From: Treating lung cancer with dynamic conformal arc therapy: a dosimetric study
Dynamic conformal arc therapy (DCAT) | |||||
---|---|---|---|---|---|
Parameter | 3D CRT | 0° | 30° | 45° | 90° |
HI | 0.144±0.018 | 0.132±0.027 | 0.133±0.027 | 0.133±0.027 | 0.134±0.028 |
DHD | 0.028±0.005 | 0.032±0.007 | 0.032±0.007 | 0.032±0.007 | 0.032±0.007 |
CN | 0.594±0.109 | 0.688±0.071 | 0.692±0.069 | 0.692±0.068 | 0.674±0.083 |
Spinal cord: D1cc (cGy) | 2888±985 | 2916±723 | 2903±717 | 2896±711 | 2971±733 |
Lung: V20Gy (%) | 23.9±9.4 | 22.9±12.1 | 23.1±12.2 | 23.3±12.3 | 23.7±12.5 |
Lung: V5Gy(%) | 53.6±17.8 | 63.3±18.9 | 63.5±18.9 | 63.5±18.9 | 63.1±18.8 |
Lung: MLD (cGy) | 1173±422 | 1270±466 | 1278±470 | 1284±473 | 1289±480 |
Oesophagus: D mean (cGy) | 1695±756 | 1750±756 | 1751±749 | 1752±748 | 1740±744 |