Table 1 Dose-volume indices between CBCT and CT

From: The feasibility of evaluating radiation dose to the heart by integrating kilovoltage-cone beam computed tomography in stereotactic body radiotherapy of early non-small-cell lung cancer patients

Parameter

CBCT, ± SEM

CT, ± SEM

t

P

V5Gy-Ra

47.7 ± 16.4%

49.0 ± 16.0%

1.5

0.17

V10Gy-R

18.9 ± 11.7%

17.4 ± 11.8%

2.3

0.03

V15Gy-R

14.4 ± 10.2%

12.8 ± 10.4%

2.3

0.02

V20Gy-R

10.5 ± 9.0%

9.0 ± 8.9%

2.3

0.04

V25Gy-R

6.6 ± 2.8%

4.5 ± 2.3%

3.6

0.01

V30Gy-R

4.2 ± 2.2%

2.5 ± 1.6%

3.7

0.01

V35Gy-R

3.2 ± 1.8%

1.7 ± 1.3%

3.9

0.01

V5Gy-Ab

333.0 ± 78.4 cm3

282.0 ± 61.7 cm3

7.9

0.0001

V10Gy-A

133.0 ± 68.6 cm3

101.0 ± 59.0 cm3

6.8

0.0001

V15Gy-A

102.0 ± 62.6 cm3

74.6 ± 53.7 cm3

6.2

0.0001

V20Gy-A

74.0 ± 55.8 cm3

52.5 ± 46.4 cm3

4.8

0.0001

V25Gy-A

46.0 ± 16.9 cm3

27.0 ± 12.0 cm3

6.2

0.0001

V30Gy-A

30.0 ± 14.4 cm3

15.0 ± 8.7 cm3

5.5

0.0001

V35Gy-A

24.0 ± 12.0 cm3

10.0 ± 7.0 cm3

5.5

0.0001

D50cm3c

19.9 ± 7.7 Gy

17.1 ± 6.3 Gy

4.5

0.0001

D100cm3

15.3 ± 6.2 Gy

12.8 ± 5.7 Gy

3.6

0.003

D150cm3

11.8 ± 5.5 Gy

9.2 ± 3.4 Gy

3.6

0.003

D200cm3

8.2 ± 2.0 Gy

6.8 ± 1.5 Gy

6.2

0.0001

D250cm3

6.4 ± 1.7 Gy

5.2 ± 2.0 Gy

6.5

0.0001

Dmean

7.0 ± 2.8 Gy

6.5 ± 2.4 Gy

3.0

0.009

Dmax

45.7 ± 13.1 Gy

45.1 ± 13.5 Gy

2.9

0.009

  1. a). V5Gy-R means the heart relative volume in the 5Gy region, V10Gy-R, V15Gy-R, V20Gy-R, V25Gy-R, V30Gy-R, and V35Gy-R are of the same mean;
  2. b). V5Gy-A means the heart absolute volume in the 5Gy region, V10Gy-A, V15Gy-A, V20Gy-A, V25Gy-A, V30Gy-A, and V35Gy-A are of the same mean;
  3. c). D50cm3 means the exposure dose in the 50 cm3, D100cm3, D150cm3, D200cm3, and D250cm3 are of the same mean.