Fig. 1 | Radiation Oncology

Fig. 1

From: Dosimetric impact of tumor treating field (TTField) transducer arrays onto treatment plans for glioblastomas – a planning study

Fig. 1

a MV-CT (left side) and kV-CT (right side) of a RW3 slab phantom equipped with a NovoCure Optune TTField array planned for a 15 × 15 cm2 0° 6 MV 200 MU photon beam. b For verification of the dose calculation, the dose in 2 cm depth was measured in the same setup with and without a TTField array at the surface of a RW3 slab phantom. The dots represent the measurement without the TTField array, the dashed line represents the measurements with TTField