Skip to main content

Table 3 Percent difference between planned and verification dose for kVCT and OMAR corrected kVCT images

From: Optimization of tomotherapy treatment planning for patients with bilateral hip prostheses

Image set kVCT OMAR corrected kVCT
Organ Bladder Rectum PTV Bladder Rectum PTV
D5 1.23 -0.20   0.95 0.19  
D10 1.03 -0.57   0.87 0.12  
D25 0.86 -0.51   0.47 0.05  
D35 0.87 -0.06   0.61 0.33  
D50 0.76 -0.05   0.76 0.43  
D1    1.38    0.86
D99    -0.77    0.08