Management of organ motion in scanned ion beam therapy

Abstract

Scanned ion beam therapy has special demands for treatment of intra-fractionally moving tumors such as lesions in lung or liver. Interplay effects between beam and organ motion can in those settings lead to under-dosage of the target volume. Dedicated treatment techniques such as gating or abdominal compression are required. In addition 4D treatment planning should be used to determine strategies for patient specific treatment planning such as an increased beam focus or the use of internal target volumes incorporating range changes.

Several work packages of the Clinical Research Units 214 and 214/2 funded by the German Research Council investigated the management of organ motion in scanned ion beam therapy. A focus was laid on 4D treatment planning using TRiP4D and the development of motion mitigation strategies including their quality assurance. This review focuses on the activity in the second funding period covering adaptive treatment planning strategies, 4D treatment plan optimization, and the application of motion management in pre-clinical research on radiation therapy of cardiac arrhythmias.

Background

Up to now one great challenge for scanned particle beam therapy is the management of intra-fractional organ motion since interplay effects between scanned beam and the moving organ can lead to deviations in the dose deposited to the clinical target volume (CTV) [1, 2]. The focus of several work packages of the Clinical Research Units (KFO) 214 and 214/2 funded by the German Research Council (DFG) was therefore related to methods and devices aiming at a safe and reliable treatment of moving organs with a scanned beam. In the following the main findings of the second funding period are summarized focusing on 4D treatment planning with an emphasis on 4D treatment plan optimization, its application for determining parameters in adaptive ion beam therapy and for pre-clinical research addressing cardiac arrhythmias. This review is by definition very biased towards the work of the funded groups. More general reviews on that topic can be found in the literature [3,4,5,6,7].

4D treatment planning

Within the pilot project of the GSI Helmholtz Centre for Heavy Ion Research in Darmstadt, Germany (GSI) on carbon beam therapy all patients’ treatment was planned with “Treatment planning for particles” (TRiP98) as initially reported by Krämer et al. [8, 9]. Starting in 2002, TRiP98 was extended by the time domain to address 4D treatment planning (TRiP4D) [10,11,12,13,14,15,16]. The software has interfaces to the GSI therapy control system (TCS) for 4D optimized treatment plans (see next section) and accepts treatment delivery log files from the beam application systems of GSI and the Heidelberg Ion-Beam Therapy Center (HIT) for calculation of the 4D delivered dose. TRiP4D is not certified for clinical use and thus remains a research tool. Apart from the applications in research studies (see below on animal experiments addressing cardiac arrhythmia) TRiP4D was used to explore treatment parameters for patients with hepato cellular carcinoma (HCC) which are treated at HIT according to the protocol of the PROMETHEUS-01 trial [17].

Treatment parameters and 4D dose reconstruction for HCC therapy

Richter et al. determined by a series of 4D dose calculations using the data of 8 HCC patients which parameters reduce best the residual interplay in treatments based on gating or abdominal compression [18]. They found that an increased pencil beam size has the biggest effect and that similar motion amplitudes yield comparable V95 target coverage for patients treated under abdominal compression. Currently, HIT uses pencil beams with a FWHM of 10 mm at 2 mm raster spacing to treat those tumors. For the first patients with HCC treated at HIT with a scanned ion beam, Richter et al. reconstructed the daily dose delivery using the beam delivery sequence and log files from the ANZAI system used for motion detection [19]. In addition to delays in the gating sensors [20] the technical implementation resulted into a temporal uncertainty between the log files of beam delivery sequence and motion detection. The dose reconstruction analysis found that 25 ms temporal uncertainty can be permitted for an acceptable accuracy in the reconstructed V95 of the clinical target volume (CTV). HIT therefore meanwhile implemented an improved data acquisition system. Despite these limitations the feasibility of daily 4D dose reconstruction could be shown but is not yet part of daily clinical workflows. A general limitation is the choice of the CT data underlying all calculations incl. deformation maps connecting the phases of the 4DCT. Typically these data are not acquired at time of treatment such that inter-fractional changes influence the accuracy of the calculations. Dose calculations based on CBCT recently reported by other groups are possible solution to overcome that issue [21, 22].

