venous blood supply with only little arterial vessels supplying the tumor [28, 29]. In these cases, SBRT is beneficial and promising, both in palliative therapy and in case of bridging to liver transplantation. But by increasing robustness and accuracy of radiotherapy, the probability of limiting complications, such as RILD and a general decrease in liver function, will further be reduced and the indications of SBRT will close up to those of selective internal radiation therapy, radiofrequency ablation and transarterial chemoembolization [30].

Abbreviations

CBCT:

Cone beam computed tomography

CT:

Computed tomography

GTV:

Gross tumor volume

ITV:

Internal target volume

RILD:

Radiatiotherapy-induced liver disease

RPM:

Respiratory position management

SBRT:

Stereotactic body radiotherapy

SPSS:

Statistical package for the social sciences

TPS:

Treatment planning system

VMAT:

Volumetric modulated arc therapy

References

  1. Bujold A, Massey CA, Kim JJ, Brierley J, Cho C, Wong RK, Dinniwell RE, Kassam Z, Ringash J, Cummings B, et al. Sequential phase I and II trials of stereotactic body radiotherapy for locally advanced hepatocellular carcinoma. J Clin Oncol. 2013;31:1631–9.

    Article  PubMed  Google Scholar 

  2. Tse RV, Hawkins M, Lockwood G, Kim JJ, Cummings B, Knox J, Sherman M, Dawson LA. Phase I study of individualized stereotactic body radiotherapy for hepatocellular carcinoma and intrahepatic cholangiocarcinoma. J Clin Oncol. 2008;26:657–64.

    Article  PubMed  Google Scholar 

  3. Habermehl D, Herfarth KK, Bermejo JL, Hof H, Rieken S, Kuhn S, Welzel T, Debus J, Combs SE. Single-dose radiosurgical treatment for hepatic metastases - therapeutic outcome of 138 treated lesions from a single institution. Radiat Oncol. 2013;8:175.

    Article  PubMed  PubMed Central  Google Scholar 

  4. Mendez Romero A, Hoyer M. Radiation therapy for liver metastases. Curr Opin Support Palliat Care. 2012;6:97–102.

    Article  PubMed  Google Scholar 

  5. Andratschke N, Parys A, Stadtfeld S, Wurster S, Huttenlocher S, Imhoff D, Yildirim M, Rades D, Rodel CM, Dunst J, et al. Clinical results of mean GTV dose optimized robotic guided SBRT for liver metastases. Radiat Oncol. 2016;11:74.

    Article  PubMed  PubMed Central  Google Scholar 

  6. Pan CC, Kavanagh BD, Dawson LA, Li XA, Das SK, Miften M, Ten Haken RK. Radiation-associated liver injury. Int J Radiat Oncol Biol Phys. 2010;76:S94–100.

    Article  PubMed  PubMed Central  Google Scholar 

  7. Herrmann E, Naehrig D, Sassowsky M, Bigler M, Buijsen J, Ciernik I, Zwahlen D, Pellanda AF, Meister A, Brauchli P, et al. External beam radiotherapy for unresectable hepatocellular carcinoma, an international multicenter phase I trial, SAKK 77/07 and SASL 26. Radiat Oncol. 2017;12:12.

    Article  PubMed  PubMed Central  Google Scholar 

  8. Dawson LA, Ten Haken RK. Partial volume tolerance of the liver to radiation. Semin Radiat Oncol. 2005;15:279–83.

    Article  PubMed  Google Scholar 

  9. Jung J, Yoon SM, Kim SY, Cho B, Park JH, Kim SS, Song SY, Lee SW, Ahn SD, Choi EK, Kim JH. Radiation-induced liver disease after stereotactic body radiotherapy for small hepatocellular carcinoma: clinical and dose-volumetric parameters. Radiat Oncol. 2013;8:249.

