Skip to main content

Table 1 Setup errors as detected by 3DUS, CBCT, and skin marks

From: A comparative assessment of prostate positioning guided by three-dimensional ultrasound and cone beam CT

  

Mean ± SD

Median

Range

≤5 mm

Lateral axis

    
 

3DUS

−0.2 ± 2.7

−0.5

−5.6 … 6.9

92%

 

Bone-match

−0.3 ± 1.3

−0.3

−3.7 … 1.7

100%

 

Skin marks

−0.4 ± 2.6

−0.3

−6.1 … 4.5

95%

Longitudinal axis

    
 

3DUS

−1.9 ± 2.3

−2.0

−10.0 … 2.9

94%

 

Bone-match

−2.1 ± 2.5

−2.3

−7.5 … 8.9

90%

 

Skin marks

−1.5 ± 3.5

−1.6

−8.4 … 7.5

78%

Vertical axis

    
 

3DUS

0.0 ± 3.0

0.2

−7.3 … 7.2

92%

 

Bone-match

0.2 ± 3.0

−0.3

−5.3 … 9.4

92%

 

Skin marks

−0.8 ± 4.3

−0.8

−10.6 … 7.5

71%

  1. In units of mm; relative to position readings by seed-match; N = 78.