Modelling of 4DCT data

One option to overcome those limitations is modelling of the daily 4DCT. The extension of such a model and its test against clinical data was reported by Wölfelschneider et al. for lung cancer [23]. The model uses the treatment planning 4DCT to calculate vector fields from the mid-ventilation image to each motion phase of the 4DCTs. In order to generate a daily 4DCT, imaging data from patient positioning such as a CBCT are used to determine the baseline shift of the tumor and motion surrogates such as the contour of the patient’s skin (which could e.g. determined by surface imaging) are used to determine the varying phase and amplitude [24] of the intra-fractional motion. The modelled 4DCTs were checked against regular follow-up 4DCTs from clinical data. Differences of (61.3 ± 16.7) HU were determined. In addition, approximately 400 scale invariant features were extracted from the modelled as well as from the ground-truth 4DCT. Median distances of (2.9 ± 3.0) mm were determined between those landmarks. The authors concluded that the modelled 4DCTs can be used for 4D dose calculations in photon therapy. For the same model Fassi et al. determined changes in water equivalent path-length of <1.9 mm [25].

4D optimized treatment planning

4D optimized treatment planning uses time as an additional degree of freedom [26, 27]. It can serve as a motion mitigation technique itself or be used in combination with other techniques such as gating, rescanning, or beam tracking to improve the treatment delivery. In those cases often a dedicated treatment control system is necessary such that organ motion, i.e. the breathing phases determined by a motion monitoring system, is incorporated into the delivery process. A review on the methodologies was reported by C. Graeff [5].

To a certain extent, time was already included in the early reports on tracking with a scanned ion beam since a look-up table (LUT) was required to provide the motion compensation parameters for all combinations of raster points/beam positions and motion phases [12, 28]. In that approach target motion was handled in the treatment plan generation process but not directly in the optimization. That requires summation over the different motion phases in the cost function which in general can be described as

$$ E\left(\overrightarrow{N}\right)=\sum \limits_{k=1}^m\sum \limits_{i=1}^v{\left[{D}_{pre}^i-{D}_{act}^{ik}\left({\overrightarrow{N}}_k\right)\right]}^2=\sum \limits_{k=1}^m\sum \limits_{i=1}^v{\left[{D}_{pre}^i-\sum \limits_{j=1}^r{c}_{ijk}{N}_{jk}\right]}^2 $$

with D pre the prescribed dose and D act the actual dose during the optimization process for voxel i, N the particle number delivered to beam spot j during motion phase k. The difference to 3D optimization is a number of motion phases m > 1 allowing that, e.g. an organ at risk (OAR) with motion related changes in distance to the planning target volume (PTV) will get favorable scores in the distant motion phases once OAR burden is incorporated in the cost function by additional terms.

Eley et al. implemented that approach for beam tracking [29]. The approach was assessed in simulations for simple geometries and lung cancer cases considering absorbed dose, only. In case of phantoms the dose to an OAR could be reduced by 53%, for patient data sets by 13%, each relative to 3D beam tracking. Target coverage was similar for the phantom and significantly improved for the patient. A full feasibility check also requires proof for deliverability. Eley et al. thus also upgraded the GSI TCS such that tracking was delivered in combination with gating, i.e. individual motion phases of the 4D treatment plan were delivered sequentially. The experiment used phantoms with a simple target/OAR setup showed the expected reduction in dose to the OAR (mean optical density of the film in the OAR reduced from 0.71 to 0.26 going from 3D to 4D tracking) but required very long irradiation times.

A similar strategy was addressed by Graeff et al. [16] but extended to RBE-effective dose and applying a strategy that resulted in faster delivery. The main idea was sector-wise division of the target volume into multiple 3D treatment plans all on a single grid of beam positions. Each of the plans corresponded to one motion state and all of them were optimized in 4D in parallel. Also Graeff et al. required a dedicated 4D TCS with gating functionality but the delivery process was more efficient than described before: during the application the beam scans along the single grid underlying all 3D treatment plans corresponding to the individual motion phases. A motion monitoring signal determines the motion phase and the particle numbers of the corresponding 3D treatment plan are delivered. As long as all individual 3D treatment plans contain beam positions in a delivered iso-energy slice (IES), delivery is continuous. Once certain 3D plans applied all positions of an IES the beam is gated in the corresponding motion phase. The approach was successfully implemented as shown in irradiations of radiographic films. Simulations using data of 9 lung cancer patients resulted in target coverages slightly lower than the stationary optimizations (median V95 of 97.9% and 99.3% for 4D–optimized and stationary, respectively).

The reported studies on 4D optimized treatment planning are at a very early stage, i.e. based on simulations and experimental studies involving phantoms. For clinical application further developments especially with respect to (commercialized) 4D treatment planning systems and 4D treatment control systems are essential.

Adaptive ion beam therapy

Adaptive treatment planning aims at improved target conformation by changing the treatment plan on a daily level such that the dose delivery matches best to the varying anatomy of the patient. Clinical examples include treatment of bladder cancer [30], prostate cancer [31], or lung lesions [32]. In several reports, a reduction in normal tissue dose at comparable target coverage is reported.

