Skip to main content

Table 3 Correlations between global gamma passing rates with various gamma criteria and the modulation indices for intensity modulated radiation therapy plans of TrueBeam STx (p ≤ 0.05)

From: Assessment of the modulation degrees of intensity-modulated radiation therapy plans

 

3%/3 mm

2%/2 mm

2%/1 mm

1%/2 mm

r s

p

r s

p

r s

p

r s

p

MapCHECK2

 MIs

− 0.241

0.016

− 0.200

0.046

–

–

–

–

 MIa

–

–

–

–

–

–

–

–

 MIc, IMRT

–

–

–

–

–

–

− 0.201

0.045

 MCS

0.453

< 0.001

0.548

< 0.001

0.509

< 0.001

0.460

< 0.001

 PA

−0.223

0.026

− 0.236

0.018

–

–

− 0.196

0.050

 PI

−0.465

< 0.001

−0.591

< 0.001

− 0.550

< 0.001

− 0.474

< 0.001

 PM

− 0.476

< 0.001

− 0.566

< 0.001

− 0.516

< 0.001

− 0.438

< 0.001

ArcCHECK

 MIs

− 0.375

< 0.001

− 0.341

0.001

−0.518

< 0.001

− 0.292

0.003

 MIa

−0.234

0.019

–

–

− 0.234

0.019

–

–

 MIc, IMRT

−0.225

0.025

–

–

−0.228

0.022

–

–

 MCS

0.431

< 0.001

0.593

< 0.001

0.451

< 0.001

0.699

< 0.001

 PA

−0.235

0.019

–

–

− 0.225

0.024

–

–

 PI

−0.340

0.001

−0.545

< 0.001

−0.328

0.001

−0.677

< 0.001

 PM

−0.369

< 0.001

−0.545

< 0.001

− 0.389

< 0.001

− 0.667

< 0.001

  1. Note: rs Spearman’s rho, MIs Modulation index considering multi-leaf collimator speed, MIa Modulation index considering multi-leaf collimator acceleration, MIc, IMRT Modulation index considering multi-leaf collimator acceleration and field aperture irregularity for intensity modulated radiation therapy, MCS Modulation complexity score, PA Plan-averaged beam area, PI Plan-averaged beam irregularity, PM Plan-averaged beam modulation