Table 1 Hounsfield Units measurements of air gaps between boluses and RANDO® phantom external surface

From: Improving 3D-printing of megavoltage X-rays radiotherapy bolus with surface-scanner

MATERIALS

Mean HU

Standard deviation HU

Maximum Gap Depth HU

Maximum Gap Depth (mm)

BolusCT

 Clear Resin

− 295

58

− 627

1–2

 Digital ABS

− 387

148

− 590

1–2

 PLA

− 390

102

− 591

1–2

 All

− 383

134

−627

1–2

BolusS

 Clear Resin

− 45

51

− 167

< 0.6

 ABS

− 191

54

− 313

< 0.6

 PLA

− 155

77

− 294

< 0.6

 All

− 130

87

−313

< 0.6

BolusW

 Wax

−469

219

− 860

2–3