One reason that adaptive treatment schemes are currently not widely adopted in ion beam therapy could be the impracticability of frequent changes of patient specific hardware such as compensator or collimator in therapy centers using passive scattering. More recently established centers typically rely on beam scanning which does not bring such hardware limitations but currently neither use adaption of treatment plans on a widespread level. In the few reports on treatment plan adaption online adaptation is typically done either using a library of treatment plans optimized e.g. based on multiple (CB)CT data of the first treatment fractions (plan-of-the-day approach) [33, 34] or by swift re-optimization of the treatment plan based on daily imaging data [35, 36].

For prostate cancer treatments using carbon beam therapy Hild et al. studied in a treatment plan comparison three different treatment approaches, namely conventional therapy using a single plan and geometrically defined margins, an offline approach using a varying number of CT datasets to form an internal target volume (ITV), and an online approach including daily re-optimization of the treatment plan [37]. They determined that adequate CTV coverage can be assured with all studied concepts if the prostate motion is below 4 mm. For larger motion, only the online approach resulted in a V95 of the CTV >95%. The offline approach with 4 datasets and especially the online approach with its reduced margins resulted in parallel in a significant reduction of the dose deposited in bladder and rectum. As also authors from other studies [38, 39] reporting online ART with treatment plan re-optimization Hild et al. stress the increased daily workload (in particular re-contouring), the computational load for the daily optimization + dose calculation, and the development of adequate quality assurance. Due to parallelization of TRiP4D the time demands could be decreased to ~ 6 min which might already be sufficiently short for a clinical application [40].

With respect to the need for treatment plan adaptation due to dosimetric influence of inter-fractional changes in lung cancer treatments, Brevet et al. used serial 4DCTs to investigate in a treatment plan comparison the target coverage in gated scanned ion beam treatments [41]. Based on 9 data sets containing 6–10 weekly 4DCTs per patient a single treatment plan was optimized based on the first 4DCT. Plan optimization used the ITV approach of Graeff et al. [15] to cover residual motion in the gating window of 25% of the breathing motion amplitude. For varying parameters of gating window and size of the beam focus, the coverage of the target (V95) and the conformity index were investigated by repeated dose calculations based on the weekly 4DCTs. In addition, the number of fields of the treatment plan was varied since a homogenization effect is expected [42], and the ITV was expanded by additional margins. Calculation of the 4D dose distributions using TRiP4D considered the inter- and intra-fractional motion component. The results show that a combination of increased beam focus size (15 mm FWHM), reduced gating window (11.9%), additional ITV-PTV margins especially addressing the beam range, and dose application by 3 fields yielded the best target coverage of the multi-week fractionated treatment scheme. A V95 coverage of the CTV of 96.5% was determined for that treatment parameter combination.

4D treatment planning for the treatment of cardiac arrhythmias

Cardiac arrhythmias and especially atrial fibrillation (AF) as the most common arrhythmia [43] is a major cause of stroke [44] and effects more than 2.3 million patients in the USA per year [45]. In patients suffering from AF the sinus rhythm of the heart is disturbed due to disorganized electrical impulses originating from the pulmonary veins or the atria which occur in parallel to the impulses of the sinoatrial node, the natural pacemaker. This leads to an irregular accumulated impulse to the ventricles that causes the dangerous irregular heart beat and predisposes stroke. To date, standard of care to treat these patients is electrical isolation of the pulmonary veins by endocardial radio frequency ablation (ERFA) or drug therapy [44]. ERFA is an invasive and typically more than 5 h long intervention [46] with only 75% success rate after 1 year [47]. Besides required improvements of success rates, reduction of severe complications is essential since these affect 6% of the patients and include peri-procedural death and stroke [47]. Last but not least treatments of AF are very expensive, leading to annual cost of €13.5 billion in the EU [48].

Initial studies indicate that radiation therapy might be a non-invasive alternative to ERFA [49, 50]. These studies were conducted with photon beam therapy and thus suffer from a lower target conformation and especially higher integral normal tissue dose than particle beam therapy. In a joint project of GSI, the Heidelberg University Clinic in Heidelberg, Germany and the Mayo Clinic in Rochester, Mn USA the feasibility of carbon beam therapy of cardiac fibrillation has been tested in an animal study after in-vitro irradiations of an explanted heart in a Langendorff setup showed that AV blocks can be achieved if sufficiently high doses are applied [51, 52].

