Table 1 Half value layers (HVL) in Al and mean or effective energies for the spectra of IORT devices using KV X-rays

From: Present state and issues in IORT Physics

Publication

System

Probe

Surrounding material

Distance from scource

Acceleration Voltage

HVL (mm Al)

Mean or effective photon energy

Other remarks

Zeiss Dosimetry manual [90]

Intrabeam

bare probe

water

10 mm fom tip

50 kV

0.64 mm

20.4 keV

beam quality T30 -T50 (PTW definition)

    

40 kV

0.48 mm

19.1 keV

≈T30

    

30 kV

0.41 mm

17.3 keV

≈T30

Beatty 1996 [17]

 

bare probe

air

 

40 keV

0.1 mm

  
   

2nd HVL

40 keV

0.16 mm

  

Armoogum 2007 [18]

 

bare probe

air

     

Beatty 1996 [17]

 

bare probe

in air behind 0.5 cm solid water

 

40 kV

0.3 mm

14.8 keV

 
  

in air behind 1 cm solid water

 

40 kV

0.71 mm

19.9 keV

 
  

1.5cm solid water

   

21.7 keV

 
  

2cm solid water

   

23.5 keV

 

Armoogum 2007 [18]

 

bare probe

in solid water

1cm

 

1.11 mm

23.5 keV

broad beam

   

5mm

 

0.54 mm

18 keV

 
   

in air

 

0.11 mm

10.75 keV

 

Avanzo 2012 [95]

 

bare probe

in solid water

1cm

50 kV

1.75 mm

  
   

2cm

 

2.14 mm

  
     

3.82 mm

  

Ebert 2009 [101]

 

bare probe

in water

0 cm

50 kV

 

19.5 keV

 
   

5 mm

  

27.3 keV

For other depths, changes in spectra were calculated using linear attenuation in water and inverse-square of distance

   

15 mm

  

31.5 keV

 
   

30 mm

  

34 keV

"

Avanzo 2012 [95]

 

bare probe

air

 

50 kV

0.11 mm

10.2 keV

 
 

3.5 cm appl.

solid water

  

0.8 mm

20.7 keV

 
 

4 cm appl.

   

0.98 mm

22.5 keV

 
 

4.5 cm appl.

   

1,1 mm

23.7 keV

 
 

5 cm appl.

   

1,23 mm

25.1 keV

 
 

bare probe

1 cm depth in solid water

  

1,75 mm

28.0 keV

 
 

bare probe

2 cm depth in solid water

  

2,14 mm

29.3 keV

 
 

bare probe

beind Tungsten rubber shield

  

3,82 mm

36.3 keV

 

Eaton 2010 [43]

 

1.5 cm appl.

air

 

50 kV

1.1 mm

 

narrow beam

 

3cm appl.

air

  

1.3 mm

 

narrow beam

 

3.5 cm appl.

air

  

0.85 mm

 

narrow beam

 

4.5 cm appl.

air

  

1.24 -1.25 mm

 

broad beam

 

5cm appl.

air

  

1.26-1.29 mm

 

broad beam

Croce et al. 2012 [28]

Papillon

   

50 kV

   

Carver et al. 2013 [98]

 

22/25 mm Appl.

   

0.64 mm

  

Liu 2008 [111]

Xoft Axxent

bare source

air

 

40 kV

0.45 ± 0.07 mm

16.8 ± 0.9 keV

 
    

2nd HVL

0.7 ± 0.1

  
    

50 kV

0.5 ± 0.2 mm

18 ± 2 keV

 
    

2nd HVL

0.9 ± 0.3 mm

  
   

water

1.5cm

40 kV

1.04 ± 0.04 mm

 Â