Skip to main content

Table 1 Systematic Σ and random σ setup error depending on surrogate used for matching kV/kV images to the reference CT

From: PTV margin definition in hypofractionated IGRT of localized prostate cancer using cone beam CT and orthogonal image pairs with fiducial markers

 

Σ (mm)

σ (mm)

Bones

vrt

1.78

1.97

 

lng

1.39

2.30

 

lat

0.77

0.85

Balloon

vrt

1.53

1.36

 

lng

3.20

3.15

 

lat

1.70

1.35

FM

vrt

0.88

0.71

 

lng

1.16

0.46

 

lat

0.46

0.37