All details of the conducted study are reported by Lehmann et al. [53]. In total 17 pigs were randomized for either irradiation of the AV node (n = 8), the right superior pulmonary vein left atrial junction, the left ventricle and to comparable sham-procedures (each n = 3). Irradiation was delivered at GSI using rasterscanned pencil beams on a horizontal beam line. A dose of 40 Gy in a single fraction was delivered to the target volume apart for the AV node group, were three different dose levels (25 Gy (n = 2), 40 Gy, 55 Gy (each n = 3)) were delivered.

Scanned particle beam delivery to the beating heart is influenced by cardiac as well as respiratory motion. To compensate the dosimetric influence of respiratory motion the ventilated and sedated animals were treated in end-exhale for all therapeutically relevant procedures (imaging, positioning, irradiation) by controlling the respirator such that a 25 s breath-hold at end-exhale was achieved. Influence of cardiac motion was addressed by rescanning and assessed by 4D treatment planning using TRiP4D as described above. 4DTP started with deformable image registration for propagation of contours and 4D dose calculation. The core part were multiple 4D dose calculations to investigate the influence of changing breathing and delivery parameters on the dose distribution. Similar studies were performed by Constantinescu et al. to study the feasibility of AF treatments for humans [54]. Within that approach margins (lateral and range) and the number of rescans were optimized such that the clinical goals were met. 15 rescans in the distal slices resulted in a sufficiently homogenous target coverage. Prior animal irradiation the treatment plans were delivered into a water tank using the robotic 4D phantom developed by Steidl et al. [55]. During delivery of these treatment plans and especially during treatment of the animals, the log files of the beam delivery sequence and the ECG trace were recorded such that the delivered 4D dose could be reconstructed [56].

Treatment outcome was judged after 24–40 weeks using among other tests electroanatomical mapping against baseline data and inspection of the macroscopic lesion. The data show that scanned carbon beam therapy allows chronic interruption of impulse propagation in the heart 13–17 weeks after a single irradiation with 40–55 Gy. No severe radiation induced side effects were seen. Concerning 4D reconstruction of the delivered dose using TRiP4D, Richter et al. showed that within 30 min after irradiation a preliminary dose assessment of individual fields was possible [56]. Target volume D95 dose levels were >95% in all but one animal for which technical reasons could be identified to explain the deviation. OAR dose differed in median by 0.1% from the planned dose. The workflow was initially developed for HCC treatments at HIT [19] but improvements allowed dose assessment briefly after dose delivery, a prerequisite, e.g. for adaptive treatment schedules.

Conclusions & outlook

Within the scope of the funded period an infrastructure was set up at GSI/HIT allowing to investigate several relevant issues concerning the management of organ motion in scanned ion beam therapy. A central item is TRiP4D, an in-house 4D treatment planning system, which allows treatment plan optimization and dose calculation in the presence of motion. Within that period simulation and experimental phantom studies on 4D optimized treatment plans and simulation studies addressing the dosimetric benefits of adaptive treatments have been performed. In addition, first patients with intra-fractionally moving hepato cellular carcinoma have been treated with a scanned carbon beam at HIT using parameters identified in pre-clinical research based on 4D dose calculations and irradiations using in-house developed motion phantoms. Treatment outcome was successfully monitored by reconstructing the delivered dose distribution. Recently, preclinical research on charged particle beam treatment of cardiac arrhythmias was based on 4D treatment planning and delivery using the GSI infrastructure.

Future investigations will lead to ion treatment of pancreatic cancer und lung cancer using a scanned beam. Tumor movement will probably have a stronger impact on the dose distribution in a target surrounded by low density tissue compared to targets in the liver.

Abbreviations

4D:

Four dimensional

AF:

Atrial fibrillation

AV node:

Atrioventricular node

CBCT:

Cone beam CT

CT:

Computed tomography

CTV:

Clinical target volume

D95:

Minimal dose, covering 95% of the volume of interest in the dose volume histogram

DFG:

German research council

ERFA:

Endocardial radio frequency ablation

EU:

European union

FWHM:

Full width at half maximum

GSI:

GSI Helmholtz centre for heavy Ion research in Darmstadt, Germany

HCC:

Hepato cellular carcinoma

HIT:

Heidelberg Ion-beam therapy center

HU:

Hounsfield unit

IES:

Iso-energy slice

ITV:

Internal target volume

KFO:

Clinical research unit

LUT:

Look-up table

OAR:

Organ at risk

PTV:

Planning target volume

TCS:

Therapy control system

TRiP4D:

extension of TRiP98 by the time domain

TRiP98:

Treatment planning for particles

V95:

Fraction of the dose volume histogram covered with more than 95% of the prescribed dose

References

  1. Bert C, Grozinger SO, Rietzel E. Quantification of interplay effects of scanned particle beams and moving targets. Phys Med Biol. 2008;53(9):2253–65.

