Skip to main content

Table 2 Dosimetric endpoint comparison for case 2

From: Voxel-based automatic multi-criteria optimization for intensity modulated radiation therapy

ROI

Endpoints

Original plan

Optimal plan

Hard constraints

PTV

D99%

42.55 Gy

42.96 Gy

40.5 Gy

PTV

D97%

43.7 Gy

43.7 Gy

43.65 Gy

PTV

V115%

0

0

1

PTV

V110%

0.22

0

10

Rectum

Dmax

47.7 Gy

47.7 Gy

51.75 Gy

Bladder

Dmax

48.1 Gy

46.5 Gy

51.75 Gy

Bowel

Dmax

49.3 Gy

47.3 Gy

51.75 Gy

Femoral Heads

Dmax

44.9 Gy

43.5 Gy

51.75 Gy

Bone

V10

88.72%

79.65%

90%

Bone

V20

74.16%

61.24%

75%

    

Soft constraints

Rectum

V30

99.97%

96.40%

60%

V45

39.71%

12.01%

50%

Bladder

V45

44.64%

13.13%

50%

Bowel

V40

7.44%

8.11%

30%

Femoral Heads

V30

30.02%

13.87%

15%

Bone

V10

88.72%

79.65%

80%

Bone

V20

74.16%

61.24%

66%