Table 2 Collimator scatter factors. Monte Carlo Sc are corrected for the MBSF, using the published factors for TrueBeam (the original simulation), and for Clinac (according to the measurements, to compare with real measurements). Measurements (on a Clinac treatment unit) refer to Mini-Phantom data acquired with a Farmer ion chamber at 10 cm depth of PMMA

From: Collimator scatter factor: Monte Carlo and in-air measurements approaches

Field size

Monte Carlo

Monte Carlo

Measurements

Sc 6MV, TrueBeam MBSF

Sc 15MV, TrueBeam MBSF

Sc 6MV, Clinac MBSF

Sc 15MV, Clinac MBSF

Sc 6MV

Sc 18MV

4 × 4 cm2

0.962 ± 0.008

0.957 ± 0.005

0.961 ± 0.008

0.955 ± 0.005

0.955 ± 0.003

0.946 ± 0.002

5 × 5 cm2

0.969 ± 0.008

0.965 ± 0.005

0.969 ± 0.008

0.964 ± 0.005

0.968 ± 0.003

0.961 ± 0.002

10 × 10 cm2

1.000 ± 0.008

1.000 ± 0.005

1.000 ± 0.008

1.000 ± 0.005

1.000 ± 0.003

1.000 ± 0.002

20 × 20 cm2

1.020 ± 0.008

1.024 ± 0.005

1.024 ± 0.008

1.029 ± 0.005

1.026 ± 0.003

1.024 ± 0.002

30 × 30 cm2

1.026 ± 0.008

1.035 ± 0.005

1.033 ± 0.008

1.044 ± 0.005

1.040 ± 0.003

1.038 ± 0.002

40 × 40 cm2

1.027 ± 0.008

1.035 ± 0.005

1.038 ± 0.008

1.050 ± 0.005

1.050 ± 0.003

1.050 ± 0.002