    Article  PubMed  Google Scholar 

  2. Phillips MH, Pedroni E, Blattmann H, Boehringer T, Coray A, Scheib S. Effects of respiratory motion on dose uniformity with a charged particle scanning method. Phys Med Biol. 1992;37(1):223–34.

    Article  CAS  PubMed  Google Scholar 

  3. Knopf AC, Stutzer K, Richter C, Rucinski A, da Silva J, Phillips J, Engelsman M, Shimizu S, Werner R, Jakobi A, et al. Required transition from research to clinical application: report on the 4D treatment planning workshops 2014 and 2015. Phys Med. 2016;32(7):874–82.

    Article  PubMed  Google Scholar 

  4. Bert C, Durante M. Motion in radiotherapy: particle therapy. Phys Med Biol. 2011;56(16):R113–44.

    Article  CAS  PubMed  Google Scholar 

  5. Graeff C. Motion mitigation in scanned ion beam therapy through 4D-optimization. Phys Med. 2014;30(5):570–7.

    Article  PubMed  Google Scholar 

  6. Kubiak T. Particle therapy of moving targets-the strategies for tumour motion monitoring and moving targets irradiation. Br J Radiol. 2016;89(1066):20150275.

    Article  PubMed  PubMed Central  Google Scholar 

  7. De Ruysscher D, Sterpin E, Haustermans K, Depuydt T. Tumour movement in proton therapy: solutions and remaining questions: a review. Cancers. 2015;7(3):1143–53.

    Article  PubMed  PubMed Central  Google Scholar 

  8. Kramer M, Scholz M. Treatment planning for heavy-ion radiotherapy: calculation and optimization of biologically effective dose. Phys Med Biol. 2000;45(11):3319–30.

    Article  CAS  PubMed  Google Scholar 

  9. Kramer M, Jakel O, Haberer T, Kraft G, Schardt D, Weber U. Treatment planning for heavy-ion radiotherapy: physical beam model and dose optimization. Phys Med Biol. 2000;45(11):3299–317.

    Article  CAS  PubMed  Google Scholar 

  10. Grozinger SO, Rietzel E, Li Q, Bert C, Haberer T, Kraft G. Simulations to design an online motion compensation system for scanned particle beams. Phys Med Biol. 2006;51(14):3517–31.

    Article  PubMed  Google Scholar 

  11. Li Q, Groezinger SO, Haberer T, Rietzel E, Kraft G. Online compensation for target motion with scanned particle beams: simulation environment. Phys Med Biol. 2004;49(14):3029–46.

    Article  PubMed  Google Scholar 

  12. Bert C, Rietzel E. 4D treatment planning for scanned ion beams. Radiat Oncol. 2007;2:24.

    Article  PubMed  PubMed Central  Google Scholar 

  13. Gemmel A, Rietzel E, Kraft G, Durante M, Bert C. Calculation and experimental verification of the RBE-weighted dose for scanned ion beams in the presence of target motion. Phys Med Biol. 2011;56(23):7337–51.

    Article  CAS  PubMed  Google Scholar 

  14. Richter D, Schwarzkopf A, Trautmann J, Kramer M, Durante M, Jakel O, Bert C. Upgrade and benchmarking of a 4D treatment planning system for scanned ion beam therapy. Med Phys. 2013;40(5):051722.

    Article  CAS  PubMed  Google Scholar 

  15. Graeff C, Durante M, Bert C. Motion mitigation in intensity modulated particle therapy by internal target volumes covering range changes. Med Phys. 2012;39(10):6004–13.

    Article  PubMed  Google Scholar 

  16. Graeff C, Luchtenborg R, Eley JG, Durante M, Bert C. A 4D-optimization concept for scanned ion beam therapy. Radiother Oncol. 2013;109(3):419–24.

    Article  PubMed  Google Scholar 

  17. Combs SE, Habermehl D, Ganten T, Schmidt J, Edler L, Burkholder I, Jäkel O, Haberer T, Debus J. Phase i study evaluating the treatment of patients with hepatocellular carcinoma (HCC) with carbon ion radiotherapy: the PROMETHEUS-01 trial. BMC Cancer. 2011;11:67.

    Article  PubMed  PubMed Central  Google Scholar 

  18. Richter D, Graeff C, Jakel O, Combs SE, Durante M, Bert C. Residual motion mitigation in scanned carbon ion beam therapy of liver tumors using enlarged pencil beam overlap. Radiother Oncol. 2014;113(2):290–5.

