Skip to main content

Table 1 Time required in each process in the DQA procedure for five TMI case

From: Fast, simple, and informative patient-specific dose verification method for intensity modulated total marrow irradiation with helical tomotherapy

 

WB phantom (minute)

Cheese phantom (minute)

p-value

DQA setup in the DQA work station

10.4 ± 1.5

22.2 ± 0.8

p < 0.0001

Phantom setup

9.2 ± 1.3

15.0 ± 1.0

p < 0.0001

Beam delivery

37.8 ± 2.3

113.4 ± 6.9

p < 0.0001

Total

57.4 ± 2.3

150.6 ± 6.1

p < 0.0001