Skip to main content

Table 2 Recorded results of patient-specific quality assurance

From: Integrating plan complexity and dosiomics features with deep learning in patient-specific quality assurance for volumetric modulated arc therapy

Measured GPR

3%/3 mm

3%/2 mm

2%/2 mm

Mean (%)

97.94

94.33

87.27

SD (%)

1.87

3.22

4.81

Case numbers

   

 GPR ≥ 95%

186

106

7

 90% ≤ GPR < 95%

15

75

58

 80% ≤ GPR < 90%

0

22

115

 GPR < 80%

0

0

21