From: A simple method for determining dosimetric leaf gap with cross-field dose width for rounded leaf-end multileaf collimator systems
6 MV
6 MV-FFF
10 MV
10 MV-FFF
DLG (mm)
5 cm- ion chamber, gD
1.43
1.15
1.66
10 cm- ion chamber, gD
1.47
1.21
1.68
5 cm- Film
1.36
1.26
1.45
1.54
10 cm- Film
1.35
1.37
1.88
1.65