Skip to main content

Table 1 Depth of maximum dose and PDD at 10 cm depth for 6MV flattened and unflattened beams (reproduced with permission from MedEuropa)

From: Dosimetric characteristics of 6MV flattening filter free and flattened beams among beam-matched linacs: a three-institutional study

Beam type

6MV FFF

6MV

Field size (cm2)

3 × 3

5 × 5

10 × 10

3 × 3

5 × 5

10 × 10

dmax (mm)

Reference

16.97

16.99

16.98

15.01

15.01

14.99

Linac 1

16.6

17.3

17.8

14.9

15.7

15.0

Linac 2

17.0

17.2

16.4

15.7

15.7

15.1

Linac 3

16.9

17.2

17.0

15.0

15.5

15.1

PDD10 (%)

Reference

60.96

62.95

65.9

60.49

62.54

65.95

Linac 1

60.58

62.91

66.03

60.21

62.86

65.86

Linac 2

60.76

62.8

65.95

60.61

62.7

65.95

Linac 3

60.93

62.88

65.96

60.57

62.46

65.95