Skip to main content

Table 1 Comparison of gamma passing rates between DoseCBCT and DosesCT (threshold = 10%, global 3D)

From: Dosimetric comparison of deformable image registration and synthetic CT generation based on CBCT images for organs at risk in cervical cancer radiotherapy

 

1%/1 mm (%)

2%/2 mm (%)

2%/3 mm (%)

3%/2 mm (%)

3%/3 mm (%)

DosepCT vs. DosedpCT

95.53 ± 3.53

99.12 ± 0.87

99.51 ± 0.61

99.52 ± 0.61

99.68 ± 0.47

DoseCBCT vs. DosedpCT

85.92 ± 7.56

97.92 ± 1.65

98.69 ± 1.24

99.31 ± 0.71

99.67 ± 0.52

DosesCT vs. DosedpCT

99.66 ± 0.34

99.98 ± 0.05

99.99 ± 0.02

99.99 ± 0.02

100 ± 0

P* value

p < 0.001

p < 0.001

p < 0.001

p < 0.001

p < 0.001

  1. *P was calculated by comparing the gamma passing rates of DoseCBCT vs. DosedpCT and DosesCT vs. DosedpCT according to paired sample t tests