    Article  PubMed  PubMed Central  Google Scholar 

  10. Cheng JC, Wu JK, Huang CM, Liu HS, Huang DY, Cheng SH, Tsai SY, Jian JJ, Lin YM, Cheng TI, et al. Radiation-induced liver disease after three-dimensional conformal radiotherapy for patients with hepatocellular carcinoma: dosimetric analysis and implication. Int J Radiat Oncol Biol Phys. 2002;54:156–62.

    Article  PubMed  Google Scholar 

  11. Huang Y, Chen SW, Fan CC, Ting LL, Kuo CC, Chiou JF. Clinical parameters for predicting radiation-induced liver disease after intrahepatic reirradiation for hepatocellular carcinoma. Radiat Oncol. 2016;11:89.

    Article  PubMed  PubMed Central  Google Scholar 

  12. Sterzing F, Brunner TB, Ernst I, Baus WW, Greve B, Herfarth K, Guckenberger M. Stereotactic body radiotherapy for liver tumors: principles and practical guidelines of the DEGRO working group on stereotactic radiotherapy. Strahlenther Onkol. 2014;190:872–81.

    Article  PubMed  Google Scholar 

  13. Sio TT, Jensen AR, Miller RC, de los Santos LEF, Hallemeier CL, Foster NR, Park SS, Bauer HJ, Chang K, Garces YI, Olivier KR. Influence of patient’s physiologic factors and immobilization choice with stereotactic body radiotherapy for upper lung tumors. J Appl Clin Med Phys. 2014;15:4931.

    PubMed  Google Scholar 

  14. Foster R, Meyer J, Iyengar P, Pistenmaa D, Timmerman R, Choy H, Solberg T. Localization accuracy and immobilization effectiveness of a stereotactic body frame for a variety of treatment sites. Int J Radiat Oncol Biol Phys. 2013;87:911–6.

    Article  PubMed  Google Scholar 

  15. Han K, Cheung P, Basran PS, Poon I, Yeung L, Lochray F. A comparison of two immobilization systems for stereotactic body radiation therapy of lung tumors. Radiother Oncol. 2010;95:103–8.

    Article  PubMed  Google Scholar 

  16. Navarro-Martin A, Cacicedo J, Leaman O, Sancho I, Garcia E, Navarro V, Guedea F. Comparative analysis of thermoplastic masks versus vacuum cushions in stereotactic body radiotherapy. Radiat Oncol. 2015;10:176.

    Article  PubMed  PubMed Central  Google Scholar 

  17. Herfarth KK, Debus J, Lohr F, Bahner ML, Fritz P, Hoss A, Schlegel W, Wannenmacher MF. Extracranial stereotactic radiation therapy: set-up accuracy of patients treated for liver metastases. Int J Radiat Oncol Biol Phys. 2000;46:329–35.

    Article  CAS  PubMed  Google Scholar 

  18. Alderliesten T, Sonke JJ, Betgen A, van Vliet-Vroegindeweij C, Remeijer P. 3D surface imaging for monitoring intrafraction motion in frameless stereotactic body radiotherapy of lung cancer. Radiother Oncol. 2012;105:155–60.

    Article  PubMed  Google Scholar 

  19. Josipovic M, Persson GF, Logadottir A, Smulders B, Westmann G, Bangsgaard JP. Translational and rotational intra- and inter-fractional errors in patient and target position during a short course of frameless stereotactic body radiotherapy. Acta Oncol. 2012;51:610–7.

    Article  PubMed  Google Scholar 

  20. Peguret N, Dahele M, Cuijpers JP, Slotman BJ, Verbakel WF. Frameless high dose rate stereotactic lung radiotherapy: intrafraction tumor position and delivery time. Radiother Oncol. 2013;107:419–22.

    Article  PubMed  Google Scholar 

  21. Eccles CL, Dawson LA, Moseley JL, Brock KK. Interfraction liver shape variability and impact on GTV position during liver stereotactic radiotherapy using abdominal compression. Int J Radiat Oncol Biol Phys. 2011;80:938–46.

