Table 2 Dosimetric constancy checks: MPC and independent checks for all evaluated parameters, all energies

From: Evaluation of the Machine Performance Check application for TrueBeam Linac

Parameter

Check

Thresh

6MV

10MV

15MV

6MV FFF

10MV FFF

BEAM

       

Output Change [%]

MPC

±2.0

0.15 ± 0.13

0.10 ± 0.29

0.06 ± 0.15

0.24 ± 0.24

0.22 ± 0.16

Output Change [%]

Indep

 

0.4 ± 0.4

0.4 ± 0.5

0.3 ± 0.6

−0.1 ± 0.4

0.5 ± 0.3

Uniformity Change [%]

MPC

±2.0

0.77 ± 0.20

0.92 ± 0.13

0.94 ± 0.21

0.88 ± 0.19

0.89 ± 0.12

Uniformity Change [%]

Indep

 

−0.1 ± 0.1

−0.2 ± 0.2

0.4 ± 0.3

0.1 ± 0.1

0.1 ± 0.1

Center Shift [mm]

MPC

±0.5

0.04 ± 0.02

0.04 ± 0.01

0.04 ± 0.01

0.15 ± 0.02

0.09 ± 0.02