Skip to main content

Table 4 Comparison of the total beam-on time in minutes per patient with 1,400 MU/min dose rate delivery

From: Dosimetric evaluation of LINAC-based single-isocenter multi-target multi-fraction stereotactic radiosurgery with more than 20 targets: comparing MME, HyperArc, and RapidArc

Beam-on time

Patient#

# of target

RA

HA

MME

1

27

4.13

4.47

8.05

2

25

3.06

3.16

8.40

3

36

3.29

3.03

6.86

4

24

2.57

2.17

7.67

5

24

3.86

5.71

6.99

6

37

3.12

2.72

8.93

7

20

2.13

1.79

4.13

8

27

2.97

3.44

7.56

9

22

3.59

3.67

6.63

10

21

2.31

8.39

2.67

Mean

26.3

3.10 ± 0.61

3.28 ± 1.08

7.36 ± 1.29