    Article  PubMed  Google Scholar 

  19. Richter D, Saito N, Chaudhri N, Hartig M, Ellerbrock M, Jakel O, Combs SE, Habermehl D, Herfarth K, Durante M, et al. Four-dimensional patient dose reconstruction for scanned ion beam therapy of moving liver tumors. Int J Radiat Oncol Biol Phys. 2014;89(1):175–81.

    Article  PubMed  Google Scholar 

  20. Steidl P, Haberer T, Durante M, Bert C. Gating delays for two respiratory motion sensors in scanned particle radiation therapy. Phys Med Biol. 2013;58(21):N295–302.

    Article  CAS  PubMed  Google Scholar 

  21. Veiga C, Alshaikhi J, Amos R, Lourenço AM, Modat M, Ourselin S, Royle G, McClelland JR. Cone-beam computed tomography and deformable registration-based “dose of the day” calculations for adaptive proton therapy. Int J Particle Ther. 2015;2(2):404–14.

    Article  Google Scholar 

  22. Kurz C, Dedes G, Resch A, Reiner M, Ganswindt U, Nijhuis R, Thieke C, Belka C, Parodi K, Landry G. Comparing cone-beam CT intensity correction methods for dose recalculation in adaptive intensity-modulated photon and proton therapy for head and neck cancer. Acta Oncol (Stockh, Sweden). 2015;54(9):1651–7.

    Article  CAS  Google Scholar 

  23. Wolfelschneider J, Seregni M, Fassi A, Ziegler M, Baroni G, Fietkau R, Riboldi M, Bert C. Examination of a deformable motion model for respiratory movements and 4D dose calculations using different driving surrogates. Med Phys. 2017;44(6):2066–76.

    Article  PubMed  Google Scholar 

  24. Wolfelschneider J, Brandt T, Lettmaier S, Fietkau R, Bert C. Quantification of an external motion surrogate for quality assurance in lung cancer radiation therapy. Biomed Res Int. 2014;2014:595430.

    Article  PubMed  PubMed Central  Google Scholar 

  25. Fassi A, Seregni M, Riboldi M, Cerveri P, Sarrut D, Ivaldi GB, de Fatis PT, Liotta M, Baroni G. Surrogate-driven deformable motion model for organ motion tracking in particle radiation therapy. Phys Med Biol. 2015;60(4):1565–82.

    Article  PubMed  Google Scholar 

  26. Nohadani O, Seco J, Bortfeld T. Motion management with phase-adapted 4D-optimization. Phys Med Biol. 2010;55(17):5189–202.

    Article  PubMed  PubMed Central  Google Scholar 

  27. Trofimov A, Rietzel E, HM L, Martin B, Jiang S, Chen GT, Bortfeld T. Temporo-spatial IMRT optimization: concepts, implementation and initial results. Phys Med Biol. 2005;50(12):2779–98.

    Article  PubMed  Google Scholar 

  28. Bert C, Saito N, Schmidt A, Chaudhri N, Schardt D, Rietzel E. Target motion tracking with a scanned particle beam. Med Phys. 2007;34(12):4768–71.

    Article  PubMed  Google Scholar 

  29. Eley JG, Newhauser WD, Luchtenborg R, Graeff C, Bert C. 4D optimization of scanned ion beam tracking therapy for moving tumors. Phys Med Biol. 2014;59(13):3431–52.

    Article  PubMed  PubMed Central  Google Scholar 

  30. Pos FJ, Hulshof M, Lebesque J, Lotz H, van Tienhoven G, Moonen L, Remeijer P. Adaptive radiotherapy for invasive bladder cancer: a feasibility study. Int J Radiat Oncol Biol Phys. 2006;64(3):862–8.

    Article  PubMed  Google Scholar 

  31. Yan D, Lockman D, Brabbins D, Tyburski L, Martinez A. An off-line strategy for constructing a patient-specific planning target volume in adaptive treatment process for prostate cancer. Int J Radiat Oncol Biol Phys. 2000;48(1):289–302.

    Article  CAS  PubMed  Google Scholar 

  32. Koay EJ, Lege D, Mohan R, Komaki R, Cox JD, Chang JY. Adaptive/nonadaptive proton radiation planning and outcomes in a phase II trial for locally advanced non-small cell lung cancer. Int J Radiat Oncol Biol Phys. 2012;84(5):1093–100.

    Article  PubMed  PubMed Central  Google Scholar 

  33. Chen W, Gemmel A, Rietzel E. A patient-specific planning target volume used in 'plan of the day' adaptation for interfractional motion mitigation. J Radiat Res. 2013;54(Suppl 1):i82–90.

