Sub-process | N | Potential failure mode | Potential causes of failure | Potential effects of failure | S | O | D | RPN |
---|---|---|---|---|---|---|---|---|
I. Call of the patient in the waiting room | 1 | The patient is called but a different one answers/ The patient is not called | Identification does not include patient’s name, surname, date of birth, photo-Patient was not informed of modifications regarding the time of the appointment, patient is late | Delivery of the treatment to the wrong patient -the radiotherapy treatment is not delivered or is administered late | 10 | 1 | 2 | 20 |
II. Verification of the patient’s identity at the treatment’s room entry by asking personal data confirmation | 2 | Patient’s identity verification by checking all the personal data not performed | Only patient’s surname check | Possibility of mistaking patients and therefore treatments | 10 | 2 | 3 | 60 |
X. Check of the correct view of the patient from the treatment workspace using adjustable video cameras | 3 | Patient is not monitored during treatment | Video cameras are not correctly oriented or functioning | Cyberknife may hit the patient without the operator noticing it. Patient may be in need and not been seen | 9 | 2 | 2 | 36 |
XII. Patient selection using personal data (Name and surname) | 4 | Wrong patient’s name-Personal data check is not performed | Patient is called without checking patients’ list-Lapse of memory | Delivery of the treatment to the wrong patient-possibility of mistaking patients and therefore treatments | 10 | 2 | 5 | 100 |
XIII. Check of the correct treatment plan and of the number of fractions as described on the report print | 5 | Delivery to the patient of a wrong plan-plan check not performed | Personal data and patient ID on the printed plan not checked-lapse of memory | Patient receives wrong irradiation-possibility of mistaking patients and therefore treatments | 10 | 2 | 3 | 60 |
XV. Check of patient’s name, surname and medical ID by flagging the appropriate box for acceptance | 6 | Patient’s personal data not checked | Automatic action- Lapse of memory | Wrong patient or treatment-possibility of mistaking patients and therefore treatments | 10 | 2 | 7 | 140 |
XVI. Check of: plan name, tracking method (XSight spine), path, number of fraction, collimator type and aperture-flag of the appropriate box for acceptance | 7 | Data check is wrong or not performed | High workload-lapse of memory | Wrong patient or treatment-possibility of mistaking patients and therefore treatments | 10 | 2 | 7 | 140 |
XVII. Accurate alignment of the patient by comparing DRR and live images: adjustment of the values and tolerance levels defined in the image parameters window-adjustment of the X Sight Spine ROI dimensions | 8 | Wrong alignment-Threshold levels of the different parameters not modified when necessary | Difficulty to visually identify spine tract in the live images-Lapse of memory, insufficient experience of the operator with the treatment system | Treatment not properly delivered-longer time to start treatment | 10 | 1 | 4 | 40 |
XIX. Setting of the most appropriate patient size | 9 | Appropriate patient size not set | Lapse of memory, insufficient experience of the operator with the delivery system | Possible collisions or errors of the PDP system slowing down treatment | 9 | 2 | 5 | 90 |