Table 3 Summary of gamma pass rates (%) for gafchromic film measurements. All plans were measured on 3 separate instances where the error depicts the standard deviation

From: Technical and dosimetric implications of respiratory induced density variations in a heterogeneous lung phantom

Range of motion (cm) CT dataset AAA AXB
(3%, 3 mm) (3%, 1 mm) (3%, 3 mm) (3%, 1 mm)
1 cm AIP 99.4 ± 0.4 86.4 ± 5.2 99.8 ± 0.1 88.0 ± 4.1
FB 98.7 ± 0.7 86.9 ± 5.0 99.8 ± 0.1 88.2 ± 3.6
MIP 95.9 ± 3.6 80.5 ± 5.7 99.7 ± 0.2 89.8 ± 3.6
PWD 97.1 ± 3.2 85.7 ± 3.5 99.7 ± 0.2 89.4 ± 5.7
2 cm AIP 98.6 ± 0.7 85.0 ± 2.3 99.0 ± 0.4 86.8 ± 3.9
FB 99.3 ± 1.2 87.1 ± 5.6 99.6 ± 0.8 86.8 ± 4.7
MIP 98.2 ± 0.2 77.8 ± 4.5 99.5 ± 0.8 85.8 ± 4.0
PWD 98.4 ± 2.5 83.2 ± 4.5 99.6 ± 0.3 89.1 ± 3.6
3 cm AIP 98.1 ± 1.6 85.4 ± 4.5 99.1 ± 0.8 91.2 ± 1.7
FB 98.4 ± 1.3 88.8 ± 3.5 99.1 ± 0.8 91.7 ± 4.6
MIP 94.2 ± 3.8 69.4 ± 5.9 98.9 ± 0.8 88.1 ± 1.5
PWD 94.1 ± 0.4 76.0 ± 5.2 99.3 ± 0.6 94.2 ± 4.2