Skip to main content

Table 2 OSL detector dose measurements for cycle one of first patient

From: A technique for pediatric total skin electron irradiation

OSL Detector Location AP + LPO + RPO Day 1 Dose (cGy) PA + LAO + RAO Day 2 Dose (cGy) Total Dose (cGy) % Diff from Prescription
Forehead 106.2 119.1 225.3 12.7
Posterior of head 152.1 96.2 248.3 24.2
Sternum 78.9 136.9 215.8 7.9
Posterior of sternum 137.8 85.9 223.7 11.9
Umbilicus 80.5 132.7 213.2 6.6
Posterior of Umbilicus 146.9 78.7 225.6 12.8
Left thigh 75.8 139.1 214.9 7.5
Posterior Left Thigh 145.9 81.5 227.4 13.7
Right thigh 77.1 133.7 210.8 5.4
Posterior Right Thigh 130.0 86.5 216.5 8.3
Right Foot 84.6 141.6 226.2 13.1
Bottom of Right Foot 135.9 82.6 218.5 9.3
Average    222.2 11.1