Table1 Gamma passing rates of the comparison between measured and calculated dose distributions for the anthropomorphic phantom

From: Feasibility of postoperative spine stereotactic body radiation therapy in proximity of carbon and titanium hybrid implants using a robotic radiotherapy device

  Collimator Phantom PTV 1/PTV 2
Plane 1 Iris Bone 100%/97%
  CFP-T implant 100%/97%
MLC Bone 99%/97%
  CFP-T implant 100%/99%
Plane 2 Iris Bone 99%/97%
  CFP-T implant 100%/99%
MLC Bone 99%/97%
  CFP-T implant 98%/97%
  1. The position of the measurement planes is indicated in Fig. 2