Skip to main content

Table 5 Average penumbra (mm) for 6MV flattened and unflattened beams (reproduced with permission from MedEuropa)

From: Dosimetric characteristics of 6MV flattening filter free and flattened beams among beam-matched linacs: a three-institutional study

Beam type

Field size

(cm2)

Right Penumbra (mm)

Left Penumbra (mm)

Reference

Linac 1

Linac 2

Linac 3

Reference

Linac 1

Linac 2

Linac 3

6MV FFF

3 × 3

4.27

4.77

4.46

4.36

4.27

4.70

4.43

4.39

5 × 5

6.39

6.40

6.11

6.13

6.39

6.38

6.12

6.13

10 × 10

7.33

7.25

7.23

7.27

7.33

7.16

7.23

7.26

6MV

3 × 3

4.39

4.75

4.57

4.46

4.39

4.72

4.52

4.47

5 × 5

6.51

6.51

6.34

6.34

6.51

6.49

6.33

6.31

10 × 10

7.53

7.48

7.48

7.42

7.53

7.46

7.42

7.43