    Article  PubMed  PubMed Central  Google Scholar 

  34. Foroudi F, Pham D, Rolfo A, Bressel M, Tang CI, Tan A, Turner S, Hruby G, Williams S, Hayne D, et al. The outcome of a multi-centre feasibility study of online adaptive radiotherapy for muscle-invasive bladder cancer TROG 10.01 BOLART. Radiother Oncol. 2014;111(2):316–20.

    Article  PubMed  Google Scholar 

  35. Nagano A, Minohara S, Kato S, Kiyohara H, Ando K. Adaptive radiotherapy based on the daily regression of a tumor in carbon-ion beam irradiation. Phys Med Biol. 2012;57(24):8343–56.

    Article  PubMed  Google Scholar 

  36. Vestergaard A, Hafeez S, Muren LP, Nill S, Hoyer M, Hansen VN, Gronborg C, Pedersen EM, Petersen JB, Huddart R, et al. The potential of MRI-guided online adaptive re-optimisation in radiotherapy of urinary bladder cancer. Radiother Oncol. 2016;118(1):154–9.

    Article  PubMed  Google Scholar 

  37. Hild S, Graeff C, Rucinski A, Zink K, Habl G, Durante M, Herfarth K, Bert C. Scanned ion beam therapy for prostate carcinoma: comparison of single plan treatment and daily plan-adapted treatment. Strahlenther Onkol. 2016;192(2):118–26.

    Article  PubMed  Google Scholar 

  38. Ahunbay EE, Peng C, Holmes S, Godley A, Lawton C, Li XA. Online adaptive replanning method for prostate radiotherapy. Int J Radiat Oncol Biol Phys. 2010;77(5):1561–72.

    Article  PubMed  Google Scholar 

  39. Zhang M, Westerly DC, Mackie TR. Introducing an on-line adaptive procedure for prostate image guided intensity modulate proton therapy. Phys Med Biol. 2011;56(15):4947–65.

    Article  CAS  PubMed  Google Scholar 

  40. Hild S, Graeff C, Trautmann J, Kraemer M, Zink K, Durante M, Bert C. Fast optimization and dose calculation in scanned ion beam therapy. Med Phys. 2014;41(7):071703.

    Article  CAS  PubMed  Google Scholar 

  41. Brevet R, Richter D, Graeff C, Durante M, Bert C. Treatment parameters optimization to compensate for Interfractional anatomy variability and Intrafractional tumor motion. Front Oncol. 2015;5:291.

    Article  PubMed  PubMed Central  Google Scholar 

  42. Knopf AC, Hong TS, Lomax A. Scanned proton radiotherapy for mobile targets-the effectiveness of re-scanning in the context of different treatment planning approaches and for different motion characteristics. Phys Med Biol. 2011;56(22):7257–71.

    Article  PubMed  Google Scholar 

  43. Feinberg WM, Blackshear JL, Laupacis A, Kronmal R, Hart RG. Prevalence, age distribution, and gender of patients with atrial fibrillation. Analysis and implications. Arch Intern Med. 1995;155(5):469–73.

    Article  CAS  PubMed  Google Scholar 

  44. Haissaguerre M, Jais P, Shah DC, Takahashi A, Hocini M, Quiniou G, Garrigue S, Le Mouroux A, Le Metayer P, Clementy J. Spontaneous initiation of atrial fibrillation by ectopic beats originating in the pulmonary veins. N Engl J Med. 1998;339(10):659–66.

    Article  CAS  PubMed  Google Scholar 

  45. Go AS, Hylek EM, Phillips KA, Chang Y, Henault LE, Selby JV, Singer DE. Prevalence of diagnosed atrial fibrillation in adults: national implications for rhythm management and stroke prevention: the AnTicoagulation and risk factors in Atrial fibrillation (ATRIA) study. JAMA. 2001;285(18):2370–5.

    Article  CAS  PubMed  Google Scholar 

  46. Jongbloed MR, Dirksen MS, Bax JJ, Boersma E, Geleijns K, Lamb HJ, van der Wall EE, de Roos A, Schalij MJ. Atrial fibrillation: multi-detector row CT of pulmonary vein anatomy prior to radiofrequency catheter ablation--initial experience. Radiology. 2005;234(3):702–9.

    Article  PubMed  Google Scholar 

  47. Cappato R, Calkins H, Chen SA, Davies W, Iesaka Y, Kalman J, Kim YH, Klein G, Packer D, Skanes A. Worldwide survey on the methods, efficacy, and safety of catheter ablation for human atrial fibrillation. Circulation. 2005;111(9):1100–5.

