Skip to main content

Table 4 Summary of target volumes dosimetric analysis from the PDP analysis. Data are reported only for the positive mis-calibrations since these represent the worst scenarios

From: Proton versus photon deep inspiration breath hold technique in patients with hodgkin lymphoma and mediastinal radiation

Parameter

reference

Δ (+ 3%)

Δ (+4%)

CTV

 Mean [Gy]

30.1 ± 0.2

29.9 ± 0.2 -0.7%

29.9 ± 0.3 -0.7%

 D95% [Gy]

29.4 ± 0.1

28.9 ± 0.3 -1.7%

28.5 ± 0.4 -3.1%

 D98% [Gy]

29.4 ± 0.1

28.4 ± 0.4 -3.4%

27.8 ± 0.6 -5.4%

PTV

 Mean [Gy]

30.0

29.7 ± 0.2 -1.0%

29.4 ± 0.3 -2.0%

 D95% [Gy]

28.8 ± 0.3

27.6 ± 0.5 -4.2%

26.7 ± 0.7 -7.3%

 D98% [Gy]

28.8 ± 0.3

26.4 ± 0.8 -8.3.%

25.3 ± 1.0 -12.2%

  1. Dx% dose received by the x% of the volume, Vx% volume receiving at least x% of the prescribed dose, CI ratio between the patient volume receiving at least 90% of the prescribed dose and the volume of the total PTV