    Article  PubMed  Google Scholar 

  22. Habermehl D, Henkner K, Ecker S, Jakel O, Debus J, Combs SE. Evaluation of different fiducial markers for image-guided radiotherapy and particle therapy. J Radiat Res. 2013;54(Suppl 1):i61–8.

    Article  PubMed  PubMed Central  Google Scholar 

  23. Habermehl D, Naumann P, Bendl R, Oelfke U, Nill S, Debus J, Combs SE. Evaluation of inter- and intrafractional motion of liver tumors using interstitial markers and implantable electromagnetic radiotransmitters in the context of image-guided radiotherapy (IGRT) - the ESMERALDA trial. Radiat Oncol. 2015;10:143.

    Article  PubMed  PubMed Central  Google Scholar 

  24. Heinz C, Gerum S, Freislederer P, Ganswindt U, Roeder F, Corradini S, Belka C, Niyazi M. Feasibility study on image guided patient positioning for stereotactic body radiation therapy of liver malignancies guided by liver motion. Radiat Oncol. 2016;11:88.

    Article  PubMed  PubMed Central  Google Scholar 

  25. Poulsen PR, Worm ES, Hansen R, Larsen LP, Grau C, Hoyer M. Respiratory gating based on internal electromagnetic motion monitoring during stereotactic liver radiation therapy: first results. Acta Oncol. 2015;54:1445–52.

    Article  PubMed  Google Scholar 

  26. Velec M, Moseley JL, Dawson LA, Brock KK. Dose escalated liver stereotactic body radiation therapy at the mean respiratory position. Int J Radiat Oncol Biol Phys. 2014;89:1121–8.

    Article  PubMed  PubMed Central  Google Scholar 

  27. Wojcieszynski AP, Rosenberg SA, Brower JV, Hullett CR, Geurts MW, Labby ZE, Hill PM, Bayliss RA, Paliwal B, Bayouth JE, et al. Gadoxetate for direct tumor therapy and tracking with real-time MRI-guided stereotactic body radiation therapy of the liver. Radiother Oncol. 2016;118:416–8.

    Article  CAS  PubMed  Google Scholar 

  28. Mendez Romero A, de Man RA. Stereotactic body radiation therapy for primary and metastatic liver tumors: from technological evolution to improved patient care. Best Pract Res Clin Gastroenterol. 2016;30:603–16.

    Article  PubMed  Google Scholar 

  29. Crane CH, Koay EJ. Solutions that enable ablative radiotherapy for large liver tumors: fractionated dose painting, simultaneous integrated protection, motion management, and computed tomography image guidance. Cancer. 2016;122:1974–86.

    Article  PubMed  PubMed Central  Google Scholar 

  30. Dreher C, Hoyer KI, Fode MM, Habermehl D, Combs SE, Hoyer M. Metabolic liver function after stereotactic body radiation therapy for hepatocellular carcinoma. Acta Oncol. 2016;55:886–91.

    Article  CAS  PubMed  Google Scholar 

Download references

Acknowledgements

None to declare.

Funding

None to declare.

Availability of data and materials

The datasets analysed during the current study are included as a supporting file.

Author information

Authors and Affiliations

Authors

Contributions

CD and DH designed the concept, collected the relevant parameters, performed the analysis, wrote and drafted the manuscript. MO helped designing the concept and contributed to the data analysis and helped drafting the manuscript. MM, SB and MD participated in the design of the study and parameter collection. SEC designed the concept, participated in the critical analysis and critically revised the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Daniel Habermehl.

Ethics declarations

Ethics approval and consent to participate

The medical ethics commission of the medical faculty (Technical University München) has approved the presented study (257/16S).

Consent for publication

The data was anonymized. The analysis was performed retrospectively. No official consent for publication is available.

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Dreher, C., Oechsner, M., Mayinger, M. et al. Evaluation of the tumor movement and the reproducibility of two different immobilization setups for image-guided stereotactic body radiotherapy of liver tumors. Radiat Oncol 13, 15 (2018). https://doi.org/10.1186/s13014-018-0962-9

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13014-018-0962-9

Keywords