Skip to main content

Table 3 The results of mechanical performance metrics for the clinical plan and KBP

From: Mechanical performance of a commercial knowledge-based VMAT planning for prostate cancer

 

Clinical Plan

KBP

p-value

Mean

Max

Min

SD

Mean

Max

Min

SD

MFA (cm2)

18.28

25.07

12.39

3.34

17.25

22.89

12.15

2.80

0.23

MAD (mm)

21.08

25.04

17.27

2.14

20.47

24.49

16.88

1.92

0.38

CAS

0.54

0.63

0.34

0.07

0.55

0.66

0.41

0.07

0.46

CLS

0.040

0.072

0.0050

0.015

0.051

0.16

0.0070

0.027

0.03

SAS2mm

0.13

0.22

0.022

0.042

0.15

0.24

0.026

0.044

0.20

SAS5mm

0.20

0.30

0.089

0.053

0.23

0.33

0.091

0.087

0.05

SAS10mm

0.32

0.48

0.15

0.071

0.35

0.48

0.22

0.063

0.07

SAS20mm

0.54

0.71

0.29

0.087

0.58

0.77

0.44

0.081

0.11

LT (mm)

458.51

550.80

362.80

41.62

418.81

499.60

362.10

29.39

< 0.01

MCSv

0.27

0.31

0.22

0.023

0.27

0.30

0.25

0.015

0.24

MU

618.24

738.80

475.70

64.77

622.12

672.60

572.40

25.58

0.76

  1. Abbreviations: MFA mean field area, MAD mean asymmetry distance, CAS cross-axis score, CLS closed leaf score, SAS small aperture score, LT leaf travel, MCSv modulation complexity score for VMAT, MU monitor unit