    Article  PubMed  Google Scholar 

  48. Fuster V, Ryden LE, Cannom DS, Crijns HJ, Curtis AB, Ellenbogen KA, Halperin JL, Le Heuzey JY, Kay GN, Lowe JE, et al. ACC/AHA/ESC 2006 guidelines for the management of patients with atrial fibrillation: full text: a report of the American College of Cardiology/American Heart Association task force on practice guidelines and the European Society of Cardiology Committee for practice guidelines (writing committee to revise the 2001 guidelines for the management of patients with atrial fibrillation) developed in collaboration with the European heart rhythm association and the Heart Rhythm Society. Europace. 2006;8(9):651–745.

    Article  PubMed  Google Scholar 

  49. Bode F, Blanck O, Gebhard M, Hunold P, Grossherr M, Brandt S, Vonthein R, Thiele H, Dunst J, Rades D. Pulmonary vein isolation by radiosurgery: implications for non-invasive treatment of atrial fibrillation. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of. Cardiology. 2015;17(12):1868–74.

    Google Scholar 

  50. Sharma A, Wong D, Weidlich G, Fogarty T, Jack A, Sumanaweera T, Maguire P. Noninvasive stereotactic radiosurgery (CyberHeart) for creation of ablation lesions in the atrium. Heart Rhythm. 2010;7(6):802–10.

    Article  PubMed  Google Scholar 

  51. Prall M, Lehmann HI, Prokesch H, Richter D, Graeff C, Kaderka R, Sonnenberg K, Hauswald H, Weymann A, Bauer J, et al. Treatment of arrhythmias by external charged particle beams: a Langendorff feasibility study. Biomed Tech (Berl). 2015;60(2):147–56.

    Article  CAS  Google Scholar 

  52. Lehmann HI, Richter D, Prokesch H, Graeff C, Prall M, Simoniello P, Fournier C, Bauer J, Kaderka R, Weymann A, et al. Atrioventricular node ablation in Langendorff-perfused porcine hearts using carbon ion particle therapy: methods and an in vivo feasibility investigation for catheter-free ablation of cardiac arrhythmias. Circ Arrhythm Electrophysiol. 2015;8(2):429–38.

    Article  CAS  PubMed  Google Scholar 

  53. Lehmann HI, Graeff C, Simoniello P, Constantinescu A, Takami M, Lugenbiel P, Richter D, Eichhorn A, Prall M, Kaderka R, et al. Feasibility study on cardiac arrhythmia ablation using high-energy heavy ion beams. Sci Rep. 2016;6:38895.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  54. Constantinescu A, Lehmann HI, Packer DL, Bert C, Durante M, Graeff C. Treatment planning studies in patient data with scanned carbon ion beams for catheter-free ablation of Atrial fibrillation. J Cardiovasc Electrophysiol. 2016;27(3):335–44.

    Article  PubMed  Google Scholar 

  55. Steidl P, Richter D, Schuy C, Schubert E, Haberer T, Durante M, Bert C. A breathing thorax phantom with independently programmable 6D tumour motion for dosimetric measurements in radiation therapy. Phys Med Biol. 2012;57(8):2235–50.

    Article  CAS  PubMed  Google Scholar 

  56. Richter D, Lehmann HI, Eichhorn A, Constantinescu A, Kaderka R, Prall M, Lugenbiel P, Takami M, Thomas D, Bert C, et al. ECG-based 4D-dose reconstruction of cardiac arrhythmia ablation with carbon ion beams: methods and validation in a porcine model. Phys Med Biol. 2017;62(17):6869–83.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

The authors acknowledge the contributions of Dr. Daniel Richter, Dr. Jens Wölfelschneider, Dr. Peter Steidl, Dr. Romain Brevet, Dr. Christian Graeff, Dr. Sebastian Hild, and Dr. John Eley who worked on the reported projects in the last years.

Funding

The presented research was funded by the German Research Council (DFG) – Clinical Research Unit (KFO) 214 and 214/2.

Availability of data and materials

Data sharing not applicable to this article as no datasets were generated or analyzed during the current study.

Author information

Authors and Affiliations

Authors

Contributions

All authors contributed to the manuscript which was initially drafted by CB. All authors approve the final version.

Corresponding author

Correspondence to Christoph Bert.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

All authors state no conflict of interest.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Bert, C., Herfarth, K. Management of organ motion in scanned ion beam therapy. Radiat Oncol 12, 170 (2017). https://doi.org/10.1186/s13014-017-0911-z

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13014-017-0911-z

Keywords