From: Optimization of tomotherapy treatment planning for patients with bilateral hip prostheses
Image Set | Water | O-MAR kVCT | Original kVCT | ||||||
---|---|---|---|---|---|---|---|---|---|
 | Measured dose (Gy) | Planned dose (Gy) | % difference | Measured dose (Gy) | Planned dose (Gy) | % difference | Measured dose (Gy) | Planned dose (Gy) | % difference |
kVCT | 2.01 | 2.03 | -1.00 | 1.99 | 2.03 | -1.91 | 1.97 | 2.04 | -3.31 |
HybridCoarse | 2.01 | 2.04 | -1.14 | 2.02 | 2.03 | -0.72 | 2.01 | 2.03 | -1.21 |
HybridNormal | 2.01 | 2.04 | -1.06 | 2.00 | 2.02 | -0.92 | 2.02 | 2.03 | -0.75 |
HybridFine | 2.01 | 2.03 | -1.18 | 2.02 | 2.03 | -0.74 | 2.01 | 2.03 | -1.21 |
MV C | 1.89 | 2.01 | -6.21 | 1.91 | 2.01 | -5.04 | Â | Â | Â |
MV N | 1.91 | 2.03 | -6.00 | 1.92 | 2.02 | -5.41 | Â | Â | Â |
MV F | 1.91 | 2.02 | -5.73 | 1.91 | 2.03 | -6.11 | Â | Â | Â |