From: Linac photon beam fine-tuning in PRIMO using the gamma-index analysis toolkit
Number of Histories (106) | Average Dose Uncertainty (%) | 3 % /3mm (%) | 2 % /2mm (%) | 1 % /1mm (%) |
---|---|---|---|---|
5 | 12.4 | 99.4 | 85.1 | 62.1 |
10 | 9.1 | 98.8 | 85.1 | 54.0 |
25 | 5.9 | 100.0 | 100.0 | 74.5 |
65.06 | 3.7 | 100.0 | 90.7 | 55.9 |