Table 6 RTOG dosimetric parameter ratios of the re-optimized Type-C plan over the original Type-B plan for the 11 re-optimized patients

From: A comprehensive dosimetric study on switching from a Type-B to a Type-C dose algorithm for modern lung SBRT

Pt #

PTV Volume (cm3)

D95% Reduction (%)

Ratio R100%

Ratio R50%

Ratio D2cm

Ratio V20

1

6.1

12.5

1.0

1.2

1.1

1.2

2

6.5

2.3

1.0

0.8

1.0

0.8

3

13.0

10.6

0.9

0.9

1.0

1.0

4

13.6

15.1

1.0

0.9

1.0

1.0

5

14.1

7.4

0.9

0.8

1.0

0.9

6

17.0

4.1

1.0

0.8

1.0

0.9

7

18.0

4.5

1.0

1.1

1.0

1.2

8

35.4

9.1

1.0

1.1

0.9

1.2

9

37.1

4.7

0.9

1.0

1.0

1.0

10

48.8

9.7

1.0

1.0

1.0

1.0

11

58.4

16.7

1.0

1.2a

1.0

1.2

  1. Also listed are the PTV volumes and PTV D95% reduction revealed by the corresponding re-calculated Type-C plans as a reference. The patients are sorted by the PTV size
  2. aPatient# 11 had a minor deviation on R50% even after Type-C re-optimization, while the corresponding original Type-B plan was fully compliant