From: Initial characterization, dosimetric benchmark and performance validation of Dynamic Wave Arc
Site | Nr. of patients | Fractionation | Minimum % of PD covering PTV | PTV Volume (ccm) | OAR | Dose constrains | Clinical delivery technique | |
---|---|---|---|---|---|---|---|---|
Prostate | 3 |  | 78Gy(2Gy/fx) | >90 % | 171.8 [129.8-222.8] | Rectum | D2% < 74.1 Gy V70.2Gy < 25 % V62.4Gy < 50 % | 5 IMRT beams |
Oligo-metastatic cases | 9 | 5 x1lesion 2x2lesions 2x3lesions | 50Gya(5Gy/fx) | 100 % | 28.0 [12.2-57.5] | Lung Liver | V20Gy < 25 % V30Gy < 30 % Dmean < 22 Gy | 8-10 CRT beams per lesion |
Centrally-located NSCLC | 9 | 6 lesions 3 lesions | 48Gyb(12Gy/fx) 60Gy(7.5Gy/fx) | >95 % >95 % | 76.2 [17.2-209.7] | Esophagus Spinal cord Heart Proximal- bronchus Lung | D2% < 30 Gy D2% < 26 Gy D2% < 34 Gy D2% < 34 Gy V20Gy < (8-10)% | 8-10 CRT beams |
LAPC | 10 |  | 48Gy(3.2Gy/fx) | >95 % | 188.5 [140.6-257.2] | Duodenum | V39Gy < 1.0 cc V36Gy < 10 cc | 6 IMRT beams |
Stomach Kidney Liver | V39Gy < 1.0 cc V36Gy < 10 cc V20Gy < 30 % Dmean < 30 Gy |